GEM™ BIOVER disposable microvascular clamps are instruments which are used for all microsurgical procedures. They are used to occlude vessels during anastomosis, which is necessary as a result of vessel damage or thrombosis.
Device Story
GEM™ Biover Microvascular Clamps are sterile, single-use, disposable instruments used in microsurgery, plastic surgery, and hand surgery. Available in single and double configurations, they are color-coded by vessel type: yellow for arteries (higher clamping force) and green for veins. Clamps are applied by surgeons or scrubbed personnel in a hospital setting to temporarily occlude blood vessels during anastomosis. The device functions mechanically to stop blood flow, facilitating vessel repair. No software or electronic components are involved.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by design identity, non-clinical performance testing, and a packaging summative human factors/usability study.
Technological Characteristics
Mechanical vascular clamps; single and double configurations. Materials: Not specified. Sterilization: Gamma radiation. Packaging: Double Tyvek pouches. No energy source, connectivity, or software.
Indications for Use
Indicated for patients undergoing microsurgical procedures requiring temporary occlusion of blood vessels (arteries or veins) during anastomosis due to vessel damage or thrombosis.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
September 14, 2022
Synovis Micro Companies Alliance, Inc. (a subsidiary of Baxter International. Inc.) Julie Carlston Senior Manager, Regulatory Affairs 439 Industrial Lane Birmingham, Alabama 35211
Re: K221843
Trade/Device Name: GEM™ Biover Microvascular Clamps Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular clamp Regulatory Class: Class II Product Code: DXC Dated: August 12, 2022 Received: August 15, 2022
Dear Julie Carlston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221843
Device Name GEMTM BIOVER Microvascular Clamps
Indications for Use (Describe)
GEM™ BIOVER disposable microvascular clamps are instruments which are used for all microsurgical procedures. They are used to occlude vessels during anastomosis, which is necessary as a result of vessel damage or thrombosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary K221843
l. Submitter Synovis Micro Companies Alliance, Inc. A Subsidiary of Baxter International Inc. 439 Industrial Lane Birmingham, AL 35211-4464 Establishment Registration Number: 1062741
Julie Carlston Synovis Micro Companies Alliance, Inc. A Subsidiary of Baxter International Inc. 2575 University Avenue West St. Paul, MN 55114-1024 Tel: 651-796-7511 Fax: 651-642-9018 julie_carlston@baxter.com
Date Prepared: September 13, 2022
#### II. Subject Device K221843
| Device Trade Name | GEM™ Biover Microvascular Clamps |
|---------------------|-----------------------------------------------------------------------------------|
| Common Name | Microvascular Clamps |
| Classification Name | Clamp, Vascular<br>Regulation Number: 21 CFR 870.4450<br>Classification: Class II |
| Product Code | DXC |
#### III. Predicate Device K011208
| Device Trade Name | Biover Microvascular Clamps |
|---------------------|-----------------------------------------------------------------------------------|
| Common Name | Microvascular Clamps |
| Classification Name | Clamp, Vascular<br>Regulation Number: 21 CFR 870.4450<br>Classification: Class II |
| Product Code | DXC |
#### IV. Device Description
The GEM™ Biover Microvascular Clamps (Clamps) are arterial and venous clamps in both single and double clamp configurations, available in varying sizes. The clamps are sterile, disposable clamps for single use. The clamps are used in microsurgery, plastic surgery, and hand surgery for the temporary occlusion of blood vessels during an anastomosis.
The clamps are available for arteries and veins. The clamping force is higher for the arterial clamps than for the veinous clamps. The clamps are identified by a product reference code and by the color of the clamp. The clamps for veins are green, and the clamps for arteries are yellow.
{5}------------------------------------------------
The clamps are packaged individually in two Tyvek pouches (double pouched). The devices in the Tyvek pouches are gamma sterilized and are placed in cartons of 10. The clamps are used once and are disposed of after use.
#### V. Intended Use and Indications for Use
The Intended Use remains the same as the predicate device (K011208).
Indication for Use
GEM BIOVER disposable microvascular clamps are instruments which are used for all microsurgical procedures. They are used to occlude vessels during anastomosis, which is necessary as a result of vessel damage or thrombosis.
The wording of the Indication for Use was modified slightly from the predicate to improve the clarity of the statement while maintaining the same intended use and scope. The indications for use statements are the same in intended clinical use and therefore there is no substantial change to the indication for use of the microvascular clamps. The similarity in the indication for use statements supports substantial equivalence.
## VI. Comparison of Technological Characteristics with the Predicate Device
The device and primary packaging of the double pouch configuration (sterile packaging) remain unchanged and are identical to the predicate device. There are no engineering or performance changes to the device or its packaging. The only technology change is a minor modification to the sales unit carton (box of 10 individual pouches) which was made to ease removal of the packaged (pouched) clamps. There is no change in sterilization.
