Terumo Capiox NX19 Oxygenator with Reservoir (east orientation ), Terumo Capiox NX19 Oxygenator with Reservoir (west orientation ), Terumo Capiox NX19 Oxygenator (east orientation ), Terumo Capiox NX19 Oxygenator (west orientation)

K172071 · Terumo Cardiovascular Systems Corporation · DTZ · Mar 29, 2018 · Cardiovascular

Device Facts

Record IDK172071
Device NameTerumo Capiox NX19 Oxygenator with Reservoir (east orientation ), Terumo Capiox NX19 Oxygenator with Reservoir (west orientation ), Terumo Capiox NX19 Oxygenator (east orientation ), Terumo Capiox NX19 Oxygenator (west orientation)
ApplicantTerumo Cardiovascular Systems Corporation
Product CodeDTZ · Cardiovascular
Decision DateMar 29, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4350
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Capiox® NX19 Oxygenator with Integrated Arterial Filter and UltraPrime™ Technology is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of adult and small adult patients during cardiopulmonary bypass surgery. The integrated arterial filter is intended to filter non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit. The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device. The hardshell reservoir is used to store blood during extra-corporeal circulation from the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The 4-liter reservoir may be used for Vacuum Assisted Drainage procedures and Post-Operative Chest Drainage Procedures. The Capiox® NX19 Oxygenator with Integrated Arterial Filter and UltraPrime™ Technology is for use with patients when the required blood flow rate will not exceed 8.0 L/min. The Capiox® NX19 Oxygenator with Integrated Arterial Filter and UltraPrime™ Technology can be used in procedures lasting up to 6 hours.

Device Story

Device is an extracorporeal cardiopulmonary bypass oxygenator with integrated arterial filter, heat exchanger, and hardshell reservoir. Input: venous and cardiotomy blood flow. Operation: blood flows through porous fiber bundle for gas exchange via diffusion; heat exchanger (PET tubes) regulates blood temperature via external water bath; filter mesh provides mechanical entrapment of emboli/particulates; defoamers remove air bubbles. Output: oxygenated, temperature-regulated, filtered blood returned to patient. Used in OR by perfusionists/surgeons. Benefits: reduced priming volume (185mL) minimizes hemodilution; PMEA coating improves biocompatibility; integrated design simplifies circuit. Clinical decision-making: provides life-sustaining gas exchange and temperature control during cardiac surgery.

Clinical Evidence

Bench testing only. No clinical data. Performance evaluations included gas transfer, pressure drop, heat exchanger efficiency, hemolysis, air removal, filtration efficiency, mechanical integrity, and blood compatibility. Reservoir performance supported by data from predicate K151791.

Technological Characteristics

Materials: Polypropylene fibers, Polyethylene Terephthalate (PET) heat exchanger tubes, PMEA (400kDa) coating. Principle: Hollow fiber diffusion for gas exchange; mechanical entrapment for filtration. Flow: up to 8 L/min. Priming volume: 185mL. Sterilization: Ethylene oxide (SAL 10^-6). Biocompatibility: ISO 10993 compliant.

Indications for Use

Indicated for adult and small adult patients requiring gas exchange during cardiopulmonary bypass surgery, with blood flow rates up to 8.0 L/min and procedure durations up to 6 hours.

