← Product Code [DWF](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF) · K042156

# CARDEON COBRA CATHETER (K042156)

_Cardeon Corp. · DWF · Oct 6, 2004 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K042156

## Device Facts

- **Applicant:** Cardeon Corp.
- **Product Code:** [DWF](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF.md)
- **Decision Date:** Oct 6, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Intended Use

The Cardeon Cobra Catheter is indicated for use in open chest surgery on cardiopulmonary bypass up to 6 hours. It is used to cannulate the aorta and provide hypothermic arch perfusate and normothermic corporeal perfusate.

## Device Story

Cobra Catheter is a vascular cannula for cardiopulmonary bypass (CPB) procedures. Device inserted into aorta by cardiac surgeons in OR; facilitates selective perfusion by delivering hypothermic perfusate to the aortic arch and normothermic perfusate to the rest of the body. Operates within standard extracorporeal perfusion circuits. By enabling independent temperature control of the brain/arch versus the body, the device aims to preserve cerebral autoregulation during cardiac surgery. Clinical benefit demonstrated through maintenance of adequate venous oxygen saturation and temperature differentials. Output is physical delivery of blood flow; monitored via standard clinical practices including arterial line pressures and temperature sensors.

## Clinical Evidence

Clinical evidence includes feasibility and randomized clinical studies. Primary endpoints focused on safety and effectiveness, specifically the delivery of adequate venous oxygen saturation and the ability to maintain differential temperatures to preserve cerebral autoregulation. Results indicated the device functions as intended and performs comparably to predicate arterial return devices.

## Technological Characteristics

Externally communicating vascular catheter for limited contact (up to 6 hours) with circulating blood. Designed for aortic cannulation in CPB circuits. Features directed flow capability for selective perfusion of the aortic arch and distal thoracic aorta. Biocompatible materials validated per 21 CFR Part 58. No electronic components, software, or energy sources.

## Regulatory Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) SUMMARY

510(k) number:

This summary of information is being provided in accordance with 21 CFR 807.92(a).

| Applicant Information: | Cardeon Corporation<br>10600 N. Tantau Avenue<br>Cupertino, CA 95014 |
|------------------------|----------------------------------------------------------------------|
| Contact Person:        | Jane Beggs<br>Regulatory Affairs<br>Cardeon Corporation              |
| Date:                  | 9 August 2004                                                        |
| Device Trade Name:     | Cobra™ Catheter                                                      |
| Device Common Name     | Catheter, cannula and tubing, vascular, cardiopulmonary              |
| Regulation No.:        | bypass 870.4210                                                      |
| Classification / Code: | Class II / DWF                                                       |

Indications for Use: The Cardeon Cobra Catheter is indicated for use in open chest surgery on cardiopulmonary bypass up to 6 hours. It is used to cannulate the aorta and provide hypothermic arch perfusate and normothermic corporeal perfusate.

Summary of Substantial Equivalence: The Cobra Catheter is substantially equivalent to currently marketed devices used to directly cannulate the aorta and return perfusion to patients undergoing general cardiac surgery. Perfusion is achieved through standard connections to the extracorporeal circuit. ). Intended use and indications for the subject and predicate devices are substantially equivalent for arterial return during CPB. Substantial equivalence is also supported through comparison with several marketed devices with the same indications for use, including arterial return cannulae with directed flow (i.e., specified flow patterns). Directed flow is a feature offered by devices cleared through premarket notification and currently marketed for CPB and peripheral vascular thrombectomy. Selective perfusion systems have been reported on extensively and safely used for over two decades. These selective perfusion systems utilize predicate cannulae and standard perfusion circuitry for cooling the brain independently from the rest of the body. Arterial line pressures, nasopharyngeal and rectal temperatures, radial arterial pressures, oxygen saturation, etc. are standard clinical practices for monitoring the patient used in predicate selective perfusion systems. Technological characteristics (design and materials) of the Cobra Catheter are substantially equivalent to arterial return devices currently marketed for CPB. Differences between the Cobra Catheter and currently marketed devices do not substantially alter arterial return during cardiopulmonary bypass.

Substantial equivalence with regard to safety and effectiveness has also been Product Testing: demonstrated through standardized in vitro and in vivo tests as well clinical study.

The Cobra Catheter is an externally communicating device in limited contact with circulating blood. Tests were conducted and inspected in accordance with Good Laboratory Practice Regulation, 21 CFR Part 58. Based on the test results, the Cobra Catheter is biologically compatible, non-toxic and safe for use in limited contact (up to 6 hours) with circulating blood.

{1}------------------------------------------------

In vitro and in vivo testing was designed to ensure device structural integrity, key device functionality and ease of use. Device in vivo performance was evaluated using standard surgical techniques and und case. Dor cardiopulmonary bypass and patient monitoring. Test results provided a high degree of confidence that the Cobra Catheter function as intended. Many of the tests allowed direct comparison to currently marketed devices.

Clinical studies were conducted to demonstrate the safety and effectiveness of the Cobra Catheter. Feasibility and randomized clinical studies demonstrated that adequate venous oxygen saturation was delivered with differential temperatures indicating the preservation of cerebral autoregulation. Thus, the Cobra Catheter is substantially equivalent to predicate arterial return devices with regard to intended use, technological characteristics and device performance.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an abstract human figure with three flowing lines representing the human form.

OCT 6 - 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Cardeon Corp. c/o Ms. Jane Beggs Regulatory Affairs 10600 N. Tantau Avenue Cupertino, CA 95014

Re: K042156

Cardeon® Cobra™ Catheter Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Catheter, Cannula, or Tubing Regulatory Class: Class II (two) Product Code: DWF Dated: August 9, 2004 Received: August 10, 2004

Dear Ms. Beggs;

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass based or to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

## Page 2 - Ms. Jane Beggs

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be actived that I broundevice complies with other requirements of the Act that I Drinas Intact a and regulations administered by other Federal agencies. You must or any I eath all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI It Fart 6077; accems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product in the will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premaired predicated device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): _________

Page 1 of 1

Device Name: Cardeon® Cobra™ Catheter

Indications for Use:

The Cardeon Cobra Catheter is indicated for use in open chest surgery on cardiopulmonary bypass up to 6 hours. It is used to provide separated perfusion of the arch and distal thoracic aorta with temperature differential, if desired.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

Over-the Counter Use __________

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number k042156

OR

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K042156](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K042156)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
