← Product Code [DTN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN) · K953821

# BIOCOR 200 HARDSHELL VENOUS RESERVIOR (K953821)

_Minntech Corp. · DTN · Mar 7, 1996 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN/K953821

## Device Facts

- **Applicant:** Minntech Corp.
- **Product Code:** [DTN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN.md)
- **Decision Date:** Mar 7, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4400
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Biocor™ 200 Hardshell Venous Reservoir is safe and effective as a storage reservoir for venous blood and for blood recovered by intrathoracic suction.

## Device Story

Hardshell venous reservoir; stores venous blood and blood recovered via intrathoracic suction during cardiopulmonary bypass. Device functions as passive storage container; integrates cardiotomy filtration. Used in clinical settings (OR) by perfusionists/surgeons. Provides reservoir capacity for extracorporeal circulation; ensures blood volume management and filtration of debris/particulates during bypass. Benefits patient by maintaining hemodynamic stability and blood integrity during cardiac surgery.

## Clinical Evidence

Bench testing only. In-vitro performance testing included volume calibration, dynamic priming volume, breakthrough volume, hold-up volume, air removal/separation, particulate filtration efficiency, and blood cell damage. Biocompatibility testing included cytotoxicity, hemolysis, acute systemic toxicity, intracutaneous reactivity, sensitization, AMES mutagenicity, in vitro hemocompatibility (PTT, PT), thrombosis formation, and complement activation. Results demonstrated non-cytotoxic, non-hemolytic, non-sensitizing, and non-toxic profile with acceptable clotting and complement activation.

## Technological Characteristics

Hardshell reservoir design; includes integral cardiotomy filter. Materials demonstrated biocompatible per ISO/FDA standards. Functions as passive storage/filtration unit for extracorporeal circuits. No energy source or software components.

## Regulatory Identification

A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.

## Special Controls

*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Affinity CVR Cardiotomy/Venous Reservoir with filter (Avecor Cardiovascular)
- Sorin HSVRF Hardshell Venous Reservoir with Integral Cardiotomy Filter (Sorin Biomedical)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K953821

# BIOCOR™ 200 HARDSHELL VENOUS RESERVOIR

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

Minntech Corporation has provided the following information to the US Food and Drug Administration to support that the Biocor™ 200 Hardshell Venous Reservoir is substantially equivalent to other devices currently in commercial distribution within the United States.

1. Comparison to Another Device in Commercial Distribution within the United States

The Biocor™ 200 Hardshell Venous Reservoir is comparable in regards to its intended use and technological characteristics to other device in commercial distribution within the United States. Minntech Corporation has compared the similarities and difference of the Biocor™ 200 Hardshell Venous Reservoir to the Affinity CVR Cardiotomy/ Venous Reservoir with filter (Avecor Cardiovascular) and the cardiopulmonary bypass indications of the Sorin HSVRF Hardshell Venous Reservoir with Integral Cardiotomy Filter (Sorin Biomedical). This comparison supported comparability with respect to intended use, design features, and product specifications.

2. Summary of Safety and Effectiveness

2.1 In-Vitro Performance Testing

Minntech Corporation has described the test plan for the Biocor™ 200 Hardshell Venous Reservoir and has provided the specific test results. The plan included a comparative analysis performance and structural integrity. Tests included volume calibration, dynamic priming volume, breakthrough volume, hold up volume, air removal/separation, particulate filtration efficiency, general performance assessment, and blood cell damage. Data support that the performance and integrity of the Biocor™ 200 Hardshell Venous Reservoir is suitable for its intended use and comparable to the Affinity CVR Cardiotomy/Venous Reservoir with filter (Avecor Cardiovascular) and the cardiopulmonary bypass indications of the Sorin HSVRF Hardshell Venous Reservoir with Integral Cardiotomy Filter (Sorin Biomedical).

P. 02 FAX NO. 16125533387 MUNNTECH Wd 89:21 18d 96-24-96 FRI 12:58 PM

{1}

## 2.2 Biocompatibility Studies

Minntech Corporation has described the test plan for the Biocor™ 200 Hardshell Venous Reservoir and has provided the specific test results. The biocompatibility studies conducted were cytotoxicity, hemolysis, acute systemic, intracutaneous, sensitization AMES mutagenicity, in vitro hemocompatibility UPTT, PT, thrombosis formation and complement activation. Under the conditions of these studies, the Biocor™ 200 Hardshell Venous Reservoir was demonstrated to be non-cytotoxic, non-hemolytic, non-sensitizing, non-toxic and non-mutagenic. The Biocor™ 200 Hardshell Venous Reservoir is hemocompatibility and has acceptable clotting time and complement activation profiles. Thrombosis formation was not detected. Based on these results, the Biocor™ 200 Hardshell Venous Reservoir is safe for use in cardiopulmonary bypass procedures.

## 3. Summary of Substantial Equivalence

Minntech Corporation has provided the above information within the 510(k) Premarket Notification to support that the Biocor™ 200 Hardshell Venous Reservoir is safe and effective as a storage reservoir for venous blood and for blood recovered by intrathoracic suction. Additionally, the Biocor™ 200 Hardshell Venous Reservoir has been shown to be comparable to other devices currently in commercial distribution. Data and information provided within this 510(k) Premarket Notification adequately support that the Biocor™ 200 Hardshell Venous Reservoir is substantially equivalent to other devices currently in commercial distribution.

P. 03
FAX NO. 16125533387
MINNTECH
MAY-24-96 FRI 12:59 PM

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN/K953821](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN/K953821)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
