← Product Code [DTZ](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTZ) · K150489

# EOS PMP, EOS PMP Integrated (K150489)

_Sorin Group Italia S.R.L. · DTZ · Jun 11, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTZ/K150489

## Device Facts

- **Applicant:** Sorin Group Italia S.R.L.
- **Product Code:** [DTZ](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTZ.md)
- **Decision Date:** Jun 11, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4350
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric

## Intended Use

EOS PMP Integrated: Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothernia during surgery. The venous reservoir with cardiotomy filter is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacum assisted) during nomal operation to assure the proper oxygenation capability of the device. The device is intended to be used for 6 hours or less. EOS PMP: Hollow Fiber Oxygenator The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The device is intended to be used for 6 hours or less.

## Device Story

EOS PMP is a hollow fiber diffusion membrane oxygenator with an integrated heat exchanger; used during cardiopulmonary bypass surgery to provide oxygenation, CO2 removal, and blood temperature control. The device processes venous blood via a polymethylpentene (PMP) fiber membrane; integrated heat exchanger allows for hypothermia or normothermia maintenance. The Integrated version includes a hardshell venous/cardiotomy reservoir to collect, defoam, and filter blood aspirated from the surgical field or drained from patient veins. Operated by perfusionists in an OR setting; output is oxygenated blood returned to the patient. Benefits include extracorporeal gas exchange and thermal regulation during cardiac procedures.

## Clinical Evidence

Bench testing only. In vitro performance testing conducted on the oxygenating module per FDA guidance for cardiopulmonary bypass oxygenators. Tests included blood trauma, biological activity, coating integrity (leaching, flaking, uniformity), blood volume capacity, oxygenating performance, pressure drops, pathway integrity, and plasma leakage. All acceptance criteria were met.

## Technological Characteristics

Hollow fiber oxygenator with integrated heat exchanger. Oxygenating fiber: Polymethylpentene (PMP), surface area ~1.4 m². Priming volume: ≤200 ml. Max flow rate: 5 L/min. Min flow rate: 0.5 L/min. Sterilization: Ethylene oxide. Single-use. Non-pyrogenic fluid path. Integrated hardshell venous/cardiotomy reservoir (in Integrated model).

## Regulatory Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

## Predicate Devices

- D905 EOS: hollow fiber oxygenator with integrated hardshell venous/cardiotomy reservoir ([K043323](/device/K043323.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 11, 2015

Sorin Group Italia S.r.1. % Barry Sall RAC, FRAPS Principal Consultant PAREXEL Consulting, LLC 195 West Street Waltham, Massachusetts 02451

Re: K150489

Trade/Device Name: EOS PMP, EOS PMP Integrated Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: April 10, 2015 Received: April 13, 2015

Dear Barry Sall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |  |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|--|
| Indications for Use                                                     |                                                                                                   |  |
| 510(k) Number (if known)                                                |                                                                                                   |  |
| K150489                                                                 |                                                                                                   |  |
| Device Name                                                             |                                                                                                   |  |
| EOS PMP                                                                 |                                                                                                   |  |
| EOS PMP Integrated                                                      |                                                                                                   |  |

#### Indications for Use (Describe)

EOS PMP Integrated: Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothernia during surgery. The venous reservoir with cardiotomy filter is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacum assisted) during nomal operation to assure the proper oxygenation capability of the device. The device is intended to be used for 6 hours or less.

Patient population: Paediatric/small adults

#### EOS PMP: Hollow Fiber Oxygenator

The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The device is intended to be used for 6 hours or less.

Patient population: Paediatric/small adults

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a callection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

PSC Publishing Servisore (101) 443-6300 . EX

{3}------------------------------------------------

| SUBMITTER:            | Sorin Group Italia<br>86, Via Statale 12 Nord<br>41037 Mirandola (MO) Italy                                                                                |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON:       | Luigi Vecchi<br>Phone: 39 0535 29811<br>Fax: 39 0535 25229                                                                                                 |
| DATE PREPARED:        | May 28, 2015                                                                                                                                               |
| DEVICE TRADE NAME:    | EOS PMP<br>EOS PMP Integrated                                                                                                                              |
| COMMON NAME:          | Hollow Fiber Oxygenator<br>Hollow Fiber Oxygenator with integrated hardshell<br>venous/cardiotomy reservoir<br>Hardshell Venous/Cardiotomy Reservoir       |
| CLASSIFICATION NAME:  | Cardiopulmonary Bypass Oxygenator/<br>Cardiopulmonary Bypass Heat Exchanger/<br>Cardiopulmonary Bypass Blood Reservoir/<br>Cardiopulmonary Bypass Defoamer |
| CLASSIFICATION CODE:  | DTZ                                                                                                                                                        |
| REGULATION NUMBER:    | 870.4350                                                                                                                                                   |
| UNMODIFIED DEVICE(S): | D905 EOS: hollow fiber oxygenator with integrated<br>hardshell venous/cardiotomy reservoir (K043323)                                                       |

#### 510(k) SUMMARY

## DEVICE DESCRIPTION:

The EOS PMP is a high efficiency hollow fiber diffusion membrane oxygenator with integrated heat exchanger.

