CARDIOPLEGIA ADAPTERS, VESSEL CANULAE, AND AORTIC ROOT CANNULAE
K972503 · Sorin Group Italia S.R.L. · DWF · Aug 26, 1997 · Cardiovascular
Device Facts
Record ID
K972503
Device Name
CARDIOPLEGIA ADAPTERS, VESSEL CANULAE, AND AORTIC ROOT CANNULAE
Applicant
Sorin Group Italia S.R.L.
Product Code
DWF · Cardiovascular
Decision Date
Aug 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4210
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The vessel cannula will be supplied with a wing or without a wing, with or without a check valve to deliver cardioplegia solution to the heart during cardiopulmonary bypass surgery and to perfuse a vein graft or access the patency of the vessel to be used as a graft during cardiopulmonary bypass surgery. The aortic root cannula is designed to deliver cardioplegia solution during cardiopulmonary bypass surgery and aspirate air from the aorta at the end of the cardiopulmonary bypass surgery. The aortic root cannula with vent line is designed to deliver cardioplegia solution and vent the heart during cardiopulmonary bypass surgery. The cardioplegia "Y" perfusion adapter is designed for use with the aortic root cannula and vessel cannula to deliver cardioplegia solution during cardiopulmonary bypass surgery. The cardioplegia "Y" recirculation adapter is designed for use for recirculating of cardioplegia solution during cardiopulmonary bypass surgery. The cardioplegia straight adapter is designed for use during cardiopulmonary bypass surgery for connecting a cannula or catheter to an administration or vent line. The cardioplegia "Y" venting adapter is designed for use during cardiopulmonary bypass surgery to deliver cardioplegia solution and vent the left heart.
Device Story
Device suite includes vessel cannulae, aortic root cannulae, and various cardioplegia adapters (straight, Y-perfusion, Y-recirculation, Y-venting). Used during cardiopulmonary bypass surgery by surgeons/clinicians. Vessel cannula delivers cardioplegia or assesses vein graft patency; aortic root cannula delivers cardioplegia and vents air/heart; adapters facilitate fluid line connections, recirculation, or venting. Flexible PVC tubing construction; luer-lock connections; suture rings/anchor hubs for secure placement. Devices enable controlled delivery of cardioplegia solution to the heart and management of air/fluid during bypass, supporting myocardial protection and surgical access.
Clinical Evidence
Bench testing only.
Technological Characteristics
Flexible polyvinyl chloride (PVC) tubing; luer-lock connectors; suture rings; anchor hubs; check valves (optional); wings (optional). Mechanical fluid delivery and venting system. No electronic components or software.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery requiring cardioplegia solution delivery, venting of the heart/aorta, or perfusion/patency assessment of vein grafts.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
Predicate Devices
Research Medical, Inc. Vasoplegia Cannula (K896778)
{0}------------------------------------------------
K072503
### Summary of Safety and Effectiveness
AUG 26 1997
## DETERMINATION OF SUBSTANTIAL EQUIVALENCE
The California Medical Laboratories, Inc. devices are substantially equivalent to the DLP and Research Medical predicate device. The California Medical Laboratories, Inc. devices have substantially equivalent intended uses as the predicate device. The California Medical Laboratories, Inc. devices have technologic characteristics which are substantially equivalent to the DLP and Research Medical predicate device.
#### COMPANY AND CONTACT PERSON
California Medical Laboratories Inc. 2681 Kelvin Avenue Irvine, California 92714
Mehmet Bicakci President
## DEVICE NAME
California Medical Laboratories Inc. Vessel Cannula with and without wing and with and without check valve, Aortic Root Cannula with and without Vent Line, Cardioplegia "Y" Perfusion Adapter, Cardioplegia "Y" Recirculation Adapter, Cardioplegia Straight Adapter, Cardioplegia "Y" Venting Adapter.
# NAME OF PREDICATE OR LEGALLY MARKETED DEVICE
The claim of substantial equivalence is based upon the following devices:
- Research Medical, Inc. Vasoplegia Cannula (K 896778) designed for the . delivery of cardioplegia solution to the heart during cardibulmonary bypass surgerv.
- DLP, Inc. Aortic Root Cannula (K790565) with and without a vent line . designed for the delivery of cardioplegia solution and venting the left heart during cardiopulmonary bypass surgery,
- DLP, Inc. "Y" Adapter-Coronary Perfusion (K791498) is designed for delivery . of cardioplegia solution during cardiopulmonary bypass surgery.
- DLP, Inc. Straight Adapter (K790564) is designed for use during . cardiopulmonary bypass surgery for connecting a cannula or catheter to an administration or vent line.
- DLP, Inc. "Y" Adapters (K790563) is designed for use during . cardiopulmonary bypass surgery to deliver cardioplegia solution and vent the heart.
## DESCRIPTION OF DEVICE
The Vessel Cannula is indicated for use in the delivery of cardiopleqia solution during cardiopulmonary bypass surgery or to perfuse a vein graft or access the patency in a harvested vein which will be used for a graft. The Vessel Cannula is a flexible, tapered, polyvinyl chloride tube with a soft distal tip. The distal tip is composed of two sizes of anchor hubs (3mm distal-3.6mm proximal) designed to accommodate vessels of various diameters. The proximal female luer-lock is designed to securely connect with a standard cardioplegia delivery line. The cannula is available with or without wings and with or without a check valve.
