In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the heart-lung machine and it's components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours.
Device Story
HLM Tubing Sets with Bioline Coating serve as conduits in extracorporeal perfusion circulation systems; connect patient to heart-lung machine components during open heart surgery. Device facilitates blood oxygenation and carbon dioxide removal. Used in clinical settings by perfusionists or surgical staff. Bioline coating applied to tubing and connectors to improve physical surface properties. Single-use only; restricted to six-hour duration. Output is the physical transport of blood within the bypass circuit; enables life support during cardiac procedures.
Clinical Evidence
Bench testing only. Evaluation included integrity, performance, biocompatibility, and sterility testing to demonstrate substantial equivalence to predicate devices.
Technological Characteristics
Extracorporeal tubing and connectors featuring Bioline surface coating. Designed for single-use in cardiopulmonary bypass circuits. Mechanical device; no energy source or software components.
Indications for Use
Indicated for patients undergoing extracorporeal circulation during cardiopulmonary bypass procedures lasting six hours or less.
Regulatory Classification
Identification
A cardiopulmonary bypass pump tubing is polymeric tubing which is used in the blood pump head and which is cyclically compressed by the pump to cause the blood to flow through the cardiopulmonary bypass circuit.
Quadrox D Diffusion Membrane Oxygenator with Bioline Coating (K071774)
Related Devices
K090533 — HLM TUBING SET WITH SOFTLINE COATING · Maquet Cardiopulmonary, AG · Dec 18, 2009
K061832 — JOSTRA HLM TUBING SETS WITH SAFELINE COATING · Maquet Cardiopulmonary, AG · Aug 11, 2006
K113845 — TUBING, CONMNECTORS, AND ACCESSORIES WIT BALANCE BIOSURFACE · Medtronic, Inc. · Jan 25, 2012
K030077 — GISH TUBING AND CONNECTORS WITH GBS COATING · Gish Biomedical, Inc. · Mar 14, 2003
K053025 — JOSTRA HLM TUBING SET · Maquet Cardiopulmonary, AG · Nov 10, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
SEP 1 2008
K080592
Maquet Cardiopulmonary AG
Phone: (011) 49 7478 921- 151
(011) 49 7478 921- 400
Hechinger Strasse 38 72145 Hirrlingen, Germany
Katrin Schwenkglenks
February 22, 2008
# 510(k) SUMMARY
SUBMITTER:
## CONTACT PERSON:
DATE PREPARED:
HLM Tubing Sets with Bioline Coating DEVICE TRADE NAME:
Fax:
COMMON/USUAL NAME Custom Tubing Pack
CLASSIFICATION NAME
Bypass Cardiopulmonary Vascular Catheter, Cannula, or Tubing; Cardiopulmonary Bypass Adaptor, Stopcock, Manifold, or Fitting; Cardiopulmonary Bypass Pump Tubing.
## PREDICATE DEVICES OR LEGALLY MARKETED DEVICES
Jostra HLM Tubing Sets (K053025)
Quadrox D Diffusion Membrane Oxygenator with Bioline Coating (K071774)
## DEVICE DESCRIPTION / INDICATONS FOR USE STATEMENT
The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile, and bulk packed. Tubing sets that are sold sterile are not to be re-sterilized by the user.
In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the heart-lung machine and it's components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours.
· Maquet Cardiopulmonary AG, Hirrlingen, Germany
{1}------------------------------------------------
#### Special 510(k) Device Modification HLM Tubing Set with Bioline Coating
The Bioline Coating improves the physical surface properties of products for the extracorporeal circulation system.
#### STATEMENT OF TECHNICAL CHARACTERISTICS COMPARISON
The HLM Tubing Set - Bioline Coated has the same intended use, design, principals of operation, and performance as the uncoated Jostra HLM Tubing Set. The only difference is the application of the Bioline Coating to the tubing and connectors.
#### DETERMINATION OF SUBSTANTIAL EQUIVALENCE
Evaluation on safety and effectiveness was executed to demonstrate that the HLM-Tubing Set with Bioline Coating described in this submission is substantially equivalent to the Jostra HLM-Tubing Set as a custom tubing pack and to the Quadrox D Diffusion Membrane Oxygenator with Bioline Coating with regards to the Bioline Coating.
The following areas have been evaluated:
- Integrity
- Performance
- Biocompatibility
- Sterility
## CONCLUSION
The data given demonstrate that the HLM-Tubing Set with Bioline Coating is substantially equivalent to the named predicate devices which hold currently market clearance.
Maquet Cardiopulmonary AG, Hirrlingen, Germany
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved lines.
SEP 1 1 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Maquet Cardiopulmonary AG c/o Ms. Katrin Schwenkglenks Regulatory Affairs Manager Hechinger Strasse 38 72145 Hirrlingen, Germany
Re: K080592
HLM Tubing Set with Bioline Coating Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing Regulatory Class: Class II (two) Product Code: DWE, DWF, DTL Dated: August 29, 2008 Received: September 4, 2008
Dear Ms. Schwenkglenks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Katrin Schwenkglenks
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
er R. Vo. Amer
o Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): _ KO80592 Device Name: HLM Tubing Set with Bioline Coating
Indications for Use:
The HLM Tubing Sets with Bioline Coating are designed to be used in extracorporeal circulation during cardiopulmonary bypass procedures lasting six hours or less.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Donna R. Wehner
---
(Division Sign-Off)
Division of Cardiovascular Devices
Page *_ of _*
10(k) Number_K080592 (Posted November 13, 2003)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.