SARNS 9000 UNIVERSAL ROLLER PUMP

K953904 · 3M Health Care, Ltd. · DWB · Apr 19, 1996 · Cardiovascular

Device Facts

Record IDK953904
Device NameSARNS 9000 UNIVERSAL ROLLER PUMP
Applicant3M Health Care, Ltd.
Product CodeDWB · Cardiovascular
Decision DateApr 19, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4370
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sarns 9000 Universal Roller Pump is indicated for use in extracorporeal circulation of blood for arterial regional perfusion, and cardiopulmonary bypass procedures only, when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Device Story

Microprocessor-controlled 2-roller peristaltic pump; provides extracorporeal blood circulation for cardiopulmonary bypass and arterial regional perfusion. Input: electrical power from Sarns 9000 Perfusion System; mechanical rotation of rollers. Operation: adjustable occlusion; flows up to 9.99 L/min; speeds up to 250 RPM. Output: controlled blood flow. Used in clinical settings (e.g., OR) by qualified perfusionists. Healthcare providers monitor pump speed/flow to manage patient circulation during bypass. Benefits: enables life-sustaining blood circulation during cardiac/vascular procedures.

Clinical Evidence

Bench testing only. No clinical data provided. Performance equivalence established through nonclinical testing and EMC evaluation.

Technological Characteristics

Microprocessor-controlled 2-roller peristaltic pump. Circuit boards utilize Surface Mount Technology (SMT) components. Powered by Sarns 9000 Perfusion System. Adjustable occlusion mechanism. Flow capacity up to 9.99 L/min; max speed 250 RPM.

Indications for Use

Indicated for patients requiring extracorporeal circulation of blood for arterial regional perfusion and cardiopulmonary bypass procedures. Intended for use by qualified, experienced perfusionists.

