K070423 · Terumo Medical Corp. · DXC · Mar 28, 2007 · Cardiovascular
Device Facts
Record ID
K070423
Device Name
TR BAND
Applicant
Terumo Medical Corp.
Product Code
DXC · Cardiovascular
Decision Date
Mar 28, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The TR Band™ is a compression device to assist hemostasis of the radial artery after a transradial procedure.
Device Story
TR Band is a pneumatic tourniquet for radial artery hemostasis post-transradial procedure. Device consists of a plastic belt with hook-and-loop fastener, two compression balloons, and an injection port. Operated by physician/clinician in a clinical setting. After introducer sheath removal, device is placed around patient's wrist; balloons are inflated via a dedicated 20ml syringe (TR Band Inflator) to apply pressure to the access site. A 5ml syringe (TR Band Air Volume Regulator) allows fine pressure adjustments. Clear plastic construction enables visual monitoring of the access site during hemostasis. Valve on injection port maintains balloon inflation. Single-use device.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 and FDA G95-1 for surface-contacting, breached/compromised skin devices (≤ 24h). Sterilization validated per ANSI/AAMI/ISO 11135-1994.
Technological Characteristics
Pneumatic tourniquet; plastic belt with hook-and-loop fastener; two compression balloons; air injection port with valve. Manual operation via syringe inflator. Biocompatible materials. Sterilized per ANSI/AAMI/ISO 11135-1994 (SAL 10^-6).
Indications for Use
Indicated for patients requiring hemostasis of the radial artery following a transradial procedure.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
Predicate Devices
Radstat Radial Artery Compression System (none)
Related Devices
K152525 — TR BAND Radial Compression Device · Terumo Medical Corporation · Nov 10, 2015
K213531 — TR BAND Radial Compression Device · Terumo Medical Corporation · Dec 20, 2021
K253070 — Namic Radial Arm Band, 23 cm (DYNJRADBAND); Namic Radial Arm Band, 26 cm (DYNJRADBANDL) · Medline Industries, LP · Nov 20, 2025
{0}------------------------------------------------
Terumo Corporation Premarket Notification - TR BandTM Section V. 510(k) Summary
# 0423
## MAR 2 8 2007
# SECTION V. 510(k) SUMMARY
#### Device Name A.
Proprietary Name: Classification Name: Common Name:
TR Band Tourniquet , Pneumatic Tourniquet
#### B. PREDICATE DEVICE
The predicate device is:
Product Name: Radstat Radial Artery Compression System 510k #: none Establishment listing number: 1721504 Manufacaturer: Merit Medical Systems
#### C. INDICATIONS FOR USE
The TR Band™ is a compression device to assist hemostasis of the radial artery after a transradial procedure.
{1}------------------------------------------------
#### D. DESCRIPTION
The TR Band is a tourniquet style device consisting of a plastic belt with hook and loop adjustable fastener on each end, two compression balloons, and an injection port. The device also contains a TR Band Inflator and an Air Volume Regulator.
The TR Band is placed around the patient's wrist and the hook and loop strap is connected. Once the introducer sheath is removed from the patient's wrist, pressure is applied to the patient's access site by inflating the compression balloons of the TR Band. Both compression balloons are filled at the same time while air is being introduced through the air injection port.
The balloons are inflated when air is injected into the air injection port by the TR Band Inflator. The TR Band inflator is a specially designed syringe for use only with the TR Band. A valve on the air injection port assures that the air remains within the compression balloons. The volume of air can be reduced or increased by use of the TR Band Inflator or the TR Band Air Volume Regulator. The TR Band Air Volume Regulator is an accessory device. The TR Band Air Volume Regulator is a smaller syringe (5 ml) compared to the TR Band Inflator (20ml syringe). This allows the physician to make fine adjustments to the pressure in the TR Band.
With the two compression balloons (large and small) the pressure is applied for efficient compression for haemostasis. The belt also has a support plate over the two compression balloons to assure that the balloons and belt conform to the contour of the wrist. The belt and compression balloons are made of clear plastic which allows the physician to view the access site during the haemostasis process.
The TR Band is for single use only.
. "
{2}------------------------------------------------
#### PRINCIPLE OF OPERATION / TECHNOLOGY E.
The TR Band is operated manually or by a manual process.
#### F. DESIGN / MATERIALS
The design of the TR Band is similar to the predicate device. Differences in design between the devices do not raise any new issues of safety and effectiveness.
#### G. SPECIFICATIONS
The TR Band is available in two sizes. The sizes correspond to the length of the band. The larger size is available to accommodate individuals with larger size wrists. The specification for the device can be seen in the Engineering drawing section.
#### H. PERFORMANCE
The performance of the TR Band is substantially equivalent to the performance of the predicate device. The equivalence was shown through bench testing.
#### I. Additional Safety Information
Sterilization conditions have been validated in accordance with ANSI / AAMI / ISO 11135-1994 to provide a Sterility Assurance Level of 10-6.
Blood contacting materials were tested in accordance with the test recommendations in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices - Part I: Evaluation and Testing." The TR Band is classified as Surface Device, Breached or compromised surface, Limited Contact (≤ 24h). The blood contacting materials were found to be biocompatible.
{3}------------------------------------------------
Terumo Corporation Premarket Notification - TR Band™ Section V. 510(k) Summary
#### J. SUBSTANTIAL EQUIVALENCE
The TR Band submitted in this 510(k) is substantially equivalent in intended use, design, principle of operation / technology, and performance to the predicate device. Differences between the devices do not raise any issues of safety or effectiveness.
#### K. SUBMITTER INFORMATION
Name and Address Terumo Medical Corporation 950 Elkton Blvd. Elkton, MD 21921
### Contact Person
Mr. Mark Unterreiner Sr. Regulatory Affairs Specialist Ph: 410-392-7213 Fax: 410-398-6079 Email: mark.unterreiner@terumomedical.com
### Date Prepared
February 9, 2007
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of human figures.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Terumo Medical Corporation % Mr. Mark Unterreiner Senior Regulatory Affairs Specialist West Plant 950 Elkton Boulevard Elkton, Maryland 21921
MAR 2 8 2007
Re: K070423
Trade/Device Name: TR Band™ Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular clamp Regulatory Class: II Product Code: DXC Dated: February 9, 2007 Received: February 13, 2007
Dear Mr. Unterreiner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Mark Unterreiner
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K050423
Device Name:__________________________________________________________________________________________________________________________________________________________________ TR Band ™
Indications For Use:
The TR Band™ is a compression device to assist hemostasis of the radial artery after a transradial procedure.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Polo
(Division Sign-Off) Division of General, Restorative, and Neurological Devi
**510(k) Number** 4070424
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.