QUADROX-I SMALL ADULT MICROPOROUS MEMBRANE OXYGENATOR; W/ BIOLINE COATING AND W/ INTEGRATED ARTERIAL FILTER W/ BIOLINE C

K093522 · Maquet Cardiopulmonary, AG · DTZ · Dec 7, 2009 · Cardiovascular

Device Facts

Record IDK093522
Device NameQUADROX-I SMALL ADULT MICROPOROUS MEMBRANE OXYGENATOR; W/ BIOLINE COATING AND W/ INTEGRATED ARTERIAL FILTER W/ BIOLINE C
ApplicantMaquet Cardiopulmonary, AG
Product CodeDTZ · Cardiovascular
Decision DateDec 7, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4350
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The membrane oxygenator QUADROX-i Small Adult is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. The blood flow rate is defined from 0.5 - 5 l/min. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature. The QUADROX-i Small Adult (HMO 51000) version with integrated arterial filter also filters out air bubbles and particles larger than 40 µm. The device's utilization period is limited to six hours. The oxygenator is suitable for delivery of the volatile anesthetics isoflurane and sevoflurane. The anesthetic gas is administered through the oxygenator's gas inlet by means of a suitable anesthetic gas vaporizer. Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.

Device Story

Device is a microporous membrane oxygenator with integrated heat exchanger and optional arterial filter; used in extracorporeal circulation during cardiac surgery. Input: venous blood from patient; Output: oxygenated, CO2-depleted, temperature-regulated blood returned to patient. Integrated filter removes air bubbles/particles >40 µm. Operated by perfusionists/physicians in OR. Device facilitates gas exchange and anesthetic delivery (isoflurane/sevoflurane) via vaporizer. Benefits: maintains patient physiological stability during bypass. BIOLINE coating applied to blood-contacting surfaces.

Clinical Evidence

Bench testing only. Evaluation included integrity, performance, biocompatibility, and sterility testing to demonstrate equivalence to predicate devices.

Technological Characteristics

Microporous membrane oxygenator; integrated heat exchanger; optional integrated arterial filter (40 µm). BIOLINE coating on blood-contacting surfaces. Flow range 0.5-5 L/min. Duration limit 6 hours. Mechanical gas exchange principle.

Indications for Use

Indicated for patients undergoing cardiac surgery requiring extracorporeal circulation during cardiopulmonary bypass. Supports blood flow rates of 0.5-5 L/min for up to 6 hours. Suitable for delivery of isoflurane and sevoflurane.

