HLS CANNULA NON-COATED, WITH BIOLINE COATING AND WITH SOFTLINE COATING

K102532 · Maquet Cardiopulmonary, AG · DWF · Feb 25, 2011 · Cardiovascular

Device Facts

Record IDK102532
Device NameHLS CANNULA NON-COATED, WITH BIOLINE COATING AND WITH SOFTLINE COATING
ApplicantMaquet Cardiopulmonary, AG
Product CodeDWF · Cardiovascular
Decision DateFeb 25, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4210
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HLS Cannula from MAQUET is intended for use by trained physicians only. The HLS Cannula may be utilized to cannulate suitable vessels to provide circulatory perfusion of organs and vessels by forming a connection with the extracorporeal circulation. Standard surgical or percutaneous insertion techniques can be employed. This product is intended for use up to six hours or less.

Device Story

HLS Cannula is a wire-reinforced, thin-wall polyurethane catheter used for extracorporeal circulation; facilitates vessel cannulation for organ/vessel perfusion. Available in various sizes/lengths; includes pre-mounted 3/8" connector and introducer for guide wire insertion (up to 0.038"). Optional percutaneous insertion kit (PIK 100/150) supports Seldinger technique. Used by trained physicians in clinical settings; provides connection between patient vasculature and extracorporeal circuit. Device is sterile, single-use, and non-pyrogenic. Performance features include kink resistance, pressure resistance (up to 6 hours at 1.5 bar), and air-tight connections. Optional BIOLINE or SOFTLINE coatings available. Clinical benefit derived from reliable vascular access for cardiopulmonary bypass procedures.

Clinical Evidence

Bench testing only. No clinical data provided. Performance evaluated via flow rate comparison against Bio-Medicus Femoral Cannula, kink resistance, tensile strength of cannula-connector and introducer-grip bonds, 6-hour pressure resistance at 1.5 bar, air-tightness under vacuum, corrosion resistance, and adhesion of depth-markings. Compliance with ISO 10993-1 (Biologic Evaluation) and ISO 10555-1 (Sterile, single-use intravascular catheters) verified.

Technological Characteristics

Wire-reinforced polyurethane cannula; 3/8" connector; optional BIOLINE or SOFTLINE coatings. Complies with ISO 10993-1 and ISO 10555-1. Single-use, sterile, non-pyrogenic. Mechanical design includes kink-resistant shaft and designated clamping section. No electronic components or software.

Indications for Use

Indicated for use by trained physicians to cannulate vessels for extracorporeal circulation to provide circulatory perfusion of organs and vessels. Intended for use up to six hours. No specific age or gender restrictions provided.

