← Product Code [DWF](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF) · K150567

# MEDTRONIC BIO-MEDICUS CANNULA INSERTION KIT (K150567)

_Medtronic, Inc. · DWF · Apr 9, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K150567

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [DWF](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF.md)
- **Decision Date:** Apr 9, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

This kit is intended for use by trained physicians only, to assist in vessel cannulation for cardiopulmonary bypass circulation. Standard surgical or percutaneous insertion techniques can be employed. This kit is intended for use for up to 6 hours.

## Device Story

Bio-Medicus Insertion Kit provides surgical tools for vessel cannulation during cardiopulmonary bypass. Kit components include Seldinger needle, guidewire, scalpel blade, stepped dilators, and catheter tip syringe. Used by trained physicians in clinical settings to facilitate cannula placement. Device assists in establishing bypass circulation; supports standard surgical or percutaneous insertion techniques. Benefits patient by enabling secure vascular access for cardiopulmonary bypass procedures.

## Clinical Evidence

Bench testing only. Verification and validation testing performed on kit components including dilator tracking, kink resistance, and luer body separation. All tests passed.

## Technological Characteristics

Kit includes mechanical surgical instruments: Seldinger needle, guidewire, scalpel blade, stepped dilators, and catheter tip syringe. No energy source or software. Sterilization requirements are identical to the predicate.

## Regulatory Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

## Predicate Devices

- Medtronic Bio-Medicus™ Cannula Insertion Kit ([K924643](/device/K924643.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines suggesting movement or connection. The profiles are arranged in a way that they appear to be emerging from a single point. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 16, 2015

Medtronic, Inc. Bruce Backlund Principal Regulatory Affairs Specialist 7611 Northland Dr. Minneapolis, MN 55428

Re: K150567

Trade/Device Name: Bio-Medicus Insertion Kit Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula, or tubing Regulatory Class: Class II Product Code: DWF Dated: March 2, 2015 Received: March 13, 2015

Dear Mr. Bruce Backlund,

This letter corrects our substantially equivalent letter dated April 9, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

M.A. Willemen

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K150567

Device Name Bio-Medicus Insertion Kit

Indications for Use (Describe)

This kit is intended for use by trained physicians only, to assist in vessel cannulation for cardiopulmonary bypass circulation. Standard surgical or percutaneous insertion techniques can be employed. This kit is intended for use for up to 6 hours.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Date Prepared:    | April 7, 2015                                                                                                                                                                                                                                                                                                                                                                             |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:        | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009                                                                                                                                                                                                                                             |
| Contact Person:   | Bruce J. Backlund<br>Principal Regulatory Affairs Specialist<br>Medtronic Perfusion Systems<br>Phone: (763) 526-2357<br>Fax: (763) 367-8360<br>Email: bruce.j.backlund@medtronic.com<br><br>Alternate Contact:<br>Susan C. Fidler<br>Sr. Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: (763) 514-9839<br>Fax: (763) 367-8360<br>Email: susan.c.fidler@medtronic.com |
| Common Name:      | Cardiopulmonary bypass vascular catheter, cannula, or tubing.                                                                                                                                                                                                                                                                                                                             |
| Proprietary Name: | Bio-Medicus™ Insertion Kit                                                                                                                                                                                                                                                                                                                                                                |
| Classification:   | Classification: Class II<br>Panel: Cardiovascular<br>Regulation: 21 CFR 870.4210<br>Product Code: DWF                                                                                                                                                                                                                                                                                     |
| Predicate Device: | Medtronic Bio-Medicus™ Cannula Insertion Kit (K924643)                                                                                                                                                                                                                                                                                                                                    |

### Device Description

Bio-Medicus™ Insertion Kits contains the necessary components to achieve insertion of a Bio-Medicus™ cannula and introducer. The included items are: a Seldinger needle, a guidewire, a scalpel blade, stepped dilators, and a catheter tip syringe.

{4}------------------------------------------------

#### Indications for Use

This kit is intended for use by trained physicians only, to assist in vessel cannulation for cardiopulmonary bypass circulation. Standard surgical or percutaneous insertion techniques can be employed. This kit is intended for use for up to 6 hours.

#### Comparison to Predicate Device

A comparison of the proposed product to the currently marketed Medtronic Bio-Medicus™ Cannula Insertion Kit (K924643) indicates the following similarities:

- Same intended use/indications
- Same operating principle
- Same fundamental technological characteristics
- Same overall design and performance
- Same materials
- Same packaging materials and design
- Same sterilization requirements
- Three (3) new models include additional dilator, guidewire and syringe sizes as shown in the following table:

| Kit Configurations                                                             |                                                                |                                                                |                                                        |
|--------------------------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|--------------------------------------------------------|
| Model 96550<br>(currently marketed<br>device K924643 S.E.<br>December 8, 1992) | Model 96551 (Subject<br>of this submission)                    | Model 96552 (Subject<br>of this submission)                    | Model 96553 (Subject<br>of this submission)            |
| 60-cc syringe                                                                  | 10-cc syringe                                                  | 10-cc syringe                                                  | 10-cc syringe                                          |
| 18 ga Seldinger style<br>needle                                                | 18 ga Seldinger style<br>needle                                | 18 ga Seldinger style<br>needle                                | 18 ga Seldinger style<br>needle                        |
| 8/10 Fr and 12/14 Fr<br>stepped dilator                                        | 8 Fr/10 Fr, 12 Fr/14 Fr,<br>and 16 Fr/18 Fr<br>stepped dilator | 8 Fr/10 Fr, 12 Fr/14 Fr,<br>and 16 Fr/18 Fr<br>stepped dilator | 8/10 Fr and 12/14 Fr<br>stepped dilator                |
| 0.038-in (0.1 cm) x 180<br>cm (70.87-in)<br>guidewire                          | 0.038-in (0.965-mm) x<br>180-cm (70.87-in)<br>guidewire        | 0.038-in (0.965-mm) x<br>100-cm (39.37-in)<br>guidewire        | 0.025-in (0.535-mm) x<br>60-cm (23.62-in)<br>guidewire |
| #11 scalpel blade                                                              | #11 scalpel blade                                              | #11 scalpel blade                                              | #11 scalpel blade                                      |

{5}------------------------------------------------

# Summary of Performance Data

Testing has demonstrated that the Bio-Medicus™ Insertion Kits are substantially equivalent to the predicate.

The following tests were conducted to demonstrate substantial equivalence of proposed Insertion kits to the current Insertion kit (1 and 4 year data).

| Component           | Verification/Validation | Results |
|---------------------|-------------------------|---------|
| Dilator             | Follows Guidewire       | Pass    |
| Dilator / Guidewire | Kink                    | Pass    |
| Dilator Luer        | Body Separation         | Pass    |

### Conclusion

Compared to the predicate device, the fundamental scientific technology, operating principles, design features and intended use are unchanged. As such, it has been demonstrated that the modifications made to the Bio-Medicus™ Insertion Kit described in this submission result in a substantially equivalent device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K150567](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K150567)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
