The CAPIOX® RX25 Hollow Fiber Oxygenator with/without Hardshell Reservoir is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integral heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device. The (detachable) hardshell reservoir is used to store blood during extra-corporeal circulation from both venous line and the cardiotomy line (via gravity or vacuum assisted venous drainage procedures). The venous section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The Reservoir may also be used for Post-Operative Chest Drainage procedures. The CAPIOX® RX25 Oxygenator with/without Hardshell Reservoir can be used in procedures lasting up to 6 hours. The CAPIOX® RX25 is for use with patients when the required blood flow rate will not exceed 7.0 l/min.
Device Story
Device is a hollow fiber oxygenator with integrated heat exchanger and detachable hardshell reservoir; used in cardiopulmonary bypass surgery. Inputs: venous blood and cardiotomy suction blood collected via gravity or vacuum. Operation: blood passes through reservoir filters/defoamers to remove air/particulates; flows through heat exchanger for temperature regulation; enters oxygenator module where gas transfer (O2/CO2 exchange) occurs via diffusion across hollow fiber membranes. Output: oxygenated, temperature-controlled blood returned to patient. Used in OR by perfusionists/surgeons. Benefits: provides life-sustaining gas exchange and temperature control during cardiac procedures. Features X-Coating to reduce platelet adhesion.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via in-vitro performance evaluations including gas transfer, hemolysis, pressure drop, mechanical integrity, priming volume, heat exchanger performance, defoaming, filtration efficiency, flow rate, and tubing connection strength.
Technological Characteristics
Materials: polycarbonate, stainless steel, polyester, polypropylene, polyethylene, and X-Coating. Principle: hollow fiber membrane diffusion for gas exchange; external water bath for heat exchange. Form factor: integrated oxygenator/heat exchanger with detachable hardshell reservoir. Sterilization: ethylene oxide. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery requiring gas exchange, blood temperature control, and blood storage/filtration. Suitable for procedures up to 6 hours with blood flow rates up to 7.0 L/min. Also indicated for post-operative chest drainage.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
{0}------------------------------------------------
K040210
MAR - 3 2004
Section II 510(k) Summary and Certification Terumo Corporation Capiox RX25 Oxygenator/Reservoir
## CAPIOX® RX25 Hollow Fiber Oxygenator with/without Hardshell Reservoir
# Submitter Information:
Submitter Information was prepared in January 2002 by: Garry A. Courtney, MBA, RAC Sr. Regulatory Affairs Specialist Terumo Cardiovascular Systems Corp. 125 Blue Ball Road Elkton, MD 21921 Telephone: 1-800-283-7866, Ext. 7420
This submission was prepared for: Ashitaka Factory of Terumo Corporation 150 Maimaigi-cho Fujinomiya city, Shizuoka Pref. Japan 418-0015
| Device Names/Classifications:<br>Proprietary Name | Classification Name | Common Name |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------|----------------------------------------|
| CAPIOX® RX25 Hollow<br>Fiber Oxygenator<br>with/without Hardshell<br>Reservoir | Cardiopulmonary Bypass<br>Oxygenator (Code: DTZ) | Oxygenator |
| | Cardiopulmonary Bypass<br>Heat Exchanger (Code: DTR) | Heat Exchanger |
| | Cardiopulmonary Bypass<br>Blood Reservoir (Code: DTN) | Blood Reservoir |
| | Cardiopulmonary Bypass<br>Defoamer (Code: DTP) | Defoamer |
| | Cardiopulmonary Bypass<br>Cardiotomy Suction Line<br>Blood Filter (Code: JOD) | Blood Filter |
| | Cardiopulmonary Bypass<br>Stopcock, Manifold, Fitting<br>(Code: DTL) | Sämpling<br>Manifold with<br>Stopcocks |
{1}------------------------------------------------
#### Predicate Device:
Predicale Device.
The device submitted in this 510(k) maintains characteristics that are substantially equivalent in I he device submitted in this 910(x) mainless of operation, materials and specifications to the following devices:
- Terumo's CAPIOX® SX25 Oxygenator w/wo Hardshell Reservoir K962667. .
#### Intended Use:
The CAPIOX® RX25 Hollow Fiber Oxygenator with/without Hardshell Reservoir is intended to The CATION® RA25 Honor Process on James on James to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery.
The integral heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.
The (detachable) hardshell reservoir is used to store blood during extra-corporeal circulation The (detachable) hardshen to servity or vacuum assisted venous assisted venous drainage If on both vellous inno and the and section that is comprised of a filter and defoamer to procedures). The reservor comains a vincomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The Reservoir may also be used for Post-Operative Chest Drainage procedures.
The CAPIOX® RX25 Oxygenator with/without Hardshell Reservoir can be used in procedures lasting up to 6 hours.