{1}------------------------------------------------
The Aortic Root Cannula is designed to deliver cardioplegia solution during cardiopulmonary bypass surgery. The cannula consists of flexible polyvinyl chloride tubing with a movable sew ring attached to a soft distal tip. The distal tip has a flange with four suture holes. An introducer needle assembly is provided to facilitate entry into the aorta.
The Cardioplegia Straight Adapter is indicated during cardiopulmonary bypass surgery for connecting a cannula or catheter to an administration and or a vent line. The straight adapter consists of flexible polyvinyl tubing, and connecting adapters in a straight configuration.
The Cardioplegia "Y" Venting Adapter is designed to be used in conjunction with cardiopulmonary bypass surgery to deliver cardioplegia solution, and venting the left heart. It may also be used as an extension line by itself. The adapter consists of flexible polyvinyl chloride tubing, "Y" connector, tubing clamps, and connecting adapters in a "Y" configuration.
The Cardioplegia "Y" Recirculation Adapter is indicated for recirculation of cardioplegia solution during cardiopulmonary bypass surgery, to keep the cardioplegia solution at constant temperature. The "Y" recirculation adapter consists of a flexible polyvinyl chloride tubing, "Y" connector, tubing clamps and connecting adapters in a "Y" configuration.
The Cardioplegia "Y" Perfusion Adapter is indicated for use in delivery of cardioplegia solution during cardiopulmonary bypass surgery. "The perfusion
adapter consists of polyvinyl chloride tubing, tubing clamps, "Y" connector, and connecting adapters in a "Y" configuration.
{2}------------------------------------------------
# STATEMENT OF INTENDED USE
The vessel cannula will be supplied with a wing or without a wing, with or without a check valve to deliver cardioplegia solution to the heart during cardiopulmonary bypass surgery and to perfuse a vein graft or access the patency of the vessel to be used as a graft during cardiopulmonary bypass surgery.
The aortic root cannula is designed to deliver cardioplegia solution during cardiopulmonary bypass surgery and aspirate air from the aorta at the end of the cardiopulmonary bypass surgery. The aortic root cannula with vent line is designed to deliver cardioplegia solution and vent the heart during cardiopulmonary bypass surgery.
The cardioplegia "Y" perfusion adapter is designed for use with the aortic root cannula and vessel cannula to deliver cardioplegia solution during cardiopulmonary bypass surgery. The cardioplegia "Y" recirculation adapter is designed for use for recirculating of cardioplegia solution during cardiopulmonary bypass surgery. TThe cardioplegia straight adapter is designed for use during cardiopulmonary bypass surgery for connecting a cannula or catheter to an administration or vent line. The cardioplegia "Y" venting adapter is designed for use during cardiopulmonary bypass surgery to deliver cardioplegia solution and vent the left heart.
# STATEMENT OF INTENDED USE OF PREDICATE/MARKETED DEVICES
The predicate devices have the same intended use as stated above.
# STATEMENT OF COMPARISON OF TECHNOLOGIC CHARACTERISTICS BETWEEN DEVICE AND PREDICATE DEVICE
The California Medical Laboratories. Inc devices have technologic characteristics which are substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, with its head and wings stylized in a flowing, curved design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
Mr. Mehmet Bicakci President California Medical Laboratories, Inc. AUG 26 1997 2681 Kelvin Avenue Irvine, California 92714 -
Re: K972503 Cardioplegia Adapters, Vessel Cannulae, and Aortic Root Cannulae Regulatory Class: II (two) Product Code: 74 DWF Dated: July 2, 1997 Received: July 3, 1997
Dear Mr. Bicakci:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Mehmet Bicakci
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) Additionally, for questions on the promotion and 594-4648. advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K 9778503
Device Name: California Medical Laboratories Inc. Vessel Cannula with and without wing and with and without check valve, Aortic Root Cannula with and without Vent Line, Cardioplegia "Y" Perfusion Adapter, Cardioplegia "Y" Recirculation Adapter, Cardioplegia Straight Adapter, Cardioplegia "Y" Venting Adapter.
Indications For Use: The vessel cannula will be supplied with a wing or without a wing, with or without a check valve to deliver cardioplegia solution to the heart during ......... cardiopulmonary bypass surgery and to perfuse a vein graft or access the patency of the vessel to be used as a graft during cardiopulmonary bypass surgery.
The aortic root cannula is designed to deliver cardioplegia solution during cardiopulmonary bypass surgery and aspirate air from the aorta at the end of the cardiopulmonary bypass surgery. The aortic root cannula with vent line is designed to deliver cardioplegia solution and vent the heart during cardiopulmonary bypass surgery.
The cardioplegia "Y" perfusion adapter is designed for use with the aortic root cannula and vessel cannula to deliver cardioplegia solution during cardiopulmonary bypass surgery. The cardioplegia "Y" recirculation adapter is designed for recirculating of cardioplegia solution during cardiopulmonary bypass surgery. The cardioplegia straight adapter is designed for use during cardiopulmonary bypass surgery for connecting a cannula or catheter to an administration or vent line. The cardioplegia "Y" venting adapter is designed for use during cardiopulmonary bypass surgery to deliver cardioplegia solution and vent the left heart.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bert R. Kimmerle
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
5103(k) Number <972502<
Prescription Use
Per 21 CFR 801.109
ાર Over-The-Counter ilse _
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.