K031150 · Global Blood Resources, LLC · DTN · Jun 23, 2003 · Cardiovascular
Device Facts
Record ID
K031150
Device Name
HEMOBAG, MODEL HBBSD2000
Applicant
Global Blood Resources, LLC
Product Code
DTN · Cardiovascular
Decision Date
Jun 23, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HEMOBAG™ is a blood circulating storage reservoir designed to collect anticoagulated whole blood salvaged from extracorporeal circuits after the patient has been disconnected from the circuit. When used in connection with the TS3 Tubing Set and a commercially available hemoconcentrator, the HEMOBAG™ serves as a reservoir for circulating and hemoconcentrating blood in a closed circuit recovery loop.
Device Story
HEMOBAG™ is a flexible storage reservoir for salvaged anticoagulated whole blood. Used in closed circuit recovery loops with TS3 Tubing Set and hemoconcentrator; facilitates removal of excess plasma water and low molecular weight solutes. Features internal baffle for mixing/flow, universal stepped arterial infusion port, quick-connect/disconnect ports, and needleless access port. Designed for gravity transfusion via integrated spike port. Operated by clinical staff in surgical/clinical settings. Benefits patient by enabling autologous blood recovery and concentration, reducing need for donor blood.
Clinical Evidence
Bench testing only. Biocompatibility evaluated per ANSI/AAMI/ISO 10993. Hemolysis and cellular depletion testing performed using anticoagulated bovine whole blood circulated at 1.2 L/min for 1 hour; results showed statistically insignificant differences compared to predicate. Functional effectiveness tested by overfilling with water (10-20% excess) and circulating at 2000 ml/min for 2 hours.
Technological Characteristics
Flexible reservoir bag with internal baffle; 2000ml capacity. Materials biocompatible per ANSI/AAMI/ISO 10993. Features universal 1/4"-3/8" stepped connector, 1/4" quick-connect/disconnect ports, needleless port, and spike port. Mechanical device; no energy source or software.
Indications for Use
Indicated for patients requiring salvage of anticoagulated whole blood from extracorporeal circuits for hemoconcentration and subsequent gravity transfusion.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Related Devices
K031151 — TS3 TUBING SET · Global Blood Resources, LLC · Jun 26, 2003
K970739 — RMI HEMOCONCENTRATOR PREFERENCE PACK · Research Medical, Inc. · Jul 8, 1997
K100507 — XRES BLOOD COLLECTION RESERVOIR · Sorin Group Italia S.R.L. · Mar 24, 2010
K241236 — XTRA Collection sets; XTRA Sequestration set X · Sorin Group Italia S.R.L. · Aug 16, 2024
K053000 — HAEMONETICS CARDIOVASCULAR PERIOPERATIVE AUTOTRANSFUSION SYSTEM (CARDIOPAT) · Haemonetics Corp. · Nov 21, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
## JUN 2 3 2003
K 031150
## 510(k) SUMMARY
The HEMOBAG™ Blood Salvage Device is a storage reservoir for blood that can be held, circulated and hemoconcentrated. It is designed for use in salvaging anticoagulated whole blood from extracorporeal circuits. When used in conjunction with a TS3 Tubing Set and a commercially available hemoconcentrator, the HEMOBAG™ serves as a hemoconcentrating reservoir for use in a closed circuit recovery loop for removal of excess plasma water and low molecular weight solutes. The richly concentrated, anticoaqulated, autologous whole blood is then quickly made available to the patient for gravity transfusion.
The HEMOBAG™ is designed with a universal (1/4" – 3/8″) stepped connector on the arterial infusion port to allow lines of varying sizes to be attached to the bag. The inlet and outlet ports of the bag are designed with quick connect/disconnect ¼" connectors to mate with the TS3 Tubing set for fluid delivery to and from the hemoconcentrator. The HEMOBAG™ is equipped with a needleless port for ease of access. The HEMOBAG™ has been designed with a spike port to facilitate gravity transfusion. A baffle has been positioned inside the bag to allow the blood from the hemoconcentrator to enter the bag and circulate toward the top of the bag and down again to the outlet port allowing proper mixing and flow during hemoconcentration.
