# LEVITRONIX CENTRIMAG PRIMARY CONSOLE (K053630)

_Levitronix, LLC · DWA · Jan 26, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWA/K053630

## Device Facts

- **Applicant:** Levitronix, LLC
- **Product Code:** [DWA](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWA.md)
- **Decision Date:** Jan 26, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4380
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Levitronix CentriMag Primary Console is indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).

## Device Story

Microprocessor-based console for CentriMag Extracorporeal Blood Pumping System; monitors system sensors; generates motor control signals; provides operator displays and alarms via alphanumeric screen. Operates on single-phase AC power with internal rechargeable battery backup. Used in clinical settings for extracorporeal circulatory support up to six hours. Healthcare providers use console to monitor pump performance and adjust parameters via menus; output informs clinical management of blood flow during bypass or circulatory support procedures.

## Clinical Evidence

Bench testing only; performance characteristics compared against established specifications and predicate device.

## Technological Characteristics

Microprocessor-based control unit; single-phase AC power; internal rechargeable battery; alphanumeric display; flow rate up to 9.9 LPM.

## Regulatory Identification

A cardiopulmonary bypass pump speed control is a device used that incorporates an electrical system or a mechanical system, or both, and is used to control the speed of blood pumps used in cardiopulmonary bypass surgery.

## Predicate Devices

- Levitronix CentriMag Console ([K020271](/device/K020271.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K053630

# Attachment 3 - 510(k) Summary

# 510(k) Summary for Levitronix CentriMag Primary Console

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807, and in particular §807.92.

#### Application Information: A.

| Date Prepared:              | December 27, 2005                                                                                                                               |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name & Address: | Levitronix LLC<br>45 First Avenue<br>Waltham, MA 02451                                                                                          |
| Contact Person:             | Farzad Parsaie<br>V.P., RA/QA<br>Ph:<br>(781) 622-5075<br>Fax:<br>(781) 622-5090<br>e-mail: fparsaie@levitronix.com                             |
| Device Information:         |                                                                                                                                                 |
| Trade or Proprietary Name:  | Levitronix CentriMag® Primary Console                                                                                                           |
| Common or Usual Name:       | Cardiopulmonary Bypass Pump<br>Console                                                                                                          |
| Classification Name:        | Class II, DWA, 21 CFR - 870.4380<br>Control, Pump Speed, Cardiopulmonary<br>Bypass                                                              |
| Performance Standard:       | Performance standards do not currently<br>exist for these devices. None<br>established under section 514 of the<br>Food, Drug and Cosmetic Act. |

### B.

#### C. Predicate Device:

Levitronix CentriMag Console (K020271)

{1}------------------------------------------------

#### Device Description D.

The CentriMag Primary Console is a microprocessor-based device. The microprocessor generates the primary Motor control signal, monitors system sensors, generates front display outputs, and provides alarm functions. The microprocessor acquires the sensor data for use in generating operator displays and alarms. An alphanumeric screen is used to display monitored data, system options, and menus. Oneator adjustable alarms and parameters are accessible via the system menus.

#### ய் Intended Use

The Levitronix CentriMag Primary Console is indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).

#### F. Technological Characteristics

The CentriMag Primary Console has Levitronix technological characteristics similar to the predicate device.

The Levitronix CentriMag Primary Console is a microprocessor-based device. The microprocessor generates the primary Motor control signal, monitors system sensors, generates front display outputs, and provides alarm functions. The microprocessor acquires the sensor data for use in generating operator displays and alarms. An alphanumeric screen is used to display monitored data.

The CentriMag Primary Console uses single phase AC power and is capable of a flow rate of up to 9.9 LPM. Each Primary Console contains a rechargeable internal battery that is capable of maintaining Primary Console functionality in the event of a loss of AC Power.

#### G. Comparison to Predicate Device

The Levitronix CentriMag Primary Console has indications for use which is substantially equivalent to the predicate device, is composed of the same or equivalent materials as the predicate device, has equivalent design features as the predicate device, and has functional characteristics whigh are the same or equivalent to those of the predicate device. Due to the

{2}------------------------------------------------

equivalency of indications for use, materials of composition, design features, and functional characteristics, the device raises no new safety or effectiveness issues.

#### H. Summary of Performance Data

The performance characteristics of the Levitronix CentriMag Primary Console were tested and compared with Levitronix performance specifications established for the device and with the commercially available predicate device.

{3}------------------------------------------------

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 6 2006

Levitronix LLC. c/o Mr. Farzad Parsale V.P., RA/QA 45 First Avenue Waltham, MA 02451

Re: K053630

CentriMag® Primary Console (with v200 Application Software onboard) Regulation Number: 21 CFR 870.4380 Regulation Name: Cardiopulmonary Bypass Pump Speed Control Regulatory Class: Class II (Two) Product Code: DWA Dated: December 27, 2005 Received: January 4, 2006

Dear Mr. Parsale:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Farzad Parsale

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Duna R. Lochner

JA Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Attachment 1 - Indications for Use Statement

Applicant: Levitronix LLC

510(k) Number (if known): __ K 6 5 3 ( 3 6

Device Name: CentriMag® Primary Console (with v2.00 Application Software onboard)

### Indications for Use:

The Levitronix CentriMag Primary Console is indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duana R. Vochner

(Division Sign-Off) Division of Cardiovascular Devices

510(K) Number_K653630

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWA/K053630](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DWA/K053630)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).