The CAPIOX® RX25 is for use with patients when the required blood flow rate will not exceed 7.0 I /min.
### Principles of Operation and Technology:
The design of the CAPIOX® RX25 Hollow Fiber Oxygenator with/without Hardshell Reservoir is such that blood is collected into the reservoir via gravity or external vacuum. Blood may enter is such that brood is confect and/or the cardiotomy inlet port. The reservoir contains filtering devices to remove particulate matter and air. Blood is then pumped from the reservoir into the devices to remove partisance by blood temperature is controlled. After the blood exits the heat exchanger, it enters the oxygenator device whereby gas transfer (introduction of oxygen and removal of carbon dioxide) occurs. After gas transfer has occurred, the blood exits the device and is pumped towards the patient.
#### Design and Materials:
Design of the CAPIOX® RX25 Hollow Fiber Oxygenator with/without Hardshell Reservoir The dosign of the Orice whereby the oxygenator and heat exchanger are joined together, while the hardshell reservoir can be detached from the device assembly.
The materials that are used in the construction of the CAPIOX® RX25 Hollow Fiber Oxygenator with/without Hardshell Reservoir include, but are not limited to, polycarbonate, stainless steel, with without Trafficity Iretting, polyester, polypropylenc, polyethylene and X-Coating.
{2}------------------------------------------------
#### Performance Evaluations:
Clinical studies are not necessary to demonstrate substantial equivalence of the subject device to the predicate devices. Substantial equivalence is demonstrated with the following in-vitro performance evaluations:
- Gas Transfer .
- Effects on Blood Components (Hemolysis) .
- Pressure Drop .
- Mechanical Integrity .
- Static Priming Volume .
- Heat Exchanger Performance .
- Defoaming .
- Filtration Efficiency .
- Flow Rate .
- Tubing Connection Strength .
#### Substantial Equivalence Comparison:
The CAPIOX® RX25 Hollow Fiber Oxygenator with/without Hardshell Reservoir is substantially equivalent to the predicate devices as follows:
- Intended Use: The intended uses of the subject device and the predicate devices (Terumo's . SX25 devices) are essentially the same. The oxygenator devices are used to provide for gas exchange between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery.
Each of the integral heat exchangers is used to warm or cool blood and/or perfusion fluid as it flows through the device.
The respective hardshell reservoirs are each used to collect and store blood during the bypass procedure. Filters are present in each device to facilitate air and particulate removal.
The new device and the predicate devices may each be used in Vacuum Assisted Venous Drainage and Post-Operative Chest Drainage procedures.
- Principles of Operation and Technology: The technology of the subject device and the . predicate devices are essentially identical. The devices operate in a manner where blood is collected into the reservoir. The blood may enter the reservoir via the venous inlet or the cardiotomy inlet. The reservoirs each contain filtering/defoaming devices that facilitate the removal of particulate matter and air. Blood is then pumped from the reservoir into the heat exchanger device whereby blood temperature is controlled with the use of an external water bath. After the blood exits the heat exchanger, it enters the oxygenator device whereby gas transfer occurs (i.e., introduction of oxygen; removal of carbon dioxide). The transfer process occurs via diffusion across the walls of the hollow fiber membranes contained within the oxygenator. After gas transfer has occurred, the blood exits the devices and is pumped towards the patient.
{3}------------------------------------------------
- Design and Materials: The design and the materials of the CAPIOX® RX25 Hollow Fiber . Design and Materials: The design and the predicate devices are essentially the Oxygenator with/Without Hardshen Reservon and they cach contain a hardshell reservoir for use same. The design of each device is similar in thar ally control, and an oxygenator for gas
collection of blood, a heat exchanger for temperature control, and any and states transfer. Such a design is common among oxygenating devices in the marketplace.
The devices are manufactured with variations of plastics, adhesives, wrethanes, The devices stainless steel, etc. The RX25 device contains X-Coating, which is a biocompatible surface coating that reduces platelet adhesion to the device. of cofety, and/or Coating has been demonstrated as safe and raises no new issues of safety and/or Coating has been demonshated as sare the rasso no the predicate SX25 Oxygenators and Reservoirs cleared by FDA in K993772.
The most notable design difference between the new device and the predicate is the The most the polypropylene fibers that comprise the oxygenator module. The predicated orientation of the porypropyithe frocts that cempion is essentially a collection of individual SX25 device is represenced by a Jroer ounder, which end of the bundle. By contrast, the new RX25 design utilizes a wound fiber design whereby single strands of fiber are new KX2> design uninzes a "wand" free" dong" and surface area of fibers that are continuously wrapped around a supporting vere. remains identical for the new RX25 device that fiber buyedly exposed to blood and gas during bypass surgery telled difference in the fiber builde and predicate SX23 device at approximatery E15 m + + 1 x = 1 safety issues with the new design.
