ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH

K022947 · Terumo Cardiovascular Systems Corp. · DTQ · Mar 20, 2003 · Cardiovascular

Device Facts

Record IDK022947
Device NameADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH
ApplicantTerumo Cardiovascular Systems Corp.
Product CodeDTQ · Cardiovascular
Decision DateMar 20, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4220
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Terumo® Advanced Perfusion System 1 is indicated for up to 6 hours of use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only, and intended to be used by a qualified perfusionist who is experienced in the operation of Sarns™ brand or similar equipment.

Device Story

Terumo Advanced Perfusion System 1 is a modular, configurable cardiopulmonary bypass heart-lung machine. System comprises a central chassis providing power/battery backup, a central control monitor with touch screen, and various plug-in modules (e.g., air bubble detection, level detection, temperature, pressure, flow, occluder, gas blender, roller pumps, centrifugal control). Operated by qualified perfusionists in clinical settings to manage patient perfusion during bypass. Central monitor allows configuration, display, and control of all connected modules; supports saving/recalling case-specific parameters. System provides real-time monitoring and control of perfusion circuit parameters, enabling perfusionists to customize system layout and functionality to meet specific procedural requirements. Benefits include enhanced control, flexibility in component selection, and streamlined parameter management compared to legacy systems.

Clinical Evidence

Bench testing only. Performance characteristics were exhaustively tested and compared against the Sarns 8000 and 9000 series systems and Delphin II modules. All new and existing performance characteristics were validated to confirm the system performs as intended according to specifications.

Technological Characteristics

Modular system with distributed network architecture. Components include central chassis, touch-screen central control monitor, and various plug-in modules (pumps, sensors, interface modules). Connectivity includes RS-232 and RS-485 interfaces. System supports configurable multi-module layouts. Computer-controlled operation with software-based parameter entry and configuration recall.

Indications for Use

Indicated for extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures for up to 6 hours. Intended for use by qualified medical professionals experienced in perfusion equipment operation. Centrifugal pump component indicated for cardiopulmonary bypass procedures only.

