DLP™ Silicone Coronary Artery Ostial Cannulae

K230960 · Medtronic, Inc. · DWF · May 4, 2023 · Cardiovascular

Device Facts

Record IDK230960
Device NameDLP™ Silicone Coronary Artery Ostial Cannulae
ApplicantMedtronic, Inc.
Product CodeDWF · Cardiovascular
Decision DateMay 4, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4210
Device ClassClass 2
AttributesTherapeutic

Intended Use

These Cannulae are intended for use in conjunction with cardiopulmonary bypass surgery up to six hours or less for delivery of cardioplegia solutions directly to the coronary arteries.

Device Story

DLP Silicone Coronary Artery Ostial Cannulae are sterile, single-use devices used during cardiopulmonary bypass surgery. The device consists of a silicone body with a soft bulb beveled tip and a locking female luer fitting. It is operated by a surgeon or perfusionist to deliver cardioplegia solutions directly into the coronary arteries. The device functions as a conduit for fluid delivery to the heart during bypass procedures. It is intended for short-term use (up to six hours).

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including product functional testing, biocompatibility assessment, and shelf-life studies to validate the change in packaging material.

Technological Characteristics

Silicone body with soft bulb beveled tip; locking female luer fitting. Single-use, sterile, nonpyrogenic. Packaging: Nylon film FFS pouch. Shelf life: 3 years.

Indications for Use

Indicated for patients undergoing cardiopulmonary bypass surgery requiring delivery of cardioplegia solutions directly to the coronary arteries for a duration of six hours or less.

Regulatory Classification

Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. May 4, 2023 Medtronic, Inc. Stephen Beier Principal Regulatory Affairs Specialist 7611 Northland Drive Minneapolis, Minnesota 55428 #### Re: K230960 Trade/Device Name: DLP™ Silicone Coronary Artery Ostial Cannulae Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, Or Tubing Regulatory Class: Class II Product Code: DWF Dated: April 4, 2023 Received: April 5, 2023 #### Dear Stephen Beier: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerelv. Kathleen M. Grunder -S Kathleen Grunder Acting Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K230960 Device Name DLP Silicone Coronary Artery Ostial Cannulae Indications for Use (Describe) These Cannulae are intended for use in conjunction with cardiopulmonary bypass surgery up to six hours or less for delivery of cardioplegia solutions directly to the coronary arteries. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | FORM FDA 3881 (6/20) | | | |----------------------|--|--| | | | | Page of Image /page/2/Picture/21 description: The image contains the text 'KXXXXXXX' on the top line and '1-25' on the second line. The text is in a simple, sans-serif font. The arrangement of the text suggests a label or identifier, possibly indicating a range or category. {3}------------------------------------------------ # 510(k) Summary | Date Prepared: | April 04, 2023 | |--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant: | Medtronic, Inc<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration No. 2184009 | | Contact Person: | Stephen Beier<br>Principal Regulatory Affairs Specialist<br>Phone: (763) 526-6751<br>Fax: (763) 367-8361<br>Email: stephen.beier2@medtronic.com | | Alternate Contact: | Diane Howell<br>Regulatory Affairs Manager<br>Phone: (651) 214-2693<br>Email: diane.d.howell@medtronic.com | ## Device Name and Classification | Trade Names: | DLPTM Silicone Coronary Artery Ostial Cannulae<br>(Models: 30315, 30317, and 30320) | |--------------------|-------------------------------------------------------------------------------------| | Common Name: | Cardiopulmonary bypass vascular catheter, cannula or tubing | | Regulation Number: | 21 CFR 870.4210 | | Product Code: | DWF | | Classification: | Class II | | Predicate Device | | | K141951 | DLP Silicone Coronary Artery Ostial Cannulae | {4}------------------------------------------------ ## Device Description The DLP™ Silicone Coronary Artery Ostial Cannulae feature a soft bulb beveled tip with a silicone body. The Cannulae terminate with a locking female luer fitting. The Cannulae are nonpyrogenic, single use, and sterile. ## Indications for Use These Cannulae are intended for use in conjunction with cardiopulmonary bypass surgery up to six hours or less for delivery of cardioplegia solutions directly to the coronary arteries. #### Comparison to Predicate Devices A comparison of the of the modified devices to the currently marketed predicate device (K141951) indicates the subject devices are substantially equivalent11 with the following similarities: - Same intended use and labeling ● - Same technological characteristics - Same operating principle ● - Same design features and device materials - Same sterilization requirements, methods, and parameters ● - Same 3-year shelf life - Same packaging configuration o The following device modifications were made to the predicate device: - Packaging material change ● <sup>1</sup> The term 'substantially equivalent' as used herein is intended to be a determination of substantial equivalency under the Federal Food, Drug, and Cosmetic Act and relates to the fact that the product can be marketed without premarket approval or reclassification. Such a determination is not intended to have any bearing whatsoever on the resolution of patent infringement suits or other patent matters." (Federal Register, Vol. 42, No. 163, Aug. 23, 1977, page 42525 and 42529). {5}------------------------------------------------ | Component | Material Changes | Design Verification and<br>Validation | Results | |----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------| | Form-fill-seal (FFS)<br>pouch material | The pouch material used in<br>the FFS packaging process<br>has undergone a change<br>from a Surlyn film pouch to<br>a Nylon film pouch. | Risk-based testing and evaluations to<br>qualify this change included product<br>functional testing, a biocompatibility<br>assessment and completion of a product<br>shelf life study. | All results<br>pass | Table 5-1: Summary of Changes from Predicate Device # Summary of Performance Data Testing has demonstrated that the DLP™ Silicone Coronary Artery Ostial Cannulae are substantially equivalent to their predicate devices. Table 5-1 describes the change that has been made to the predicate devices' packaging and the design verification and validation activities that were completed to evaluate the new packaging material. # Conclusion Medtronic has demonstrated that the modified DLP™ Silicone Coronary Artery Ostial Cannulae are substantially equivalent to the predicate devices based on the fundamental scientific principles, operating principles, design features and intended use being unchanged from the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...