← Product Code [DWF](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF) · K171308

# Tubing Pack (K171308)

_Medtronic, Inc. · DWF · Jun 2, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K171308

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [DWF](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF.md)
- **Decision Date:** Jun 2, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.

## Device Story

Medtronic Tubing Packs consist of pre-assembled coated and uncoated tubing, connectors, and accessories used to form extracorporeal cardiopulmonary bypass (CPB) circuits. These packs facilitate the connection of primary perfusion devices within the bypass circuit. The device is used in clinical settings during CPB surgical procedures, operated by perfusionists or surgical staff. The tubing and components provide the necessary pathway for blood circulation outside the body during surgery. The current modification involves a change in the packaging tray material, supplier, and manufacturing process (thermoformed to injection molded). The device remains sterile and biocompatible, maintaining the same fundamental design, operating principle, and intended use as previously cleared Medtronic tubing packs. The output is a sterile, ready-to-use circuit component that supports patient life-support during cardiac surgery.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including process characterization, shelf-life validation, packaging performance testing, sterilization validation, and biocompatibility assessment of the new packaging tray materials.

## Technological Characteristics

Tubing pack components include coated/uncoated tubing and connectors. Packaging tray material changed from PETG copolyester (Eastman Spectar 14471) to Styrene Methyl Methacrylate Acrylic Copolymer (Resirene CET 265). Sterilization via ethylene oxide. The device is a passive fluid pathway component; no software or energy source is integral to the tubing pack function.

## Regulatory Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

## Predicate Devices

- Uncoated Tubing and Connector Components ([K151523](/device/K151523.md))

## Reference Devices

- Cortiva - Coated Tubing Packs ([K891687](/device/K891687.md))
- Trillium BioPassive Surface Coated Tubing Packs ([K012538](/device/K012538.md))
- Balance BioSurface Coated Tubing Packs ([K122811](/device/K122811.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other.

June 2, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Medtronic, Inc. Renee Cveykus Principal Regulatory Affairs Specialist 8200 Coral Sea Street NE Mounds View, Minnesota 55112

Re: K171308

Trade/Device Name: Tubing Pack Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, Or Tubing Regulatory Class: Class II Product Code: DWF Dated: May 2, 2017 Received: May 3, 2017

Dear Renee Cveykus:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Fernando
Aguel -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K171308

Device Name Tubing Pack

Indications for Use (Describe)

This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness

| Date Prepared:                | May 2, 2017                                                                                                                                                                        |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name and Address: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009                                      |
| Contact Person:               | Renee L. Cveykus<br>Principal Regulatory Affairs Specialist<br>Medtronic Perfusion Systems<br>Phone: (763) 505-3059<br>Fax: (763) 367-0401<br>Email: renee.l.cveykus@medtronic.com |
| Alternate Contact:            | Jake Roeller<br>Sr. Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: (763) 526-0404<br>Fax: (763) 514-9521<br>Email: jake.w.roeller@medtronic.com               |

#### Proprietary Name:

| Models         | Description |
|----------------|-------------|
| See Appendix 2 | Tubing Pack |

## Device Name and Classification:

| Trade Name:           | Tubing Pack                                                    |
|-----------------------|----------------------------------------------------------------|
| Common Name:          | Catheter, cannula and tubing, vascular, cardiopulmonary bypass |
| Classification Name:  | Cardiopulmonary bypass vascular catheter, cannula, or tubing.  |
| Classification Panel: | Cardiovascular                                                 |
| Regulation Number:    | 21 CFR 870.4210                                                |
| Product Code:         | DWF                                                            |
| Classification:       | Class II                                                       |

{4}------------------------------------------------

## Predicate Device:

K151523 Uncoated Tubing and Connector Components

# Device Description

Medtronic Tubing Packs are used in cardiopulmonary bypass for connecting the primary devices of the bypass circuit. The functionality and intended use of these devices are the same as those for the coated (Cortiva - Coated Tubing Packs(K891687), Trillium BioPassive Surface Coated Tubing Packs (K012538), and Balance BioSurface Coated Tubing Packs(K122811) and uncoated tubing, connectors, and accessories that are FDA cleared.

### Indications for Use

There is no change to the intended use of the devices/components. The current Indications for Use statement for these devices is listed below:

The Medtronic Tubing Pack is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.

