COBE CENTURY PERFUSION PUMP

K960974 · Cobe Cardiovascular, Inc. · DWB · Jul 14, 1997 · Cardiovascular

Device Facts

Record IDK960974
Device NameCOBE CENTURY PERFUSION PUMP
ApplicantCobe Cardiovascular, Inc.
Product CodeDWB · Cardiovascular
Decision DateJul 14, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4370
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The COBE Century Perfusion Pump is intended for use in cardiopulmonary surgical procedures requiring pumping of fluids in an extracorporeal circuit for periods up to 6 hours.

Device Story

COBE Century Perfusion Pump is a peristaltic roller pump for extracorporeal circuits during cardiopulmonary surgery. Device uses tube-occluding rollers to move fluids; positive displacement provides flow rates up to 10 LPM. Operated by perfusionists in clinical settings; features front panel with membrane switches and displays. System includes chassis with control cards and power supply. Device supports master/slave operation with another pump; integrates with COBE Computerized Perfusion Controller (CPC) for pressure/volumetric/pulse control and COBE Air Emboli Protection System (AEPS) for safety. Modifications from predicate include improved user interface and enhanced overspeed protection circuit with component failure detection. Output allows clinicians to maintain extracorporeal circulation; benefits patient by providing controlled, reliable blood flow during bypass.

Clinical Evidence

No clinical testing was performed. Evidence is based on bench testing of functional and user interface specifications.

Technological Characteristics

Peristaltic roller pump; accommodates PVC and silicone rubber tubing (1/8" to 5/8" ID). Features membrane switch interface, digital displays, and overspeed protection circuit. Connectivity includes master/slave synchronization and integration with external controllers (CPC, AEPS). Powered by internal supply. Software-controlled speed set-point circuitry with safety monitoring.

Indications for Use

Indicated for patients undergoing cardiopulmonary surgical procedures requiring extracorporeal fluid pumping for up to 6 hours.

