K050195 · Gish Biomedical, Inc. · DTN · Mar 17, 2005 · Cardiovascular
Device Facts
Record ID
K050195
Device Name
V-BAG WITH GBS COATING
Applicant
Gish Biomedical, Inc.
Product Code
DTN · Cardiovascular
Decision Date
Mar 17, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Circulatory Technology Inc. V-BAG™ with GBS® Coating is a disposable soft shell venous reservoir that accepts venous and cardiotomy blood and facilitates the removal of air bubbles during surgical procedures requiring extracorporeal support for up to six hours. Federal Law (US) restricts this device to sale by or on the order of a physician.
Device Story
Disposable soft-shell venous reservoir; accepts venous and cardiotomy blood during extracorporeal circulation. Device consists of two welded polyvinyl chloride films forming an expandable chamber; internal polyester screen (105µ pore size) separates inlet from outlet. Principle of operation: blood flows across wet screen, which allows liquid passage while rejecting air bubbles; bubbles rise to top and are removed via purge line. Vertical channels along tube diameter facilitate upward bubble movement to purge port. Used in clinical surgical settings; operated by perfusionists or surgical staff. Provides air removal and blood storage; supports extracorporeal circulation for up to six hours. Benefits include reduced air embolism risk during bypass procedures.
Clinical Evidence
Bench testing only. Device subjected to extensive safety, performance, and validation testing to ensure functional requirements and specifications are met. No clinical trial data provided.
Technological Characteristics
Soft-shell reservoir; materials: polyvinyl chloride (PVC) film, polyester screen (105µ pore size). Features GBS® heparin coating. Rigid top-plate with inlet, outlet, and gas purge ports. Standalone device for extracorporeal circuits. Sterilization method not specified.
Indications for Use
Indicated for patients undergoing surgical procedures requiring extracorporeal support for up to six hours, including Venous Drainage (VAVD). Contraindicated in patients with known or suspected heparin sensitivity due to the heparin-based GBS® coating.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
## MAR 17 2005
K 050195
# Summary
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR part 807.92.
- 1. Company making the submission:
| Company | |
|------------|------------------------|
| Name: | Gish Biomedical, Inc. |
| Address: | 22942 Arroyo Vista |
| | Rancho Santa Margarita |
| | CA 92688-2600 |
| Telephone: | 949-635-6240 voice |
| | 949-635-6294 fax |
| Contact: | Edward F. Waddell |
| | Director RA/QA |
- 2. Device:
| Proprietary Name: | Circulatory Technology Inc. V-BAG™ with GBS® Coating. |
|----------------------|-------------------------------------------------------|
| Common Name: | Soft Venous Reservoir |
| Classification Name: | Reservoir, blood, cardiopulmonary bypass |
- 3. Predicate Devices:
Circulatory Technology Inc. V-BAG™. K001822 and Gish Soft Venous Reservoir SVR with GBS™ Coating, K024065,
- Classifications Names & Citations: 4.
- 21 CFR 870.4400, Reservoir, blood, cardiopulmonary bypass, Class II, DTN, Cardiovascular. 5.
- 6. Description:
The V-BAG™ with GBS® Coating is an expandable blood chamber having a polyester screen with a pore size of 105u separating the inlet from the outlet of the blood chamber. The inlet with a pore the blood chamber at its top and extends to the bottom of the screen. Blood flows table oncert the became the screen, and to the outlet port. The screen, once wet, allows mom the mid liguid but rejects most air bubbles float up and are removed through the purge line. Channels formed along the outside diameter of the tubes placed vertically, running parge inforcement of the inlet section to the top of the blood chamber, provide pathways for hrom the to move upward to the purge port. The inlet, outlet and gas purge tubes are threaded through, bonded to, and sealed within a rigid top-plate.
The V-Bag™, is made by welding two polyvinyl chloride films to form an expandable blood chamber with a front and back wall. A polyester screen with a pore size of 105µ is sealed between the two walls along their vertical sides as well as along the front wall at the bottom thereby forming a pouch between the screen and the front wall. The inlet tube enters the blood chamber at its top and extends diagonally to the bottom of the pouch. Blood flows from across the screen to the outlet port. The screen, once wet, allows passage of liquid but rejects most air bubbles. Bubbles in the pouch float up and are removed through the purge line.
