← Product Code [DTM](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM) · K242092

# MICRO Arterial Filters (K242092)

_Sorin Group Italia S.R.L. · DTM · Nov 18, 2024 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM/K242092

## Device Facts

- **Applicant:** Sorin Group Italia S.R.L.
- **Product Code:** [DTM](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM.md)
- **Decision Date:** Nov 18, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4260
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The devices are recommended for use on the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filters are used to trap and remove gaseous emboli as well as particulate debris that may be introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer period is not advised.

## Device Story

Single-use arterial line blood filter; used in extracorporeal circuits during cardiopulmonary bypass procedures. Input: blood flow from arterial line. Operation: mechanical filtration via silicon net to trap/remove gaseous emboli and particulate debris. Output: filtered blood returned to patient. Used in clinical settings (OR) by perfusionists/clinicians. Benefits: reduces risk of emboli/debris reaching patient during bypass. Modified from predicate devices to remove DEHP plasticizer from PVC tubing and update silicon valve diaphragm formulation.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical verification and validation testing, including compliance with applicable national and international standards for arterial filters.

## Technological Characteristics

Materials: PVC (DEHP-free), silicon filtering net, silicon valve diaphragm. Principle: mechanical filtration of blood. Form factor: arterial line filter. Sterilization: ethylene oxide. Single-use, non-pyrogenic fluid path.

## Regulatory Identification

A cardiopulmonary bypass arterial line blood filter is a device used as part of a gas exchange (oxygenator) system to filter nonbiologic particles and emboli (blood clots or pieces of foreign material flowing in the bloodstream which will obstruct circulation by blocking a vessel) out of the blood. It is used in the arterial return line.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions.”

## Predicate Devices

- DIDECO ADULT ARTERIAL FILTERS D732 MICRO 20 W/20, D734 MICRO 40 W/40 MICRON SCREEN ([K952270](/device/K952270.md))
- DIDECO MICRO 40 PH.I.S.I.O. ADULT ARTERIAL FILTER ([K040184](/device/K040184.md))
- D731 MICRO 20 AND D733 MICRO 40 PEDIATRIC ARTERIAL FILTERS ([K041061](/device/K041061.md))
- D731 MICRO 27 PH.I.S.I.O. AND D733 MICRO 40 PH.I.S.O. ARTERIAL FILTERS ([K112525](/device/K112525.md))
- DIDECO NEWBORN/INFANT ARTERIAL FILTERS: D735 MICRO 20 AND D736 MICRO 40 ([K961869](/device/K961869.md))
- D735 MICRO 20 AND D736 MICRO 40 NEWBORN-INFANT ARTERIAL FILTERS ([K033987](/device/K033987.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

November 18, 2024

Sorin Group Italia S.r.1. Luigi Vecchi Director, Regulatory Affairs CP Disposables Via Statale 12 Nord, 86 Mirandola (Modena), 41037 Italy

# Re: K242092

Trade/Device Name: MICRO Arterial Filters Regulation Number: 21 CFR 870.4260 Regulation Name: Cardiopulmonary Bypass Arterial Line Blood Filter Regulatory Class: Class II Product Code: DTM Dated: October 18, 2024 Received: October 18, 2024

# Dear Luigi Vecchi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

# Nicole M. Gillette -S

Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K242092

Device Name

MICRO Arterial Filters

### Indications for Use (Describe)

"The devices are recommended for use on the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass.

The filters are used to trap and remove gaseous emboli as well as particulate debris that may be introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer period is not advised."

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for LivaNova. The logo is purple and consists of the word "LivaNova" in a sans-serif font. Below the word "LivaNova" is the tagline "Health innovation that matters" in a smaller, orange font. The logo is simple and modern, and the colors are eye-catching.

