VASCULAR SOLUTIONS D-STAT CLAMP ACCESSORY

K050146 · Vascular Solutions, Inc. · DXC · May 27, 2005 · Cardiovascular

Device Facts

Record IDK050146
Device NameVASCULAR SOLUTIONS D-STAT CLAMP ACCESSORY
ApplicantVascular Solutions, Inc.
Product CodeDXC · Cardiovascular
Decision DateMay 27, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Intended Use

The D-Stat Clamp Accessory is indicated for use with the Compressar Universal System (Advanced Vascular Dynamics) and the Femoral Artery Vascular Clamp (Pressure Products) compression devices or as a stand alone device to assist in the control of bleeding following catheterization or cannulation procedures. Following achieving hemostasis the D-Stat Dry Bandage may be detached from the D-Stat Clamp Accessory and left in place for up to 24 hours and is indicated for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes.

Device Story

D-Stat Clamp Accessory consists of lyophilized pad containing thrombin, sodium carboxymethylcellulose, and calcium chloride; pad secured to compressible translucent pad and plastic base. Device functions as topical hemostat; used with femoral access compression devices (Compressar Universal System, Femoral Artery Vascular Clamp) or as standalone. After hemostasis, lyophilized pad detaches from base; covered with adhesive bandage; left in place up to 24 hours. Used in clinical settings by healthcare providers to assist bleeding control post-procedure. Benefits include accelerated hemostasis and protection of vascular access sites.

Clinical Evidence

Bench testing only. Testing included assessment of the ability to separate the lyophilized pad from the Clamp Accessory Base and biocompatibility assessment of new materials.

Technological Characteristics

Lyophilized pad containing thrombin, sodium carboxymethylcellulose, and calcium chloride. Secured to compressible translucent pad and plastic base. Mechanical separation mechanism for pad detachment. No performance standards developed under section 514.

Indications for Use

Indicated for patients undergoing catheterization or cannulation procedures requiring control of bleeding at vascular access sites or around percutaneous catheters/tubes.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 27 2005 Image /page/0/Picture/1 description: The image shows a handwritten alphanumeric string. The string is "KOS0146". The characters are written in black ink on a white background. ## 510(K) SUMMARY | Common/Usual Name: | Topical Hemostat/Vascular Clamp Accessory | |----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Trade Name: | D-Stat Clamp Accessory | | Classification Name: | Vascular Clamp, 21CFR 870.4450, product code DXC | | Manufacturer: | Vascular Solutions, Inc.<br>6464 Sycamore Court<br>Minneapolis, Minnesota 55369 | | Establishment Registration: | 2134812 | | Contact: | Linda Busklein<br>Sr. Regulatory Affairs Associate<br>(763) 656-4217 phone<br>(763) 656-4250 fax | | Performance Standards: | No performance standards have been developed under section<br>514 for this device. | | Device Description: | The D-Stat Clamp Accessory consists of a lyophilized pad<br>containing thrombin, sodium carboxymethylcellulose, and<br>calcium chloride secured to a compressible translucent pad and<br>plastic base. The device is designed for attachment to several<br>commercially available femoral access compression devices or as<br>a standalone device. After hemostasis has been achieved, the<br>lyophilized pad may be removed from the Clamp Accessory<br>base, covered with a provided adhesive bandage, and left in place<br>for up to 24 hours. | | Intended Use: | The D-Stat Clamp Accessory is indicated for use with the<br>Compressar Universal System (Advanced Vascular Dynamics)<br>and the Femoral Artery Vascular Clamp (Pressure Products)<br>compression devices or as a stand alone device to assist in the<br>control of bleeding following catheterization or cannulation<br>procedures. Following achieving hemostasis the D-Stat Dry<br>Bandage may be detached from the D-Stat Clamp Accessory and<br>left in place for up to 24 hours and is indicated for the control of<br>surface bleeding from vascular access sites and percutaneous<br>catheters or tubes. | | Summary of Non-Clinical Testing: | Tests conducted included assessment of the ability to separate<br>the lyophilized pad from the Clamp Accessory Base and a<br>biocompatibility assessment of new materials. | | Predicate Devices: | Vascular Solutions D-Stat Clamp Accessory (K040730)<br>Vascular Solutions D-Stat Dry Hemostatic Bandage (K030836) | | Conclusions: | The D-Stat Clamp Accessory is substantially equivalent to the<br>currently marketed D-Stat Clamp Accessory and the D-Stat Dry<br>Hemostatic Bandage based on a comparison of the indications<br>for use and the technological characteristics of the device. | {1}------------------------------------------------ Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, superimposed over a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 2 7 2005 Vascular Solutions, Inc. c/o Ms. Linda Busklein Sr. Regulatory Affairs Associate 6464 Sycamore Court Minneapolis, MN 55369 Re: K050146 D-Stat Clamp™ Accessory Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II (Two) Product Code: DXC Dated: April 13, 2005 Received: April 14, 2005 Dear Ms. Busklein: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) i You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Ms. Linda Busklein Please be advised that FDA's issuance of a substantial equivalence determination does not muan that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as sct forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.huml. Sincerely yours, Dunker R Kirchner Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Statement | 510(k) Number: | K050146 | |----------------|-------------------------------------------| | Device Name: | Vascular Solutions D-Stat Clamp Accessory | Indications for Use: The D-Stat Clamp Accessory is indicated for use with the Compressar Universal System (Advanced Vascular Dynamics) and the Femoral Artery Vascular Clamp (Pressure Products) compression devices or as a stand alone device to assist in the control of bleeding following catheterization or cannulation procedures. Following achieving hemostasis the D-Stat Dry Bandage may be detached from the D-Stat Clamp Accessory and left in place for up to 24 hours and is indicated for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes." Prescription Use _ x AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Duna R. Vachner (Division Sign-Off) Division of Cardiovascular Devices Page 1 of ____________________________________________________________________________________________________________________________________________________________________ Number k050146
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...