- Comparisons of the performance of the CAPIOX® RX25 Hollow Fiber Performance: . Performance: Comparisons of the performation of the predicate devices were conducted.
Oxygenator with/without Hardshell Reservoir and the producted, professmance Oxygenator with/Winfour Hardshen reserver and clinically significant performance differences between the devices.
#### Substantial Equivalence Summary:
Substantial Equivatence Sammary.
In summary, the CAPIOX® RX25 Hollow Fiber Oxygenator with/without Hardshell Reservoir In summary, the CAYION® RX23 Yours Provent in intended use, principle of operation the and the predicale devices are substancials, and performance. Any noted differences between the and technology, design and materials, and possible in the person safety and effectiveness.
#### Additional Safety Information:
- Sterilization conditions have been validated in accordance with AAM guidelines to provide . Sterility Assurance Level (SAL) of 10%. Terumo further asserts that the ethylene oxide a Sterlify Assurance Dever (DAD) of Your residue limits at the time of product distribution.
- Terumo conducted biocompatibility studies as recommended in the FDA General Program . I erumo conducted blocompatibutity statics as reational Standard 150 10993, "Diological Memorandulin #G9591 (3/1773). Oso J : Evaluation and Testing." [External Communicating Evaluation of Medical Devices – Far T. Branachen and Cost Contact Duration]. The blood contacting materials were found to be biocompatible.
- Terumo conducted studies for materials characterization, including physico-chemical profiles . of aged and nonaged cevices.
{4}------------------------------------------------
- The polymer coating material that is applied to the blood-contacting surfaces of the device in . The polymer coating material that is applica to the orders and the golds.
was also evaluated in an in-vivo animal study. No adverse conditions were noted.
#### Conclusion:
Conclusion:
In summary, the CAPIOX® For the local incincincincincincinci has of one and technology, design and In summary, the CAPIOX® EX2. Hollow Fiber Of operation and technology, design and
substantially equivalent in intended use, principles of operation and technology, design and substantially equivalent in intended use, principies of openition and working of the only of the Oxygenator
materials, and performance to the prodicated devices, Terumo SX25 materials, and performance to the presentation of K993772).
. .
{5}------------------------------------------------
Public Health Service
Image /page/5/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 3 2004
Terumo Cardiovascular Systems Corp. c/o Mr. Garry A. Courtney Sr. Regulatory Affairs Specialist 125 Blue Ball Road Elkton, MD 21921
Re: K040210
Capiox® RX25 Oxygenator/Reservoir Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II (two) Product Code: DTZ Dated: January 28, 2004 Reccived: January 30, 2004
Dear Mr. Courtney:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rovewed your becamined the device is substantially cquivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use thated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Garry A. Courtney
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA 3 issuation of a budenia complies with other requirements of the Act that IDA has made a determinations administered by other Federal agencies. You must of any Pederal Statutes and regulations daminders of registration and listing (21 comply with an the Act 3 requirements, 01); good manufacturing practice requirements as set CrK Part 807), labeling (21 CFR Part 820); and if applicable, the electronic forum in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product laulation control provisions (over device as described in your Section 510(k) I ms letter will anow you to oegin maxically your and equivalence of your device to a legally premaired notheadon: "The PDF Intelligstification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at no 10 - 30 - 301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to (201) 394-4039. Also, please note the regation may obtain. Other general information on your prematics nonneation - (210) NY fare obtained from the Division of Small Manufacturers, responsionnies under the Are They over out its toll-free number (800) 638-2041 or (301) 443-6597 incernational and Oonbainttp://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K040210
Device Name:
Capiox® RX25 Oxygenator/Reservoir
#### Indications For Use:
The CAPIOX® RX25 Hollow Fiber Oxygenator with/without Hardshell Reservoir is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery.
The integral heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.
The (detachable) hardshell reservoir is used to store blood during extra-corporcal circulation from both venous line and the cardiotomy line (via gravity or vacuum assisted venous drainage nom over vollows internairs a venous section that is comprised of a filter and defoamer to procedured). The 1000 removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The Reservoir may also be used for Post-Operative Chest Drainage procedures.
The CAPIOX® RX25 Oxygenator with/without Hardshell Reservoir can be used in proccdures lasting up to 6 hours.
The CAPIOX® RX25 is for use with patients when the required blood flow rate will not exceed 7.0 I /min.
X = Prescription Use ____ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sig. Off) | |
|-----------------------|-------------|
| Division of Cardiovas | Page 1 of 1 |
| 510(k) Number | K040 210 |
|---------------|----------|
|---------------|----------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.