BMR 4500 FILTERED VENOUS RESERVOIR AND BMR -4500 GOLD FILTERED VENOUS RESERVOIR WITH DURAFLO TREATMENT
Applicant
Baxter Healthcare Corp
Product Code
DTN · Cardiovascular
Decision Date
Jan 22, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4400
Device Class
Class 2
Indications for Use
The BMR™-4500 Filtered Venous Reservoir is indicated for use during cardiopulmonary bypass surgery in the extracorporeal circuit to collect, store, filter and defoam the patient's blood. The BMR™-4500 GOLD® Hard Shell Venous Reservoir with Duraflo® treatment is indicated for use during cardiopulmonary bypass surgery in the extracorporeal circuit to collect, store, filter and defoam the patient's blood. The BMR-4500 GOLD® is further intended for use when a heparin treated blood path is desired.
Device Story
Hard shell venous reservoir used during cardiopulmonary bypass; collects, stores, filters, and defoams patient blood. Operates within extracorporeal circuit; handles blood flows up to 7.0 LPM (venous inlet) or 5.0 LPM (cardiotomy inlet). GOLD® version features Duraflo® complex surface treatment to enhance blood compatibility; non-treated version also available. Device includes rotatable venous/cardiotomy inlets and supplemental luer fittings. Used by perfusionists/clinicians in surgical settings. Provides mechanical blood management to support bypass procedures; benefits include filtration and defoaming to maintain blood integrity during extracorporeal circulation.
Clinical Evidence
No clinical data. Safety and efficacy established via in-vitro performance testing and historical experience with similar devices. In-vitro testing included venous/cardiotomy blood damage, filtration efficiency, breakthrough/residual volumes, defoaming capability, valve patency, and heparin leaching, conducted per AAMI OXY-D and AAMI AT6R-D guidelines. Biocompatibility confirmed per ISO-10993-1.
Technological Characteristics
Hard shell venous reservoir; materials include biocompatible polymers (compliant with ISO-10993-1). GOLD® version features Duraflo® complex surface treatment. Features rotatable venous and cardiotomy inlets and luer fittings. Mechanical operation; no energy source required. Designed for flows up to 7.0 LPM. Non-software device.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery requiring extracorporeal blood collection, storage, filtration, and defoaming. The GOLD® version is indicated when a heparin-treated blood path is desired.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Bentley HSRTM-4000 Venous Reservoir with Cardiotomy Autotransfusion Filter
Submission Summary (Full Text)
{0}------------------------------------------------
K974155
| Submitter: | JAN 22 1998 | Baxter Healthcare Corporation<br>Cardio Vascular Group<br>Bentley Division<br>17511 Armstrong Avenue<br>Irvine California 92714 USA |
|---------------------------------------|-------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | | Tom Humphrey<br>Phone: 714 756 4292<br>Fax: 714 756 4021 |
| Device Trade Name: | | BMRTM-4500 Filtered Venous Reservoir<br>and BMRTM-4500 GOLD® Filtered<br>Venous Reservoir with Duraflo®<br>Treatment |
| Common Name: | | Cardiopulmonary bypass venous reservoir<br>with cardiotomy filter. |
| Classification: | | Class II (Reference 21 CFR 870.4400) |
| Predicate or Legally Marketed Device: | | Bentley HSRTM-4000 Venous Reservoir<br>with Cardiotomy Autotransfusion Filter. |
| Date Prepared: | | 11/3/97 |
# 510(k) Summary, Safety and Effectiveness
### Device Description:
The BMR™-4500 GOLD® Filtered Venous Reservoir with Duraflo® Treatment is used during cardiopulmonary bypass surgery to collect, store, filter and defoam the patient's blood. All blood contacting surfaces of the device, with the exception of luer connectors, stopcock manifold and sample lines, are treated with Bentley's Duraflo® complex. This treatment is intended to enhance the blood compatibility of non-biological surfaces in the extracorporeal circuit.
{1}------------------------------------------------
Bentley's BMR™-4500 GOLD® will also be offered in a non-treated version. Designated BMR™-4500™, this device will not contain the Duraflo® complex. The indications for use for BMR™-4500™ are identical to the treated version of the device.
Bentley's BMR™-4500™ and BMR™-4500 GOLD® are intended to be used for blood flows of up to 7.0 LPM through the venous inlet, up to 5.0 LPM through the cardiotomy inlet or up to 7.0 LPM combined flow.
#### Indications for Use:
The BMR™-4500 Filtered Venous Reservoir is indicated for use during cardiopulmonary bypass surgery in the extracorporeal circuit to collect, store, filter and defoam the patient's blood.
