TRIVEX SYSTEM

K032387 · Smith & Nephew, Inc. · DWQ · Oct 29, 2003 · Cardiovascular

Device Facts

Record IDK032387
Device NameTRIVEX SYSTEM
ApplicantSmith & Nephew, Inc.
Product CodeDWQ · Cardiovascular
Decision DateOct 29, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4885
Device ClassClass 2
AttributesTherapeutic

Intended Use

The proposed Smith & Nephew TriVex™ System is indicated for use in ambulatory phlebectomy procedures for the resection and ablation of varicose veins.

Device Story

The Smith & Nephew TriVex™ System is a surgical device used in ambulatory phlebectomy procedures. It consists of a control unit on a pedestal stand, dual peristaltic pumps, an illuminator, a resector handpiece, and footswitches. The system is operated by a physician to assist in the resection and ablation of varicose veins. By providing illumination and mechanical resection capabilities, the device facilitates the removal of diseased veins, potentially benefiting patients by improving clinical outcomes in phlebectomy procedures.

Clinical Evidence

Bench testing only. The device meets biocompatibility requirements per ISO 10993-1 and electrical safety standards per UL 60601-1 and IEC 60601-1.

Technological Characteristics

System includes control unit, dual peristaltic pumps, illuminator, resector handpiece, and footswitches. Complies with ISO 10993-1 for biocompatibility and UL/IEC 60601-1 for electrical safety (BF type equipment).

Indications for Use

Indicated for use in ambulatory phlebectomy procedures for resection and ablation of varicose veins.

Regulatory Classification

Identification

An external vein stripper is an extravascular device used to remove a section of a vein.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Endoscopy Smith & Nephew, Inc 150 Minuteman Road Andover, MA 01810 978 749 1000 978 749 1599 Fax www.smith-nephew.com OCT 2 9 2003 K032387(pg 1 of 2) # SECTION IV # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based. ### Date Prepared: July 31, 2003 ## A. Submitter's Name: Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road Andover, MA 01810 ## B. Company Contact: Karen Provencher Regulatory Affairs Specialist Phone: 978-749-1365 Fax: 978-749-1443 #### C. Device Name | Trade Name: | Smith & Nephew TriVex <sup>TM</sup> System | |----------------------|--------------------------------------------| | Common Name: | Varicose Vein Ablation System | | Classification Name: | Stripper, Vein, External | #### D. Predicate Devices - The Smith & Nephew TriVex™ System is substantially equivalent in Intended Use and Fundamental Scientific Technology to the following legally marketed device in commercial distribution: The Dyonics® Varicose Vein Ablation System - K990723, Subcutaneous Illuminator - K991323, Smith & Nephew Xenon Light Source -K994084, and H.K. Surgical Klein Pump - K012044. Smith & Nephew TriVex™ System Page 29 of 308 {1}------------------------------------------------ #### E. Description of Device The Smith & Nephew TriVex™ System consists of a control unit mounted on a pedestal stand, dual peristaltic pumps, an illuminator, resector handpiece and footswitches. #### F. Intended Use The proposed Smith & Nephew TriVex™ System is indicated for use in ambulatory phlebectomy procedures for the resection and ablation of varicose veins. # G. Comparison of Technological Characteristics The Smith & Nephew TriVex™ System is substantially equivalent in design. materials, function and intended use to the following devices cleared for commercial distribution: - The Dyonics® Varicose Vein Ablation System - K990723 - Subcutaneous Illuminator - K991323 - Smith & Nephew Xenon Light Source K994084 ● - H.K. Surgical Klein Pump - K012044 # G. Summary Performance Data The Smith & Nephew TriVex™ System meets the biocompatibilty requirements of ISO 10993-1 and electrical safety standards for BF type equipment as stated in UL 60601-1 and IEC 60601-1. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. The text is written in a sans-serif font. Public Health Service OCT 2 9 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Karen Provencher Regulatory Affairs Specialist Smith & Nephew , Inc. Endoscopy Division 150 Minuteman Road Andover, Massachusetts 01810 Re: K032387 Trade/Device Name: Smith & Nephew TriVex™ System Regulation Number: 21 CFR 870.4885 Regulation Name: External vein stripper Regulatory Class: II Product Code: DWQ Dated: July 31, 2003 Received: August 4, 2003 Dear Ms. Provencher: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Karen Provencher This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost to Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use 510(k) Number (if known): KC32387 Smith & Nephew TriVex™ System_ Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications For Use: Indicated for use in ambulatory phlebectomy procedures for resection and ablation of varicose veins. Muriam C. Provost (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number *K632387* (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ Smith & Nephew TriVex™ System Page 14 of 308
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...