Regulatory Classification

Identification

A cardiopulmonary bypass heart-lung machine console is a device that consists of a control panel and the electrical power and control circuitry for a heart-lung machine. The console is designed to interface with the basic units used in a gas exchange system, including the pumps, oxygenator, and heat exchanger.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 1622947 TERUMO CARDIOVASCULAR SYSTEMS CORPORATION MAR 2 0 2003 #### ન ર 510(k) SMDA Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) Number is Submitter's Name: Terumo Cardiovascular Systems Corporation Submitter's Address: 6200 Jackson Road, Ann Arbor, Michigan 48103-9300 Contact Person: Mark A. Bur (734) 741-6117 Phone Number: FAX Number: (734) 663-5062 Summary Date: July 31, 2002 ### Device Trade Name: Terumo® Advanced Perfusion System 1 Device Classification Name: Cardiopulmonary bypass heart-lung machine (21 CFR 870.4220) #### Predicate Devices: Sarns™9000 Heart-Lung Console (K871131) Sarns™ 8000 Modular Perfusion System (K915183) Sarns™ 8000 Roller Pump (K953901) Sarns™ 9000 Universal Roller Pump (K953904) Flow Sensor Centrifugal Control Module (K950739) Delphin II/Stand-Alone Centrifugal Module (K935978) #### Indication for Use The Terumo® Advanced Perfusion System 1 is indicated for up to 6 hours of use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only, and intended to be used by a qualified perfusionist who is experienced in the operation of Sarns™ brand or similar equipment. #### Device Description: The Terumo® Advanced Perfusion Svstem 1 is a highly configurable system with a distributed network architecture that allows the user to customize the number and types of system components, which can then be configured, displayed, and controlled from a central monitor. Within reason, the system should allow users to define and configure a system to meet their individual requirements, and as needed, adapt the configuration as their needs change. The basic Terumo® Advanced Perfusion System 1 consists of a Chassis platform which provides operating power and battery back-up power for all other devices of the system that enable, facilitate or improve the intended use of the product and integrated functions, the Central Control Monitor which through its software and display touch screen facilitates the configuration and control of all other components, the Air Bubble Detect Module which is used to detect gross air bubbles in the arterial line of the perfusion circuit with the Terumo® Air Sensor, and the Level Detect Module which is used to monitor blood levels within a hard shell reservoir with the Terumo® Level Sensors. The following modules may also be used to enhance the abilities of the Terumo® Advanced Perfusion System 1: Temperature Module, Pressure Module, Flowmeter Module, Occluder Module, RS-232 Interface Module, RS-485 Interface Module, CDI™ 100 Interface Module, CDI™ 500 Interface Module, 6" diamter race Roller Pump, 4" diameter race Roller Pump, Centrifuqal Control Module, and a 33" or 15" Lamp. This design allows perfusionists to select the number and types of the {1}------------------------------------------------ RUMO CARDIOVASCULAR SYSTEMS CORPORATION accessory devices needed to control and monitor patient perfusion during cardiopulmonary bypass. Technological Characteristics: The Terumo® Advanced Perfusion System 1 is a combination of the technologies from the current legally marketed devices, Sarns™ 8000 Modular Perfusion System, Sarns™9000 Heart-Lung Console, Sarns™ 8000 Roller Pump, Sarns™ 9000 Universal Roller Pump, Flow Sensor Centrifugal Control Module, and Delphin II/Stand-Alone Centrifugal Module and present day technology for computer controlled systems. The main differences between the predicate devices, Sarns™ 8000 Modular Perfusion System, Sarns™9000 Heart-Lung Console, and Delphin II/Stand-Alone Centrifygal Module, and the Terumo® Advanced Perfusion System 1 are: - · individual modules applications compared to the subsystem modules in the 8000 system. - · enhanced parameter entry for all modules via the Central Control Monitor compared to the minimal parameter user entry of the 9000 system via its touch screen. - · ability to control Electronic Gas Blender, Roller Pumps, and Centrifygal Control Module of the system via the Central Control Monitor or individual module controls compared to the individual controls required of the 8000, 9000, and Delphin standalone system. - ability to setup multiple case configurations parameters on the Central Control Monitor software and recall the configuration parameters for use in a future cases compared to only one past systems setup and parameter setup each time, - · ability to configure a multi-module system for use of the users choice desired components rather than be constricted to only the few components available and the manufacturers platform layout on past systems. # Non-clinical Performance: The performance characteristics of the Terumo® Advanced Perfusion System 1 were exhaustively tested and compared with the performance characteristics of the currently marketed Sarns™ 8000 Modular Perfusion System, Sarns™9000 Heart-Lung Console, Sarns™ 8000 Roller Pump, Sarns™ 9000 Universal Roller Pump, Flow Sensor Centrifugal Control Module, and Delphin II/Stand-Alone Centrifugal Module. All new and existing performance characteristics of the Terumo® Advanced Perfusion System 1 have been validated. # Conclusion from Conducted Testing: The Terumo® Advanced Perfusion System 1 perform as intended according to its performance specifications. The Terumo® Advanced Perfusion System 1 is substantially equivalent to its predicated devices. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 2 0 2003 Terumo Cardiovascular Systems Corporation c/o Mr. Mark A. Bur 6200 Jackson Road Ann Arbor, MI 48103-9300 Re: K022947 Advance Perfusion System 1 Regulation Number: 21 CFR 870.4220 Regulation Name: CPB Heart Lung Machine Console Regulatory Class: Class II (two) Product Code: DTQ Dated: December 20, 2002 Received: December 23, 2002 ### Dear Mr. Bur: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Mark A. Bur Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours, N.D. Zuckerman, M.D. ram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K022947 510(k) Number (if known): Terumo® Advanced Perfusion System 1 Device Name: Indications For Use: The Terumo® Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The centrifuqal pump is indicated for use in cardiopulmonary bypass procedures only. Mark A. Burr Mark A. Bur Regulatory Affairs Specialist Terumo Cardiovascular Systems (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) N. Cole Th 510(k) Number K022947 Prescription Use X OR Over-The-Counter Use (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...