← Product Code [DXC](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC) · K162988

# PreludeSYNC Radial Compression Device (K162988)

_Merit Medical Systems, Inc. · DXC · Jan 25, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC/K162988

## Device Facts

- **Applicant:** Merit Medical Systems, Inc.
- **Product Code:** [DXC](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC.md)
- **Decision Date:** Jan 25, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The PreludeSYNC is a compression device used to assist in gaining hemostasis of arterial percutaneous access sites.

## Device Story

PreludeSYNC is a sterile, single-use radial compression device; consists of a soft wristband with hook-and-loop fastener, clear curved backer plate for puncture site visualization, and an inflatable bulb. Used in clinical settings by healthcare providers following catheterization. The device is placed around the patient's wrist; the inflatable bulb is positioned over the puncture site. Inflation and deflation are managed via a check valve and tubing using an accompanying 20ml syringe. Adjustable compression assists in achieving and maintaining hemostasis of the radial or ulnar artery. The clear backer plate allows clinicians to monitor the puncture site during the compression process. Benefits include controlled, adjustable pressure to stop bleeding at the access site while maintaining visibility.

## Clinical Evidence

Bench testing only. Performance data included dimensions, tensile strength of joints, submerged air leak test, inflation bulb burst test, hook-and-loop detachment test, integrity test, band absorbency, backer plate transparency, ink adherence, and simulated use. Biocompatibility testing included cytotoxicity, sensitization, irritation, and acute systemic toxicity per ISO 10993 standards.

## Technological Characteristics

Sterile, single-use wristband with hook-and-loop fastener and clear curved backer plate. Features inflatable bulb, check valve, and tubing for manual inflation/deflation via 20ml syringe. Available in 24cm and 29cm sizes. Complies with ISO 594-1/2 (Luer fittings), ISO 11607-1 (packaging), ISO 11135 (EO sterilization), and ISO 10993 (biocompatibility). ASTM standards include D4169-09 (shipping) and F1980-11 (accelerated aging).

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Predicate Devices

- Vasc™ Band Hemostat ([K142359](/device/K142359.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 25, 2017

Merit Medical Systems, Inc. Ileana Davis Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, Utah 84095

Re: K162988

Trade/Device Name: PreludeSYNC Radial Compression Device Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: December 22, 2016 Received: December 30, 2016

Dear Ms. Davis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv.

M.A. Hillemann

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162988

#### Device Name

PreludeSYNC Radial Compression Device

#### Indications for Use (Describe)

The PreludeSYNC is a compression device used to assist in gaining hemostasis of arterial percutaneous access sites.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5.0 510(k) Summary

| General<br>Provisions                | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date Prepared:<br>Registration Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4187<br>(801) 316-4065<br>Ileana Davis, Regulatory Affairs Specialist<br>26 October 2016<br>1721504 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device                       | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | PreludeSYNC™ Radial Compression Device<br>Radial Compression Device<br>Vascular Clamp<br>II<br>DXC<br>870.4450<br>Cardiovascular                                                                  |
| Predicate<br>Device                  | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Vasc™ Band Hemostat<br>Vascular Clamp<br>K142359<br>Lepu Medical Technology (Beijing) Co., Ltd.<br>This predicate has not been subject to a design-related recall in the US.                      |
| Device<br>Description                | The PreludeSYNC™ Radial Compression Device is a sterile, single use<br>disposable device used to assist in gaining and maintaining hemostasis<br>of the radial and ulnar artery following catheterization procedures. It<br>consists of a soft wristband with a secure hook and loop fastener and a<br>clear curved backer plate that provides optimal visualization of the<br>puncture site and ease of placement. The inflatable bulb delivers<br>adjustable compression of the puncture site. A check valve and tubing<br>allow for easy inflation and deflation with the accompanying 20ml<br>syringe inflator.<br>PreludeSYNC is available in a variety of graphic designs and in two<br>band sizes: regular (24cm) and long (29cm). |                                                                                                                                                                                                   |
| Indications for<br>Use               | The PreludeSYNC is a compression device used to assist in gaining<br>hemostasis of arterial percutaneous access sites.<br>The Indications for Use statement for the PreludeSYNC is not identical<br>to the predicate device; however, the differences do not alter the<br>intended therapeutic use of the device nor do they affect the safety and<br>effectiveness of the device relative to the predicate. Both the subject                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                   |
| Comparison to<br>Predicate<br>Device | and predicate devices have the same intended use of providing<br>compression to assist hemostasis of percutaneous access sites.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                   |
|                                      | The technological characteristics of the subject PreludeSYNC™ Radial<br>Compression device are substantially equivalent to those of the<br>predicate Vasc™ Band Hemostat.<br>At a high level, the subject and predicate devices are based on the<br>following same technological elements:                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                   |
|                                      | Clinical use Labeling Basic design Principle of operation Performance Band sizes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                   |
|                                      | The following technological difference exists between the subject and<br>predicate devices:  Materials                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                   |
|                                      | No performance standards have been established under Section 514 of<br>the Food, Drug and Cosmetic Art for these devices. Performance                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                   |

{4}------------------------------------------------

-

rug and Cosmetic Act for testing of the subject PreludeSYNC™ Radial Compression Device was conducted based on the risk analysis and based on the requirements of the following international standards:

- ISO 594-1 Conical fittings with 6% (Luer) taper for syringes. ● needles and certain other medical equipment – Part 1: General requirements
- . ISO 594-2 - Conical fittings with 6% (Luer) taper for syringes. needles and certain other medical equipment – Part 2: Lock fittings
- ISO 2233:2001 Packaging Complete, filled transport packages . and unit loads - Conditioning for testing

Performance Data

- . ISO 11607-1:2006 - Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems, and packaging systems
- . ASTM D4169-09 - Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM F1980-11 Standard Guide for Accelerated Aging of Sterile . Barrier Systems for Medical Devices
- . ISO 11135:2014 – Sterilization of health-care products – Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
- . ISO 10993-1:2009 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-7:2008 – Biological evaluation of medical devices – Part 7: Ethylene oxide residuals

{5}------------------------------------------------

- ISO 10993-10:2010 Biological evaluation of medical devices ● Part 10: Tests for irritation and skin sensitization
- . ISO 10993-11:2006 – Biological evaluation of medical devices – Part 11: Tests for systemic toxicity

The following performance data were provided in support of the substantial equivalence determination:

# Performance Testing - Bench

- Dimensions ●
- Tensile of Joints ●
- Submerged Air Leak Test ●
- Inflation Bulb Burst Test ●
- Hook-and- Loop Detachment Test
- Integrity Test ●
- Band Absorbency
- Backer Plate Transparency ●
- Ink Adherence ●
- Simulated Use

## Biocompatibility testing

The biocompatibility evaluation for the PreludeSYNC™ Radial Compression Device was conducted in accordance with the FDA Guidance Document "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" June 16, 2016, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following:

- Cytotoxicity ●
- Sensitization ●
- Irritation
- Acute Systemic Toxicity ●

Based on the indications for use, design, safety and performance testing, the subject PreludeSYNC™ Radial Compression Device Summary of meets the requirements that are considered essential for its intended Substantial use and is substantially equivalent to the predicate device, the Vasc™ Equivalence Band Hemostat, K142359.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC/K162988](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC/K162988)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
