K223161 · Abiomed, Inc. · DTZ · Feb 23, 2023 · Cardiovascular
Device Facts
Record ID
K223161
Device Name
OXY-1 System
Applicant
Abiomed, Inc.
Product Code
DTZ · Cardiovascular
Decision Date
Feb 23, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OXY-1 System is intended to be used for extracorporeal circulation. The OXY-1 System pumps, oxygenates and removes carbon dioxide from blood during cardiopulmonary bypass up to 6 hours in duration
Device Story
The OXY-1 System provides extracorporeal circulation for cardiopulmonary bypass support. The system comprises a disposable pump and oxygenator, a pump driver, blood tubing, and a console for pump control and gas flow management. It operates by pumping blood through a hollow fiber oxygenator (polymethylpentene membrane) to facilitate gas exchange (oxygenation and CO2 removal). The device is used in clinical settings; operated by trained healthcare professionals. The console provides real-time monitoring of blood flow, pump RPM, and alarm status (bubble, low battery, system failure). An emergency hand crank is provided for manual drive. The system simplifies bedside operation and reduces equipment footprint. Clinical benefit is derived from maintaining patient oxygenation and CO2 removal during short-term bypass procedures.
Clinical Evidence
Bench testing only. Performance data includes hemolysis testing (average NIH 0.0210 mg/100L), pressure drop (average 38 mmHg at 5 Lpm), and gas transfer testing conducted per FDA guidance for cardiopulmonary bypass oxygenators.
Indicated for patients requiring extracorporeal circulation, oxygenation, and carbon dioxide removal during cardiopulmonary bypass procedures for up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
February 23, 2023
Abiomed Inc. Ken Ryder Sr. Director, Regulatory Affairs 22 Cherry Hill Drive Danvers, Massachusetts 01923
Re: K223161
Trade/Device Name: OXY-1 System Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ, KFM, DWA, DWF Dated: January 25, 2023 Received: January 27, 2023
Dear Ken Ryder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223261
Device Name OXY-1 System
Indications for Use (Describe)
The OXY-1 System is intended to be used for extracorporeal circulation. The OXY-1 System pumps, oxygenates and removes carbon dioxide from blood during cardiopulmonary bypass up to 6 hours in duration
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 201 Subpart D) </div> | <div> <span></span>Over-The-Counter Use (21 CFR 300) </div> |
|---------------------------------------------------------------------------|----------------------------------------------------------------|
|---------------------------------------------------------------------------|----------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Abiomed OXY-1 System 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807.92.
## A. Application Information:
| Date Prepared: | October 06, 2022 |
|-----------------------------|--------------------------------------------|
| Submitter's Name & Address: | ABIOMED, Inc. |
| | 22 Cherry Hill Drive |
| | Danvers, MA 01923 |
| Contact Person: | J. Kenneth Ryder |
| | Senior Director, Global Regulatory Affairs |
| | Ph: 978-646-1707 |
| | E-mail: kryder@abiomed.com |
#### B. Device Information:
| Trade or Proprietary Name: | OXY-1 System |
|----------------------------|-------------------------------------------------------------------------|
| Common or Usual Name: | OXY-1 System |
| FDA Classification: | Class II, 21 CFR 870.4350 |
| Regulation Description: | Cardiopulmonary bypass oxygenator |
| Product Code: | DTZ - Oxygenator, Cardiopulmonary Bypass |
| | KFM - Pump, Blood, Cardiopulmonary Bypass, Non-Roller<br>Type |
| | DWA - Control, Pump Speed, Cardiopulmonary Bypass |
| | DWF - Catheter, Cannula and Tubing, Vascular,<br>Cardiopulmonary Bypass |
#### C. Predicate Device:
The primary predicate device was the OXY-1 System, which is cleared under K200109.
#### D. Device Description:
The Abiomed OXY-1 System provides extracorporeal circulation for cardiopulmonary bypass support for up to six hours in duration. The OXY-1 System includes:
- A Disposable pump and oxygenator .