The BMR™-4500 GOLD® Hard Shell Venous Reservoir with Duraflo® treatment is indicated for use during cardiopulmonary bypass surgery in the extracorporeal circuit to collect, store, filter and defoam the patient's blood. The BMR-4500 GOLD® is further intended for use when a heparin treated blood path is desired.
#### Technology Comparison
With exception to the addition of rotatable venous and cardiotomy inlets and supplemental luer fittings, the BMR™-4500, BMR™-4500 GOLD® and HSR™-4000 are essentially the same. The only difference between the BMR™-4500 and BMRTM. 4500 GOLD® is the inclusion of Bentley's Duraflo® complex in the "GOLD®" version.
{2}------------------------------------------------
## Test Summary, In-vitro
The untreated BMR™-4500 and Duraflo® treated BMR™-4500 GOLD® Reservoirs were evaluated in-vitro for functional performance. The results were compared with those of the untreated HSR™-4000. The number of devices tested was selected based on AAMI guidelines (AAMI OXY-D). AAMI standards for autologous transfusion devices (AAMI AT6R-D) were also considered in the performance evaluation. The parameters under which the device was tested were considered worst case conditions.
Performance evaluation included the following tests:
- Venous and Cardiotomy Blood Damage .
- . Filtration Efficiency
- . Venous and Cardiotomy Breakthrough Volume
- . Venous and Cardiotomy Residual Volume
- . Defoaming Capability
- Patency of One-Way Valve .
- . Heparin Leaching
The test results demonstrated no remarkable differences between the proposed vs. predicate populations.
### Test Summary, In-vivo:
Clinical testing was not conducted. Safety and efficacy has been established via in-vitro analyses and historical experience with similar devices.
{3}------------------------------------------------
#### Test Summary, Biocompatibility:
All components of the BMR™-4500 and BMR™-4500 GOLD® were found to be biocompatible in compliance with ISO-10993-1.
## Rational for Substantial Equivalence Determination
The coated and uncoated versions of Bentley's BMR™-4500 are either functionally identical or substantially equivalent to the predicate device, Bentley's HSR™-4000. The primary raw materials, principal of operation, developmental technology, functionality, intended use and labeling claims are essentially the same for all products referenced under this notification. The only notable differences between the proposed vs. predicate device are the autotransfusion function of the HSR™-4000 and addition of Bentley's Duraflo Filtered Venous Reservoir complex to the blood contacting surfaces of the GOLD® version of the device.
Bentley's Duraflo® Treatment is utilized as a biocompatible surface treatment on several devices manufactured and distributed by Baxter Healthcare corporation, Bentley Division. Numerous Duraflo treated devices have been previously "cleared" by FDA.
The results of the above referenced analyses demonstrate no remarkable differences between the proposed and predicate populations. In addition to in-vitro assessment, the device materials, intended use and technological characteristics of the proposed vs. predicate devices are largely identical. The test results and strong similarities in raw materials and intended use between the BMR™-4500, BMR™-4500 GOLD® and HSR™-4000, combined with Bentley's historical experience in developing similar products, provide strong rational for substantial equivalence between the proposed and predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that form the head, body, and wings. The text is in a simple, sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Tom Humphrey Sr. Requlatory Affairs Specialist Baxter Healthcare Corporation 17221 Red Hill Avenue (Irvine) P.O. Box 11150 Santa Ana, California 92711-1150
JAN 2 2 1998
Re : K974155 BMR™-4500 Filterted Venous Reservoir and BMR™-4500 GOLD® Filtered Venous Reservoir with Duraflo® Treatment Requlatory Class: II (Two) Product Code: DTN Dated: November 3, 1997 November 4, 1997 Received:
Dear Mr. Humphrey:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such Existing major requlations affecting your device additional controls. can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Mr. Tom Humphrey
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,
Thomas J. Callahon
Thomas J. CallaMan, Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
DIN II -870. 4400 -
Venous
Reservoir
510(k) Number (if known): k974155
Device Name: BMR™-4500 and BMR™-4500 GOLD® Filtered Venous Reservoir with Duraflo® Treatment --
Indications For Use:
The BMR™-4500 Filtered Venous Reservoir is indicated for use during cardiopulmonary bypass surgery in the extracorporeal circuit to collect, store, filter and defoam the patient's blood.
The BMR™-4500 GOLD® Hard Shell Venous Reservoir with Duraflo® treatment is indicated for use during cardiopulmonary bypass surgery in the extracorporeal circuit to " collect, store, filter and defoam the patient's blood. The BMR-4500 GOLD® is further intended for use when a heparin treated blood path is desired.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office OF Device Evaluation (ODE)
Bura G. Remmersle
Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
K974155 510(k) Number_
Prescription Use
OR
Over-The-Counter Use_
(Per 21 CFR 801.109) --------------
x
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.