PORT ACCESS SYSTEMS ENDODIRECT ARTERIAL CANNULA MODEL ED24 (OLD CODE WAS-DFK24)
Applicant
Edwards Lifesciences, LLC
Product Code
DWF · Cardiovascular
Decision Date
Jan 14, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4210
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EndoDirect arterial cannula is indicated for patients undergoing cardiopulmonary bypass. The cannula is intended to deliver oxygenated blood for cardiopulmonary bypass for a duration of less than 6 hours. The EndoDirect arterial cannula also allows the hemostatic introduction and removal of vascular catheters such as the Port Access EndoClamp aortic catheter. The EndoDirect arterial cannula is intended for introduction and use through a thoracic trocar or incision. The AutoIncisor Introducer is intended for use with Port Access arterial cannulae. It is intended for incising the aorta and introducing the cannula into the aorta.
Device Story
EndoDirect arterial cannula is a flexible plastic tube for cardiopulmonary bypass; delivers oxygenated blood to patient; facilitates hemostatic introduction/removal of vascular catheters (e.g., EndoClamp). Features: beveled tip with end hole, side perfusion holes, stabilizer ring, wirewound shaft for kink resistance, barbed connection, hemostasis valve, tip orientation mark, lubricious coating. Used in surgical settings (OR) by physicians. AutoIncisor introducer used to incise aorta and introduce cannula. Device benefits include safe blood delivery and vascular access during bypass procedures.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Flexible plastic tube; wirewound shaft; lubricious coating; beveled tip with end hole and side perfusion holes; stabilizer ring; barbed connection (3/8 in.); hemostasis valve. Mechanical device; no software or energy source.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass requiring oxygenated blood delivery for <6 hours; allows hemostatic introduction/removal of vascular catheters (e.g., EndoClamp aortic catheter) via thoracic trocar or incision.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
Predicate Devices
Edwards Lifesciences DirectFlow Arterial Cannula
Related Devices
K233895 — EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A, EZC24TA, EZF21A, EZF21TA, EZF24A, EZF24TA, EZS21A, EZS21TA, EZS24A and EZS24TA); OptiSite Arterial Perfusion Cannulae (OPTI16, OPTI18, OPTI20 and OPTI22); EndoReturn Arterial Cannulae (ER21B and ER23B); · Edwards Lifesciences · Feb 5, 2024
K974736 — HEARTPORT DIRECT AORTIC RETURN CANNULA WITH INTRODUCER · Heartport, Inc. · Jul 15, 1998
K955121 — HEARTPORT ENDOARTERIAL RETURN CANNUAL · Heartport, Inc. · May 10, 1996
K990772 — DIRECTFLOW ARTERIAL CANNULA · Heartport, Inc. · Mar 23, 1999
K994243 — DIRECTFLOW KIT, 24 FR, SOFTCLAMP KIT, 24FR, STRAIGHTSHOT KIT, 23 FR, STRAIGHT TIP, STRAIGHTSHOT KIT, 23 FR, ANGLED TIP · Heartport, Inc. · May 5, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
K093730
510(k) Notification
## 510(k) Summarv
# JAN 1 4 2010
| Submitter: | Edwards Lifesciences Research Medical, Inc. | |
|----------------------|--------------------------------------------------------------------------------------------------------------------|--|
| Contact Person: | Spencer Walker, Associate II, Regulatory Affairs<br>6864 South 300 West<br>Midvale, UT 84047<br>801-565-6103 | |
| Date Prepared: | December 2, 2009 | |
| Trade Name: | PORT ACCESS Systems EndoDirect Arterial Cannula | |
| Classification Name: | Catheter, Cannula and Tubing, Vascular, Cardiopulmonary Bypass<br>21 CFR Part 870.4210, Product Code DWF, Class II | |
| Predicate Device: | Edwards Lifesciences DirectFlow Arterial Cannula | |
### Device Description:
Edwards EndoDirect arterial cannula is a flexible plastic tube intended to provide a means of safely delivering oxygenated blood for cardiopulmonary bypass procedures and allows for the hemostatic introduction and removal of vascular catheters.
The EndoDirect arterial cannula has the following features: a beveled tip with an end hole, side perfusion holes, and a stabilizer ring for suture and tourniquet, a wirewound shaft for flexibility and kink resistance, a barbed connection site with 3/8 in. acceptance, a hemostasis valve, and a mark indicating tip orientation. A lubricious coating is applied to the surface of the cannula body to facilitate ease of insertion and retraction of the EndoClamp aortic catheter.
## Intended Use:
#### EndoDirect Arterial Cannula
The EndoDirect arterial cannula is indicated for patients undergoing cardiopulmonary bypass. The cannula is intended to deliver oxygenated blood for cardiopulmonary bypass for a duration of less than 6 hours. The EndoDirect arterial cannula also allows the hemostatic introduction and removal of vascular catheters such as the Port Access EndoClamp aortic catheter. The EndoDirect arterial cannula is intended for introduction and use through a thoracic trocar or incision.
AutoIncisor Introducer: The AutoIncisor introducer is intended for use with Port Access arterial cannulae. It is intended for incising the aorta and introducing the cannula into the aorta.
#### Comparative Analysis:
It has been demonstrated that the proposed arterial cannula is comparable to the predicate device in intended use and other labeling, fundamental scientific technology, material type, principles of operation and functional performance evaluations.
Functional/Safety Testing: The functional data indicate that the proposed device performs in a substantially equivalent manner when compared with the predicate device.
#### Conclusion:
The EndoDirect arterial cannula is substantially equivalent to the cited predicate device.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
JAN 1 4 2010
Edwards Lifesciences, LLC c/o Spencer Walker Regulatory Affairs Associate II 6864 South 300 West Midvale, UT 84047
### Re: K093730
Edwards Lifesciences EndoDirect Arterial Cannula - ED24 Regulation Number: 21 CFR 870.4210 Regulation Name: Catheter, Cannula and Tubing, Vascular, Cardiopulmonary Bypass Regulatory Class: Class II Product Code: DWF Dated: December 2, 2009 Received: December 3, 2009
#### Dear Mr. Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Spencer Walker
Please be advised that FDA's issuance of a substantial equivalence determination does not niean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
i R. Lacher
/ Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known):
PORT ACCESS Systems EndoDirect Arterial Cannula Device Name:
#### Indications for Use:
The EndoDirect arterial cannula is indicated for patients undergoing cardiopulmonary bypass. The cannula is intended to deliver oxygenated blood for cardiopulmonary bypass for a duration of less than 6 hours. The EndoDirect arterial cannula also allows the hemostatic introduction and removal of vascular catheters such as the Port Access EndoClamp aortic catheter. The EndoDirect arterial cannula is intended for introduction and use through a thoracic trocar or incision.
The AutoIncisor Introducer is intended for use with Port Access arterial cannulae. It is intended for incising the aorta and introducing the cannula into the aorta.
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | |
|------------------------|--|
| (21 CFR 807 Subpart C) | |
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Durra D. La James
ion Sign-Off Division of Cardiovascular Devices
510(k) Number_K093733
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.