●

# Comparison to Predicate Devices

When compared to the predicate devices, the Tubing Packs have the same:

- o Intended Use
- Performance ●
- Technological characteristics ●
- Base materials

Method of sterilization and sterility assurance level

- Operating principle ●
- Design features ●
- Shelf life ●

# Packaging Change - New Trays

Medtronic Tubing Packs consist of coated and uncoated tubing, connectors and various medical devices that are pre-assembled into user-specified extracorporeal cardiopulmonary bypass perfusion circuits. Tubing Packs are intended for use within the extracorporeal blood pathway during cardiopulmonary bypass surgical procedures. Medtronic Tubing Packs are currently packaged into qualified clear tray and sterile packaging systems. The packaging trays for the Tubing Pack is changing as part of continual process improvement efforts. The trays are available in three sizes and include the Universal tray, the MIDI Deep tray and MIDI Low tray, for the remainder of this submission they will be referred to as tray(s) and includes all three models. The tray size for a particular tubing pack is based upon the pack contents (size and volume of the pack components). The tray modification includes material, supplier, mold and slight dimension changes. All other packaging components remain unchanged (peel pouches, ties, bands and shipper). There is no change to the sterilization and test methods

<sup>1</sup> Brand Name Change from Carmeda to Cortiva.

{5}------------------------------------------------

used for the new trays (Sterilization, Biocompatibility, Stability and Performance testing). The packaging tray changes are summarized in Tables 1.

| Feature                                     |           | Current:                                              | Future:                                                                              | Model Number                     | Change |
|---------------------------------------------|-----------|-------------------------------------------------------|--------------------------------------------------------------------------------------|----------------------------------|--------|
| Tray                                        |           |                                                       |                                                                                      |                                  |        |
| Dimension                                   | Universal | 27.75Lx23.79Wx21.77H                                  | 27.78Lx23.79Wx21.72H                                                                 | 999505-04 old<br>M057602T001 new | Yes    |
|                                             | MIDI Deep | 29.92Lx16.02Wx8.21H                                   | 29.83Lx 15.73W x 8.33H                                                               | 999505-02 old<br>M972494A001 new | Yes    |
|                                             | MIDI Low  | 29.92Lx16.02Wx6.36H                                   | 29.75Lx 15.68W x 6.37H                                                               | 999505-03 old<br>M972495A001 new | Yes    |
| Supplier                                    |           | Nelipak                                               | C&J Tech                                                                             | N/A                              | Yes    |
| Technology (Mold)                           |           | Thermoformed Molded                                   | Injection Molded                                                                     | N/A                              | Yes    |
| Material                                    |           | Resin<br>Eastman Spectar<br>PETG 14471<br>Copolyester | Resin<br>Resirene<br>CET® 265<br>Styrene Methyl<br>Methacrylate Acrylic<br>Copolymer | N/A                              | Yes    |
| Additional Packaging Components             |           |                                                       |                                                                                      |                                  |        |
| Lid                                         |           | Tyvek 1073B coated<br>with Oliver Tolas<br>TPT-0260   | Tyvek 1073B coated<br>with Oliver Tolas<br>TPT-0260                                  | See Appendix 2                   | No     |
| Peel Pouches                                |           | Tyvek 1073B / 48 GA<br>PET, 28 LBS LDPE (RLE-<br>005) | Tyvek 1073B / 48 GA<br>PET, 28 LBS LDPE (RLE-<br>005)                                | See Appendix 2                   | No     |
| Straps, velcros, and tie<br>bands, foam,etc |           | Various                                               | Various                                                                              | See Appendix 2                   | No     |
| Shipper Box                                 |           | Various                                               | Various                                                                              | See Appendix 2                   | No     |
| Sterilization                               |           | Ethylene oxide<br>sterilization                       | Ethylene oxide<br>sterilization                                                      | N/A                              | No     |
| Testing                                     |           | Biocompatibility,<br>Stability and<br>Performance     | Biocompatibility,<br>Stability and<br>Performance                                    | N/A                              | No     |

#### Table 1: Package Component Details

### Summary of Performance and Biological Testing

Medtronic conducted the following packaging system retesting for new tray:

- Process Characterization
- Shelf Life
- Process Validation
- ●
- Packaging Performance
- Sterilization
- Biocompatibility

### Conclusion

Medtronic has demonstrated that the Tubing Pack with the new tray used in extracorporeal cardiopulmonary bypass perfusion systems is substantially equivalent to the predicate device based upon design, test results, and indications for use. The fundamental scientific principle, labeling and the intended use are unchanged because of this device modification.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K171308](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K171308)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