# 510(k) Summary

(in accordance with 21 CFR 807.92)

510(k) Number: K242092

| Applicant Information<br>Applicant: | SORIN GROUP ITALIA S.R.L.<br>Via Statale 12 Nord, 86<br>Mirandola, MO 41037 Italy                               |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Contact Person:                     | Martina Carlini<br>Regulatory Affairs Specialist<br>Tel: +39 0535 29811<br>e-mail: martina.carlini@livanova.com |
| Application Correspondent:          | SORIN GROUP ITALIA S.R.L.<br>Via Statale 12 Nord, 86<br>Mirandola, MO 41047 Italy                               |
| Contact Person:                     | Martina Carlini<br>Regulatory Affairs Specialist<br>Tel: +39 0535 29811<br>e-mail: martina.carlini@livanova.com |
| Date Prepared:                      | November 14th 2024                                                                                              |

#### II. Subject Device Identification

| Device Trade Name:    | MICRO Arterial Filters                               |
|-----------------------|------------------------------------------------------|
| Classification Name:  | Filter, blood, cardiopulmonary bypass, arterial line |
| Regulation Number:    | 21 CFR 870.4260                                      |
| Product Code:         | DTM                                                  |
| Classification:       | Class II                                             |
| Classification Panel: | Cardiovascular                                       |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for LivaNova, a medical technology company. The logo consists of the company name in a stylized font, with the word "LivaNova" in purple. Below the name is the tagline "Health innovation that matters" in a smaller, orange font. The logo is simple and modern, and it conveys the company's focus on health and innovation.

#### III. Predicate Device Primary Predicate: K040184

The MICRO Arterial Filters are substantially equivalent to the following cleared predicate devices. Both modified and unmodified models have fundamental scientific technology and intended use:

|                                       | 510(k) Number:        | K952270                                                                                        | K040184                                                                         |
|---------------------------------------|-----------------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
|                                       | Device Trade Name:    | DIDECO ADULT<br>ARTERIAL FILTERS<br>D732 MICRO 20 W/20,<br>D734 MICRO 40 W/40<br>MICRON SCREEN | DIDECO MICRO 40<br>PH.I.S.I.O. ADULT<br>ARTERIAL FILTER                         |
| MICRO 40<br>A Start<br>(adult)        | Classification Name:  | Filter, blood,<br>cardiopulmonary<br>bypass, arterial line                                     | Filter, blood,<br>cardiopulmonary<br>bypass, arterial line                      |
|                                       | Regulation Number:    | 21 CFR 870.4260                                                                                | 21 CFR 870.4260                                                                 |
|                                       | Product Code:         | DTM                                                                                            | DTM                                                                             |
|                                       | Classification Panel: | Class II<br>Cardiovascular                                                                     | Class II<br>Cardiovascular                                                      |
|                                       | 510(k) Number:        | K041061                                                                                        | K112525                                                                         |
|                                       | Device Trade Name:    | D731 MICRO 20 AND<br>D733 MICRO 40<br>PEDIATRIC ARTERIAL<br>FILTERS                            | D731 MICRO 27<br>PH.I.S.I.O. AND D733<br>MICRO 40 PH.I.S.O.<br>ARTERIAL FILTERS |
| MICRO 40<br>SA Start<br>(small adult) | Classification Name:  | Filter, blood,<br>cardiopulmonary<br>bypass, arterial line                                     | Filter, blood,<br>cardiopulmonary<br>bypass, arterial line                      |
|                                       | Regulation Number:    | 21 CFR 870.4260                                                                                | 21 CFR 870.4260                                                                 |
|                                       | Product Code:         | DTM                                                                                            | DTM                                                                             |
|                                       | Classification Panel: | Class II<br>Cardiovascular                                                                     | Class II<br>Cardiovascular                                                      |
|                                       | 510(k) Number:        | K961869                                                                                        | K033987                                                                         |
|                                       | Device Trade Name:    | DIDECO<br>NEWBORN/INFANT<br>ARTERIAL FILTERS:<br>D735 MICRO 20 AND<br>D736 MICRO 40            | D735 MICRO 20 AND<br>D736 MICRO 40<br>NEWBORN-INFANT<br>ARTERIAL FILTERS        |
| MICRO 40 I<br>Start<br>(infant)       | Classification Name:  | Filter, blood,<br>cardiopulmonary<br>bypass, arterial line                                     | Filter, blood,<br>cardiopulmonary<br>bypass, arterial line                      |
|                                       | Regulation Number:    | 21 CFR 870.4260                                                                                | 21 CFR 870.4260                                                                 |
|                                       | Product Code:         | DTM                                                                                            | DTM                                                                             |
|                                       | Classification Panel: | Class II<br>Cardiovascular                                                                     | Class II<br>Cardiovascular                                                      |

#### IV. Device Description

MICRO Arterial Filters are single-use, non-toxic, no pyrogenic fluid path devices and supplied sterile and individually packaged. They are devices made of plastic material

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the LivaNova logo. The logo is purple and consists of the word "LivaNova" in a stylized font. Below the logo is the tagline "Health innovation that matters" in a smaller, orange font. The logo is simple and modern, and the tagline conveys the company's mission.

