K123551 · Innovative Trauma Care, Inc. · DXC · May 14, 2013 · Cardiovascular
Device Facts
Record ID
K123551
Device Name
ITCLAMP 50
Applicant
Innovative Trauma Care, Inc.
Product Code
DXC · Cardiovascular
Decision Date
May 14, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The ITClamp™ is a trauma clamp device for the temporary control of severe bleeding in the extremities, axilla and inguinal areas.
Device Story
iTClamp is a sterile, single-use, self-locking surgical clamp designed for emergency hemorrhage control. Device features suture needles, plastic shell, and adjustable locking mechanism. Operation: needles penetrate skin to evert edges between pressure bars; locking mechanism secures device, creating fluid-tight seal over wound. Mechanism: seals skin to create contained, pressurized hematoma, facilitating stable clot formation until surgical repair. Used in trauma settings; operated by medical personnel. Benefits: rapid, temporary control of severe bleeding in junctional areas (extremities, axilla, inguinal) where traditional tourniquets may be ineffective.
Clinical Evidence
Bench testing included extreme temperature evaluation, tensile/compression testing, torque, and needle bend/pull tests. Biocompatibility testing per ISO 10993 (cytotoxicity, acute systemic toxicity, muscle implantation, pyrogenicity) confirmed safety. Animal study (porcine lethal junctional bleed model) showed statistically significant improvements in survival, survival time, and blood loss compared to controls and standard gauze packing; no tissue damage observed via histology after 3 hours.
Technological Characteristics
Self-locking surgical clamp; components include suture needles, plastic shell, and locking/release mechanism. Materials biocompatibility verified per ISO 10993. Sterile, single-use. Mechanical operation; no energy source or software.
Indications for Use
Indicated for temporary control of severe bleeding in extremities, axilla, and inguinal areas in patients requiring emergency hemorrhage control.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
K140805 — ITCLAMP 50 · Innovative Trauma Care, Inc. · Oct 9, 2014
K132651 — ITCLAMP 50 · Innovative Trauma Care, Inc. · Oct 28, 2013
K150813 — iTClamp50 · Innovative Trauma Care, Inc. · Apr 30, 2015
K141685 — CROC · Combat Medical Systems, LLC · Jan 26, 2015
K102025 — COMBAT READY CLAMP · Combat Medical Systems, LLC · Aug 11, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
#### 5.0 510(k) Summary
# MAY 1 4 2013
In accordance with: 21 CFR 807.87(h) and (21 CFR 807.92) the:510(k) Summary for the iTClamp™ device is provided below.
| Device Common Name: | Vascular Clamp |
|-------------------------------|---------------------------------------------------------------------------------------|
| Device Proprietary Name: | iTClampTM |
| Submitter: | Innovative Trauma Care, Inc.<br>3463 Magic Dr., Suite 120<br>San Antonio, Texas 78229 |
| Contact: | Richard Waite<br>Sr. Director of Quality Assurance/ Regulatory Affairs, |
| | 3463 Magic Dr., Suite 120 |
| | San Antonio, TX 78229 |
| | Office: 210-582-5850 ext. 1932 |
| | FAX: 210-582-5851 |
| | Cell: 214-662-9277 |
| | rwaite@innovativetraumacare.com |
| Date Prepared: | November 16, 2012 |
| Classification<br>Regulation: | 870.4450 |
| Panel: | Cardiovascular |
| Product Code: | DXC |
| Predicate Device | K102025 - Combat Ready Clamp<br>Combat Medical Systems LLC |
#### Indication for Use
The ITClamp™ is a trauma clamp device for the temporary control of severe bleeding in the extremities, axilla and inguinal areas.
{1}------------------------------------------------
#### Device Description
The ITClamp™ is a clamp device that quickly controls critical bleeding by closing the skin to create a temporary, contained hematoma until surgical repair. The ITClamp is a self-locking surgical clamp with suture needles that penetrate the skin to evert the skin edges between pressure bars of the device and anchor it to the skin to reduce slippage and leakage. Pressure is evenly distributed across the bars, which seal the skin over a wound. An adjustable locking mechanism increases or decreases pressure across the wound to achieve a fluid tight seal.
The ITClamp™ controls bleeding by sealing the skin closed to apply direct pressure and create a temporary pool of blood (hematoma) under pressure. This permits formation of a stable clot until the patient can receive medical care and/or surgical repair.
The device is provided sterile and is for single use.
The ITClamp™ consists of the following components:
- 1) Suture needles
- 2) Plastic Shell
- 3) Locking mechanism
- 4) Lock release mechanism
## Biocompatibility Testing
Biocompatibility testing was performed according to ISO 10993 including, cytotoxicity, acute system toxicity, muscle implantation and pyrogenicity. Results confirmed that the materials used in the ITClamp™ are safe for its intended use.