The device provides oxygenation and carbon dioxide removal from patient's blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia, or aids in the maintenance of normothermia during surgery.

The device can be operated at flow rates up to 5 liters per minute (I/min).

The device can be connected with, but not limited to, the D905 EOS reservoir (K043323) that collects, defoams, and filters patient's blood.

The EOS PMP is a modified version of the currently marketed D905 EOS (K043323).

{4}------------------------------------------------

## INDICATION FOR USE:

The intended uses for the device models are:

## EOS PMP Integrated: Hollow Fiber Oxyqenator with Integrated Hardshell Venous/Cardiotomy Reservoir

The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The venous reservoir with cardiotomy filter is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacuum assisted) during normal operation to assure the proper oxygenation capability of the device is intended to be used for 6 hours or less.

## EOS PMP: Hollow Fiber Oxygenator

The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The device is intended to be used for 6 hours or less.

# TECHNOLOGICAL CHARACTERISTICS:

The EOS PMP has the same fundamental technological characteristics, principles of operation and control mechanisms as the unmodified device.

The EOS PMP can be operated at a minimum blood flow rate of 0.5 LPM rather than 1 LPM and is provided with a different oxygenating fiber consisting of polymethylpentene (PMP, surface area approx. 1.4 m²) rather than the polypropylene fiber used for the unmodified device (PP, surface area approx. 1.1 m²). As a result the static priming volume is ≤200 ml rather than ≤210 ml.

No other changes have been made to the device. The venous/cardiotomy reservoir has not been modified and is not subject of the present 510(k).

Except for the change of the oxygenating fiber, the EOS PMP utilizes the same materials as the unmodified device.

No change to the intended use has been made as a result of these modifications.

The EOS PMP and the D905 EOS unmodified device share fundamental technological characteristics except for modifications that do not affect the basic device function. These differences do not raise any new issues of safety and effectiveness.

The devices are substantially equivalent on the basis of operating principles and basic function.

There are no differences in packaging design between modified and unmodified device.

The EOS PMP is ethylene oxide sterilized and has a non-pyroqenic fluid path. It is for single use only.

{5}------------------------------------------------

#### NON CLINICAL TEST RESULTS:

Applicable tests were conducted in accordance with the requirements of ISO 10993-1.

The current draft Guidance "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and its quidelines was also considered limited to external communicating device, circulating blood, limited contact duration.

## IN VITRO TEST RESULTS:

In vitro testing was conducted on the oxygenating module that is the only device element subject to modifications with respect to the unmodified device. The tests were performed to demonstrate unmodified device substantial equivalency and compliance to safety and effectiveness requirements.

The tests were conducted in accordance with the relevant requirements of "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff, November 13, 2000".

For venous/cardiotomy reservoir, the performance data previously submitted and already cleared are cross referenced.

The following table lists the performance tests conducted. The EOS PMP successfully met all acceptance criteria for each of the following tests.

| TEST | TEST CLASSIFICATION    | TEST TITLE                                        |
|------|------------------------|---------------------------------------------------|
| 1    | Functional/Performance | Blood trauma                                      |
| 2    | Functional/Performance | Biological activity                               |
| 3    | Functional/Performance | Leaching of coating                               |
| 4    | Functional/Performance | Flaking of coating                                |
| 5    | Functional/Performance | Uniformity of coating                             |
| 6    | Functional/Performance | Blood volume capacity                             |
| 7    | Functional/Performance | Oxygenating performance/blood side pressure drops |
| 8    | Functional/Performance | Blood, water, gas pathway integrity               |
| 9    | Functional/Performance | Plasma leakage                                    |

#### CONCLUSIONS:

The results of in vitro studies demonstrated that the modified oxygenator performs in a manner substantially equivalent to the unmodified oxygenator device with respect to the relevant functional parameters.

Test results of this study demonstrate the EOS PMP is substantially equivalent to the D905 EOS unmodified device with respect to device function.

Additional testing has also demonstrated the effectiveness of production techniques to assure that the device is sterile and non-pyrogenic.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTZ/K150489](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTZ/K150489)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
