ENCLOSE ANASTOMOSIS ASSIST DEVICE

K023682 · Novare Surgical Systems, Inc. · DXC · Jan 16, 2003 · Cardiovascular

Device Facts

Record IDK023682
Device NameENCLOSE ANASTOMOSIS ASSIST DEVICE
ApplicantNovare Surgical Systems, Inc.
Product CodeDXC · Cardiovascular
Decision DateJan 16, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ENclose™ Anastomosis Assist Device is intended for use by cardiac surgeons in place of partial occluding clamps during coronary artery bypass grafting (CABG) procedures requiring one to three proximal anastomoses in ascending aortas free of atheromatous disease.

Device Story

The ENclose™ Anastomosis Assist Device is a surgical instrument used intraoperatively by cardiac surgeons during coronary artery bypass grafting (CABG). It functions as a vascular clamp to occlude the ascending aorta, replacing traditional partial occluding clamps. The device is designed to facilitate the creation of one to three proximal anastomosis sites. It is provided sterile for single-patient use only. By providing a controlled clamping mechanism, it assists the surgeon in maintaining a clear surgical field for graft attachment, potentially improving procedural efficiency and safety in patients with aortas free of atheromatous disease.

Clinical Evidence

Clinical testing conducted in Canada and Europe compared the performance of the ENclose™ device for one to three proximal anastomoses against the predicate device cleared for single anastomosis. Results demonstrated that the multi-vessel procedure is no more difficult and is as safe as the single anastomosis procedure.

Technological Characteristics

Vascular clamp; materials include medical grade plastic, stainless steel, and silicone. Sterilization via e-beam radiation. Single-patient use. Mechanical design identical to predicate.

Indications for Use

Indicated for cardiac surgeons performing coronary artery bypass grafting (CABG) requiring one to three proximal anastomoses in ascending aortas free of atheromatous disease.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # JAN 1 6 2003 | Summary of Safety Information | NOVARE SURGICAL SYSTEMS, INC. | K023682<br>JANUARY 15, 2003 | |----------------------------------------|-------------------------------------------------|-----------------------------| | Premarket Notification, Section 510(k) | | | | Regulatory Authority: | Safe Medical Devices Act of 1990, 21 CFR 807.92 | | | Device Name: | ENclose™ Anastomosis Assist Device | | | Common Name(s): | Vascular Clamp | | | Classification Name(s): | Vascular Clamp | | | Manufacturer: | | | | Name: | Novare Surgical Systems, Inc. | | | Number: | 2954739 | | | Address: | 10231 Bubb Road<br>Cupertino, CA 95014 | | | Classification(s): | | | Sec. 870.4450 Vascular clamp. (a) Identification. A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily. (b) Classification. Class II (performance standards). | Device Class: | Class II for the requested indications | |-----------------------|----------------------------------------| | Classification Panel: | Cardiovascular Devices Panel | | Product Code(s): | DXC | ### Equivalent Predicate Device: The ENclose™ Anastomosis Assist Device is substantially equivalent to the following legally marketed vascular clamp: #### 1. InCluder™ Vascular Clamp - Novare Surgical Systems, Inc., K010694 Sec. 870.4450 Vascular clamp. (a) Identification. A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily. (b) Classification. Class II (performance standards). | Device Class: | Class II | |------------------|----------| | Product Code(s): | DXC | The new anastomosis device is identical to the previous device. Equivalency can be seen with respect to place of manufacture, physical appearance, functional testing, material certification, surgical technique and indications for use. A feature comparison chart follows this section. | FEATURE | ENclose™ Anastomosis Assist Device | Includer™ Vascular Clamp -K010694 | SE? | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-----| | Indications for Use: | The ENclose™ Anastomosis Assist Device is intended<br>for use by cardiac surgeons in place of partial occluding<br>clamps during coronary artery bypass grafting (CABG)<br>procedures requiring one to three proximal anastomoses<br>in ascending aortas free of atheromatous disease. | Identical, except single anastomosis | YES | | Materials: | Medical grade plastic, stainless steel and silicone | Identical | YES | | Design: | Clamp | Identical | YES | | Sterilization: | Radiation (e-beam) | Identical | YES | | Manufacturer: | Novare Surgical Systems, Inc. | Identical | YES | | Product Code: | DXC | Identical | YES | | K - Number: | K023682 | K010694 | YES | {2}------------------------------------------------ #### Device Description: The change is limited to a request for multiple use. The modified ENclose™ Anastomosis Assist Device is a intraoperative device used to assist in the creation of one or more proximal anastomosis sites in the execution of coronary artery bypass grafting (CABG). Clinical Testing Summary. Clinical testing conducted in Canada and Europe demonstrate the functional equivalence of the added ENclose™ Anastomosis Assist Device. The one to three vessel use ENCLOSE™ Anastomosis Assist Devices' level of performance is substantially equivalent to the existing legally marketed ENCLOSE™ Anastomosis Assist Device (InCluder™, K010694) cleared for single vessel use. The collected clinical data demonstrate that one to three proximal anastomoses are no more difficult and are just as safe as the existing single anastomosis procedure performed using the previously cleared ENCLOSE™ Anastomosis Assist Device. Please refer to the product instructions for use for a more complete listing of indications, warnings, cautions and precautions. ### Company Contact: Mr. Kerry Pope Novare Surgical Systems, Inc. 10231 Bubb Road Cupertino, CA 95014 408.873.3161 ### Submission Correspondent: Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax #### Snecial Controls: Special controls were published by FDA. The following special controls are believed to apply to the marketing of class II devices: - (i) Compliance with material standards, - (ii) Compliance with biocompatability standard, and Compliance with specified labeling requirements. Novare Surgical Systems, Inc. also meets appropriate general controls authorized under Sections 501, 502, 510, 516, 518, 519, and 520 of the Food, Drug, and Cosmetic Act. #### Sterility Processing: The ENclose™ Anastomosis Assist Device is provided sterile and is designed strictly for single patient use. The device may not be cleaned, reprocessed or reused under any circumstances. The product must be handled, stored and placed into use in such a way that it is protected from inadvertent damage or contamination. When used, the product must be placed into use following accepted surgical sterile technique. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, with flowing lines representing movement or connection. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 6 2003 Novare Surgical Systems, Inc. c/o Mr. David Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 Re: K023682 Enclose™ Anastomosis Assist Device Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II (two) Product Code: DXC Dated: October 31, 2002 Received: November 1, 2002 Dear Mr. Schlerf: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. David Schlerf Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Nalgtleh Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 1 - of of of l K023682 510(k) Number: ENclose™ Anastomosis Assist Device Device Name(s): Indication(s) for Use: The ENclose™ Anastomosis Assist Device is intended for use by cardiac surgeons in place of partial occluding clamps during coronary artery bypass grafting (CABG) procedures requiring one to three anastomoses in ascending aortas free of atheromatous disease. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE) Nolen Tull (Division Sign-Off) Division of Cardiovascular Devices **510(k) Number** K023682 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...