25
{1}------------------------------------------------
- 7. Indications for use:
The Circulatory Technology Inc. V-BAG™ with GBS® Coating is a disposable soft shell The Circulatory Technology Inc. V-DAS With ODO Obailighters the removal of air venous reservoir mat accepts venous and caracomy presal support for up to six hours. When
bubbles during surgical procedures requiring extracorporal for up to six hours. IVAV bublies during surgical procedures requiring oxicusively and opportunities and the enous Drainage (VAVD).
- Contraindications: 8.
For heparin coated devices, heparin has been reported, on rare occasions, to induce For heparin coated devices, hepanif has been reports and supers are routinely
thrombocytopenia. Since patients undergoing cardiopulmonary bypass are routinely thrombocytopenia. Since patients undergoing careformedia on this device is
systemically heparinized, and although the amount of heparing which a exerciaed when web systemically neparmized, and almough the arriven caution should be exercised when using very small in oompanse with known or suspected heparin sensitivity.
- 9. Comparison:
The Circulatory Technology Inc. V-BAG™ with GBS® Coating has the same device The Circulatory Technology inc. V-BAG, Circulatory Technology, Inc., K001822.
- 10. Test Data:
The Circulatory Technology Inc. V-BAG™ with GBS® Coating has been subjected to The Circulatory Technology Inc. V-DAO - with SBO-Socking has a subsition for the systems
extensive safety, performance, and validations prior to release. Final testing for th includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications.
- 11. Literature Review:
A review of literature pertaining to the safety and effectiveness has been conducted. A review of literature perfailing to the safety of Circulatory Technology Inc. V-BAG™ with GBS® Coating.
- 12. Conclusions:
The conclusion drawn from these tests is that Circulatory Technology Inc. V-BAG™ with GBS® Coating is equivalent in safety and efficacy to its predicated devices.
Edward Waddell
Edward Wadde Director Regulatory Affairs Gish Biomedical. Inc
Date: Jan 25, 2005
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
MAR 1 7 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Gish Biomedical, Inc. c/o Mr. J. Harvey Knauss Delphi Consulting Group 11874 South Evelyn Circle Houston, TX 77071-3404
Re: K050195
V-BAG™ with GBS® Coating Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II (two) Product Code: DTN Dated: March 1, 2005 Received: March 2, 2005
Dear Mr. Knauss:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 310(x) premained is substantially equivalent (for the indications referenced above and nave decemblicated predicate devices marketed in interstate for use stated in the enclosure) to regarly manated provice Americal Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food . Drya commerce prior to May 26, 1976, the enaonine with the provisions of the Federal Food. Drug, devices that have been reclassified in accereanse with a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premaining of the Act, and Cosmelle Act (Act) that do not require approvisions of the general controls provisions of the Act. The You may, therefore, market the act include requirements for annual registration, listing of general controls provisions of the free herals of the more of the manage inst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 a0070) als. Existing major regulations affecting your device can may be subject to such additional centrely of Title 21, Parts 800 to 898. In addition, FDA may be found in the Code of Peachar Ing your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Mr. J. Harvey Knauss
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reaso or as made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny reather statuates as a s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of It Part 80 73 regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) rramarket notification. The FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
parna R. Values
E
L
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use Statement
510(k) Number K050195
Circulatory Technology Inc. V-BAG™ with GBS® Coating Device Name:
Indications for use:
The Circulatory Technology Inc. V-BAG™ with GBS® Coating is a disposable soft shell The Circulatory Technology inc. V BAO the cardiotomy blood and facilitates the removal of air venous reservoir that accepts vehous and our and survive arracorporeal support for up to six hours.
bubbles during surgical procedures requiring extracorporeal support for up bubbles during surgical procedures roquiring oxiraces virus Drainage (VAVD).
#### Prescription Device:
Federal Law (US) restricts this device to sale by or on the order of a physician.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ 21 CFR 807 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dvina R. Vechner
Division Sign-Off) ision of Cardiovascular Devices
umber E050195
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.