HEMOSTASIS VALVE

K072556 · Merit Medical Systems, Inc. · DTL · Nov 20, 2007 · Cardiovascular

Device Facts

Record IDK072556
Device NameHEMOSTASIS VALVE
ApplicantMerit Medical Systems, Inc.
Product CodeDTL · Cardiovascular
Decision DateNov 20, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4290
Device ClassClass 2

Intended Use

The Hemostasis Valve is intended to maintain a fluid-tight seal around interventional and diagnostic devices during use.

Device Story

Mechanical hemostasis valve; maintains fluid-tight seal around interventional/diagnostic devices. Components: clear polycarbonate body, silicone valve, polyacetal actuation lever, polycarbonate/EPDM o-ring rotating male Luer lock connector, female Luer lock. Used in clinical settings by healthcare professionals during interventional procedures. Device functions as a passive mechanical seal; no electronic or software components. Benefits patient by preventing blood loss and maintaining system integrity during catheter-based procedures.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Mechanical valve; materials include polycarbonate, silicone, polyacetal, and EPDM. Features rotating male Luer lock and female Luer lock. Manual actuation via lever. No energy source, software, or connectivity.

Indications for Use

Indicated for maintaining a fluid-tight seal around interventional and diagnostic devices during clinical use.

Regulatory Classification

Identification

A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### SECTION 5: 510(k) SUMMARY NOV 2 0 2007 Merit Medical Systems, Inc. 1600 West Merit Parkway South Jordan, UT 84095 | CONTACT: | Shirley Hyink | |----------------------------|----------------------------------------------------------------------------------| | DATE PREPARED: | September 7, 2007 | | TRADE OR PROPRIETARY NAME: | Hemostasis Valve | | CLASSIFICATION NAME: | Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting<br>21CFR 870.4290 | | PREDICATE DEVICE: | Merit Passage® Hemostasis Valve (K925419) | DEVICE DESCRIPTION: The Hemostasis Valve is a valve consisting of a clear polycarbonate body, silicone valve, polyacetal actuation lever, and polycarbonate and EPDM o-ring rotating male Luer lock connector and female Luer lock. Image /page/0/Picture/5 description: The image shows a detailed technical drawing of a mechanical device, possibly a type of valve or actuator. The device features a cylindrical body with various attachments and levers. The drawing style suggests it is a blueprint or schematic, emphasizing the mechanical components and their arrangement. INTENDED USE: The Hemostasis Valve is intended to maintain a fluid-tight seal around interventional and diagnostic devices during use. TECHNOLOGICAL CHARACTERISTICS: The Hemostasis Valve and the predicate device Passage are similar in design and technology. All of the component materials found in the Hemostasis Valve have been used in legally marketed Merit devices and were found safe for their intended use. The prior use of component materials of the Hemostasis Valve in legally marketed devices, a similar indication for use, the performance data provided, and the biocompatibility statement support the safety and effectiveness of this hemostasis valve for the indicated use. {1}------------------------------------------------ Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or other bird, with its wings spread. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # NOV 2 0 2007 Merit Medical Systems, Inc. c/o Ms. Shirley Hyink Manager, Regulatory Affairs 1600 West Merit Parkway South Jordan, UT 84095 Re: K072556 Hemostasis Valve Regulation Number: 21 CFR 870.4290 Regulation Name: Cardiopulmonary Bypass Adaptor, Stopcock, Manifold, or Fitting Regulatory Class: Class II Product Code: DTL Dated: September 7, 2007 Received: September 10, 2007 Dear Ms. Hyink: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ #### Page 2 - Ms. Shirley Hyink Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, uma R. lochner \Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### SECTION 4: INDICATIONS FOR USE KOZ2556 510(k) Number (if known): Device Name: Hemostasis Valve Indications for Use: The Hemostasis Valve is indicated to maintain a fluid-tight seal around interventional and diagnostic devices during use. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) suma R. Jaimes (Division Sign-Off) (Division of Cardiovascular Devices 310(k) Number K072556 Premarket Notification Hemostasis Valve Merit Medical Systems, Inc. ,9
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...