(mainly PVC) and a silicon filtering net and they are recommended for use in the arterial line of an extracorporeal circuit during any procedure that requires cardiopulmonary bvpass.

These filters are used to trap and remove gaseous emboli that may be introduced through the arterial line and thev can be used up to 6 hours.

The MICRO Arterial Filters are the modified version of the following currently marketed devices:

a) the D734 Micro 40 included in the Dideco Micro 20 & 40 Adult Arterial Filter (K952270) and in the Dideco MICRO 40 Ph.I.S.I.O. Adult Arterial Filter (K040184); b) the D736 Micro 40 included into the Dideco Newborn/Infant Arterial Filters, 20/40 Micron (K961869) and in the D735 MICRO 20, Dideco D735 Micro 20 Newborn-Infant Arterial Filter with 20 micron screen and for D736 MICRO 40, DIdeco D736 Micro 40 Newborn-Infant Arterial Filter with 40 micron screen (K033987);

c) the D733 Micro 40 included in the D731 Micro 20 and D733 Micro 40 Pediatric Arterial Filters (K041061) and in the D731 MICRO 27 Ph.I.S.I.O. and D733 MICRO 40 Ph.I.S.I.O. Arterial Filters (K112525).

#### V. Indications for Use

The devices are recommended for use on the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass.

The filters are used to trap and remove gaseous emboli as well as particulate debris that may be introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer period is not advised.

#### VI. Summary of Technical Characteristics

The MICRO Arterial Filters have the same fundamental technological characteristics, principles of operation and control mechanisms as the unmodified devices.

The tubing in the purge/recirculation line made of PVC materials were changed within the modified MICRO Arterial Filters in order to remove the diethylhexyl phthalate (DEHP) used as plasticizer currently present in the device.

Moreover, the silicon formulation of the valve's diaphragm in the purge/recirculation line for MICRO Arterial Filters was changed.

Additionally, some technical specifications for MICRO Arterial Filters were updated and tested to better reflect the state of the art of the devices.

No other design changes have been made to the devices.

The devices are ethylene oxide sterilized and have a non-pyrogenic fluid path. They are for single use only.

#### VII. Non-Clinical Performance Data

Sorin Group Italia S.r.l. has conducted extensive verification and validation testing of the MICRO Arterial Filters, as disposables used to trap and remove gaseous emboli that may be introduced through the arterial line of an extracorporeal circuit during any procedure that requires cardiopulmonary bypass.

The MICRO Arterial Filters comply with all the applicable voluntary standards related to Arterial Filters. The devices passed all the testing in accordance with national and international standards.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the LivaNova logo. The logo is purple and consists of the word "LivaNova" in a sans-serif font. Below the logo is the tagline "Health innovation that matters" in a smaller, orange font. The logo is simple and modern, and the tagline conveys the company's mission.

#### VIII. Clinical Performance Data

No clinical testing was conducted in support of the MICRO Arterial Filters, as the indications for use and technical characteristics are equivalent to those of the predicate devices, which have been on the market for several years with proven safety and efficacy of use. The non-clinical testing summarized in this submission supports the substantial equivalence of the subject devices with the predicate devices when used according to their intended use.

#### IX. Statement of Substantial Equivalence

Based on equivalent intended use and technological characteristics, as well as on equivalent performance testing, the MICRO Arterial Filters can be deemed to be substantially equivalent to their predicate devices:

- the unmodified D734 Micro 40, cleared under K952270 and K040184; -
- the unmodified D736 Micro 40, cleared under K961869 and K033987; -
- the unmodified D733 Micro 40, cleared under K041061 and K112525. ।

The MICRO Arterial Filters, as designed and manufactured, do not raise new questions regarding safety and effectiveness as compared to their predicate devices and are determined to be substantially equivalent to their predicate devices listed above.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM/K242092](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM/K242092)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
