← Product Code [DTM](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM) · K022071

# EMBOL-X AORTIC FILTERS,ETRA-SMALL,SMALL,MEDIUM,LARGE,EXTRA LARGE,MODELS CF26000.CF29000,CF32000,CF35000,CF4000 (K022071)

_Embol-X, Inc. · DTM · Feb 14, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM/K022071

## Device Facts

- **Applicant:** Embol-X, Inc.
- **Product Code:** [DTM](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM.md)
- **Decision Date:** Feb 14, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4260
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Intended Use

The EMBOL-X Aortic Filter is indicated for use with the EMBOL-X Aortic Cannula in cardiac surgery procedures to capture and remove particulate emboli from the ascending aorta and heart during and following cross clamp removal in patients aged 60 years and older undergoing first time non-emergent CABG or isolated aortic or mitral valve repair/replacement.

## Device Story

The EMBOL-X Aortic Filter is a surgical device used during cardiopulmonary bypass. It consists of a distal heparin-coated mesh filter on a flexible frame, a locking cartridge housing for attachment to an aortic cannula side port, and a proximal plunger mechanism. The device is introduced into the aorta via the cannula; the filter basket captures particulate emboli as blood passes through. It remains in situ for up to 60 minutes. Operated by surgeons in an OR setting, the device aims to reduce embolic debris, potentially improving patient outcomes by preventing post-operative complications like stroke or organ damage. The filter is withdrawn using the plunger mechanism before procedure completion.

## Clinical Evidence

The ICEM 2000 trial was a prospective, multi-center, 1:1 randomized controlled study of 1289 patients (645 treatment, 644 control) undergoing CABG or valve procedures. Primary safety endpoint (composite of death, stroke, renal insufficiency, MI, GI complications, limb ischemia) showed equivalence (17.1% treatment vs 18.9% control, p<0.001). Primary effectiveness endpoint (particulate capture) was met in 96.8% of filters (p<0.001).

## Technological Characteristics

Device comprises a heparin-coated mesh filter, locking cartridge, and plunger mechanism. Uses conventional medical-grade materials. Single-use, sterile. Designed for integration with the EMBOL-X Aortic Cannula via a side port. Mechanical filtration principle.

## Regulatory Identification

A cardiopulmonary bypass arterial line blood filter is a device used as part of a gas exchange (oxygenator) system to filter nonbiologic particles and emboli (blood clots or pieces of foreign material flowing in the bloodstream which will obstruct circulation by blocking a vessel) out of the blood. It is used in the arterial return line.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions.”

## Predicate Devices

- Medtronic PercuSurge GuardWire Plus Temporary Occlusion and Aspiration System ([K003992](/device/K003992.md))
- Edwards Lifesciences AF-1025D/AF-1040D Duraflo Arterial Blood Filter ([K820044](/device/K820044.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

The following 510(k) Summary of Safety and Effectiveness information is provided in accordance with the requirements of 21 CFR §807.92 and SMDA 1990.

| 510(k) Number:       | K022071                                                                             |              |
|----------------------|-------------------------------------------------------------------------------------|--------------|
| Date Prepared:       | February 13 2003                                                                    | FEB 1 4 2003 |
| Applicant:           | EMBOL•X, Inc.                                                                       |              |
| Address:             | 645 Clyde Avenue, Mountain View, CA 94043                                           |              |
| Phone Number:        | 650/390-0280                                                                        |              |
| Fax Number:          | 650/390-0282                                                                        |              |
| Contact Person:      | Jean Chang                                                                          |              |
| Trade Name:          | EMBOL•X® Aortic Filter                                                              |              |
| Common Name:         | Arterial Line Blood Filter                                                          |              |
| Classification Name: | Filter, Blood, Cardiopulmonary Bypass, Arterial Line; 21 CFR §870.4260;<br>Class II |              |

The EMBOL • X Aortic Filter device consists of three primary components: Device 1.) a distal heparin-coated mesh filter, mounted onto a flexible frame to Description: form a filter basket for particulate emboli capture and retention; 2.) a locking cartridge housing for attachment to the EMBOL•X Aortic Cannula side port, permitting access to the aorta and to ensure correct orientation of the filter during use; and 3.) a proximal syringe-like plunger mechanism to deploy and withdraw the distal basket into and from the aorta, via the cannula, during surgery. The filter is introduced surgically into the aorta via the previously placed cannula, and particulates are captured and removed as blood passes through the filter basket. The filter may remain in situ for up to 60 minutes. The EMBOL.X Aortic Filter utilizes conventional medical grade materials and processes, and is provided packaged, labeled, and sterile, intended for single-use.

