← Product Code [DXC](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC) · K963503

# CTS MIDCAB CORONARY SHUNT (K963503)

_Cardiothoracic Systems, Inc. · DXC · Apr 3, 1997 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC/K963503

## Device Facts

- **Applicant:** Cardiothoracic Systems, Inc.
- **Product Code:** [DXC](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC.md)
- **Decision Date:** Apr 3, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Intended Use

The CTS MIDCAB Coronary Shunt is intended to be used during minimally invasive direct coronary artery bypass (“MIDCAB”) procedures in conjunction with or without the CTS MIDCAB™ Access Platform and Stabilizer. The CTS Coronary Shunt is designed to help reduce blood in the operative field by temporary occlusion of the artery and to provide blood flow distal to the arteriotomy. This CTS Coronary Shunt is not an implant and is removed prior to completion of the surgery.

## Device Story

Hollow tube device with tapered atraumatic tips and dual seals; used during MIDCAB procedures to provide temporary coronary artery occlusion and distal blood flow. Operated by surgeons in the OR; inserted into the coronary artery to reduce blood in the operative field. Includes a tether for insertion/removal and a separate cylinder-based Shunt Remover for withdrawal post-procedure. Device is non-implantable and removed before surgery completion. Benefits include improved visualization and blood flow maintenance during bypass grafting.

## Clinical Evidence

Bench testing only. Functional performance testing confirmed the device's ability to occlude arteries and maintain distal blood flow. Biocompatibility testing confirmed material safety.

## Technological Characteristics

Hollow tube with tapered atraumatic tips and dual seals; includes tether and shunt remover cylinder. Materials passed biocompatibility testing. Non-implantable, single-use device.

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Predicate Devices

- Bio-Vascular, Inc. Flo-Rester
- Research Medical, Inc. Yacoubian Clamp External Coronary Artery Occluder
- Research Medical, Inc. Carotid Artery Shunts
- Research Medical, Inc. Vacu-Sponge Surgical Sponge

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) Premarket Notification
CTS MIDCAB Coronary Shunt
K963503
APR - 3 1997

# 510(k) Summary
## CardioThoracic Systems, Inc.
### CTS MIDCAB™ Coronary Shunt
#### 510(k) Notification K 963503

## GENERAL INFORMATION

**Manufacturer:** CardioThoracic Systems, Inc.
10600 N. Tantau Avenue
Cupertino, California
(408) 342-1700
(408) 342-1717 FAX
Est. Reg. No. (awaiting issuance)

**Contact Person:** Michael J. Billig
Vice President, Regulatory, Quality, and Clinical Research

**Date Prepared:** August 31, 1996

## DEVICE DESCRIPTION

**Classification:** Class II

**Trade Name:** CTS MIDCAB Coronary Shunt

**Generic/Common Name:** Vascular clamp 870.4450
Surgical vessel dilator 870.4475
Blood access device and accessories 876.5540

## PREDICATE DEVICES

1. Bio-Vascular, Inc. Flo-Rester
2. Research Medical, Inc. Yacoubian Clamp External Coronary Artery Occluder
3. Research Medical, Inc. Carotid Artery Shunts
4. Research Medical, Inc. Vacu-Sponge Surgical Sponge

## INTENDED USE

The CTS MIDCAB Coronary Shunt is intended to be used during minimally invasive direct coronary artery bypass (“MIDCAB”) procedures in conjunction with or without the CTS MIDCAB™ Access Platform and Stabilizer. The CTS Coronary Shunt is designed to help reduce blood in the operative field by temporary occlusion of the artery and to provide blood flow distal to the arteriotomy. This CTS Coronary Shunt is not an implant and is removed prior to completion of the surgery.

18

{1}

510(k) Premarket Notification
CTS MIDCAB Coronary Shunt

## PRODUCT DESCRIPTION

The CTS MIDCAB Coronary Shunt is a hollow tube with atraumatic tips on each end and two seals which are utilized for proximal and distal occlusion of the artery. The atraumatic tips are tapered for easy insertion and removal from the artery. Each tip contains holes to provide for blood flow into and out of the CTS Shunt. The thread or tether is attached to a tab and is used to aid insertion and removal of the CTS Shunt. The blood is occluded and flows through the Shunt which reduces the blood in the operative field. Also included with the CTS MIDCAB Coronary Shunt is a Shunt Remover, which is a cylinder which can be used to withdraw the CTS Shunt following completion of the procedure.

## SUBSTANTIAL EQUIVALENCE

The CTS MIDCAB Coronary Shunt is substantially equivalent to predicate devices currently being marketed. The marketed predicate devices are identified above. The CTS MIDCAB Coronary Shunt is substantially equivalent to the predicate devices with regard to intended use, function, physical characteristics, materials and sterilization method.

All necessary testing was performed on the CTS MIDCAB Coronary Shunt to ensure the product is substantially equivalent to the predicate devices and to ensure that the CTS Coronary Shunt does not have any differences which have a significant effect on safety and effectiveness.

## FUNCTIONAL PERFORMANCE TESTING

Functional testing was conducted on the CTS MIDCAB Coronary Shunt to ensure that the Shunt would function according to its intended use instructions. All testing conducted confirmed the acceptability of the CTS Shunt to perform as intended to help reduce blood in the operative field by temporary occlusion of the artery while providing blood flow distal to the arteriotomy.

## BIOCOMPATIBILITY EVALUATION

The biocompatibility testing was conducted on the CTS MIDCAB Coronary Shunt and Shunt materials to ensure the acceptability of the CTS MIDCAB Coronary Shunt when used as directed. The CTS Shunt and Shunt materials passed the necessary biocompatibility tests.

## SUMMARY

As contained in this 510(k) summary, all necessary testing was conducted on the CTS MIDCAB Coronary Shunt to ensure that the device is safe and effective when used in accordance to its intended use.

19

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC/K963503](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC/K963503)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
