K142673 · Medtronic, Inc. · DWF · Nov 18, 2014 · Cardiovascular
Device Facts
Record ID
K142673
Device Name
Bio-Medicus Adult Cannulae and Introducer
Applicant
Medtronic, Inc.
Product Code
DWF · Cardiovascular
Decision Date
Nov 18, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4210
Device Class
Class 2
Attributes
Therapeutic
Intended Use
These devices are to be used by a trained physician only. Cannulae are used to cannulate vessels, perfuse vessels or organs, and/or connect with accessory extracorporeal equipment. The cannula introducer is intended to facilitate proper insertion and placement of the appropriately-sized cannula within the vessel for cardiopulmonary bypass. The Bio-Medicus™ cannula (18-cm [7.09in] tip length models) may be used in either the femoral position as an arterial delivery cannula or in the jugular position as a venous return cannula. This product is intended for use up to 6 hours.
Device Story
Device consists of polyurethane wirewound cannulae with multiport distal tips and non-phthalate PVC introducers; used for cardiopulmonary bypass. Cannulae facilitate vessel cannulation, organ perfusion, and connection to extracorporeal circuits. Introducers provide elongated taper for vessel dilation and placement. Features include radiopaque repositionable suture rings, insertion depth marks, and securement clips. Large venous models include tubing adapters for circuit compatibility. Used in clinical settings (e.g., OR) by trained physicians. Provides physical conduit for blood flow; output is the established extracorporeal circuit connection. Benefits include facilitated insertion, improved vessel access, and securement during bypass procedures up to 6 hours.
Clinical Evidence
No human clinical data. Evidence based on pre-clinical bench testing and animal studies. Bench testing included blood trauma, sterilization, biocompatibility, packaging, pressure drop, cannula life, and securement clip testing. Results demonstrated performance equivalent or superior to predicate devices.
Technological Characteristics
Polyurethane wirewound cannula body; stainless steel wirewinding/reinforced baskets (venous); non-phthalate PVC introducer; polycarbonate/PVC connectors; silicone hemostasis cap. Radiopaque suture ring. Sterile, nonpyrogenic, single-use. Dimensions: 12.5-in to 30-in lengths. Mechanical device; no software or energy source.
Indications for Use
Indicated for patients requiring cardiopulmonary bypass; used by trained physicians to cannulate vessels, perfuse vessels/organs, or connect to extracorporeal equipment. Contraindicated in patients with severe peripheral atherosclerosis or severe arterial dissection.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
Predicate Devices
Medtronic Bio-Medicus Femoral Cannula and Introducer (K884129)
K180453 — Bio-Medicus Adult Cannulae and Introducer · Medtronic, Inc. · Mar 22, 2018
K143083 — Bio-Medicus Pediatric Cannula and Introducers · Medtronic, Inc. · Mar 25, 2015
K140208 — FERM-FLEX AND FEM-FLEX II ( FEMORAL ACCESS ARTERIAL CANNULAE) FEMTRAK,VFEM, AND FEM-FLEXII · Edwards Lifesciences, LLC · Mar 5, 2014
K060101 — ESTECH AORTIC EASY FLOW CANNULA · Endoscopic Technologies, Inc. · May 10, 2006
K180456 — Bio-Medicus Adult Cannulae Kit · Medtronic, Inc. · Mar 22, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2014
Medtronic, Inc. Bruce Backlund Principal Regulatory Affairs Specialist 8200 Coral Sea St. NE Mounds View, Minnesota 55112
Re: K142673
Trade/Device Name: Bio-Medicus Adult Cannulae and Introducer Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing Regulatory Class: Class II Product Code: DWF Dated: September 17, 2014 Received: September 19, 2014
Dear Mr. Backlund.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrenner
for
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K142673
Device Name Bio-Medicus Adult Cannula and Introducer
#### Indications for Use (Describe)
These devices are to be used by a trained physician only. Cannulate vessels, perfuse vessels or organs, and/or connect with accessory extracorporeal equipment. The cannula introducer is intended to faciliate proper insertion and placement of the appropriately-sized cannula within the vessel for cardiopulmonary bypass. The Bio-Medicus" cannula (18-cm [7.09-in] tip length models) may be used in either the femoral position as an arterial delivery cannula or in the jugular position as a venous return cannula. This product is intended for use up to 6 hours.