#### SAFETY TESTING
The sterile HEMOBAG™ was subjected to biological evaluation (biocompatibility testing) using the guidelines provided in ANSI/AAMI/ISO 10993, Vol. 4 for limited exposure medical devices that have contact with circulating blood. The device successfully passed all of these tests.
The HEMOBAG™ was evaluated to determine whether or not it introduced more trauma to the blood than the predicate device. The evaluation was done by determining the cellular depletion and hemolysis of anticoaqulated bovine whole blood that had been circulated through the HEMOBAG™ at a rate of approximately 1.2 liters per minute for a 1-hour period. An identical evaluation was performed on the anticoagulated bovine whole blood that had been circulated through the predicate device in the same manner. The results of these tests proved that the difference between the amount of cellular damage noted in the blood that had been circulated through the HEMOBAG™ and that which had been circulated through the predicate device was statistically insignificant. Cellular damage was measured by change in hematocrit, platelet depletion, white blood cell count, plasma free hemoglobin and calculated % hemolysis. Blood in the test circuits was evaluated and found to have demonstrated acceptable hematological parameters during the test period. Since the test parameters used represent a scenario that is far worse than parameters that will be seen in proper clinical use, this device is considered safe with respect to cellular damage and hemolysis when used under abnormal stressful conditions.
#### EFFECTIVENESS
The function and purpose of this device is to accommodate and hold up to 2000ml of anticoagulated whole blood salvaged from an extracorporeal circuit after the patient has been removed. The device also functions as a reservoir for the same blood as it is hemoconcentrated by ultrafiltration for ultimate gravity transfusion to the patient. The HEMOBAG™ has been designed using materials and processes that will allow the flexible bag to hold up to 2000ml of fluid circulating at a rate of 300 – 500ml per minute for a period of approximately 10 – 15 minutes. The functionality of the HEMOBAG™ was successfully tested by overfilling the bag with warm water by 10 -- 20% and allowing the bag to hang for a period of 2 hours and then water to circulate through the bag at a rate of 2000 ml per minute for a period of 2 hours. Since the test parameters used represent a scenario far worse than those seen in clinical use, the device is considered effective with respect to its ability to serve as a reservoir for holding, circulating and hemoconcentrating up to 2000ml of anticoagulated whole blood. The rounded corners and baffle inside the bag allow mixing and proper flow of the blood through the HEMOBAG™ during hemoconcentration.
### APPENDIX D
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with a circular border containing text, though the text is too blurry to read. Inside the circle is a stylized symbol consisting of three curved lines that resemble a bird in flight or a flowing design. The logo is black and white and appears to be a seal or emblem of some kind.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 3 2003
Global Blood Resources, LLC Mr. Keith Samolyk, CCP President & CEO 998 Windsor Avenue Windsor, CT 06096
Re: K031150
Trade/Device Name: HEMOBAGTM Regulation Number: 21 CFR 870.4400 Regulation Name: Blood Reservoir Regulatory Class: Class II (two) Product Code: DTN Dated: April 7, 2003 Received: April 10, 2003
Dear Mr. Samolyk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Keith A. Samolyk, CCP.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Afgealngpl
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
JUN 2 3 2003
# INTENDED USE
The HEMOBAG™ is a blood circulating storage reservoir designed to collect anticoagulated whole blood salvaged from extracorporeal circuits after the patient has been disconnected from the circuit. When used in connection with the TS3 Tubing Set and a commercially available hemoconcentrator, the HEMOBAG™ serves as a reservoir for circulating and hemoconcentrating blood in a closed circuit recovery loop.
C. Wyndham for B. Zuckerman
ivision of Cardiovascular Devices
KO31150 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.