#### Performance Testing - Bench
Results from the performance testing were provided. The laboratory performance testing was conducted to verify suitability of the design characteristics of the device. The testing consisted of the following: Extreme Temperature Evaluation, Tensile and Compression Test, Torque, and Needle (Bend and Pull). Results of the testing confirmed the ITClamp™ met design requirements.
#### Performance Testing - Animal
Animal testing was performed in a porcine model to assess ITClamp™ performance in a lethal junctional bleed model. The ITClamp™ resulted in statistically significant improvements in survival, survival time, and blood loss when compared with controls or packing with standard gauze. Tissue damage was not observed by histological examination after 3 hours of application.
{2}------------------------------------------------
### Performance Testing- Distribution/Aging/Packaging Integrity
Distribution Challenge Testing and Accelerated Aging Studies were completed to ensure that product integrity is maintained for the intended use. Due to the complexity and non-standard packaging system, microbial aerosol testing was performed to prove the structural integrity of the packaging unit. All testing preformed resulted in no nonconformance(s),
#### Substantial Equivalence
A comparison of the ITClamp to the predicate devices is provided below. Like the predicate devices, the ITClamp is intended to control bleeding through the application of direct pressure. Also like the predicate Combat Ready Clamp, the ITClamp is intended to provide emergency control of severely bleeding vascular wounds until medical and/or surgical repair can be obtained. The ITClamp has different technology than the predicate device. The ITClamp applies pressure to the severely bleeding wound by sealing off the skin surrounding the wound, which leads to blood pooling under pressure and eventual clotting. The predicate device, the Combat Ready Clamp, also applies direct pressure but this is achieved by a applying an external mechanical clamp above the inguinal area to stop blood flow into the area. Although the technology of the proposed device differs from the predicate, it does not raise any new or different types of safety or effectiveness questions.
The available performance data demonstrate that the ITClamp is safe and performs effectively in achieving hemostasis for severely bleeding wounds. The ITClamp, therefore, is substantially equivalent to other devices regulated under 21 CFR 870.4450.
{3}------------------------------------------------
510(k) Premarket Notification
u
iTClamp™
| | New Device | Predicate Device |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K123551 | K102025 |
| Classification / Procode | 870.4450 / DXC | 870.4450 / DXC |
| Device Name | ITClampTM | Combat Ready Clamp |
| Manufacturer | Innovative Trauma Care,<br>Inc. | Combat Medical Systems |
| Indication for Use | The ITClampTM is a<br>trauma clamp device for<br>the temporary control of<br>severe bleeding in the<br>extremities, axilla and<br>inguinal areas. | The Combat Ready Clamp is<br>indicated for use in the<br>battlefield to control difficult<br>bleeds in the inguinal area. |
| Device Design | ITClamp is a self-locking<br>surgical clamp with suture<br>needles that penetrate the<br>skin to evert the skin edges<br>between pressure bars of<br>the device and anchor it to<br>the skin to reduce slippage<br>and leakage. Pressure is<br>evenly distributed across<br>the bars, which seal the<br>skin over a wound. An<br>adjustable locking<br>mechanism increases or<br>decreases pressure across<br>the wound to achieve a<br>fluid tight seal or wound<br>closure as desired. | The CRoC (Combat Ready<br>Clamp) is designed to control<br>difficult bleeds in the inguinal<br>region. The CRoC can be<br>used in a tactical environment<br>where traditional methods of<br>hemorrhage control are not<br>possible and standard<br>tourniquets cannot be applied.<br>It provides compression to<br>large vessels and direct<br>pressure to difficult wounds<br>thus controlling hemorrhage<br>and eliminating the need for<br>manual pressure. It is an<br>expandable aluminum clamp<br>that is durable, collapsible,<br>light weight, and easily<br>removed. |
| Device Operation | Application of pressure by<br>applying clamp to<br>temporarily seal wound<br>site | Application of pressure by<br>applying mechanical clamp |
| Picture | Image: ITClamp | Image: Combat Ready Clamp |
ﺎ ﻳﻌﺘﺒﺮ ﺍﻟ
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all capital letters and is evenly spaced around the circle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 14, 2013
Innovative Trauma Care, Inc. C/O Richard Waite 3463 Magic Dr., Suite 120 San Antonio, Texas 78229
Re: K123551
Trade/Device Name: iTClamp Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: April 5, 2013 Received: April 9, 2013
Dear Mr. Waite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Bram D. Zittokerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
#### Indications for Use Statement 4.0
510(k) Number (if known):
Device Name: ITClamp™
## Indications For Use:
The ITClamp™ is a trauma clamp device for the temporary control of severe bleeding in the extremities, axilla and inguinal areas.
Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bram D. Zuckerman -S
2013.05.14 14:22:39 -04'00'
Page 1 of _1
Page 12
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.