XTD THROMBECTOMY CATHETER, XX CM, MODEL# CT000500X; XTD CONTROL CONSOLE, MODEL # CT0006001; XTD COLLECTION BOTTLE, MODEL
K013473 · Xtrak Medical, Inc. · MCW · Mar 6, 2002 · Cardiovascular
Device Facts
Record ID
K013473
Device Name
XTD THROMBECTOMY CATHETER, XX CM, MODEL# CT000500X; XTD CONTROL CONSOLE, MODEL # CT0006001; XTD COLLECTION BOTTLE, MODEL
Applicant
Xtrak Medical, Inc.
Product Code
MCW · Cardiovascular
Decision Date
Mar 6, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For breaking apart and removing thrombus from hemodialysis access grafts.
Device Story
XTD™ Thrombectomy System consists of flexible, rotating catheter with spiral conveyor shaft, control console, and collection bottle. Device operates by rotating agitator to macerate thrombus within hemodialysis access graft; rotating spiral conveyor shaft conveys macerated thrombus through sheath to disposable collection bottle, assisted by negative pressure. Used by clinicians in clinical setting to clear hemodialysis access grafts. Output is physical removal of thrombus, restoring graft patency. Benefits patient by clearing obstructions in dialysis access.
Clinical Evidence
Performance testing included mechanical strength, fatigue strength, electrical safety, biocompatibility, in-vitro performance, in-vivo (animal) testing, and human clinical testing. Results demonstrated conformance to specifications and substantial equivalence to predicate devices.
Technological Characteristics
System comprises a flexible, rotating catheter with a spiral conveyor shaft, control console, and collection bottle. Operates via mechanical maceration and negative pressure-assisted aspiration. Materials are stated to be suitable for intended use and consistent with previously cleared products. No specific ASTM standards or software algorithm classes are detailed.
Indications for Use
Indicated for breaking apart and removing thrombus from hemodialysis access grafts.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
Possis AngioJet Rapid Thrombectomy System (RSC61-9)
Arrow International, Inc. (no specific model provided)
Related Devices
K021641 — XTD THROMBECTOMY CATHETER, XX CM · Xtrak Medical, Inc. · Jul 26, 2002
K083370 — D-CLOT THROMBECTOMY CATHETER SYSTEM, MODEL: DC 4006 · Artegraft, Inc. · Dec 12, 2008
K072195 — THROMCAT THROMBECTOMY CATHETER SYSTEM · Kensey Nash Corp. · Aug 23, 2007
K982363 — OASIS THROMBECTOMY CATHETER SYSTEM · Boston Scientific Corp · Feb 23, 1999
K230356 — Aspirex Thrombectomy System · C.R. Bard, Inc. · Mar 31, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
6 2002 MAR
K013473
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
XTD™ Thrombectorny System Xtrak Medical, Inc. January 17, 2002
#### General Information
Classification:
Class II
XTD™ Thrombectomy System Trade Name: Xtrak Medical, Inc. Sponsor: 26H Keewaydin, Dr. Salem, NH 03079 (603) 896-6416 Tel: (603) 893-7708 Fax: Gary Boseck, Ph.D. Contact: President
## Identification of Predicate or Legally Marketed Devices
ldentification of Predicate of Legally Manufactured by Possis Medical, Possis Anglosel Rapid Thrombecomy Oystem (RSC61-9), Manufactured by Arrow International, Inc.
#### Intended Use
mtended Use
#### Device Description
Device Description
The XTD™ Thrombectomy System consists of the XTD™ Thrombectomy Catheter has a flexible , rd The XTD™ Thrombectorly System Consists of the XTD™ catheter has a flexible, rotating Control Console, and ATD - Oollection BF plastic sheatist end of the spiral shaft spiral conveyor shart contained within a chipitator. As the agitator rotates it macerated thrombus within the graft and the rotating spiral conveyor shaft conveys the macerated thrombus within the gran and the rolating spiral of researc, to a disposable Collection Bottle.
thrombus through the sheath, assisted by negative pressure, to a disposation thromous inrough the sheath, assisted by nogative process. It as a many of the catheter operation.
#### Materials
Materials
All materials used in the manufacture of the XTD™ Catheter are suitable for their intended use All materials used in the manufacture of previously cleared products.
### Performance Testing
Perrormance Testing The ATD™ Thrombectonly Oystem oumponents nave been to the predicate devices.
their specifications and to support claims of substantial equivalence to the predicate devices. This testing includes the following:
- Mechanical Strength Testing .
- Fatigue Strength Testing .
- Electrical Safety Testing ●
- Biocompatibility Testing ●
- In-vitro Performance Testing ●
- In-vivo (animal) Testing ●
- Human Clinical Testing .
{1}------------------------------------------------
Test results have demonstrated conformance of the XTD™ components to their specification Test results have demonstrated confidmance of the ATD - compensent is the results and effective as the legally marketed predicated devices.
### Summary of Substantial Equivalence
Sammary of Gubountial Equiv Thrombectomy System is substantially equivalent to the legally marketed predicate devices. This claim of equivalence is supported by the identical intended use of the devices and their common technological characteristics.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design of an eagle or bird-like figure. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird-like figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### 6 2002 MAR
Xtrak Medical, Inc. c/o Gary Boseck, Ph.D. 26H Keewaydin Dr Salem, NH 03079
Re: K013473
XTD™ Thrombectomy System Regulation Number: 870.4875 Regulation Name: Intraluminal artery stripper Regulatory Class: II (two) Product Code: MCW Dated: January 17, 2002 Received: January 18, 2002
Dear Dr. Boseck:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your bection of the device is substantially equivalent (for the indications felerenced above and nave acterimentaly marketed predicate devices marketed in interstate for use stated in the encrosure) to regally manat date of the Medical Device Amendments, or to connine.co. phor to May 20, 1978, as can and and and the Federal Food, Drug, devices mat have occh reclassinod in assess approval of a premarket approval application (PMA). and Costicule Act (71ct) that to novice, subject to the general controls provisions of the Act. The I ou may, therefore, market the act include requirements for annual registration, listing of general controls provisions of ractice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 a00 ro) as . Existing major regulations affecting your device can may oc subject to suen additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I least be advised that I Dri 3 lisualite of our device complies with other requirements of the Act that I Dri has made a acterimalistions administered by other Federal agencies. You must
{3}------------------------------------------------
### Page 2 - Gary Boseck, Ph.D.
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFK Fart 807), fabeling (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QS) regulations (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin maxioning your maxicant of your device to a legally prematication. The PDA mixing of cassification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice to: your witto diagnostic devices), please contact the Office of additionally 21 CTTC Far 009.10 Additionally, for questions on the promotion and advertising of Compliance at (301) 594-1500. Traditions) , 1977-194-4639. Also, please note the your devices, prease commonding by reference to premarket notification" (21CFR Part 807.97). I guilation chitited, "Misolanants of responsibilities under the Act may be obtained from the Other general miormation on your responsional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Dork Tills
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Ver/ 3 - 4/24/96
Applicant: XTRAK MEDICAL, INC.
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: XTD™ Thrombectomy System
Indications For Use:
For breaking apart and removing thrombus from hemodialysis access grafts.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Respiratory Devige (Per 21 CFR 801.109)
510(k) Number K013473
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.