# MODIFICATION TO CARDEON ASCENDING BALLOON CANNULA (ABC) (K022042)

_Cardeon Corp. · DWF · Sep 3, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K022042

## Device Facts

- **Applicant:** Cardeon Corp.
- **Product Code:** [DWF](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF.md)
- **Decision Date:** Sep 3, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Cardeon® Ascending Balloon Cannula is intended to perfuse the aorta during open chest (sternotomy) procedures on cardiopulmonary bypass up to 6 hours. The occlusion balloon isolates the heart from the ascending aorta when inflated. The ABC also allows delivery of antegrade cardioplegia and venting of the aortic root.

## Device Story

Cardeon Ascending Balloon Cannula (ABC) is a vascular cannula for cardiopulmonary bypass (CPB) procedures. Device features an integrated occlusion balloon; fluid inflation of balloon isolates heart from ascending aorta, preventing blood backflow into surgical field. Additional lumens facilitate antegrade cardioplegia delivery and aortic root venting. Used in OR by cardiac surgeons during sternotomy. Connects to standard CPB circuits. Provides aortic perfusion, cardioplegia delivery, and venting in a single device, simplifying surgical workflow and reducing need for separate cross-clamps.

## Clinical Evidence

Bench testing and in vitro/in vivo performance testing demonstrated safety and effectiveness comparable to predicate arterial return cannulae and aortic occlusive balloons. No clinical trial data reported.

## Technological Characteristics

Vascular cannula with integrated inflatable occlusion balloon and auxiliary lumens for cardioplegia and venting. Connects to standard CPB circuits. Mechanical device; no software or electronic components.

## Regulatory Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) SUMMARY

## K022042 510(k) number:

This summary of information is being provided in accordance with 21 CFR 807.92(a).

| Applicant Information:                | Cardeon Corporation<br>10161 Bubb Road<br>Cupertino, CA 95014              |
|---------------------------------------|----------------------------------------------------------------------------|
| Contact Person:                       | Jane Beggs<br>Regulatory Affairs<br>Cardeon Corporation                    |
| Date Summary Prepared:                | 23 August 2002                                                             |
| Device Trade Name:                    | Cardeon® Ascending Balloon Cannula (ABCTM)                                 |
| Device Common Name<br>Regulation No.: | Catheter, cannula and tubing, vascular, cardiopulmonary<br>bypass 870.4210 |
| Classification / Code:                | Class II / DWF                                                             |

Indications for Use: The Cardeon® Ascending Balloon Cannula is intended to perfuse the aorta during open chest (sternotomy) procedures on cardiopulmonary bypass up to 6 hours. The occlusion balloon isolates the heart from the ascending aorta when inflated. The ABC also allows delivery of antegrade cardioplegia and venting of the aortic root.

Summary of Substantial Equivalence: The Ascending Balloon Cannula is substantially equivalent to currently marketed devices used to directly cannulate and perfuse from a general cardiac surgery patients undergoing coronary artery bypass grafting or valve replacement and/or repairs on CPB via sternotomy. Aortic occlusion is achieved by the fluid inflation of an integrated balloon thereby preventing the backflow of blood into the surgical field. Additional lumens allow delivery of antegrade cardioplegia or venting of the aortic root. The subject and predicate devices use standard attachments for connection(s) to cardiopulmonary bypass circuit. Substantial equivalence is supported through comparison with several marketed devices with the same indications for use, including arterial cannulae integral occlusive balloon and external cross clamps. Differences between the Ascending Balloon Cannula and other devices do not raise any new issues of safety and effectiveness.

Based on comparisons to currently marketed devices and performance testing of the subject device. the Cardeon Ascending Balloon Cannula is substantially equivalent to predicate devices with regard to intended use, indications for use, device performance and technological characteristics,

Product Testing: The determination of substantial equivalence was also based on an assessment of device biocompatibility, in vitro and in vivo performance. Results of product testing demonstrated that the Ascending Balloon Cannula functions as safely and effectively as predicate arterial return cannulae and aortic occlusive balloons.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized eagle or bird symbol, depicted with flowing lines to represent its wings and body.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

3 2002 SEP

Cardeon Corporation c/o Ms. Jane Beggs VP. Regulatory Affairs 10161 Bubb Road Cupertino, CA 95014

Re: K022042

Trade Name: Cardeon® Ascending Balloon Cannula (ABCTM) Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Catheter, Cannula, Tubing Regulatory Class: Class II (two) Product Code: DWF Dated: June 13, 2002 Received: June 14, 2002

Dear Ms. Beggs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Ms. Jane Beggs

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Noa Bram D. Zuckerman, M.D.

ABram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): ΚύλλΟΥ Σ

Page 1 of 1

Device Name: Cardeon Ascending Balloon Cannula (ABC™)

Indications for Use:

The Cardeon Ascending Balloon Cannula is intended to perfuse the aorta during open chest (sternotomy) procedures on cardiopulmonary bypass up to 6 hours. The occlusion balloon isolates the heart from the ascending aorta when inflated. The ABC also allows delivery of antegrade cardioplegia and venting of the aortic root.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X OR Over-the Counter Use *_*

(Division Sign-Off)
Division of Cardiovascular and Respiratory Devices

510(k) Number K022042

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K022042](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWF/K022042)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com