INSPIRE 8 DUAL HOLLOW FIBER OXYGENATOR WITH WITH INTEGRATED HARDSHELL VENOUS/CARDIOTOMY
K121229 · Sorin Group Italia S.R.L. · DTZ · Jul 23, 2012 · Cardiovascular
Device Facts
Record ID
K121229
Device Name
INSPIRE 8 DUAL HOLLOW FIBER OXYGENATOR WITH WITH INTEGRATED HARDSHELL VENOUS/CARDIOTOMY
Applicant
Sorin Group Italia S.R.L.
Product Code
DTZ · Cardiovascular
Decision Date
Jul 23, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
INSPIRE 8M: Hollow Fiber Oxygenator The INSPIRE 8M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8M is intended to be used for 6 hours or less. INSPIRE HVR DUAL: Hardshell Venous/Cardiotomy Reservoir The INSPIRE HVR DUAL is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR DUAL can be used post-operatively for chest drainage. INSPIRE HVR DUAL is intended to be used for 6 hours or less.
Device Story
INSPIRE 8 DUAL is an integrated cardiopulmonary bypass system comprising a microporous hollow fiber membrane oxygenator (INSPIRE 8M), a heat exchanger, and a hardshell venous/cardiotomy reservoir (INSPIRE HVR DUAL). Used in surgical settings by perfusionists/surgeons to support patients during cardiopulmonary bypass. Oxygenator facilitates gas exchange (O2/CO2) and blood temperature regulation; reservoir collects, defoams, and filters venous and suctioned blood. Reservoir also supports post-operative chest drainage. Device operates at flow rates up to 8 L/min. Integrated design simplifies extracorporeal circuit setup. Benefits include effective gas exchange, thermal control, and blood management during surgery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing only. In vitro testing performed per ISO 10993-1, ISO 7199, ISO 15675, and ISO 15674. Testing included structural integrity, gas transfer performance, heat exchange efficiency, air handling, defoaming efficiency, filtration efficiency, hemolysis, and blood compatibility. All tests passed.
Technological Characteristics
Microporous hollow fiber membrane oxygenator with integrated heat exchanger and hardshell venous/cardiotomy reservoir. Materials are biocompatible per ISO 10993-1. Ethylene oxide sterilized; single-use. Flow rate capacity up to 8 L/min. Non-pyrogenic fluid path.
Indications for Use
Indicated for adult and small adult patients undergoing surgical procedures requiring cardiopulmonary bypass for up to 6 hours. Provides gas exchange, blood temperature control, venous/suction blood collection, defoaming, and filtration. Reservoir also indicated for post-operative chest drainage.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Predicate Devices
D903 AVANT 2 Ph.I.S.I.O. Adult Hollow Fiber Oxygenator with Ph.I.S.I.O. coating (K033323)
Related Devices
K122844 — INSPIRE 8F DUAL HOLLOW FIBER OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND HARDSHELL RESERVOIR · Sorin Group Italia S.R.L. · Oct 17, 2012
K121909 — INSPIRE 6 DUAL HOLLOW FIBER OXYGENATOR WITH INTEGRATED HARDSHELL VENOUS/CARDIOTOMY RESERVIOR · Sorin Group Italia S.R.L. · Jul 26, 2012
K122254 — INSPIRE 8 HOLLOW FIBER OXYGENATOR WITH INTEGRATED HARDSHELL VENOUS/CARDIOTOMY RESERVOIR · Sorin Group Italia S.R.L. · Aug 24, 2012
K113626 — INSPIRE 6 HOLLOW FIBER OXYGENATOR WITH INTEGRATED HARDSHELL VENOUS/CARDIOTOMY RESERVOIR · Sorin Group Italia S.R.L. · Apr 5, 2012
K130433 — INSPIRE 8 HOLLOW FIBER OXYGENATOR WITH INTEGRATED HARDSHELL RESERVOIR, INSPIRE 8F HOLLOW FIBER OXYGENATOR WITH INTEGRATE · Sorin Group Italia S.R.L. · Mar 20, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
K12/229
INSPIRE 8 DUAL Sorin Group Italia S.r.I. Traditional 510(k) April 24, 2012
| 510(k) SUMMARY | |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | JUL 2 3 2012 |
| SUBMITTER: | Sorin Group Italia S.r.l.<br>86, Via Statale 12 Nord<br>41037 Mirandola (MO) Italy |
| CONTACT PERSON: | Luigi Vecchi<br>Phone: 011 39 0535 29811<br>Fax: 011 39 0535 25229 |
| DATE PREPARED: | April 24, 2012 |
| DEVICE TRADE NAME: | INSPIRE 8 DUAL hollow fiber oxygenator with<br>integrated hardshell venous/cardiotomy<br>reservoir |