In addition to the minor change to the carton, updates were made to the Instruction Manual and package labeling to reflect the new tradename of the device.
| Attribute | Subject of this submission | Predicate Device | Comparison |
|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | GEM BIOVER Microvascular<br>Clamps | BIOVER Microvascular Clamps | - |
| Manufacturer<br>(Specification<br>Developer) | SYNOVIS MICRO COMPANIES<br>ALLIANCE, INC. (A Subsidiary Of<br>Baxter International Inc.)<br>439 Industrial Ln<br>BIRMINGHAM, AL 35211 | Biover AG<br>Muliweg 2<br>Hergiswil Nidwalden, CH 114-002 | Change due to<br>acquisition of<br>Biover AG |
| Contract<br>Manufacturer | DIADEM CORPORATION<br>582-7, Ubaga-Cho<br>Hitachiomiya-Shi<br>Ibaraki, JP 319-2251 | DIADEM CORPORATION<br>582-7, Ubaga-Cho<br>Hitachiomiya-Shi<br>Ibaraki, JP 319-2251 | Same |
| Attribute | Subject of this submission<br>GEM BIOVER Microvascular<br>Clamps | Predicate Device<br>BIOVER Microvascular Clamps | Comparison |
| Intended Use | GEM BIOVER disposable<br>microvascular clamps are used<br>in microsurgical procedures to<br>occlude blood vessels during<br>anastomosis. | BIOVER disposable<br>microvascular clamps are used<br>in microsurgical procedures to<br>occlude blood vessels during<br>anastomosis. | Same |
| Indications for Use | Disposable microvascular<br>clamps are instruments which<br>are used for all microsurgical<br>procedures. They are used to<br>occlude vessels during<br>anastomosis, which is<br>necessary as a result of vessel<br>damage or thrombosis. | BIOVER disposable<br>microvascular clamps are<br>used for end to end<br>anastomotic procedures for<br>arteries and veins. The<br>single version is placed to<br>stop blood flow for a certain<br>time (less than 16 hours).<br>The double version is used<br>for the repair of ruptured<br>arteries and veins. | Equivalent. Update<br>of tradename. The<br>differences in<br>wording are<br>clarifying the<br>intended use of the<br>device and do not<br>constitute a<br>change in their<br>clinical use |
| General Description | | | |
| Product Use<br>Environment | Hospital | Hospital | Same |
| End Users of<br>Product | Surgeons, scrubbed persons,<br>and nurses | Surgeons, scrubbed persons,<br>and nurses | Same |
| Clamp Design | 2 Types of Clamps are available<br>1) Clamps for arteries are<br>yellow (single (1) and<br>double (2))<br>2) Clamps for veins are<br>green (single (1) and<br>double (2)) | 2 Types of Clamps are available<br>1) Clamps for arteries are<br>yellow (single (1) and<br>double (2))<br>2) Clamps for veins are green<br>(single (1) and double (2)) | Same |
| Vessel Diameter | Artery Clamps (yellow):<br>< 1 mm,<br>1 mm - 2 mm,<br>2mm - 4mm<br><br>Vein Clamps (green):<br>0.2 – 1 mm<br>< 1 mm,<br>1 mm-2 mm,<br>2 mm | Artery Clamps (yellow):<br>< 1 mm,<br>1 mm - 2 mm,<br>2mm - 4mm<br><br>Vein Clamps (green):<br>0.2 – 1 mm<br>< 1 mm,<br>1 mm-2 mm,<br>2 mm | Same |
| Contraindications | None known. | None known | Same |
| Labeling | | | |
| Attribute | Subject of this submission<br>GEM BIOVER Microvascular Clamps | Predicate Device<br>BIOVER Microvascular Clamps | Comparison |
| Labels and<br>Instruction Manual<br>for<br>Monitor and Power<br>Supply | See section 13 - Proposed<br>Labeling | See section 13 - Proposed<br>Labeling | Equivalent.<br>Labeling was<br>modified to reflect<br>the new legal<br>manufacturer.<br>Branding updated<br>to reflect new<br>tradename.<br>Indication for Use<br>statement updated<br>to reflect the<br>clarified indication<br>for use statement |
| Packaging | Each clamp is packaged in a<br>double Tyvek pouch<br>configuration. 10 pouches<br>packed in an outer carton. | Each clamp is packaged in a<br>double Tyvek pouch<br>configuration. 10 pouches<br>packed in an outer carton. | Same. Minor<br>change to the<br>opening geometry<br>of the outer carton<br>of the sales unit<br>has no impact on<br>the device |
### Table 1: Predicate and subject device comparison
{6}------------------------------------------------
{7}------------------------------------------------
#### VII. Performance Data
The label design change and the outer carton sales unit design change were assessed and validated in a Packaging Summative Human Factors/Usability Study. Historical complaint data were reviewed and indicate no use-related concerns, thus, there are no use-related risks or complaints that trigger a need for further Human Factors validation.
#### VIII. Conclusions
The GEM Biover microvascular clamps remain unchanged from the predicate device. Performance testing has been completed to evaluate safety and effectiveness of the subject device as it compares to its predicate. The conclusions drawn from the risk-benefit assessment and nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device. The proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.