Regulatory Classification

Identification

A roller-type cardiopulmonary bypass blood pump is a device that uses a revolving roller mechanism to pump the blood through the cardiopulmonary bypass circuit during bypass surgery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K960974 JUL 14 1997 # 510(k) SUMMARY SUBMITTER: COBE Cardiovascular, Inc. 14401 W. 65th Way Arvada, CO 80004 CONTACT PERSON: Mary L. Armstrong Phone: (303) 467-6521 Fax: (303) 467-6525 DATE PREPARED: March 31, 1997 DEVICE TRADE NAME: COBE® Century™ Perfusion Pump COMMON NAME: Peristaltic perfusion blood pump CLASSIFICATION NAME: Roller-type cardiopulmonary bypass blood pump and Cardiopulmonary bypass heart-lung machine console PREDICATE DEVICE: COBE Precision Blood Pump DEVICE DESCRIPTION: The COBE Century Perfusion Pump is the principal component of the COBE Century Perfusion System. It is a peristaltic-type roller pump that functions by tube-occluding rollers that move along a piece of tubing. As the pump head rotates, the rollers draw fluid through the tubing. Pumped fluids are contained within the tubing and have no contact with the pump. The pump has positive displacement over a wide range of flow rates and delivery pressures. The COBE Century Perfusion Pump consists of: - a front panel containing displays and operational switches - a pump head with the roller pump mechanism and shields - a chassis containing control cards, a power supply, and connectors for external devices COBE® is a registered name of COBE Laboratories, Inc. Century™ is a trademark of COBE Laboratories, Inc. {1} # INDICATIONS FOR USE: The COBE Century Perfusion Pump is intended for use in cardiopulmonary surgical procedures requiring pumping of fluids in an extracorporeal circuit for periods up to 6 hours. # TECHNOLOGICAL CHARACTERISTICS: The COBE Century Perfusion Pump is a modification of its predicate, the COBE Precision Blood Pump. As such, it retains most of the technological characteristics of its predicate. Through software and hardware modifications, the pump has an improved user interface and an improved overspeed protection circuit. Updating of the user interface involved replacing most of the rotary switches with membrane switches, replacing the front panel that holds the switches, and providing additional displays. Improving the overspeed protection circuit involved increasing overspeed sensitivity at low speeds, adding a fixed, maximum speed limit, and implementing a safety function which detects a component failure in the speed set-point circuitry. The COBE Century Perfusion Pump and its predicate are equipped with universal tubing clamps, which allow the pumps to accommodate a wide range of tubing sizes and types with a minimum of hardware changes. For non-continuous pumping applications only, polyvinylchloride (PVC) tubing of 1/8 inch to 1/4 inch inside diameter (ID) by 1/16 inch wall minimum can be accommodated. For continuous pumping, PVC tubing of 1/4 inch (ID) by 3/32 inch wall to 5/8 inch ID by 1/8 inch wall can be accommodated. For continuous and non-continuous pumping, silicone rubber tubing with 3/16 inch ID by 1/16 inch wall to 5/8 inch ID by 1/8 inch wall can be accommodated. Flow rate (Liters Per Minute) through the pumps is based on pump speed (Revolutions Per Minute) and tubing ID. Flow rates can range from .01 LPM (3/16 inch ID and 1 RPM) to over 10 LPM (for larger sizes of tubing and higher speeds). For some sizes of tubing, two segments of tubing can be placed in the pump head at the same time. The COBE Century Perfusion Pump is not—nor is it intended to be—interchangeable with its predecessor, the COBE Precision Blood Pump. However, the COBE Century Perfusion Pump and its predicate device are each compatible with certain other devices: 1) The COBE Century Perfusion Pump can be run in master/slave mode with another COBE Century Perfusion Pump. Similarly, the COBE Precision Blood Pump can be run in master/slave mode with another COBE Decision Blood Pump. 2) The COBE Century Perfusion Pump can be operated in conjunction with the COBE Computerized Perfusion Controller (CPC). Similarly, the COBE Precision Blood Pump can be operated in conjunction with the COBE Computerized Perfusion Controller (CPC). The CPC can be used to pressure-limit either pump, volumetrically control either pump, and/or pulse either pump. 3) The COBE Century {2} Perfusion Pump can be used with the COBE Air Emboli Protection System (AEPS). Similarly, the COBE Precision Blood Pump can be used with the COBE Air Emboli Protection System (AEPS). The AEPS is designed to provide protection against air embolism during cardiopulmonary bypass surgery by stopping the pump if specific conditions are sensed. ## NONCLINICAL TEST RESULTS: The COBE Century Perfusion Pump was tested to assure that it met its functional specifications and its user interface specification. It successfully met all specifications at the system test level, where every system requirement was addressed by a corresponding test or tests. ## CLINICAL TEST RESULTS: No clinical testing was performed. ## CONCLUSIONS: The COBE Century Perfusion Pump is a modification of the COBE Precision Blood Pump. Through hardware and software modifications, the Century Perfusion Pump has an improved user interface and an improved overspeed protection circuit. The COBE Century Perfusion Pump is substantially equivalent to the COBE Precision Blood Pump. {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Mary L. Armstrong Regulatory Affairs Manager COBE Cardiovascular, Inc. 14401 W. 65th Way Arvada, Colorado 80004-3599 JUL 14 1997 Re: K960974 COBE® Century™ Perfusion Pump Regulatory Class: II (Two) Product Code: 74 DWB Dated: April 11, 1997 Received: April 14, 1997 Dear Ms. Armstrong: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4} Page 2 - Ms. Mary L. Armstrong This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html." Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5} PG. 10/16 ## 2. Indications For Use 510(k) Number (If known): K960974 Device Name: COBE® Century™ Precision Blood Pump Indications For Use: The COBE Century™ Precision Blood Pump is intended for use in cardiopulmonary surgical procedures requiring pumping of fluids in an extracorporeal circuit for periods up to 6 hours. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K960974 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...