Regulatory Classification

Identification

A roller-type cardiopulmonary bypass blood pump is a device that uses a revolving roller mechanism to pump the blood through the cardiopulmonary bypass circuit during bypass surgery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} # Section 2 - Summary & Certification ## 2.1 510(k) Summary A 510(k) Summary of Safety and Effectiveness is on the next page. K953904 APR 19 1996 3M Confidential Page 5 of 541 {1} 3M Health Care Sarns™ and CDI™ Products 6200 Jackson Road PO Box 1247 Ann Arbor, MI 48106-1247 USA 800 521 2818 Tel 313 663 4145 Tel 313 663 7981 Fax 3M K353904 August 9, 1995 # 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K353904. This 510(k) is for the Sarns 9000 Universal Roller Pump, a "Roller-type cardiopulmonary bypass blood pump" and "Cardiopulmonary bypass pump speed control". The new Sarns 9000 Universal Roller Pump is substantially equivalent to the existing Sarns 9000 Universal Roller Pump (K871131). The Sarns 9000 Universal Roller Pump is a microprocessor controlled 2-roller peristaltic pump with adjustable occlusion. The pump is capable of flows up to 9.99 L/min (depending on pump head tubing used) at pump speeds up to 250 RPM. The Sarns 9000 Universal Roller Pump is dependent on the Sarns 9000 Perfusion System to provide power. The Sarns 9000 Universal Roller Pump is indicated for use in extracorporeal circulation of blood for arterial regional perfusion, and cardiopulmonary bypass procedures only, when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment. The Sarns 9000 Universal Roller Pump has different technological characteristics from the predicate device. The design and the manufacturing process have been changed. The modified roller pump is manufactured using some Surface Mount Technology (SMT) components and processes to produce the circuit boards which contain the pump electronics. The electronic packaging of some of the components on these circuit boards is SMT instead of Through-Hole Technology (THT). Also, the software has undergone minor changes to remap the I/O addresses. The Sarns 9000 Universal Roller Pump has the same performance as the predicate device. Nonclinical tests were performed on the device to determine substantial equivalence. Electromagnetic Compatibility (EMC) tests were also performed to ensure equivalent or better results. Cheryl Rosenberg (313) 663-4145 Phone (313) 663-5062 Fax Page 6 of 541 {2} 3M Confidential Page 7 of 541 ## 2.2 Safety Literature Search I certify that I have conducted a reasonable search of all information known or otherwise available to me about the types and causes of safety and/or effectiveness problems that have been reported for blood roller pumps. I further certify that I am aware of the types of problems to which blood roller pumps are susceptible and that the following list of safety and/or effectiveness problems about blood roller pumps is complete and accurate: 1.) Air Emboli 2.) Hemolysis 3.) Loss of speed control 4.) Random pump stops Below is a bibliography of the materials upon which the above summary is based: 1.) Customer Field Input Log, various dates, Sarns, 3M Health Care 2.) Intermittent Failure of a Stockert/Shiley Multiflow Roller Pump Module, The Journal of Extra-Corporeal Technology, Volume 25, Number 3, 1993 3.) Medical Device Report M566956, M186992, M186044, M138277, various dates, Sarns, 3M Health Care 4.) Medical Device Report M494536, filed 4/21/94, Sarns, 3M Health Care 5.) Medical Device Report M444060, M123912, M111730, M101274, various dates, Sarns, 3M Health Care 6.) Medical Device Report M435457, M242746, M225052, M184140, M138914, M174191, various dates, Sarns, 3M Health Care 7.) Medical Device Report M377761, filed 2/26/93, Sarns, 3M Health Care 8.) Medical Device Report M125141, received 6/24/86, Cobe Labs, Inc. 9.) Medical Device Report M269021, M244886, M244883, M225527, M210562, M210551, M210545, M129763, various dates, Shiley, Inc. 10.) Medical Device Report M266724, M245762, M243742, M221182, various dates, Shiley, Inc. 11.) Medical Device Report M253241, M250682, M225194, M221859, various dates, Shiley, Inc. 12.) Medical Device Report M235614, M233235, various dates, Shiley, Inc. 13.) Medical Device Report M220822, M212362, M212346, M210217, M166630, M164988, M137248, various dates, Shiley, Inc. 14.) Medical Device Report M210544, received 8/17/90, Shiley, Inc. 15.) Medical Device Report M218408, received 12/10/90, Shiley, Inc. 16.) Medical Device Report M218946, received 12/17/90, Shiley, Inc. 17.) Medical Device Report M162834, received 1/5/89, Shiley, Inc. 18.) Medical Device Report M162828, M162829, M162830, M162831, M162832, M162833, M162836, M162838, M162839, M162840, M162841, M164234, M164858, M164860, M164850, M164934, M169675, various dates, Shiley, Inc. 19.) Medical Device Report M151200, received 2/26/88, Shiley, Inc. 20.) Medical Device Report M113582, received 11/5/92, Shiley, Inc. {3} # Bibliography (continued) 21.) **Medical Device Report** M149951, M154495, M158908, M162835, M162837, M166744, M169035, M181659, M183581, M183582, M201061, M203578, M204035, M204038, M204528, M206003, M206119, M206314, M206208, M209511, M214496, M218406, M218555, M218949, M218952, M218954, M220245, M220248, M220824, M221189, M221190, M222866, M225528, M225533, M232442, M231336, M231337, M231338, M235611, M235947, M238231, M248275, M243327, M244887, M245205, M245936, M245940, M247740, M247741, M249816, M249823, M252531, M262466, M268638, M269022, M269713, M272031, various dates, Shiley, Inc. 22.) **Medical Device Report** M364517, M311533, M310489, M308281, M306710, M306708, M306205, M283826, M282507, M273690, various dates, Sorin Biomedical, Inc. 23.) **Medical Device Report** M345060, M317126, various dates, Sorin Biomedical, Inc. 24.) **Medical Device Report** M344176, received 9/2/92, Sorin Biomedical, Inc. 25.) **Medical Device Report** M324153, received 9/30/92, Sorin Biomedical, Inc. 26.) **Medical Device Report** M321079, received 8/13/92, Sorin Biomedical, Inc. 27.) **Medical Device Report** M285491, received 4/29/92, Sorin Biomedical, Inc. 28.) **Medical Device Report** M283413, received 4/15/92, Sorin Biomedical, Inc. Printed name of person required to submit 510(k): Cheryl Rosenberg Signature of person required to submit 510(k): Cheryl Rosenberg Title of person submitting 510(k): Advanced Regulatory Affairs Coordinator Name of Company: 3M Health Care Date: August 9, 1995 3M Confidential Page 8 of 541
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...