| COMMON NAMES: | Hollow Fiber Oxygenator with integrated<br>hardshell venous/cardiotomy reservoir<br>Hollow Fiber Oxygenator<br>Hardshell Venous/Cardiotomy Reservoir |
| CLASSIFICATION NAMES: | Cardiopulmonary Bypass Oxygenator/<br>Cardiopulmonary Bypass Heat Exchanger/<br>Cardiopulmonary Bypass Blood Reservoir/<br>Cardiopulmonary Bypass Defoamer/ |
| PREDICATE DEVICE: | D903 AVANT 2 Ph.I.S.I.O. Adult Hollow Fiber<br>Oxygenator with Ph.I.S.I.O. coating (K033323) |
#### DEVICE DESCRIPTION:
The INSPIRE 8 DUAL is consisting of an oxygenator, integrated with a heat exchanger (INSPIRE 8M), and a hardshell venous/cardiotomy reservoir (INSPIRE HVR DUAL). The reservoir is connected to the gas exchange module by means of a molded fitting joint.
The INSPIRE 8 DUAL is a high efficiency microporous hollow fiber membrane oxygenator, integrated with a heat exchanger, connected to a hardshell venous/cardiotomy reservoir.
The device can be operated at flow rates up to 8 liters per minute (I/min).
The hollow fiber membrane oxygenator provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia or aids in the maintenance of
{1}------------------------------------------------
| INSPIRE 8 DUAL | Traditional 510(k) |
|---------------------------|--------------------|
| Sorin Group Italia S.r.I. | April 24, 2012 |
normothermia during surgery. The integrated hardshell reservoir collects, defoams, filters venous and suction blood, and can be used post-operatively for chest drainage. The INSPIRE 8 DUAL is a modified version of the currently marketed integrated oxygenator/hardshell venous cardiotomy reservoir system D903 AVANT 2 Ph.I.S.I.O (hereinafter referred to as D903 AVANT).
## INDICATION FOR USE: ·
The intended use for the two elements that constitute the integrated device are:
## INSPIRE 8M: Hollow Fiber Oxygenator
The INSPIRE 8M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8M is intended to be used for 6 hours or less.
## INSPIRE HVR DUAL: Hardshell Venous/Cardiotomy Reservoir
The INSPIRE HVR DUAL is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR DUAL can be used post-operatively for chest drainage. INSPIRE HVR DUAL is intended to be used for 6 hours or less.
## TECHNOLOGICAL CHARACTERISTICS:
The INSPIRE 8 DUAL has the same fundamental technological characteristics, principles of operation and control mechanisms as the predicate device.
Sorin believes that the INSPIRE 8 DUAL is substantially equivalent to the D903 AVANT on the basis of operating principles and basic function.
The INSPIRE 8 DUAL and the D903 AVANT share the same fundamental technological characteristics except for some modifications that do not affect the basic device function. Any differences do not raise any new issues of safety and effectiveness.
The INSPIRE 8 DUAL is ethylene oxide sterilized and has a non-pyrogenic fluid path. It is for sinale use only.
## NON CLINICAL TEST RESULTS:
Applicable tests were carried out in accordance with the requirements of ISO 10993-1 and the FDA May 1st, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing of materials.
## IN VITRO TEST RESULTS:
In vitro testing was carried out in accordance with the relevant requirements of "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff" issued on November 13, 2000, "Guidance for Extracorporeal Blood Circuit Defoamer 510(k) Submissions; Final Guidance for Industry and FDA" issued on November 29, 2000 ISO 7199 "Cardiovascular implants and artificial organs - Blood-gas exchangers (oxygenators)"; ISO 15675 "Cardiovascular implants and artificial organs — Cardiopulmonary bypass systems — Arterial blood line
{2}------------------------------------------------
| INSPIRE 8 DUAL | Traditional 510(k) |
|---------------------------|--------------------|
| Sorin Group Italia S.r.I. | April 24, 2012 |
filters" and ISO 15674, "Cardiovascular implants and artificial organs - Hard-shell cardiotomy/venous reservoir systems (with/without filter) and soft venous reservoir bags".