Regulatory Classification

Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # 093527 AQUET GETINGE GROUP ### 510(k) Summary יי {as required by 21 CFR 807.92(c)] | Submitter | MAQUET Cardiopulmonary AG<br>Hechinger Strasse 38<br>72145 Hirrlingen<br>Germany | DEC - 7 2009 | |--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Contact Person | Frank Moehrke<br>Phone: 011 49 7478 921 229<br>Fax: 011 49 7478 921 8667 | | | Date Prepared | November 10, 2009 | | | Device Trade Name | QUADROX-i Small Adult Microporous<br>Membrane Oxygenator with and without<br>Integrated Arterial Filter with BIOLINE Coating | | | Common/Usual Name | Oxygenator with integrated heat exchanger and<br>optional integrated arterial filter | | | Classification Names | Cardiopulmonary bypass oxygenator<br>(21 CFR 870.4350 - Product Code: DTZ)<br>Cardiopulmonary bypass heat exchanger<br>(21 CFR 870.4240 - Product Code: DTR)<br>Cardiopulmonary bypass arterial line blood filter<br>(21 CFR 870.4260 - Product Code: DTM) | | | Legally Marketed Devices | QUADROX-i Small Adult Microporous Membrane<br>Oxygenator with and without Integrated Arterial<br>Filter with Softline Coating (K090689), | | | | QUADROX-i Adult Microporous Membrane<br>Oxygenator with and without Integrated Arterial<br>Filter with BIOLINE Coating (K000511) | | #### Device Description The QUADROX-i Small Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with BIOLINE Coating is a blood-gas exchanger with integrated heat exchanger and optional integrated arterial blood filter. 014 {2}------------------------------------------------ # GETINGE GROUP #### Indications for Use The membrane oxygenator QUADROX-i Small Adult is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. The blood flow rate is defined from 0.5 - 5 Vmin. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and requiates blood temperature. The QUADROX-i Small Adult (HMO 51000) version with integrated arterial filter also filters out air bubbles and particles larger than 40 µm. The device's utilization period is limited to six hours. The oxygenator is suitable for delivery of the volatile anesthetics isoflurane and sevoflurane. The anesthetic gas is administered through the oxygenator's gas inlet by means of a suitable anesthetic gas vaporizer. Responsibility for deciding whether to use an oxygenator rests solely with the attending physician. #### Statement of Technical Comparison The QUADROX-i Small Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with BIOLINE Coating is identical to the QUADROX-i Small Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with Softline Coating with the only exception that the QUADROX-i Small Adult Oxygenators with BIOLINE Coating have been coated with BIOLINE Coating instead of Softline Coating. However, the BIOLINE Coating is the same as with the QUADROX-i Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with BIOLINE Coating. Besides this difference the QUADROX-i Small Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with BIOLINE Coating is the same in design, intended use, method of operation, components, packaging, and fundamental scientific technology as compared to the QUADROX-i Small Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with Softline Coating. #### Determination of Substantial Equivalence Evaluation and testing on safety and effectiveness was executed to demonstrate that the QUADROX-i Small Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with BIOLINE Coating described in this submission is substantially equivalent to the QUADROX-i Small Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with Softline Coating as oxygenator and to the QUADROX-i Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with BIOLINE Coating regarding the BIOLINE Coating. {3}------------------------------------------------ ## MAQU GETINGE GROUP The following areas have been tested and / or evaluated: - Integrity - - Performance - - Biocompatibility - - Sterility - #### Conclusion The data given demonstrate that the QUADROX-i Small Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filler with BIOLINE Coating is substantially equivalent to the named predicate devices which hold currently market clearance. {4}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of all Americans. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is presented in black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 DEC - 7 2009 Maquet Cardiopulmonary AG c/o Mr. Frank Moehrke Regulatory Affairs Manager Hechinger Strasse 38 72145 Hirrlingen Germany Re: K093522 > QUADROX-i Small Adult Microporous Membrane Oxygenator With and without Integrated Arterial Filter with BIOLINE Coating Regulation Number: 21 CFR 870.4350 Regulation Name: Oxygenator Cardiopulmonary Bypass Regulatory Class: Class II (two) Product Code: DTZ, DTR, DTM Dated: November 10, 2009 Received: November 13, 2009 Dear Mr. Moehrke: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. Frank Moehrke Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Dina R. Vahner Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Ko93522 Device Name: #### QUADROX-i Small Adult Microporous Membrane Oxygenator with and without Integrated Arterial Filter with BIOLINE Coating Indications for Use: The membrane oxygenator QUADROX-i Small Adult is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. The blood flow rate is defined from 0.5 - 5 I/min. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature. The QUADROX - i Small Adult (HMO 51000) version with integrated arterial filter also filters out air bubbles and particles larger than 40 um. The device's utilization period is limited to six hours. The oxygenator is suitable for delivery of the volatile anesthetics isoflurane and sevoflurane. The anesthetic gas is administered through the oxygenator's gas inlet by means of a suitable anesthetic gas vaporizer. Responsibility for deciding whether to use an oxygenator rests solely with the attending physician. AND/OR Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) bune R. Ladner (Division Sign-Off) (Division Sign-Sil, vascular Devices 510(k) Number k093522 Page 1 of 1 013
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...