Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing Set (SR180C)
Applicant
Smart Reactors
Product Code
DWF · Cardiovascular
Decision Date
Feb 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4210
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Circuit Tubing Set is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures for ≤ 6 hours.
Device Story
Extracorporeal cardiopulmonary bypass tubing set; used in CPB surgical procedures; operated by perfusionists. Device consists of 180cm DOP-free PVC tubing with proprietary 'Camouflage' coating. Tubing facilitates blood flow within extracorporeal circuit. Single-use; sterile; EtO sterilized. Provides biocompatible blood-contacting surface for duration of bypass procedure (≤ 6 hours).
Clinical Evidence
Bench testing only. Testing performed per ISO 15676:2016, ISO 18193:2021, EN 13868:2002, and ISO 11658:2012. Biocompatibility testing conducted per ISO 10993 series. Device met all internal and international performance requirements.
Technological Characteristics
DOP-free PVC tubing; 9.5mm inner diameter; 180cm length. Proprietary 'Camouflage' surface coating. EtO sterilization. Compliant with ISO 10993 (biocompatibility), ISO 15676:2016, ISO 18193:2021, EN 13868:2002, and ISO 11658:2012.
Indications for Use
Indicated for adult patients requiring cardiopulmonary bypass (CPB) surgical procedures for durations up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
K030077 — GISH TUBING AND CONNECTORS WITH GBS COATING · Gish Biomedical, Inc. · Mar 14, 2003
K993189 — CARDIOPULMONARY BYPASS PUMP TUBING - 1/4 INCH TUBING, CARDIOPULMONARY BYPASS PUMP TUBING - 3/8 INCH TUBING · Terumo Corp. · Mar 1, 2000
K213540 — Quantum SuperPAC Tubing Set, Quantum SuperPAC Cardioplegia Set · Qura S.R.L · May 20, 2022
K022857 — PUMP TUBING WITH X-COATING · Terumo Cardiovascular Systems Corporation · Sep 19, 2002
K080592 — HLM TUBING SET WITH BIOLINE COATING · Maquet Cardiopulmonary, AG · Sep 11, 2008
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 23, 2026
Smart Reactors
% Robin Stephens
CEO
Psephos Biomedica
Sussex Innovation Centre
Science Park Square
Falmer, BN1 9SB
United Kingdom
Re: K260043
Trade/Device Name: Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing Set (SR180C)
Regulation Number: 21 CFR 870.4210
Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, Or Tubing
Regulatory Class: Class II
Product Code: DWF
Dated: January 6, 2026
Received: January 7, 2026
Dear Robin Stephens:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
2K260043 - Robin Stephens
Page
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
3K260043 - Robin Stephens
Page
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kathleen M. Grunder -S
for Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260043 | ? |
| Please provide the device trade name(s). | | ? |
| Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing Set (SR180C) | | |
| Please provide your Indications for Use below. | | ? |
| The Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Circuit Tubing Set is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures for ≤ 6 hours. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary: K260043
Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing
Submission Sponsor: Smart Reactors
1st Floor, Block 15, Galway Technology Park
Parkmore Road, Galway, H91 AY0Y, Ireland
Company Phone No.: +353-872-037-556
Applicant Contact: Mark Brassil
Email: mark@smartreactors.com
Correspondent Contact: Mr Robin Stephens
Email: rstephens@psephos.com
Date Prepared: February 23, 2026
Device Identification
Trade/Proprietary Name: Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing SR180C
Common/Usual Name: Tubing Sets
Classification Name: Cardiopulmonary bypass vascular catheter, cannula, or tubing
Regulation Number: 21 CFR 870.4210
Product Code: DWF
Device Class: II
Predicate Device(s): Tubing Pack (K171979)
## Indication for Use Statement
The Indications for Use are the same as the predicate and are as follows:
The Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Circuit Tubing is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures for ≤ 6 hours.
This device is intended for use in the adult patient population.
## Device Description
The Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing is a tubing pack intended for use during cardiopulmonary bypass. The tubing is DOP free polyvinylchloride (PVC) with an inner diameter of 9.5mm coated with Smart Reactors' proprietary Camouflage coating provided in a length of 180cm.
Page 1 of 3
{5}
Page 2 of 3
# Technological Characteristics and Product Feature Comparison
The subject device, the Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing, is substantially equivalent to the primary predicate device in terms of:
- indications for use
- materials
- technological characteristics
- performance criteria
- packaging and sterilization of devices
It does not raise any new questions compared to the predicate devices as outlined in the comparison table below.
| Device | Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing | Medtronic, Inc. Tubing Pack K171979 |
| --- | --- | --- |
| Product Name | Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing | Medtronic, Inc. Tubing Pack |
| Regulation No. | 870.4210 | 870.4210 |
| Regulation Name | Cardiopulmonary bypass vascular catheter, cannula, or tubing | Cardiopulmonary bypass vascular catheter, cannula, or tubing |
| Product Code | DWF | DWF |
| Classification | Class II | Class II |
| Indications for Use | The Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Circuit Tubing is a tubing pack indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures. | The Medtronic, Inc. Tubing Pack is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures. |
| Target Population | Adult patients requiring cardiopulmonary bypass procedures for periods up to 6 hours | Patients requiring cardiopulmonary bypass procedures for periods up to 6 hours |
| Target User | Perfusionist | Perfusionist |
| Biocompatibility Requirements | In compliance with ISO 10993 series | In compliance with ISO 10993 series |
| Main Contacting Materials | DOP-free Polyvinyl Chloride (PVC) coated with Camouflage coating. | DOP-free polyvinyl chloride (PVC) Polycarbonate (PC) Cortiva, Trillium and Balance coating |
| Single Use | Yes | Yes |
| Sterile Condition | Sterile | Sterile |
| Description of Sterilization method | EtO sterilization process | EtO sterilization process |
| Shelf Life | 2 years | 2 Years |
{6}
Page 3 of 3
# Testing Summary
The device passed all performance bench testing in accordance with internal requirements, national standards and international standards. Specific standards utilized and tests performed included:
- ISO 15676:2016 "Cardiovascular implants and artificial organs – Requirements for single use tubing packs for cardiopulmonary bypass and extracorporeal membrane oxygenation (ECMO)".
- ISO 18193:2021 "Cardiovascular implants and artificial organs - Cannulae for extracorporeal circulation"
- EN 13868:2002 "Catheters - Test methods for kinking of single lumen catheters and medical tubing"
- ISO 11658:2012 "Cardiovascular implants and extracorporeal systems – Blood/tissue contact surface modifications for extracorporeal perfusion systems"
# Biocompatibility
A biological risk assessment has been completed for the subject device. To support this, biocompatibility testing was successfully completed in accordance with the ISO 10993 series of standards. The device was found to be biocompatible.
# Statement of Substantial Equivalence
Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing has the same intended use and indications for use, and equivalent technological characteristics compared to the predicate device. Therefore, the device is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.