Regulatory Classification

Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the date FEB 25 2011, followed by the word "MAQUET" in large, bold letters. Below that, the words "GETINGE GROUP" are printed in smaller letters. Above the word "MAQUET" is the number K102532. ## 510 (K) Summary [as required by 21 CFR 807.92(c) ] Maquet Cardiopulmonary AG Submitter: Hechinger Strasse 38 72145 Hirrlingen Germany Contact Person: Dr. Ingrid Richter Phone: +49 7478 921-337 +49 7478 921-8667 Fax: E-mail: ingrid.richter@maquet-cp.com August 31, 2010 Date Prepared: HLS Cannula non-coated, with BIOLINE Coating Device Trade Name: and with SOFTLINE Coating HLS Femoral Cannula Common/Usual name: Catheter, Cannulae and Tubing, Vascular, Classification names: Cardiopulmonary Bypass Predicate Devices: Bio-Medicus Femoral Cannula, Medtronic USA, cleared under K924642, QUADROX-i Adult Microporous Membrane Oxygenator with and without Arterial Filter with SOFTLINE Coating, MAQUET Cardiopulmonary AG, cleared under K082117, QUADROX-i Adult Microporous Membrane Oxygenator with and without Arterial Filter with BIOLINE Coating, MAQUET Cardiopulmonary AG, cleared under K090511. {1}------------------------------------------------ #### Device Description: The HLS Cannula from MAQUET is a wire-reinforced, thin-wall cannula made of polyurethane. The transparent proximal section has no reinforcement and can be clamped. Each cannula is supplied with a pre-mounted 3/8" connector and an introducer that allows a guide wire up to 0.038" to be inserted. The cannula is available with an optional percutaneous insertion kit for the Seldinger technique and an optional BIOLINE or SOFTLINE Coating. The HLS Cannula comes in a range of sizes and lengths. The HLS Cannula is a sterile and non-pyrogenic device, for single use only and is not to be re-sterilized by the user. The insertion kit from MAQUET comprises various components which permit access to the vessel. One insertion kit is available for the arterial cannula (PIK 100) and one for the venous cannula (PIK 150). #### Indications for Use: The HLS Cannula from MAQUET is intended for use by trained physicians only. The HLS Cannula may be utilized to cannulate suitable vessels to provide circulatory perfusion of organs and vessels by forming a connection with the extracorporeal circulation. Standard surgical or percutaneous insertion techniques can be employed. This product is intended for use up to six hours or less. The insertion kit is used for preparing the cannulation of vessels for extracorporeal circulation. Surgical procedures can be performed under visual control or with percutaneous cannulation using the Seldinger technique. #### Statement of Technical Comparison: HLS Cannula non-coated, with BIOLINE Coating and with SOFTLINE Coating from MAQUET Cardiopulmonary AG, have a comparable intended use as the Bio-Medicus Femoral Cannulae (K924642). Both are intended to be used to cannulate vessels, perfuse vessels or organs up to 6 hours. Standard surgical or percutaneous insertion techniques can be used. Furtheron the indications for use of the SOFTLINE Coating of the QUADROXi Adult with SOFTLINE Coating is the same as with the HLS Cannula with SOFTLINE Coating and the indications for use of the BIOLINE Coating of the QUADROX-i Adult with BIOLINE Coating is the same as with the HLS Cannula with BIOLINE Coating. The HLS Cannulae from MAQUET have similar technical characteristics, performance specification and materials as the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Maquet Getinge Group. The word "MAQUET" is in large, bold, sans-serif font on the top line. Below that, "GETINGE GROUP" is in a smaller, sans-serif font. Both cannulae are made of polyurethane and are protected from kinking - to a greater or lesser extent - by wire reinforcement. The standard adult sizes are equipped with 3/8" connectors; all feature a designated clamping section. #### Non-clinical Testing: Performance testing has resulted in data that demonstrate that the HLS Cannulae performs according to the applied performance standards. The following performance characteristics of the HLS Cannulae were compared with the predicate device to determine substantial equivalence: #### Evaluation of flow rate For a relevant flow of 0 - 6 liters /minute the pressure drop of arterial and venous HLS Cannulae was measured in comparison to the predicate devices Bio-Medicus Femoral Cannulae from Medtronic. The performance data of MAQUET cannulae as compared to the MEDTRONIC predicate cannulae are substantially equivalent. To evaluate the safety and effectiveness of the HLS Cannula additional testing was performed: #### Kink resistance This test is to verify that the wire re-inforced cannulae shafts will not kink. #### Tensile test of connection of cannula-connector This test is to verify the bonding stability between cannulae and connector. #### Tensile test of connection of introducer grip This test is a verification of bonding of introducer and grip. #### Verification of six hours pressure resistance This test is to verify the durability of all connections of HLS arterial and venous Cannulae for a period of 6 hours at 1.5 bar. #### Air tightness This test is to verify that all connections of the cannulae are air-tight if put to a specified vacuum. #### Evaluation of Corrosion This test is to evaluate if cannulae corrode. #### Evaluation on printed depth-markings of cannulae The test is to verify that the adhesion of the printing is sufficient. {3}------------------------------------------------ # MAQUET The HLS Cannulae non coated, with BIOLINE Coating and with SOFFENEROUP Coating have been tested or evaluated for compliance to ISO 10993-1 Biologic Evaluation of Medical Devices as well as to ISO 10555-1 "Sterile, sincle-use intravascular catheters - Part 1 general requirements". The products met these requirements. #### Determination of Substantial Equivalence MAQUET Cardiopulmonary AG has considered indications for use, design, materials, performance characteristics, and effectiveness of the proposed HLS Cannulae non-coated, with BIOLINE Coating and with SOFTLINE Coating as compared to the predicate devices. Based on the review and testing provided in this submission MAQUET Cardiopulmonary AG believes that the HLS Cannula non-coated, with BIOLINE Coating and with SOFTLINE Coating is substantially equivalent to the named predicate devices that are FDA cleared. #### Conclusion The HLS Cannulae non coated, with BIOLINE Coating and with SOFTLINE Coating are substantially equivalent to the named predicate devices which currently hold market clearance. {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Maquet Cardiopulmonary AG Whitney Torning Director of Regulatory Affairs 45 Barbour Pond Drive Wayne, NJ 07470 FEB 2 5 2011 Re: K102532 Trade Name: HLS Cannula Non-coated, with BIOLINE Coating, and with SOFTLINE Coating Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing Regulatory Class: Class II Product Code: DWF Dated: January 28, 2011 Received: February 2, 2011 Dear Ms. Torning: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {5}------------------------------------------------ Page 2 - Ms. Whitney Torning found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely Fours, Sincerely yours, Bram D. Zuckerman, M.D. m D. Directof Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use 510(k) Number (if known): ≤102532 Device Name: HLS Cannula Indications for Use: The HLS Cannula from MAQUET is intended for use by trained physicians only. The HLS Cannula may be utilized to cannulate suitable vessels to provide circulatory perfusion of organs and vessels by forming a connection with the extracorporeal circulation. Standard surgical or percutaneous insertion techniques can be employed. This product is intended for use up to six hours or less. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) signature Page 1 of 1 (Division Sigh-Off) of Cardiovas r Devices
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...