Regulatory Classification

Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a blue square and the full name written out to the right of the square. March 29, 2018 Terumo Cardiovascular Systems Corporation Adam Pickholtz Manager, Regulatory Affairs 125 Blue Ball Road Elkton, Maryland 21921 Re: K172071 Trade/Device Name: Terumo Capiox NX19 Oxygenator with Reservoir (east orientation ), Terumo Capiox NX19 Oxygenator with Reservoir (west orientation ), Terumo Capiox NX19 Oxygenator (east orientation ). Terumo Capiox NX19 Oxygenator (west orientation) Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary bypass oxygenator Regulatory Class: Class II Product Code: DTZ, DTR, DTM, DTN Dated: February 21, 2018 Received: February 26, 2018 Dear Adam Pickholtz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Nicole G. Ibrahim -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the Terumo logo. The logo consists of a red swoosh above the word "TERUMO" in green, block letters. The swoosh is curved and points to the right. # Section 4 - Indications for Use Capiox® NX19 Oxygenator with Integrated Arterial Filter 510(k) Number (if known): K172071 #### Capiox® NX19 Oxygenator with Integrated Arterial Filter Device Name: # Indications for Use: The Capiox® NX19 Oxygenator with Integrated Arterial Filter and UltraPrime™ Technology is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of adult and small adult patients during cardiopulmonary bypass surgery. The integrated arterial filter is intended to filter non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit. The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device. The hardshell reservoir is used to store blood during extra-corporeal circulation from the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The 4-liter reservoir may be used for Vacuum Assisted Drainage procedures and Post-Operative Chest Drainage Procedures. The Capiox® NX19 Oxygenator with Integrated Arterial Filter and UltraPrime™ Technology is for use with patients when the required blood flow rate will not exceed 8.0 L/min. The Capiox® NX19 Oxygenator with Integrated Arterial Filter and UltraPrime™ Technology can be used in procedures lasting up to 6 hours. | Prescription Use __XX__<br>(Part 21 CFR 801 Subpart D) | OROver-The-Counter Use<br>(Part 21 CFR 807 Subpart C) | |-------------------------------------------------------------------------------|-------------------------------------------------------| | (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF<br>NEEDED) | | Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Terumo. The logo consists of a red swoosh above the word "TERUMO" in green. The swoosh starts thin on the left and gets thicker as it moves to the right, arching over the word. The word "TERUMO" is in a bold, sans-serif font. K172071 #### Submitter Information This submission was prepared in June 2017 by: Garry A. Courtney, MBA, RAC Director, Regulatory Affairs Terumo Cardiovascular Systems Corporation 125 Blue Ball Road Elkton, MD 21921 Telephone: 1-800-262-3304, Ext. 7486 Fax: 410-392-7171 This submission was prepared for: Terumo Cardiovascular Systems Corporation 125 Blue Ball Road Elkton, MD 21921 Facility Registration No. 1124841 | Device Name &Classifications<br>Proprietary Name | Classification Name | Common Name | |------------------------------------------------------------|------------------------------------------------------------------|-----------------| | Capiox® NX19 Oxygenator<br>with Integrated Arterial Filter | Cardiopulmonary Bypass<br>Oxygenator (Code: DTZ) | Oxygenator | | | Cardiopulmonary Bypass<br>Blood Reservoir (Code: DTN) | Blood Reservoir | | | Cardiopulmonary Bypass<br>Heat Exchanger (Code: DTR) | Heat Exchanger | | | Cardiopulmonary Bypass<br>Arterial Line Blood Filter (Code: DTM) | Arterial Filter | #### Predicate Device The device submitted in this 510(k) maintains characteristics that are substantially equivalent in intended use, design, technology/principles of operation, materials and specifications to the following devices: Terumo® Capiox® FX25 Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir - K151791 # Reference Device: Sorin Inspire 8F Oxygenator with Integrated Arterial Filter and Hardshell Reservoir-K121536 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the Terumo company logo. The logo consists of a red swoosh above the company name, which is written in green, block letters. The swoosh is positioned above and to the left of the word "TERUMO." # Intended Use The Capiox® NX19 Oxygenator with Integrated Arterial Filter and UltraPrime™ Technology is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of adult and small adult patients during cardiopulmonary bypass surgery. The integrated arterial filter is intended to filter non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit. The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device. The hardshell reservoir is used to store blood during extra-corporeal circulation from the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The 4-liter reservoir may be used for Vacuum Assisted Drainage procedures and Post-Operative