Type of Use (Select one or both, as applicable)
■ Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFFEMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.•
FORMFDA 3881(1114t
Page 1 of 1
I SC Publishier Sers ices (32) | J =J 14 | 14 | 14 | 14 |- | 7-411 利
Form Approved; OMB No_0910-0120 Expiration Date= January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
| Date Prepared: | September 17, 2014 |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establish Registration Number: 2184009 |
| Contact Person: | Bruce Backlund<br>Principal Regulatory Affairs Specialist<br>Phone: (763) 526-2357<br>Fax:<br>(763) 367-8361<br>E-mail:bruce.j.backlund@medtronic.com |
| Trade Name:<br>Common Name:<br>Classification Name: | Bio-Medicus™ Adult Cannulae and Introducer<br>Cardiopulmonary bypass vascular catheter, cannula, or tubing<br>Catheter, Cannula and Tubing, Vascular, Cardiopulmonary<br>Bypass |
| Classification: | Class II, 21 CFR 870.4210 |
| Product Code: | DWF |
| Name of Predicate Devices: | Medtronic Bio-Medicus Femoral Cannula and Introducer<br>(K884129) |
Avalon Elite Bi-Caval Dual Lumen Catheter (K081820).
## Device Description:
The adult venous and arterial cannulae consist of a polyurethane wirewound body with a multiport distal tip. The adult venous cannula also has a multi-port body built with stainless steel wirewinding and reinforced baskets. All adult arterial cannulae come with a vented 3/8-in vented connector and all of the adult venous come with a non-vented 3/8-in connector. The overall length of the cannulae is 12.5-in for the adult arterial/jugular and 25.6-in, 26.5-in and 30in for the adult venous. The Adult cannulae come with a repositionable suture ring that is radiopaque. Insertion depth marks aid in positioning the cannula.
{4}------------------------------------------------
The one-piece non-phthalate PVC introducers included with these cannulae have an elongated taper for dilation over a longer distance.
The Adult Cannula and Introducer package contains a securement clip which can be used near the insertion site for securement of the cannula body which aides in securement as well as an aide for physician organization. The large venous models (Sizes 23 - 29 Fr) also contain a 3/8-in x ½-in tubing adapter with an attached length of 3/8-in tubing to allow connection to a ½-in cardio pulmonary bypass circuit.
Sterile, nonpyrogenic, single use.
# Intended Use:
These devices are to be used by a trained physician only. Cannulae are used to cannulate vessels, perfuse vessels or organs, and/or connect with accessory extracorporeal equipment. The cannula introducer is intended to facilitate proper insertion and placement of the appropriately-sized cannula within the vessel for cardiopulmonary bypass. The Bio-Medicus™ cannula (18-cm [7.09in] tip length models) may be used in either the femoral position as an arterial delivery cannula or in the jugular position as a venous return cannula. This product is intended for use up to 6 hours.
# Contraindications:
Alone, the cannula and introducer are not medical treatment devices. The cannula introducer is only to be use with the appropriately-sized Bio-Medicus cannula. These devices are not intended for use, except as indicated above. Do not use if the patient has severe peripheral atherosclerosis or severe arterial dissection.
# Comparison to Predicate Devices:
A comparison of the Medtronic Bio-Medicus Adult Arterial and Venous Cannulae to the predicate device indicates the following similarities:
- . Same intended use
- Same technological characteristics
- . Same operating principle
- . Similar design features, exceptions include improved transition between the introducer body and cannula tip to increase ease of cannula insertion, venous models include additional drainage opening along the cannula body, reduction in the overall cannula body length within the non-wirewound section and added a radiopaque suture ring.
- . Similar materials, exceptions include the introducer material which has changed from Polyurethane to Polyvinyl Chloride (PVC), connector material from Polycarbonate to PVC and hemostasis cap material from Styrene to Silicone.
- . Same shelf life
{5}------------------------------------------------
### Summary of Performance Data
Pre-clinical bench testing was used to verify the performance characteristics of this device. Animal testing was also completed to establish substantial equivalence with the predicate devices.
The following performance tests were conducted:
- . Blood Trauma Testing
- . Sterilization Testing
- Biocompatibility Testing .
- Packaging Testing ●
- Pressure Drop Testing .
- . Cannula Life Testing
- Cannula and Introducer Testing ●
- Cannula Testing ●
- Introducer Testing ●
- Securement Clip Testing .
#### Conclusion:
The primary predicate device, Bio-Medicus™ Percutaneous Cannula & Introducer Set, was used to compare the fundamental technological characteristics of the proposed Bio-Medicus Adult Cannula and Introducer. When comparative testing was completed, the proposed Bio-Medicus Adult Cannulae and Introducer performed as well or better than the legally marketed devices.
The data included in this submission is sufficient to provide reasonable assurance that the proposed Bio-Medicus Adult Arterial and Venous Cannulae are substantially equivalent to the legally marketed predicate devices, Bio-Medicus Percutaneous Cannula and Introducer Set (K884129) and Avalon Elite Bi-Caval Dual Lumen Catheter (K081820).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.