- A Pump Driver (blood pump) ●
- Blood tubing .
- A Console for controlling the pump and managing gas flow .
These components are designed to operate together, simplify operation, and reduce the overall equipment footprint at the bedside.
{4}------------------------------------------------
#### E. Indications for Use:
#### INDICATIONS FOR USE:
The OXY-1 System is intended to be used for extracorporeal circulation. The OXY-1 System pumps, oxygenates and removes carbon dioxide from blood during cardiopulmonary bypass up to 6 hours in duration.
## F. Technological Characteristics Comparison of Subject and Predicate Devices:
The subject device, the OXY-1 System (Configuration 2) is identical to the predicate device Intended Use/Indications for Use, general system components, sterilization, and mechanism of action. Differences between the subject device and the predicate device were all determined to be minor, have no adverse impact on safety or effectiveness, and raise no different questions of safety or effectiveness compared to the predicate device. The minor differences include the oxygenator effective surface area and the presence of a green indicator light on the pump driver.
| | Subject Device | Predicate Device |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | OXY-1 System (Configuration 2) | OXY-1 System (Configuration 1) |
| Oxygenator | | |
| Membrane | Polymethylpentene | Polymethylpentene |
| Housing | Polycarbonate | Polycarbonate |
| Potting | Urethane | Urethane |
| Heat Exchanger | No | No |
| Membrane Type | Hollow Fiber | Hollow Fiber |
| Effective Membrane<br>Surface Area | 2.3 m² | 2.4 m² |
| Gas Pathway | Two Gas Pathways | Two Gas Pathways |
| Oxygenator Geometry | Cylindrical | Cylindrical |
| Tubing Connectors | 3/8" | 3/8" |
| Priming Volume<br>(Oxygenator Only) | 285 ml | 285 ml |
| Sterile | SAL 10^-6 | SAL 10^-6 |
| Sterilization method | Ethylene Oxide | Ethylene Oxide |
| Performance Specifications (Oxygenator) | | |
| Blood Flow Rate | 0.5-5.0 Lpm | 0.5-5.0 Lpm |
| Gas Transfer | Tested per "Guidance for<br>Cardiopulmonary Bypass<br>Oxygenators 510(k) Submissions"<br>Dated November 13, 2000 | Tested per "Guidance for<br>Cardiopulmonary Bypass<br>Oxygenators 510(k) Submissions"<br>Dated November 13, 2000 |
| Hemolysis | Average NIH of 0.0210 mg/100L | Average NIH of 0.0210 mg/100L |
| Pressure Drop (at 5 Lpm) | Average of 38 mmHg | Average of 38 mmHg |
| Coupling | Magnetic | Magnetic |
| Energy Source | Battery / Mains | Battery / Mains |
| <b>Other Features</b> | | |
| Sweep Gas Source | Internal / External | Internal / External |
| Sighing Gas Source | Internal / External | Internal / External |
| <b>Software</b> | | |
| Constant RPM Mode | Yes | Yes |
| Constant Flow Mode | No | No |
| <b>User Interface/ Display</b> | | |
| Blood Flow | Yes | Yes |
| Pump RPM Settings | Yes | Yes |
| Internal Sweep Gas | Yes | Yes |
| Alarm Limit Settings | Yes | Yes |
| Blood Flow | Yes | Yes |
| Bubble | Yes | Yes |
| Low Battery | Yes | Yes |
| System Failure Alarms | Yes | Yes |
| Bubble Sensor | Yes | Yes |
| <b>Pump Driver</b> | | |
| LED Indicator Light | Yes | No |
| <b>Controller & Driver</b> | | |
| Flow Sensor | Yes | Yes |
| Emergency Drive Unit | Hand Crank | Hand Crank |
| RPM Range | 0-4500 RPM | 0-4500 RPM |
| <b>Blood Tubing</b> | | |
| Tubing | Polyvinyl Chloride | Polyvinyl Chloride |
{5}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.