{1}------------------------------------------------

- The EMBOL-X Aortic Filter is indicated for use with the EMBOL-X Intended Use: Aortic Cannula in cardiac surgery procedures to capture and remove particulate emboli from the ascending aorta and heart during and following cross clamp removal in patients aged 60 years and older undergoing first time non-emergent CABG or isolated aortic or mitral valve repair/replacement.
	- Substantial equivalence is derived from a composite of characteristics Predicate from multiple predicate devices. The EMBOL.X Aortic Filter is Devices: substantially equivalent in intended use, clinical application, principle of operation, design and materials, sterility and biocompatibility, and performance to the Medtronic PercuSurge GuardWire Plus Temporary Occlusion and Aspiration System (K003992) and/or the Edwards Lifesciences AF-1025D/AF-1040D Duraflo (heparin treated) Arterial Blood Filter (K820044).
- The EMBOL.X Aortic Filter has similar intended use, design intent, Technological Characteristics: principle of operation, materials, sterility and biocompatibility, accessory requirements, and labeling as that of the predicate devices. Any noted differences between the devices (specific indications for use, method of device delivery, specific physical dimensions and geometry) do not raise new types of safety or effectiveness questions, do not introduce new technological issues, and therefore do not impact the substantial equivalence of the EMBOL•X Aortic Filter.
- The results of biocompatibility, in-vitro (bench), and pre-clinical (animal) Non-Clinical Test tests demonstrate that the EMBOL·X Aortic Filter is sterile, Results: biocompatible, meets established internal performance specifications, and satisfies the requirements of relevant external standards and applicable FDA Guidance.
- Data to support the EMBOL•X Aortic Filter (and associated EMBOL•X Summary of Aortic Cannula) were obtained from the EMBOL.X ICEM 2000 trial. Clinical Studies: The purpose of this prospective, multi-center, 1:1 randomized, controlled equivalency study was to demonstrate the safety and effectiveness of the EMBOL•X Aortic Filter (and the associated EMBOL•X Aortic Cannula) in capturing particulate emboli during first-time non-emergent coronary artery bypass graft (CABG) or aortic/mitral valve repair/replacement procedures utilizing cardiopulmonary bypass. This study was conducted at 22 sites within the United States and Canada, and was comprised of 1289 patients, of which 645 were randomized to the EMBOL•X Aortic Filter, and 644 were randomized to control. An independent Clinical Events Committee (CEC) adjudicated the major clinical endpoints and events.

{2}------------------------------------------------

## Summary of Clinical Studies (continued):

The primary safety measure was a composite endpoint comprised of the following post-operative clinical events, measured from the time of randomization (the operation) through hospital discharge or 30 days. whichever occurred first: Neurologic deficit (mild and severe); Renal insufficiency (with and without dialysis); Gastrointestinal (GI) complications; Perioperative Myocardial Infarction (MI); Limbthreatening peripheral embolism (Limb Ischemia); and Death. The primary effectiveness endpoint was the capture of particulate emboli by the filter (Treatment arm only) in at least 75% of the filtered patients, with particulate debris visually confirmed by light microscopy.

The patient population of this study was limited to patients undergoing first time, non-emergent Coronary Artery Bypass Grafting (CABG), aortic valve replacement or mitral valve repair or replacement only, aged 60 years and older. Of the 1289 patients studied. 927 (71.9%) were male. and 1042 (80.8%) were 65 years of age or older; 65 (5.0%) had a LVEF < 30%, and 388 (30.1%) had a prior MI. None of the demographic or medical history differences between the randomized groups achieved statistical significance (p<0.05).

18 of 22 enrolling centers performed peri-procedural ultrasound imaging (TEE or EPI), resulting in an analysis of a subset (910, or 70.6%) of the total enrolled patients.

The primary safety endpoint was met, with the EMBOL . X Aortic Filter arm composite event rate statistically equivalent to that of standard treatment (17.1% vs. 18.9%, p<0.001). In addition, differences in the stratified events between the randomized groups did not achieve statistical significance (p<0.05). The primary effectiveness endpoint was also met, with particulate capture demonstrated in 96.8% of all filters analyzed (p<0.001).

The following tables summarize the results of the ICEM trial.

{3}------------------------------------------------

| Event                               | Treatment<br>(N=645) | Control<br>(N=644) | P-Value |
|-------------------------------------|----------------------|--------------------|---------|
| Death                               | 10 (1.6)             | 11 (1.7)           | 0.82    |
| Neurologic deficit (Stroke/TIA)     | 18 (2.8)             | 18 (2.8)           | 1.00    |
| Renal insufficiency (RI)            | 40 (6.2)             | 52 (8.1)           | 0.19    |
| RI (w/o dialysis)                   | 33 (5.1)             | 43 (6.7)           | 0.23    |
| RI (dialysis)                       | 7 (1.1)              | 9 (1.4)            | 0.61    |
| Myocardial infarction (MI)          | 67 (10.4)            | 64 (9.9)           | 0.79    |
| Q Wave MI                           | 21 (3.3)             | 18 (2.8)           | 0.63    |
| CK-MB Elevation                     | 46 (7.1)             | 46 (7.1)           | 0.99    |
| Gastrointestinal complications (GI) | 5 (0.8)              | 5 (0.8)            | 1.00    |
| Limb ischemia                       | 3 (0.5)              | 3 (0.5)            | 1.00    |
| Any event                           | 110 (17.1)           | 122 (18.9)         | 0.38    |

