K961181 · Naltiac Medical, Inc. · DTL · Jun 21, 1996 · Cardiovascular
Device Facts
Record ID
K961181
Device Name
NMI VEIN GRAFT PERFUSION SET
Applicant
Naltiac Medical, Inc.
Product Code
DTL · Cardiovascular
Decision Date
Jun 21, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4290
Device Class
Class 2
Indications for Use
The NMI Vein Graft Perfusion Set is intended to be used during a cardiopulmonary bypass procedure to connect multiple coronary cannulae to a single infusion line.
Device Story
Device connects multiple coronary cannulae to single infusion line during cardiopulmonary bypass; facilitates fluid delivery to multiple graft sites. Comprised of PVC alloy tubing, PVC quad fitting, polypropylene pinch clamp, PVC female luer, and ABS male luer connectors. Used in clinical OR settings by perfusionists or surgeons. Provides fluid path for perfusion; allows individual flow control via pinch clamps. Benefits patient by enabling efficient, simultaneous perfusion of multiple coronary grafts during bypass surgery.
Clinical Evidence
Bench testing only. Evaluated pressure drop at 4°C and 40°C (100-500 ml/min flow rates), leak testing at 10 psi, luer connection compliance (ANSI/HIMA MD70.1-1983), tubing bond strength (>0.4 lb), package integrity (ASTM F1140-88), and shipping simulation (ASTM 40169). Biocompatibility per Tripartite Guidance; non-pyrogenic per USP LAL test. Accelerated aging validated for one year.
Indicated for patients undergoing cardiopulmonary bypass procedures requiring connection of multiple coronary cannulae to a single infusion line.
Regulatory Classification
Identification
A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K972503 — CARDIOPLEGIA ADAPTERS, VESSEL CANULAE, AND AORTIC ROOT CANNULAE · Sorin Group Italia S.R.L. · Aug 26, 1997
K971041 — CHASE CORONARY OSTIAL PERFUSION CANNULA · Chase Medical, Inc. · May 13, 1997
K964200 — CHASE MULTIPLE PERFUSION SET · Chase Medical, Inc. · Apr 16, 1997
K982368 — MANIFOLD WITH EXTENSION SET · Interpore Cross Intl. · Jul 31, 1998
K151523 — Uncoated Tubing and Connector Components · Medtronic, Inc. · Jul 16, 2015
Submission Summary (Full Text)
{0}
JUN 21 1998
K961181
# Summary of Safety and Effectiveness
## NMI VEIN GRAFT PERFUSION SET
## I. General Information
A. Generic Name: Vein Graft Perfusion Set
B. Trade Name of Device: NMI Vein Graft Perfusion Set
C. Applicant's Name and Address: Naltiac Medical, Inc.
17194 Preston Rd, Suite 123-206
Dallas, TX 75248
D. Pre-market Notification Number: Not yet assigned
## II. Indications for Use
The NMI Vein Graft Perfusion Set is intended to be used during a cardiopulmonary bypass procedure to connect multiple coronary cannulae to a single infusion line.
## III. Device Description
The NMI Vein Graft Perfusion Set is fabricated with PVC alloy tubing, PVC Quad fitting, polypropylene pinch clamp, PVC female luer connector, and ABS male luer connector. The features of the device include: one 3 1/4 inlet leg with female luer port and white pinch clamp, three 9 3/4 in. outlet legs with red pinch clamps terminating in male luer lock connectors, one 3 1/4 in. outlet leg with blue pinch clamp terminating in a male luer lock, and soft flexible tubing on legs, .090 I.D., and 3/16 O.D.
## IV. Device Classification: Class II
## V. Safety and Effectiveness
Substantial Equivalence: The device has been shown to be substantially equivalent to the DLP Multiple Perfusion Set (K791498).
## VI. Other Safety and Effectiveness Data
Materials: Fluid contact of materials of construction comply with Tripartite Biocompatibility Guidance for external devices, blood path direct, short term use.
Sterilization: Validated 100% Ethylene Oxide sterilization cycle (Overkill Method) SAL 10⁻⁶
Pyrogenicity: Non-Pyrogenic per USP Pyrogen test (LAL)
## Functional Testing
Pressure Drop/Saline
NMI Cannulae average pressure drops range at flow rates between 100 to 500 ml/min = 18 to 125 mmHg @ 4°C
DLP Cannulae average pressure drops range at flow rates between 100 and 500 ml/min = 15 to 180 mmHg @ 4°C.
{1}
Pressure Drop/Saline Continued
NMI Cannuale average pressure drops range at flow rates between 100 and 500 ml/min = 6 to 109 mmHg @ 40°C
DLP Cannulae average pressure drops range at flow rates between 100 and 500 ml/min = 10 to 140 mmHg @ 40°C
Leak Test Requirements:
No leaks at 10 psi on NMI device at 4°C and at 40°C
Luer Connections:
Meets ANSI/HIMA MD70.1-1983 for Medical Materials Luer Tape Fittings
Tubing Bond Strength:
Exceeds 0.4 lb tensile strength at 4°C and at 40°C
Package Integrity:
Tyvek/Polymylar passed burst test per ASTM F1140-88
Shipping & Distribution Testing:
Passed Distribution Simulation Test per ASTM 40169 Standard
Accelerated Aging:
One (1) year with no affects on performance characteristics
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.