VIOLA clampless proximal anastomosis sealing system is indicated for use by cardiac surgeons during coronary artery bypass grafting procedures to maintain hemostasis and facilitate the completion of proximal anastomoses to the aorta without application of an aortic clamp.
Device Story
VIOLA is a sterile, single-use system for clampless proximal anastomosis during CABG. Components include a 4 mm aortic punch, a sealing assembly (sealing element, catheter, fixation strap, handle), and a silicone boundary marker stencil. Procedure: surgeon marks aortic site, creates needle hole, deploys sealing element, punches anastomosis hole, and performs anastomosis. Device supports up to 4 anastomoses per patient. Modified version features ergonomic handle with operating button and straight distal end. Used in OR by cardiac surgeons to avoid aortic clamping, potentially reducing trauma/complications associated with traditional clamping.
Clinical Evidence
Bench testing only. Risk analysis (FMECA) identified necessary verification tests: bond strength and simulated use. All tests met acceptance criteria.
Technological Characteristics
Sterile, single-use mechanical system. Components: 4 mm aortic punch, sealing assembly (sealing element, catheter, fixation strap, handle), silicone boundary marker. Ergonomic handle with operating button and straight distal end. No energy source or software.
Indications for Use
Indicated for cardiac surgeons performing coronary artery bypass grafting (CABG) procedures to maintain hemostasis and facilitate proximal anastomoses to the aorta without an aortic clamp.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
K080169 — HEARTSTRING III PROXIMAL SEAL SYSTEM · Guidant Corporation, Cardiac Surgery · Feb 8, 2008
K033979 — CORLINK AUTOMATED ANASTOMOTIC SYSTEM · By-Pass Makafim , Ltd. · Mar 31, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
February 28, 2023
Vascular Graft Solutions, Ltd. c/o Janice Hogan Partner Hogan Lovells US LLP 1735 Market St. 23rd Floor Philadelphia, Pennsylvania 19103
Re: K230248
Trade/Device Name: Viola Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular clamp Regulatory Class: Class II Product Code: DXC Dated: January 30, 2023 Received: January 30, 2023
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rohini Retarekar -S
for Katherine Trivedi Acting Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230248
Device Name VIOLA
Indications for Use (Describe)
VIOLA clampless proximal anastomosis sealing system is indicated for use by cardiac surgeons during coronary artery bypass grafting procedures to maintain hemostasis and facilitate the completion of proximal anastomoses to the aorta without application of an aortic clamp.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:10pt">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:10pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K230248
# 510(k) SUMMARY
### Vascular Graft Solutions, Ltd.'s VIOLA
#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Vascular Graft Solutions, Ltd. 24 Raoul Wallenberg Street Tel-Aviv, Israel 6971921 Phone: +972-54-3073050 Facsimile: +972-3-6024966
Contact Person: Orit Yarden Date Prepared: 30 January 2023
#### Name of Device
VIOLA
#### Common or Usual Name
Vascular Clamp
#### Classification
21 CFR 870.4450, Class II, Product Code DXC
# Predicate Device
Vascular Graft Solutions, Ltd.'s VIOLA (K203307)
#### Intended Use / Indications for Use
VIOLA clampless proximal anastomosis sealing system is indicated for use by cardiac surgeons during coronary artery bypass grafting procedures to maintain hemostasis and facilitate the completion of proximal anastomoses to the aorta without application of an aortic clamp.
#### Device Description
The VIOLA is a sterile, single-use device designed to maintain hemostasis and facilitate the completion of multiple proximal anastomoses during a coronary artery bypass grafting procedure, without application of an aortic clamp. The VIOLA system can be used to create up to 4 anastomoses within the same patient.
This modified version of VIOLA includes an ergonomic handle with operating button and a straight distal end.
{4}------------------------------------------------
The VIOLA is comprised of:
- 1. A concentric 4 mm aortic punch with a detachable handle.
- 2. A sealing assembly comprising of a sealing element (which is available in two diameters), a catheter, fixation strap and a handle.
- 3. A silicone boundary marker stencil for marking the maximal suture line around the aortic incision.
The steps to create a sealed anastomosis hole that enables the surgeon to perform a clampless anastomosis include (1) the creation of a small "needle hole" (performed in the center of the boundary marker), (2) insertion and deployment of the sealing element, (3) creation of an anastomosis hole using the VIOLA's punch, and (4) performing the anastomosis.
# Performance Data
A risk analysis was performed using the Failure Mode Effect and Criticality Analysis (FMECA) method to assess the impact of the changes. The risk analysis determined the following design verification tests were necessary to repeat: Bond Strength Testing and Simulated Use Testing. These tests passed successfully all acceptance criteria.
# Substantial Equivalence
The modified VIOLA has the same intended use, indications for use, and principles of operation, and similar technological characteristics, as the predicate VIOLA (K203307). The minor differences in the modified VIOLA's technological characteristics do not raise different questions of safety or effectiveness. Performance data demonstrates that the modified VIOLA is as safe and effective as the predicate VIOLA.
# Conclusions
The VIOLA is substantially equivalent to its predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.