← Product Code [DXC](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC) · K143078

# Bookwalter Gordon Adjustable Compression Element (K143078)

_Symmetry Surgical · DXC · May 14, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC/K143078

## Device Facts

- **Applicant:** Symmetry Surgical
- **Product Code:** [DXC](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC.md)
- **Decision Date:** May 14, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Intended Use

The Bookwalter Gordon Adjustable Compression Element is indicated for use to provide hemostasis of the femoral vascular access site, during and following catheterization or cannulation procedures.

## Device Story

The Bookwalter Gordon Adjustable Compression Element is a reusable, stainless steel mechanical compression device designed to replace manual hand pressure for achieving hemostasis at femoral vascular access sites. The device consists of a center block and adjustable finger subassemblies mounted on an articulating arm, which is secured to a surgical table. Clinicians position the device over the access site and adjust the individual fingers to apply targeted compressive pressure. Once positioned, the device maintains continuous compression without further manual intervention. It is intended for use in clinical settings during and after catheterization or cannulation procedures. By providing stable, consistent pressure, the device reduces the need for manual staff intervention, potentially improving workflow efficiency while ensuring effective hemostasis for the patient.

## Clinical Evidence

No clinical data. Evidence is based on bench testing, including mechanical performance (maximum compression force, expansion/contraction under load, 2X force load), steam sterilization efficacy, dry time validation, and cytotoxicity testing.

## Technological Characteristics

Constructed of 17-4 PH stainless steel. Consists of left, right, and center finger subassemblies mounted on an adjustable table-mounted arm. Operates via mechanical compression. Reusable; sterilized via steam. No software or electronic components.

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Predicate Devices

- CompressAR SuperComfort Disc and StrongArm SuperComfort System ([K040615](/device/K040615.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract design of three human profiles facing to the right, with flowing lines extending from the bottom of the profiles.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 14, 2015

Symmetry Surgical c/o Mr. David Furr Regulatory Affairs Consultant FDC Services 8708 Capeheart Cove Austin, TX 78733

Re: K143078

Trade/Device Name: Bookwalter Gordon Adjustable Compression Element Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: April 1, 2015 Received: April 3, 2015

Dear Mr. Furr,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.A. Willhelmen

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| <center>DEPARTMENT OF HEALTH AND HUMAN SERVICES</center> <center>Food and Drug Administration</center> <center><b>Indications for Use</b></center> |                                                                                                                                                                                                                  | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) Number (if known)                                                                                                                           |                                                                                                                                                                                                                  |                                                                                                   |
| Device Name                                                                                                                                        | Bookwalter Gordon Adjustable Compression Element                                                                                                                                                                 |                                                                                                   |
| Indications for Use (Describe)                                                                                                                     | The Bookwalter Gordon Adjustable Compression Element is indicated for use to provide hemostasis of the femoral             vascular access site, during and following catheterization or cannulation procedures. |                                                                                                   |
| Type of Use (Select one or both, as applicable)                                                                                                    | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                                                        |                                                                                                   |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Pursuant to 21 CFR 807.92

### K143078

### Date: May 13, 2015

| 1. | Submitted By:      | Symmetry Surgical Inc.<br>3034 Owen Drive<br>Antioch, TN 37013<br>1-800-251-3000                |
|----|--------------------|-------------------------------------------------------------------------------------------------|
| 2. | Contact:           | David C. Furr<br>FDC Services, LLC<br>8708 Capehart Cove<br>Austin, Texas 78733<br>512-906-9654 |
| 3. | Product:           | Bookwalter Gordon Adjustable Compression Element<br>DXC - Class II (21 CFR 870.4450)            |
| 4. | Common/Trade Name: | Vascular Clamp<br>Bookwalter Gordon Adjustable Compression Element                              |

## Description:

The Bookwalter Gordon Adjustable Compression Element is an adjustable metal compression device that provides an alternative to applying direct hand pressure to obtain hemostasis during vascular catheterization procedures. The device is secured to a surgical table by means of a compatible clamp which supports an articulating arm that allows positioning of the "metal hand" over the vascular access site. The hand has "fingers" that can be adjusted separately to apply compressive pressure to particular locations around the vascular access site. Once in place the Adjustable Compression Element remains securely in position thereby providing ongoing compression without intervention by the surgical staff although it can be adjusted at any time if so desired. The device is reusable and can be cleaned and sterilized in the hospital setting. The Adjustable Compression Element is constructed of high strength stainless steel for rigidity and durability during use. The system also has an available sterilization container to ensure effective reprocessing.

{4}------------------------------------------------

## Intended Use:

The Bookwalter Gordon Adjustable Compression Element is indicated for use to provide hemostasis of the femoral vascular access site, during and following catheterization or cannulation procedures.

## Technological Characteristics:

The Bookwalter Gordon Adjustable Compression Element consists of a Left Finger Subassembly, a Right Finger Subassembly, and a Center Block Subassembly which can be mounted onto an adjustable table mounted arm. The principal material of construction is 17-4 PH SS, which is hardened stainless steel. The adjustable arm and finger paddles can be positioned independently and locked into place to achieve the correct level of compression. Hemostasis is achieved through mechanical compression.

In comparison to the predicate device, the use of mechanical compression for hemostasis is the same. Both devices are mounted on an arm and feature adjustable height and pressure settings. The arm components are constructed of rigid materials and table mounted at the patient's side. The Bookwalter Gordon Adjustable Compression Element has a re-usable finger assembly as compared to a disposable circular disc or finger assembly on the predicate device.

The product has been evaluated to ensure that it meets mechanical requirements, biocompatibility needs, cleaning, sterilization and drying time.

Testing included the following:

Mechanical Performance, including:

- . Maximum compression force
- Expanding & contracting under full load .
- 2X force load ●

Steam Sterilization Efficacy Validation Dry Time Validation Cytotoxicity

The mechanical performance testing for maximum compression force, expanding and contracting under full load, and 2X force load test demonstrate that the subject device is capable of continuous compression without bending, or loosening. This testing demonstrates that the product is capable of performing compression and thereby hemostasis in the same manner as the predicate device.

{5}------------------------------------------------

## Substantial Equivalence:

The design and function of the Bookwalter Gordon Adjustable Compression Element has been determined to be substantially equivalent to the CompressAR SuperComfort Disc and StrongArm SuperComfort System from Advanced Vascular Dynamics (K040615). Both products have the same indications for use and perform in a similar manner.

## Conclusion:

Based on results of mechanical testing, basic design and function, the Bookwalter Gordon Adjustable Compression Element has been determined to be substantially equivalent to the CompressAR SuperComfort Disc and StrongArm SuperComfort System. The device provides sufficient compression to achieve hemostasis and can be mounted, positioned, and adjusted in a manner similar to the predicate. Validation testing for sterilization of all re-usable components has been completed successfully and cleaning rationale provided. Instructions provide user information for use, cleaning, and re-use.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC/K143078](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DXC/K143078)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