In vitro testing was carried out to demonstrate both the substantial equivalence with the predicate device and also to comply with safety and effectiveness requirements.
Testing supplied in the 510(k) premarket notification includes performance tests and physical/mechanical integrity tests that demonstrate compliance with performance specifications.
The tests that were performed are listed in the following summary table. The INSPIRE 8 DUAL passed each test mentioned in the table below.
| TEST | TEST<br>CLASSIFICATION | TEST TITLE |
|------|------------------------|--------------------------------------------------------------------------|
| 1 | Physical/Mechanical | Oxygenating module structural integrity |
| 2 | Physical/Mechanical | Reservoir structural integrity |
| 3 | Physical/Mechanical | Oxygenating module blood, water, gas<br>pathway integrity |
| 4 | Physical/Mechanical | Reservoir blood pathway integrity |
| 5 | Functional/Performance | Oxygenating module blood volume capacity |
| 6 | Functional/Performance | Reservoir blood rest volume |
| 7 | Functional/Performance | Oxygenating module gas transfer<br>performance/blood side pressure drop |
| 8 | Functional/Performance | Oxygenating module heat exchange<br>performance/water side pressure drop |
| 9 | Functional/Performance | Reservoir air handling |
| 10 | Functional/Performance | Reservoir break-through time and volume |
| 11 | Functional/Performance | Reservoir defoaming efficiency |
| 12 | Functional/Performance | Reservoir dynamic priming volume / Hold-up |
| 13 | Functional/Performance | Reservoir filtration efficiency - venous<br>section |
| 14 | Functional/Performance | Reservoir filtration efficiency - cardiotomy<br>section |
| 15 | Functional/Performance | Reservoir flow rate capacity |
| 16 | Functional/Performance | Reservoir pressure drop |
| 17 | Functional/Performance | Integrated device hemolysis |
| 18 | Functional/Performance | Integrated device blood compatibility |
| 19 | Functional/Performance | Oxygenating module leaching of coating |
| 20 | Functional/Performance | Reservoir leaching of coating |
{3}------------------------------------------------
| TEST | THEST<br>CLASSIFICATION | TEST TITLE | |
|------|-------------------------|------------------------------------------|--|
| 21 | Functional/Performance | Integrated device flaking of coating | |
| 22 | Functional/Performance | Oxygenating module uniformity of coating | |
| 73 | Functional/Performance | Reservoir uniformity of coating | |
## CONCLUSIONS:
The results of in vitro studies demonstrate that the INSPIRE 8 DUAL performs in a manner substantially equivalent to the D903 AVANT predicate device with respect to the relevant functional parameters. Test results of this study demonstrate the INSPIRE 8 DUAL is equivalent to the predicate device with respect to device function.
Additional testing has also demonstrated the effectiveness of production techniques to assure that the device is sterile and non-pyrogenic.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
# JUL 2 3 2012
Sorin Group Italia s.r.1. c/o Mr. Barry Sall Principal Consultant Parexel International Consulting 195 West Street Waltham, WA 02451
Re: K121229
Trade/Device Name: Inspire 8 DUAL Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: April 24, 2012 Received: April 24, 2012
Dear Mr. Sall:
We have reviewed your Section 510(k) premarket notification of intent to market the device we neve rollowed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 ass stated in the Med 28, 1976, the enactment date of the Medical Device Amendments, or to conninered print to trial 2011-17, in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic (110.) that the novice, subject to the general controls provisions of the Act. The r ou may, merelore, manot of the Act include requirements for annual registration, listing of general controls provisions of practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elabilited been of . Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Mr. Barry Sall
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be har nou a determination that your device complies with other requirements of the Act that I Drinas intact and regulations administered by other Federal agencies. You must or uny I eath all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Or K F at 607), hoo!ing (21 CFR 803); good manufacturing practice requirements as set do rece related daty systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the qualify of oversions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire speethe ad 100 ros JVA /CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of amall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INSPIRE 8 DUAL Sorin Group Italia S.r.I.
Indications for Use:
### INSPIRE 8M: Hollow Fiber Oxygenator
The INSPIRE 8M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8M is intended to be used for 6 hours or less.
## INSPIRE HVR DUAL: Hardshell Venous/Cardiotomy Reservoir
The INSPIRE HVR DUAL is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR DUAL can be used post-operatively for chest drainage. INSPIRE HVR DUAL is intended to be used for 6 hours or less.
× Prescription Use _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_ 220
68
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.