AFFINITY BLOOD PUMP SYSTEM

K964017 · Avecor Cardiovascular, Inc. · DTQ · Aug 5, 1997 · Cardiovascular

Device Facts

Record IDK964017
Device NameAFFINITY BLOOD PUMP SYSTEM
ApplicantAvecor Cardiovascular, Inc.
Product CodeDTQ · Cardiovascular
Decision DateAug 5, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4220
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Affinity™ Blood Pump System is intended to be used in an extracorporeal perfusion circuit to pump blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Device Story

Affinity Blood Pump System is an extracorporeal perfusion device used during cardiopulmonary bypass (CPB) procedures. System components include Model 2000 Affinity Pump Console, rotor, holder, pump chamber, and connecting tubing. Operates on peristaltic principle: polyurethane pump chamber fills passively via reservoir fluid height; pump rotor rollers exert peristaltic action against chamber to generate forward blood flow. Console controls variable speed motor and monitors flow rate, RPM, pressure, timed events, and battery status. Used in clinical settings by perfusionists/clinicians to maintain circulation during surgery. Output provides controlled blood flow to support patient hemodynamics. Benefits include reliable extracorporeal circulation during routine bypass procedures.

Clinical Evidence

No clinical data. Evidence consists of in vitro bench testing, including operating characteristics, cellular damage (hemolysis, platelet and white cell response), material biocompatibility, sterility, pyrogenicity, and sterilant residuals. Software validation and verification were also performed.

Technological Characteristics

Peristaltic pump system. Materials: Polyurethane pump chamber, polycarbonate connectors, PVC protective caps. Console features: variable speed motor control, monitoring for flow, RPM, pressure, and battery. Power: 100-120/220-240 Vac. Standards: IEC 601, UL 544. CE marked.

Indications for Use

Indicated for patients undergoing cardiopulmonary bypass procedures requiring extracorporeal blood circulation for up to 6 hours.

Regulatory Classification

Identification

A cardiopulmonary bypass heart-lung machine console is a device that consists of a control panel and the electrical power and control circuitry for a heart-lung machine. The console is designed to interface with the basic units used in a gas exchange system, including the pumps, oxygenator, and heat exchanger.

Predicate Devices

Submission Summary (Full Text)

