K120543 · Cardiacassist, Inc. · DWF · May 30, 2012 · Cardiovascular
Device Facts
Record ID
K120543
Device Name
TANDEMHEART FEMORAL ARTERIAL CANNULA SET
Applicant
Cardiacassist, Inc.
Product Code
DWF · Cardiovascular
Decision Date
May 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4210
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Femoral Arterial Cannula and Introducer Set is intended to cannulate vessels, perfuse vessels or organs and/or connect with accessory extracorporeal circulatory support equipment for a duration of six hours or less. The cannula introducer is intended to facilitate proper insertion and placement of the cannula within the vessel for extracorporeal circulatory support. These devices are to be used by a trained physician only.
Device Story
TandemHeart Femoral Arterial Cannula Set comprises 17 Fr. wire-reinforced polyurethane cannula and 12 Fr. tapered polyurethane introducer; used for vessel cannulation and extracorporeal circulatory support connection. Cannula features distal tip taper, multiple side holes, radiopaque markers, and proximal vented barbed fitting for 3/8" tubing; includes suture wing for patient securement and stop component to limit insertion depth. Introducer includes hemostasis cap to minimize blood loss during insertion. Operated by trained physicians in clinical settings. Device facilitates blood flow to/from extracorporeal equipment; radiopaque components allow fluoroscopic visualization during placement. Single-use, sterile device.
Clinical Evidence
Bench testing only. In-vitro performance testing included hemolysis, system capacity, flow vs. pressure drop (HQ), kink, leak, and deflection testing. Results confirmed device meets design requirements and performs equivalently to the predicate.
Technological Characteristics
Wire-reinforced polyurethane cannula; polyurethane introducer; soft durometer Tecoflex polyurethane suture wing and stop component. Radiopaque markers/body for fluoroscopy. 17 Fr. cannula, 12 Fr. introducer. Vented barbed fitting for 3/8" tubing. Single-use, sterile. No software or electronic components.
Indications for Use
Indicated for patients requiring vessel cannulation, organ/vessel perfusion, or connection to extracorporeal circulatory support equipment for durations of six hours or less. For use by trained physicians only.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
Predicate Devices
Medtronic Bio-Medicus Femoral Cannula and Introducer (K924642)
Submission Summary (Full Text)
{1}------------------------------------------------
MAY 3 0 2012
## 510(k) Summary
Date: May 17, 2012
## Applicant
CardiacAssist. Inc. 240 Alpha Drive Pittsburgh, PA 15238 Telephone: 412-963-7770 Fax: 412-963-0800
### Contact person
Katie Dillon Title: Regulatory Affairs Manager Phone: 412-963-7770 x266 e-mail: kdillon@cardiacassist.com
#### Device
| Trade/Proprietary Name: | TandemHeart Femoral Arterial Cannula Set |
|-------------------------|------------------------------------------------------------------------------------------------------|
| Common Name: | Femoral Arterial Cannula and Introducer |
| Classification Name: | Cardiopulmonary bypass vascular catheter, cannula, or<br>tubing. (21 CFR 870.4210. Product Code DWF) |
#### Predicate Device
Medtronic Bio-Medicus Femoral Cannula and Introducer (K924642)
### Device Description
The TandemHeart Femoral Arterial Cannula Set consists of two components, as shown in Figure 1: a 17 Fr. Femoral Arterial Cannula and a 12 Fr. Introducer. The device is intended to cannulate vessels, perfuse vessels or organs, and/or connect with accessory extracorporeal circulatory support equipment. The product is intended to be single patient, single use, sterile device.
The cannula has multiple side holes in addition to the tip opening for unimpeded flow of blood at the distal end and a vented barbed fitting at the proximal end to enable the connection of 3/8" tubing. Radiopaque markers are embedded at the distal tip of the cannula, and the cannula body is wire-reinforced for visualization under fluoroscopy. Insertion depth markings have been incorporated in the cannula body working length from 5 to 17 cm measured from the distal tip.
{2}------------------------------------------------
The cannula includes a suture wing to provide a means for securing the cannula to the patient and a stop component to minimize cannula insertion depth beyond its working length. Printing on the proximal region of the cannula indicates the area where a clamp should be applied if needed during the set-up or removal process.
