← Product Code [KFM](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/KFM) · K031300

# IBC FLOPUMP WITH GBS COATING (K031300)

_Gish Biomedical, Inc. · KFM · May 14, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/KFM/K031300

## Device Facts

- **Applicant:** Gish Biomedical, Inc.
- **Product Code:** [KFM](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/KFM.md)
- **Decision Date:** May 14, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4360
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The IBC FloPump with GBS™ Coating centrifugal blood pump is indicated for use only with the Medtronic Bio-Medicus Bio Console® to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support at flow rates of up to seven (7.0) LPM for periods up to six (6.0) hours. Federal Law (US) restricts this device to sale by or on the order of a physician.

## Device Story

Centrifugal blood pump utilizing constrained forced-vortex pumping principle; blood enters vortex created by rotating smooth-surfaced cones; energy transferred to blood as pressure and velocity; accelerates blood toward pump outlet. Fabricated from thromboresistant polycarbonate; mounted on precision bearings. Non-occlusive hemodynamic design promotes laminar flow; reduces blood trauma during extracorporeal circulation. Used in clinical settings for cardiopulmonary bypass; operated by medical professionals. Couples to Medtronic Bio-Medicus Bio Console® pump speed controller for hardware control. Provides circulatory support up to 7.0 LPM for up to 6 hours.

## Clinical Evidence

Bench testing only. No clinical data provided. Device subjected to extensive safety, performance, and validation testing to ensure compliance with functional requirements and industry safety standards.

## Technological Characteristics

Centrifugal blood pump; polycarbonate construction; non-leaching heparin-based GBS™ coating. Constrained forced-vortex pumping principle; non-occlusive design. Dimensions/form factor not specified. Standalone pump unit requiring external Medtronic Bio-Medicus Bio Console® for speed control and operation. Biocompatible, thromboresistant materials.

## Regulatory Identification

A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.

## Special Controls

*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
*Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
*Classification.* Class III (premarket approval).(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- IBC FloPump ([K983272](/device/K983272.md))
- Vision Hollow Fiber Oxygenator with GBS™ coating ([K023381](/device/K023381.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR part 807.92.

- 1. Company making the submission:

|            | Company                    | or | Correspondent (contract):   |
|------------|----------------------------|----|-----------------------------|
| Name:      | Gish Biomedical, Inc.      |    | Delphi Consulting Group     |
| Address:   | 22942 Arroyo Vista         |    | 11874 South Evelyn Circle   |
|            | Rancho Santa Margarita, CA |    | Houston, TX 77071-3404      |
|            | 92688-2600                 |    |                             |
| Telephone: | 949-635-6240 voice         |    | 832-285-9423 voice          |
|            | 949-635-6294 fax           |    | 775-429-9524 fax            |
|            |                            |    | harvey@delphiconsulting.com |
| Contact:   | Edward F. Waddell          |    | J. Harvey Knauss            |
|            | Director RA/QA             |    | Consultant                  |

## 2. Device:

| Proprietary Name:    | IBC FloPump with GBS™ Coating                    |
|----------------------|--------------------------------------------------|
| Common Name:         | Centrifugal Pump                                 |
| Classification Name: | Nonroller-type cardiopulmonary bypass blood pump |

- 3. Predicate Devices:
IBC FloPump. International Biophysics Corporation, K983272. Pump. Vision Hollow Fiber Oxygenator with GBS™ coating, K023381. Coating.

- 4. Classifications Names & Citations:
- 21 CFR 870.4360, Nonroller-type cardiopulmonary bypass blood pump, Class III, 5. KFM, Cardiovascular.
- 6. Description:

The IBC FloPump with GBS™ Coating (a non leaching heparin based coating) centrifugal blood pump has been designed to utilize the constrained forced-vortex pumping principle. In action, blood is allowed through a vortex created by the rotation of a series of smooth surfaced rotating cones. Energy is transferred from the cones to the blood in the form of pressure and velocity as the blood is gently accelerated toward the outlet of the pump.

The IBC FloPump with GBS™ Coating centrifugal blood pump is fabricated from sturdy, thromboresistant, biocompatible polycarbonate materials which are mounted on precision bearings for dependable operation. The non-occlusive hemodynamic design of the pump promotes laminar flow, allowing for improved blood handling

{1}------------------------------------------------

capabilities and decreasing trauma which may be associated with extracorporeal circulatory support during cardiopulmonary bypass.

The IBC FloPump with GBS™ Coating centrifugal blood pump couples to a safetylocked magnetic drive unit called the Medtronic Bio-Medicus Bio Console® purnp speed controller, which serves as the hardware component of the extracorporeal blood delivery system.

- 7. Indications for use:
The IBC FloPump with GBS™ Coating centrifugal blood pump is indicated for use only with the Medtronic Bio-Medicus Bio Console® to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support at flow rates of up to seven (7.0) LPM for periods up to six (6.0) hours.

- 8. Contraindications:
For heparin coated devices, heparin has been reported, on rare occasions, to induce thrombocytopenia. Since patients undergoing cardiopulmonary bypass are routinely systemically heparinized, and although the amount of heparin contributed by this device is very small in comparison to the typical dose given, caution should be exercised when using this device in patients with known or suspected heparin sensitivity.

- 9. Comparison:
The IBC FloPump with GBS™ Coating has the same device characteristics as the predicate devices.

- 10. Test Data:
The IBC FloPump with GBS™ Coating has been subjected to extensive safety, performance, and validations prior to release. Final testing for the systems includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications. Safety tests have further been performed to ensure the device complies with applicable industry and safety standards.

- 11. Literature Review:
A review of literature pertaining to the safety and effectiveness has been conducted. Appropriate safeguards have been incorporated in the design of IBC FloPump with GBS™ Coating.

- 12. Conclusions:
The conclusion drawn from these tests is that IBC FloPump with GBS™ Coating is equivalent in safety and efficacy to its predicated devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 4 2003

Gish Biomedical, Inc. c/o Mr. J. Harvey Knauss Delphi Consulting Group 11874 South Evelyn Circle Houston, TX 77071

Re: K031300

IBC FloPump with GBS™ Coating Regulation Number: 21 CFR 870.4360 Regulation Name: Nonroller type cardiopulmonary bypass blood pump Regulatory Class: Class II (two) Product Code: KFM Dated: Undated Received: April 24, 2003

Dear Mr. Knauss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced.above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. J. Harvey Knauss

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

K. Zuckerman, M.D.

ram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number K_031306

Device Name: IBC FloPump with GBS™ Coating

## Indications for use:

The IBC FloPump with GBS™ Coating centrifugal blood pump is indicated for use only with the Medtronic Bio-Medicus Bio Console® to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support at flow rates of up to seven (7.0) LPM for periods up to six (6.0) hours.

## Prescription Device:

Federal Law (US) restricts this device to sale by or on the order of a physician.

PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use | <span></span>        |
|------------------|----------------------|
| OR               | Over-The-Counter Use |

(Division Sign-Off)
Division of Cardiovascular Devices

| 510(k) Number | K031300 |
|---------------|---------|
|---------------|---------|

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/KFM/K031300](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/KFM/K031300)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
