These Cannulae are intended for use in conjunction with cardiopulmonary bypass surgery up to six hours or less for delivery of cardioplegia solutions directly to the coronary arteries.
Device Story
DLP Coronary Ostial Perfusion Cannulae are sterile, single-use devices used during cardiopulmonary bypass surgery. The device consists of a basket-style, soft silicone, or spherical tip attached to a malleable stainless steel tube, terminating in a locking female luer fitting. The cannula is used by a surgeon to deliver cardioplegia solutions directly into the coronary arteries. The device facilitates myocardial protection during surgery. It is a passive delivery tool; it does not process data or provide automated outputs.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed that material modifications (silicone, polypropylene, barium sulfate) met performance requirements for burst strength, tensile pull-off force, and torsional twist-off force. Biocompatibility assessments confirmed no adverse impact from material changes.
Technological Characteristics
Materials: Polyvinyl Chloride (PVC), Polypropylene, Stainless steel, Silicone, Synthetic Barium Sulfate. Design: Basket-style, soft silicone, or spherical tip with malleable stainless steel tube. Connectivity: None. Energy Source: None (passive). Sterilization: Sterile, single-use.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery requiring delivery of cardioplegia solutions directly to the coronary arteries for up to six hours.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
{0}------------------------------------------------
# 510(k) Summary
| Date Prepared: | May 2, 2013 | JUL 1 0 2013 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Submitter: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009 | |
| Contact Person: | Kevin T. Lam<br>Senior Regulatory Affairs Specialist<br>Medtronic Perfusion Systems<br>Phone: 763.526.2360<br>Fax: 763.367.8360<br>Email: kevin.t.lam@medtronic.com | |
| Alternate Contact: | Susan Fidler<br>Senior Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: 763.514.9839<br>Fax: 763.367.8360<br>Email: susan.c.fidler@medtronic.com | |
| Device Name and Classification | | |
| Trade Name: | DLP® Coronary Ostial Perfusion Cannulae | |
| Trade Name: | DLF Coronary Ostial Perfusion Cannulae | |
| | Models: 30010, 30011, 30012, 30014, 30050, 30055, 30315,<br>30317, 30320, 30110, 30112, 30114, 30155, 30212, and 30255. | |
| Common Name: | Cardiopulmonary bypass vascular catheter, cannula, or tubing | |
| Regulation Number: | 21 CFR 870.4210 | |
| Product Code: | DWF | |
| Product Classification: | Class II | |
## Predicate Devices
K034058
Spherical Tip Coronary Ostial Cannulae
# Device Description
These coronary Cannulae consist of basket-style tip, soft (silicone) tip, or spherical tip which is attached to a malleable stainless steel tube, or a soft bulb beveled tip with an integral silicone body. The Cannulae terminate with a locking female luer fitting. The French diameter is
{1}------------------------------------------------
measured at the base of the tip adjacent to the sealing flange or across the diameter for the soft bulb beveled tip style. The Cannulae are nonpyrogenic, single use, and sterile.
#### Indications for Use
These Cannulae are intended for use in conjunction with cardiopulmonary bypass surgery up to six hours or less for delivery of cardioplegia solutions directly to the coronary arteries.
## Comparison to Predicate Devices
A comparison of the modified product to the currently marketed predicate products (K034058) indicates the following similarities:
- Same intended use ■
- Same technological characteristics
- Same operating principle
- 트 Same design features
- 트 Same base materials (Polyviny) Chloride (PVC), Polypropyiene, and Stainless steel)
- P Same shelf life
## Summary of Performance Data
Verification and validation testing has demonstrated that the DLP Coronary Ostial Perfusion Cannula are substantially equivalent to the predicates.
| Modifications | Verification/Validation | Results |
|------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------|
| Silicone to new Silicone due to<br>supplier obsolescence. | Burst strength relative to predicate<br>material. | Pass |
| Organic to synthetic barium<br>sulfate. | Tip component tested for tensile<br>pull-off force. | Pass |
| Organic to synthetic barium<br>sulfate. | Biocompatibility Assessment. | Pass - no impact to<br>biocompatibility |
| Polypropylene to new<br>Polypropylene due to supplier<br>obsolescence. | Hub tested for Tensile Pull-Off<br>Force and Torsional Twist-Off<br>Force. | Pass |
| Polypropylene to new<br>Polypropylene due to supplier<br>obsolescence. | Biocompatibility Assessment. | Pass - no impact to<br>biocompatibility |
The following performance tests were conducted:
### Conclusion
Medtronic has demonstrated that the modifications made to the DLP® Coronary Ostial Perfusion Cannulae product family described in this submission result in a substantially equivalent device because the fundamental scientific principle, operating principle, design features and intended use are unchanged from the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
July 10, 2013
Medtronic CardioVascular, Inc. c/o Kevin Lam 8200 Coral Street NE Mailstop MVS83 Mounds View, MN 55112
K131269 Re:
> Trade/Device Name: DLP Coronary Ostial Perfusion Cannula Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing Regulatory Class: Class II Product Code: DWF Dated: June 13, 2013 Received: June 14, 2013
Dear Mr. Lam:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Kevin Lam
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
M. A. Hillemann
for
Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Statement of Indications for Use
510(k) Number: K131269
Device Name: DLP® Coronary Ostial Perfusion Cannulae
Indications for Use:
These Cannulae are intended for use in conjunction with cardiopulmonary bypass surgery up to six hours or less for delivery of cardioplegia solutions directly to the coronary arteries.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.A. Lillehamm
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.