The 12 Fr. introducer is provided to facilitate placement of the arterial cannula, within the target vessel, and is designed with a tapered distal tip. The introducer proximal end contains a luer hub to aid in the removal of the introducer. The introducer body is also constructed of a radiopaque material for visualization under fluoroscopy.
The introducer includes a hemostasis cap that provides the interface between the cannula proximal connector and introducer body. The hemostasis cap aids in minimizing blood loss during the insertion of the cannula/introducer assembly into the target vessel.
Image /page/1/Figure/4 description: The image shows a diagram of a 17 Fr TandemHeart Femoral Arterial Cannula and a 12 Fr Introducer. The cannula has several labeled parts, including a clamp area, vented barbed connector, suture wing, stop, insertion depth markers, and a wire-reinforced area. The introducer has an introducer hub with a luer and a hemostasis cap, and the cannula tip has radiopaque discs.
Figure 1: TandemHeart Femoral Arterial Cannula Set
#### Intended Use
The Femoral Arterial Cannula and Introducer Set is intended to cannulate vessels, perfuse vessels or organs and/or connect with accessory extracorporeal circulatory support equipment for a duration of six hours or less. The cannula introducer is intended to facilitate proper insertion and placement of the cannula within the vessel for extracorporeal circulatory support. These devices are to be used by a trained physician only.
{3}------------------------------------------------
#### Comparison of Technological Characteristics
The TandemHeart Femoral Arterial Cannula Set and the predicate Medtronic Bio-Medicus Femoral Cannula and Introducer have the same intended use and both consist of a wire-reinforced, polyurethane arterial cannula and a polyurethane introducer. The TandemHeart Femoral Arterial Cannula Set contains two additional components located on the exterior proximal portion of the cannula: a suture wing to provide a means of securing the cannula to the patient and a stop component to minimize insertion of the cannula beyond its working length. The suture wing and stop component are constructed of soft durometer Tecoflex polyurethane, a commonly used material for interventional medical devices. and were tested for biocompatibility. The TandemHeart Femoral Arterial Cannula Set's cannula includes a distal tip taper, while the predicate has a straight tip. The TandemHeart Arterial cannula also has fewer side holes than the predicate device.
### Performance Data
Non-clinical performance testing was performed to demonstrate substantial equivalence. The performance testing included in-vitro hemolysis testing, in-vitro system capacity testing, flow vs. pressure drop (HQ), kink testing, leak testing, and deflection testing. All performance tests were conducted on both the TandemHeart Femoral Arterial Cannula Set and the predicate Medtronic Bio-Medicus Femoral Cannula and Introducer. Based on the performance test results, the TandemHeart Femoral Cannula Set was found to meet the established design input requirements as well as perform substantially equivalent to the predicate Medtronic Bio-Medicus Femoral Cannula and Introducer.
#### Conclusions
The TandemHeart Femoral Arterial Cannula Set is substantially equivalent to the Medtronic Bio-Medicus Femoral Cannula and Introducer in design characteristics, performance, and intended use. The performance testing demonstrated that despite the geometric differences, including a distal tip taper on the TandemHeart device and fewer side holes than the predicate device, the TandemHeart Femoral Arterial Cannula Set performed substantially equivalent to the predicate device. The two additional components on the TandemHeart Femoral Arterial Cannula Set's cannula body, the suture wing and stop, were found to be biocompatible.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 3 0 2012
CardiacAssist, Inc. c/o Ms. Katie Dillon Regulatory Affairs Specialist 240 Alpha Drive Pittsburgh, PA 15238
#### Re: K120543
Trade/Device Name: TandemHeart Femoral Arterial Cannula Set Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula, or tubing Regulatory Class: Class II Product Code: DWF Dated: May 17, 2012 Received: May 18, 2012
Dear Ms. Dillon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 – Mr. Katie Dillon
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Cm A. Killeber
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use
510(k) Number: K120543
Device Name: TandemHeart Femoral Arterial Cannula Set
Indications for Use:
The Femoral Arterial Cannula and Introducer is intended to cannulate vessels, perfuse vessels or organs and/or connect with accessory extracorporeal circulatory support equipment for a duration of six hours or less. The cannula introducer is intended to facilitate proper insertion and placement of the cannula within the vessel for extracorporeal circulatory support. These devices are to be used by a trained physician only.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
MG. Stilleben
n Sian-Off) Division of Cardiovascular Devices
K120543 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.