The Levitronix CentriMag Primary Console is indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
Device Story
CentriMag Primary Console is a microprocessor-based controller for the CentriMag Extracorporeal Blood Pumping System. Device monitors system sensors, generates motor control signals, and provides operator displays and alarm functions via an alphanumeric screen. Console operates on single-phase AC power with an internal rechargeable battery for backup. Used in clinical settings to manage blood flow rates up to 9.9 LPM during extracorporeal circulatory support. Healthcare providers use the console to adjust pump parameters and monitor system status through menus. Output allows clinicians to maintain circulatory support during surgical procedures, benefiting patients by providing temporary extracorporeal bypass.
Clinical Evidence
Bench testing only. Performance characteristics were tested and compared against Levitronix performance specifications and the predicate device.
Technological Characteristics
Microprocessor-based pump speed control console. Features alphanumeric display, sensor monitoring, and alarm functions. Power: single-phase AC with internal rechargeable battery backup. Flow rate capacity: up to 9.9 LPM. Connectivity: standalone console.
Indications for Use
Indicated for patients requiring extracorporeal circulatory support via the CentriMag Extracorporeal Blood Pumping System for up to six hours, including cardiopulmonary bypass or procedures not requiring complete bypass (e.g., valvuloplasty, mitral valve reoperation, vena cava/aorta surgery, liver transplants).
Regulatory Classification
Identification
A cardiopulmonary bypass pump speed control is a device used that incorporates an electrical system or a mechanical system, or both, and is used to control the speed of blood pumps used in cardiopulmonary bypass surgery.
Predicate Devices
Levitronix CentriMag Console Software Version 2.02 (K081221)
{0}------------------------------------------------
KOr3340
### NOV 2 6 2008 Attachment 3 – 510(k) Summary
# 510(k) Summary for Levitronix CentriMag Primary Console
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807, and in particular \$807.92.
#### A. Application Information:
| Date Prepared: | November 11, 2008 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name & Address: | Levitronix LLC<br>45 First Avenue<br>Waltham, MA 02451 |
| Contact Person: | Susan K. Hamann<br>Regulatory Affairs Manager<br>Ph: (781) 466-6553<br>Fax: (781) 622-5090<br>e-mail: <span style="text-decoration:underline">shamann@levitronix.com</span> |
| Device Information: | |
| Trade or Proprietary Name: | Levitronix CentriMag® Primary Console |
| Common or Usual Name: | Cardiopulmonary Bypass Pump<br>Console |
| Classification Name: | Class II, DWA, 21 CFR - 870.4380<br>Control, Pump Speed, Cardiopulmonary<br>Bypass |
| Performance Standard: | Performance standards do not currently<br>exist for these devices. None |
#### C. Predicate Device:
Levitronix CentriMag Console Software Version 2.02 – 510(k) (K081221)
established under section 514 of the Food, Drug and Cosmetic Act.
в.
{1}------------------------------------------------
#### D. Device Description
The CentriMag Primary Console is a microprocessor-based device. The microprocessor generates the primary Motor control signal, monitors system sensors, generates front display outputs, and provides alarm functions. The microprocessor acquires the sensor data for use in generating operator displays and alarms. An alphanumeric screen is used to display monitored data, system options, and menus. Operator adjustable alarms and parameters are accessible via the system menus.
### ட். Intended Use
The Levitronix CentriMag Primary Console is indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
### ட Technological Characteristics
Levitronix CentriMag Primary Console has technological The characteristics similar to the predicate device.
The Levitronix CentriMag Primary Console is a microprocessor-based device. The microprocessor generates the primary Motor control signal, monitors system sensors, generates front display outputs, and provides alarm functions. The microprocessor acquires the sensor data for use in generating operator displays and alarms. An alphanumeric screen is used to displav monitored data.
The CentriMag Primary Console uses single phase AC power and is capable of a flow rate of up to 9.9 LPM. Each Primary Console contains a rechargeable internal battery that is capable of maintaining Primary Console functionality in the event of a loss of AC Power,
#### G. Comparison to Predicate Device
The Levitronix CentriMag Primary Console has indications for use which is substantially equivalent to the predicate device, is composed of the same or equivalent materials as the predicate device, has equivalent design features as the predicate device, and has functional characteristics which are the same or equivalent to those of the predicate device. Due to the
{2}------------------------------------------------
equivalency of indications for use, materials of composition, design features, and functional characteristics, the device raises no new safety or effectiveness issues.
#### H. Summary of Performance Data
The performance characteristics of the Levitronix CentriMag Primary Console were tested and compared with Levitronix performance specifications established for the device and with the commercially available predicate device.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 6 2008
Levitronix LLC c/o Susan Hamann Regulatory Affairs Manager 45 First Avenue Waltham, MA 02451
> K083340 Levitronix CentriMag Primary Console Regulation Number: 21 CFR 870.4380 Regulation Name: Cardiopulmonary Bypass Pump Speed Control Regulatory Class: Class II Product Code: DWA Dated: November 11, 2008 Received: November 12, 2008
Dear Ms. Hamann:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Susan Hamann
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
uma R. Vachus
A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Attachment 1 - Indications for Use Statement
| Applicant: | Levitronix LLC |
|---------------------------|---------------------------------------------------------------|
| 510(k) Number (if known): | K083340 |
| Device Name: | CentriMag Primary Console (with V3.00 Application Software) |
## Indications for Use:
The Levitronix CentriMag Primary Console is indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R. Vmmmm
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_k o83340
Page 1 of 1
Levitronix LLC
Special 510(k): Device Modification CentriMag® Primary Console
Page 25 of 138
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.