### Table 1. Major Adverse Events [Number (%)]

Numbers are for all randomized patients

Death: Death for any cause

Stroke: Central neurologic deficit persisting for > 24 hours

Transient neurologic deficit (TIA): An ischemic event of the central nervous system that causes a neurologic deficit persisting for < 24 hours

Renal insufficiency: Increase of serum creatinine to > 2.0 mg/dl or a 50% or greater increase over abnormal baseline prior to procedure

Renal insufficiency (dialysis): The new requirement for dialysis

O-Wave MI: New pathological Q-Waves in 2 or more contiguous leads

Non Q-Wave MI: CPK > 5x normal and CK-MB > 5x above the upper limit of normal for the institution, in the absence of new Q-Waves

Gastro-Intestinal Complications: include GI bleeding requiring transfusion; Pancreatitis with abnormal amylase/lipase requiring NG suction therapy; Cholecystitis requiring cholecystectomy or drainage; Mesenteric ischemia requiring exploration

Limb-threatening Peripheral Embolism: Acute onset of diminished pulse, altered pallor (discoloration, either hypo- or hyper-), and pain as evidence of limb-threatening peripheral ischemia

### Table 2.

#### Echocardiographically Evident Endothelial Disruptions Observed through Imaging

| Treatment    | Control     | P-Value |
|--------------|-------------|---------|
| n/N (%)      | n/N (%)     |         |
| 42/456 (9.2) | 9/454 (2.0) | < 0.001 |

# Table 3.

Composite Endpoint Events Stratified by Echocardiographically Evident Endothelial Disruptions Observed through Imaging

| Arm       | Events in Patients With<br>Endothelial Disruption<br>n/N (%) | Events in Patients Without<br>Endothelial Disruption<br>n/N (%) | P-Value |
|-----------|--------------------------------------------------------------|-----------------------------------------------------------------|---------|
| Treatment | 2/42 (4.8)                                                   | 74/414 (17.9)                                                   | 0.03    |
| Control   | 1/9 (11.1)                                                   | 81/445 (18.2)                                                   | 1.0     |

{4}------------------------------------------------

## Table 4. Particulate Capture Effectiveness

| Attribute                                                                    | Result      |
|------------------------------------------------------------------------------|-------------|
| Number EMBOL•X Filters Deployed                                              | 618         |
| Number (%) Filters Which Captured ≥ 1 Particle                               | 598 (96.8%) |
| Lower 95% Confidence Bound on Percent of Filters Which Captured ≥ 1 Particle | 95.3%       |

Conclusions drawn from Study

The results from the clinical investigation demonstrate that the EMBOL . X Aortic Filter, when used in patients undergoing first time, non-emergent Coronary Artery Bypass Grafting (CABG), aortic valve replacement or mitral valve repair or replacement, does not pose any additional risk to the treated patient population when compared to that of the current standard treatment of no filtration, and the Aortic Filter was effective in capturing particulates. The data demonstrate that the EMBOL·X Aortic Filter is compatible with conventional CPB procedures. It was therefore concluded that the EMBOL. X Aortic Filter is safe and effective when used as indicated in the Instructions for Use.

- Review of the device in vitro pre-clinical studies, combined Summary: with the results of the clinical investigation, provides valid scientific evidence and reasonable assurance that the EMBOL.X Aortic Filter is safe and effective for its intended use. Comparison of the product attributes supports the conclusion that the device is substantially equivalent to the commercially marketed predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines above them that resemble a bird's wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 4 2003

EMBOL-X, Inc. c/o Ms. Jean Chang Chief Operating Officer and Executive Vice President 645 Clyde Ave. Mountain View, CA 94043-2208

Re: K022071 Trade Name: EMBOL-X® Aortic Filter Regulation Number: 21 CFR 870.4260 Regulation Name: Filter, Blood, Cardiopulmonary Bypass, Arterial Line Regulatory Class: Class II (two) Product Code: DTM Dated: November 22, 2002 Received: November 25, 2002

Dear Ms. Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Ms. Jean Chang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsma/dsmamain.html

Sincerely yours,

Kala Tiller

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K022071

> EMBOL•X Aortic Filter Device Name:

Indications For Use: The EMBOL-X Aortic Filter is indicated for use with the EMBOL-X Aortic Cannula in cardiac surgery procedures to capture and remove particulate emboli from the ascending aorta and heart during and following cross clamp removal in patients aged 60 years and older undergoing first time non-emergent CABG or isolated aortic or mitral valve repair/replacement.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

DaCattun

Prescription Use
(Per 21 CFR §801.109) X

OR Over-The-Counter Use

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM/K022071](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTM/K022071)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