{0} K 904017 # SECTION II ## AVECOR CARDIOVASCULAR INC. ### AFFINITY BLOOD PUMP SYSTEM ### SUPPORTING SUMMARY FOR 510 (k) NOTIFICATION 1. Date Prepared: September 30, 1996 2. Submitter: AVECOR Cardiovascular Inc. 13010 County Road 6 Plymouth, MN 55441 3. Contact: Dennis E. Steger Director Regulatory Affairs/ Quality Assurance (800) 328-3320 4. Device Name: Cardiopulmonary Bypass Roller-Type Pumps, Consoles and Bypass Pump Tubing have been classified by the Cardiovascular Device Classification Panel as Class II devices, as proposed in Section 870.4220/4370/4380/4390, Title 21 of the Code of Federal Regulations 5. Device Description & Comparison to Predicate Device: To establish the substantial equivalency of the AVECOR Affinity Pump System to the Stockert Instrumente (Sorin) Pump design, specifications, materials, indications for use, and performance characteristics were compared. ## DESIGN The AVECOR Affinity Blood Pump System is intended for use in an extracorporeal perfusion circuit to pump blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The AVECOR Cardiovascular Affinity Pump System consists of the Model 2000 Affinity Pump Console, Affinity Pump Rotor, Affinity Pump Holder, Affinity Pump Chamber, 5/8" Connecting Tube and an AVECOR venous reservoir with 5/8" outlet. The AVECOR Affinity Blood Pump Console and components shall conform with all the applicable requirements of IEC 601, UL 544 and will display the European Community CE marking. The Affinity Pump Console, Model 2000 provides a system for controlling a variable speed pump motor and additional systems for monitoring flow rate, pump speed (RPM), pressure, timed events and battery status. {1} The AVECOR Affinity Pump has been designed to operate on the principles of a peristaltic pump for extracorporeal use. The pump consists of a polyurethane chamber incorporating polycarbonate inlet and outlet connectors, which fills passively due to the filling pressure exhibited by the fluid height in a reservoir. The peristaltic action of the rollers against the filled pump chamber results in forward flow. Flow rate is affected by the combination of reservoir volume, RPM's of the pump rotor and the total circuit resistance. The Stockert Instrumente (Sorin) Pump Console Model No. 10-10-00, provides a system for controlling a variable speed pump motor and additional systems for monitoring flow rate, pump speed (RPM), pressure and timed events. The Stockert Instrumente (Sorin) Pump is designed to operate on the principles of a peristaltic pump for extracorporeal use. The pump consists of a polyvinyl chloride tubing segment with polycarbonate connectors at each end. Flow rate is affected by the RPM's, of the pump rotor, total circuit resistance and the size of the tubing segment used. ## SPECIFICATIONS & MATERIALS A detailed comparison of specifications and materials can be found in Table 1. Specifications for the Stockert (Sorin) console are taken from Stockert's labeling. | | Affinity Console | Stockert Console | | --- | --- | --- | | Input Power | • 100 or 110-120 Vac, 50-60 Hz, • 220-240 Vac, 50-60 Hz | • 100 or 110-120 Vac, 50-60 Hz • 220-240 Vac, 50-60 Hz | | Internal Batteries | Yes | No (External Battery Pack Optional) | | Case Dimensions | 13" H x 7" W x 17" D | 12" H x 7" W x 19" D | | Weight | 50 lb. | 48 lb. | {2} # Pump Chamber Comparison of Materials | | Affinity | Stockert (Sorin) | | --- | --- | --- | | Chamber | Polyurethane | Polyvinyl Chloride | | Connectors | Polycarbonate | Polycarbonate | | Protective Caps | PVC | PVC | ## INDICATIONS FOR USE The Affinity Blood Pump System and the Stockert Instrumente (Sorin) Pump System are indicated for use in an extracorporeal perfusion circuit to pump blood during cardiopulmonary bypass procedures. ## PERFORMANCE CHARACTERISTICS The following tests were conducted by AVECOR to establish the safety and effectiveness of the Affinity Blood Pump System. | Test Description | Location | | --- | --- | | In Vitro Operating Characteristics | Section III | | In Vitro Cellular Damage | Section III | | Material Biocompatibility | Section IV | | Sterility | Section V | | Pyrogenicity | Section V | | Sterilant Residuals | Section V | | Software Validation & Verification | Appendix E | Protocols describing the test methods utilized in this application may be found in the appropriate Appendix. The results indicate that the Affinity Blood Pump System is substantially equivalent in performance to the Stockert Instrumente (Sorin) Pump System. ## BLOOD TRAUMA SUMMARY STATEMENT Blood trauma testing shows that in all aspects evaluated (hemolysis, platelet and white cell response) the Affinity Blood Pump System and the Stockert Instrumente (Sorin) Pump System are equivalent. {3} # CONCLUSION WE HAVE CONSIDERED DESIGN, MATERIALS, INDICATIONS FOR USE, PERFORMANCE CHARACTERISTICS, ALONG WITH A DIRECT COMPARISON OF TEST RESULTS BETWEEN THE AFFINITY AND STOCKERT PUMP SYSTEMS IN OUR APPLICATION. IT IS BASED UPON THE REVIEW OF ALL OF THESE PARAMETERS THAT AVECOR CARDIOVASCULAR INC. DETERMINES THAT THE AFFINITY BLOOD PUMP SYSTEM IS SUBSTANTIALLY EQUIVALENT TO THE CURRENTLY MARKETED STOCKERT INSTRUMENTS (SORIN) BLOOD PUMP SYSTEM. {4} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 Mr. Dennis E. Steger Director Regulatory Affairs/ Quality Assurance Avecor Cardiovascular, Inc. 7611 Northland Drive Minneapolis, Minnesota 55428 AUG - 5 1997 Re: K964017 AFFINITY™ Blood Pump System Regulatory Class: II (Two) Product Code: 74 DTQ Dated: May 9, 1997 Received: May 12, 1997 Dear Mr. Steger: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {5} Page 2 - Mr. Dennis E. Steger This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html." Sincerely yours, Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6} 870.4370 - DWB II - Roller-type CPB Blood Pump 510(k) Number (if known): K964017 Device Name: AFFINITY Blood Pump System Indications For Use: The Affinity™ Blood Pump System is intended to be used in an extracorporeal perfusion circuit to pump blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. ![img-0.jpeg](img-0.jpeg) Dennis E. Steger Director Regulatory Affairs/ Quality Assurance AVECOR Cardiovascular Inc. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-1.jpeg](img-1.jpeg) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K964017 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...