VISION CARDIOPLEGIA DELIVERY SYSTEM

K964571 · Lifestream Int'L, Inc. · DTR · Aug 8, 1997 · Cardiovascular

Device Facts

Record IDK964571
Device NameVISION CARDIOPLEGIA DELIVERY SYSTEM
ApplicantLifestream Int'L, Inc.
Product CodeDTR · Cardiovascular
Decision DateAug 8, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4240
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Vision Cardioplegia Delivery System is intended to mix, cool, warm, and deliver oxygenated blood cardioplegia solution during cardiopulmonary bypass.

Device Story

The Vision Cardioplegia Delivery System is a cardiopulmonary bypass heat exchanger used during cardiac surgery. It functions by mixing, cooling, and warming oxygenated blood cardioplegia solution for delivery to the patient's heart. Operated by perfusionists or surgical staff in an operating room setting, the device manages thermal regulation of the cardioplegia solution to maintain myocardial protection during bypass. The system is designed to provide controlled temperature delivery, flow resistance management, and debubbling capabilities. By regulating the temperature of the cardioplegia solution, the device assists in arresting the heart and protecting the myocardium during the surgical procedure.

Clinical Evidence

Bench testing only. Comparative performance data against the Sorin BCD Vanguard confirmed similar rates of hemolysis, heat exchange effectiveness, warming/cooling performance, internal flow resistance, maximum pressure tolerance, and debubbling capacity.

Technological Characteristics

Cardiopulmonary bypass heat exchanger. Functions via heat exchange between blood and water circuits. Designed for integration into cardiopulmonary bypass circuits. Mechanical device; no software or electronic components described.

Indications for Use

Indicated for patients undergoing cardiopulmonary bypass requiring the mixing, cooling, warming, and delivery of oxygenated blood cardioplegia solution.

Regulatory Classification

Identification

A cardiopulmonary bypass heat exchanger is a device, consisting of a heat exchange system used in extracorporeal circulation to warm or cool the blood or perfusion fluid flowing through the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K964571 AUG 8 1997 510(k) Summary SURGIMEDICS®/TMP® Vision Cardioplegia Delivery System (per 21 CFR 807.92) 1. Date Prepared: November 11, 1996 2. Sponsor/Applicant: SURGIMEDICS® /TMP® 2828 N. Crescent Ridge Drive The Woodlands Research Forest The Woodlands, TX 77381 3. Contact Name: Russell Jones, Manager of Regulatory Affairs Telephone Number: 713-363-4949 4. Device Name: Proprietary Name: Vision Cardioplegia Delivery System Common/Usual Name: Cardioplegia Heat Exchanger Classification Name: Cardiopulmonary Bypass Heat Exchanger 5. Device Classification: Cardioplegia heat exchangers are similar to cardiopulmonary bypass heat exchangers (21 CFR 870.4240, Product Code DTR) which have been classified under Section 513 of the Act as Class II by the Cardiovascular Devices Panel. 6. Predicate Devices: The SURGIMEDICS®/TMP® Vision Cardioplegia Delivery System is substantially equivalent to the following legally marketed cardioplegia heat exchangers: - Sorin Biomedical Buckberg BCD Vanguard (No K number identified) - Sorin Biomedical BCD Advanced Blood Cardioplegia System (K925369) - Avecor Cardiovascular MYOtherm Cardioplegia System (K904171) SURGIMEDICS®/TMP® Vision Cardioplegia Delivery System - 510(k) Summary 11/13/96 {1} # 7. Device Description and Substantial Equivalence: Intended Use: The Vision Cardioplegia Delivery System is intended for mixing, cooling, warming, and delivery of oxygenated blood cardioplegia solution during cardiopulmonary bypass. Substantial Equivalence: The Vision Cardioplegia Delivery System is substantially equivalent to the above listed cardioplegia heat exchangers based on their intended use, design, materials, and principles of operation. Operational and Technological Characteristics: Information provided in this Premarket Notification to support the determination of substantial equivalence for the Vision Cardioplegia Delivery System includes descriptive information about the intended use, operation, and technological characteristics, as well as comparative bench testing which characterizes device performance. When the performance of the Vision Cardioplegia Delivery System is compared to the Sorin BCD Vanguard Cardioplegia System, the SURGIMEDICS®/TMP®, both devices showed similar rates of hemolysis, heat exchange effectiveness, warming, cooling, internal resistance to flow, maximum water and blood side pressure tolerance, and debubbling capacity. SURGIMEDICS®/TMP® Vision Cardioplegia Delivery System - 510(k) Summary 11/13/96 {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 AUG 8 1997 Surgimedics, Inc. c/o Ms. Rosina Robinson Staff Consultant Medical Device Consultants, Inc. 49*Plain Street North Attleboro, Massachusetts 02760 Re: K964571 Surgimedics®/TMP® Vision Cardioplegia Delivery System (BCD) Regulatory Class: II (Two) Product Code: 74 DTR Dated: June 5, 1997 Received: June 6, 1997 Dear Ms. Robinson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3} Page 2 - Ms. Rosina Robinson This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "dsmo@fdadr.cdrh.fda.gov." Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} 870.4246 - DTR II - CPB Heat Exchanger/Delivery System 510(k) Number (if known): K964571 Device Name: Vision Cardioplegia Delivery System Indications For Use: The SURGIMEDICS®/TMP® Vision Cardioplegia Delivery System is intended to mix, cool, warm, and deliver oxygenated blood cardioplegia solution during cardiopulmonary bypass. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K964571 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...