Chest Drainage Procedures. The Capiox® NX19 Oxygenator with Integrated Arterial Filter and UltraPrime™ Technology is for use with patients when the required blood flow rate will not exceed 8.0 L/min. The Capiox® NX19 Oxygenator with Integrated Arterial Filter and UltraPrime™ Technology can be used in procedures lasting up to 6 hours. The intended use and product indications of the new NX19 Oxygenator device are consistent with those of the predicate FX25 Advance Oxygenator device. # Principles of Operation and Technology The Capiox® NX19 Oxygenator utilizes porous fiber technology to facilitate the transfer of gases between a blood-phase environment and a gas-phase environment for the intent of satisfying the gas exchange needs of adult and small adult patients during cardiopulmonary bypass surgery. A fiber bundle offers the porous membrane surface to sufficiently permit the movement of gases through the walls of the hollow fibers via diffusion. The Capiox® NX19 Oxygenator device has an integrated heat exchanger that is comprised of Polyethylene Terephthalate tubes arranged in a cylindrical shape. The circulating blood contacts the exterior surface of these tubes. Temperature-controlled water distributed from an external water bath fills the interior of the tubes. The heat exchanger material permits heat transfer to occur across the walls of the heat exchanger tubes to achieve the necessary temperature of the circulating blood. The Capiox® NX19 Oxygenator device can be operated at flow rates up to 8 liters per minute (1/min) and at a minimum flow rate of 2.0 liters per minute (1/min) or 0.5 liters per minute (1/min) for up to 2 hours. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the Terumo company logo. The logo consists of a red curved line above the word "TERUMO" in green. The curved line is thicker on the left and tapers to a point on the right. The word "TERUMO" is in a bold, sans-serif font. The integrated arterial filter of the Capiox® NX19 Oxygenator device relies upon mechanical entrapment of particulates and emboli within the filter mesh as a means to remove those particulates from the blood. The operation and technology employed by the new NX19 Oxygenator device are consistent with the operation and technology of the predicate FX25 Advance Oxygenator device. # Design and Materials The primary differences in the design and materials between the Capiox® NX19 Oxygenator and the predicate device are: - · Reduction in the diameter of the polypropylene oxygenator fibers vs. the predicate; - · Higher Molecular weight of Polymethoxyethyl acrylate (PMEA) solution vs. the predicate; - · Difference in the materials of construct of the heat exchanger plastic vs stainless steel; - · The NX19 has a lower priming volume vs. the predicate device. # Reduction in diameter of the oxygenator fibers: The Capiox® NX19 Oxygenator utilizes an oxygenator fiber that has a diameter approximately 40% smaller than that of the predicate device. This smaller diameter allows for a larger percentage of the blood in the oxygenator to be in contact with the surface area of the fibers. By increasing this contact, similar gas exchange performance can be achieved with a lower blood volume. The smaller fiber diameter allows the NX19 oxygenator to have a surface of approximately 1.9m² compared to the approximately 2.5m2 surface area of the predicate device while still achieving equivalent gas exchange performance. # Higher Molecular weight of Polymethoxyethyl acrylate (PMEA) solution: The X-Coating™ solution utilized to coat the NX19 Oxygenator varies from the solution that is used to coat the predicate device. The NX19 X-Coating™ solution is Polymethoxyethyl acrylate (PMEA). This solution has a molecular weight of 400kDa. The solution for the predicate device is also PMEA solution, but has a lower molecular weight. The predicate X-Coating™ has a molecular weight of 85kDa. This increased molecular weight with the NX19 device allows for increased stability of the coating when applied to the polypropylene fibers of the oxygenator bundle. #### Difference in the materials of construct of the heat exchanger – plastic vs stainless steel: The material and the surface area of the heat exchanger in the Capiox® NX19 Oxygenator varies from the predicate device. The heat exchanger tubes in the NX19 Oxygenator are made of Polyethylene Terephthalate (PET) plastic material differs from the stainless steel bellows in the heat exchanger of the predicate device. The surface area of the PET heat exchanger tubes is 0.44m² in the NX19, which is a larger surface area than the 0.2m² in the predicate FX25. This increased surface area ensures that adequate (equivalent) heat exchange is achieved. #### The NX19 has a lower priming volume vs. the predicate device: The prime volume of the Capiox® NX19 Oxygenator is 185mL, which is lower than the 260mL priming volume of the predicate FX25 device. The reduced prime volume helps reduce the amount of blood that needs to be removed from a patient which, in turn, reduces the need for hemo-dilution. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Terumo. The logo consists of a red swoosh above the word "TERUMO" in green, block letters. The swoosh is angled upwards from left to right. Both the new NX19 Oxygenator and the predicate FX25 Oxygenator are comprised of the same generic materials, and they both feature common design characteristics. The notable differences in materials and design (as identified above) do not raise any issues of safety or effectiveness. Both devices are deemed to be clinically acceptable. #### Performance: Terumo Cardiovascular Systems has conducted in-vitro performance evaluations for the purpose of demonstrating that the Capiox® NX19 Oxygenator is substantially equivalent to the predicate the Capiox® FX25 Advance Oxygenator. Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the subject device to the predicate device. Substantial equivalence is demonstrated with the following in-vitro performance evaluations: - Gas Transfer and Pressure Drop Performance . - Heat Exchanger Performance . - Hemolysis Performance . - Air Removal Performance . - Capnography Assessment . - Connection Strength of Ports o - Filtration Efficiency o - Mechanical Integrity . - Luer Port Assessment o - . NX Blood Compatibility In addition to the verification testing listed above conducted on the oxygenator, the reservoir testing included in K151791 supports the increased maximum flow rate of the reservoir. The reservoir to be offered with the NX19 oxygenator is the EXACT SAME RESERVOIR that is offered with the FX25 and was cleared in K151791. The verification testing listed below was conducted on the reservoir and presented in K151791. This testing is also presented in Appendix D of this submission: - Testing of Air Handling . - Testing of Hemolysis . - Testing of Venous Defoaming o - . Testing of Pressure Drop - Testing of Clotting . - . Testing of Connection Strength The Terumo Cardiovascular Systems Corporation Capiox® NX19 Oxygenator exhibits performance features that are deemed "substantially equivalent" to the predicate Capiox® FX25 Advance Oxygenator - K151791. In addition to the above listed predicate device, Terumo has identified the Sorin Inspire 8F Oxygenator with integrated arterial filter and hardshell reservoir (K121536) as a reference device for performance testing. Because this device shares similar technological characteristics to the subject device in regards to the heat exchanger materials, Terumo felt it was appropriate to use the Inspire 8F as a reference device when setting the acceptance criteria for performance testing. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the Terumo logo. The logo consists of a red swoosh above the word "TERUMO" in green. The swoosh is curved and points to the right. The word "TERUMO" is in a bold, sans-serif font. The Terumo Cardiovascular Systems Corporation Capiox® NX19 Oxygenator exhibits performance features that are deemed "substantially equivalent" to the reference device. #### Labeling The subject Capiox® NX19 Oxygenator device and the predicate Capiox® FX25 Advance device are offered to the user with adequate Instructions for Use and other product labeling as required by regulation. While the labeling for the NX19 will differ from that of the FX25 with respect to content, the same required elements are present for both devices. Less information will be present on the product label for the NX19 as compared to the FX25. However, the information removed from the product label will be incorporated into a new pouch label. The contents of the box labels will remain the same. Refer to Section 12 - Labeling for more details. #### Substantial Equivalence Statement The Terumo Cardiovascular Systems Corporation Capiox® NX19 Oxygenator device (with 4liter Reservoir) and the predicate Capiox® FX25 Advance device (with 4-liter Reservoir) are substantially equivalent in intended use, principles of operation and technology, design, materials, performance and labeling. The noted differences between the subject devices and the predicate devices do not raise new issues of safety and effectiveness. #### Additional Safety Information - Sterilization conditions have been validated to provide a Sterility Assurance Level (SAL) of o 106. Terumo further asserts that the ethylene oxide residues will not exceed the maximum residue limits at the time of product distribution. - Terumo maintains biocompatibility studies as recommended in the FDA Guidance Document . (6/16/16): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process." [External Communicating Devices, Circulating Blood, Limited Exposure (≤ 24 hours) Contact Duration]. The blood contacting materials were found to be biocompatible. #### Conclusion In summary, the Terumo Cardiovascular Systems Corporation Capiox® NX19 Oxygenator device with 4-liter Reservoir is substantially equivalent to the predicate (K151791) Capiox® FX25 Advance device with 4-liter Reservoir with respect to intended use, principles of operation and technology, design, materials, performance, and labeling. It is further noted that any recognized differences do not raise any new issues of patient/user safety or product effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...