K202557 · Spectrum Medical , Ltd. · DRY · Oct 2, 2020 · Cardiovascular
Device Facts
Record ID
K202557
Device Name
Quantum Workstation 12 Elite
Applicant
Spectrum Medical , Ltd.
Product Code
DRY · Cardiovascular
Decision Date
Oct 2, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4330
Device Class
Class 2
Indications for Use
The intended use of the Quantum Workstation 12" Elite is for the non-invasive continuous monitoring of oxygen saturation and hematocrit / hemoglobin concentration of the blood and in an extracorporeal circuit. When using its range of accessories, the Quantum Workstation 12" Elite is configured to measure and display the following measurements: - · SaO2 Arterial Saturation (%) - · SvO2 Venous Saturation (%) - · Hb Hemoglobin (g/L and gm/dl) - · Hct Calculated Hematocrit (%) - · Blood Flow 2 channels with arterial and venous flow differentials - · Pressure & Temperature 4 channels The Quantum Workstation 12" Elite provides monitoring information to trained clinicians and can be configured by them to set parameter specific alarms. The workstation's monitoring and alarm functionality does not directly control patient care. The User makes clinical iudgments regarding the treatment of the patient as a result of information displayed by the workstation.
Device Story
Quantum Workstation 12" Elite is an on-line cardiopulmonary bypass blood gas monitor; provides non-invasive continuous monitoring of oxygen saturation (SaO2, SvO2), hemoglobin, hematocrit, blood flow, pressure, and temperature in extracorporeal circuits. Device consists of a pole-mounted 12.1" high-definition touch screen; connects to various sensors (Hb/SO2, flow, pressure/temperature). Operates via AC mains with lithium-ion battery backup. Used in clinical settings by trained clinicians; displays real-time readings and trends; allows configuration of parameter-specific alarms. Does not control patient care; clinicians use displayed information to make treatment decisions. Enhanced functionality includes control of centrifugal or roller pumps and bubble detection. Benefits include real-time physiological monitoring during cardiopulmonary bypass to assist clinical decision-making.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by non-clinical testing including electrical safety, electromagnetic compatibility (EMC), hardware testing, and software verification and validation.
Technological Characteristics
Pole-mounted 12.1" landscape HD touch screen; AC mains powered with dual lithium-ion battery backup (2-hour min). Connectivity: LAN/Ethernet, USB 2.0, serial port, and proprietary Spectrum Medical ports. Sensors for Hb/SO2, blood flow, pressure/temperature. SD card for memory storage. Software-based monitoring and alarm system.
Indications for Use
Indicated for non-invasive continuous monitoring of oxygen saturation, hematocrit, and hemoglobin concentration in blood within an extracorporeal circuit for patients undergoing procedures requiring such monitoring. Intended for use by trained clinicians.
Regulatory Classification
Identification
A cardiopulmonary bypass on-line blood gas monitor is a device used in conjunction with a blood gas sensor to measure the level of gases in the blood.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 2, 2020
Spectrum Medical Ltd Colleen Powell Director of Regulatory Affairs Harrier 4, Meteor Business Park, Cheltenham Road East Gloucester, Gloucetershire GL2 9QL United Kingdom
Re: K202557
Trade/Device Name: Quantum Workstation 12" Elite Regulation Number: 21 CFR 870.4330 Regulation Name: Cardiopulmonary Bypass On-Line Blood Gas Monitor Regulatory Class: Class II Product Code: DRY Dated: September 1, 2020 Received: September 3, 2020
Dear Colleen Powell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202557
Device Name Quantum Workstation 12" Elite
#### Indications for Use (Describe)
The intended use of the Quantum Workstation 12" Elite is for the non-invasive continuous monitoring of oxygen saturation and hematocrit / hemoglobin concentration of the blood and in an extracorporeal circuit. When using its range of accessories, the Quantum Workstation 12" Elite is configured to measure and display the following measurements:
- · SaO2 Arterial Saturation (%)
- · SvO2 Venous Saturation (%)
- · Hb Hemoglobin (g/L and gm/dl)
- · Hct Calculated Hematocrit (%)
- · Blood Flow 2 channels with arterial and venous flow differentials
- · Pressure & Temperature 4 channels
The Quantum Workstation 12" Elite provides monitoring information to trained clinicians and can be configured by them to set parameter specific alarms.
The workstation's monitoring and alarm functionality does not directly control patient care. The User makes clinical iudgments regarding the treatment of the patient as a result of information displayed by the workstation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| 区 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Spectrum Medical. The word "Spectrum" is written in blue, followed by a red line that resembles a heartbeat. The word "Medical" is written in blue after the red line. The logo is simple and professional, and it is likely used to represent a medical company.
Quantum Workstation 12" Elite 510(k) Summary K202557 Page 1 of 3
# 510(k) Summary
## I. SUBMITTER
| Name: | Spectrum Medical Ltd |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Harrier 4, Meteor Business Park<br>Cheltenham Road East Gloucester<br>GL2 9QL<br>United Kingdom |
| Contact Person: | Colleen Powell, Director of Regulatory Affairs |
| Phone: | +44 (0) 1242 650 120 |
| Fax: | +44 (0) 8452 808 127 |
| Date Summary Prepared: | September 1, 2020 |
| II. DEVICE | |
| Proprietary Name: | Quantum Workstation 12" Elite |
| Common Name: | Blood Gas Monitor |
| Classification Name: | Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass (21 CFR<br>870.4330) |
| Regulatory Class: | II |
| Product Code: | DRY |
| Panel: | Office of Health Technology 2 (OHT2 Cardiovascular Devices) /<br>Division of Health Technology 2B (Circulatory Support, Structural and<br>Vascular Devices) |
#### III. PREDICATE DEVICE
The predicate device for this submission is Spectrum Medical Ltd.'s Quantum Workstation 12.1" (K181923). This predicate has not been subject to a design-related recall. Three reference devices manufactured by Spectrum Medical Ltd. were used, and these included the Quantum Diagnostic Module (K173591), Quantum Smart Occluders (K190282) and Quantum Pump Console (K173834).
#### IV. DEVICE DESCRIPTION
The Quantum Workstation 12" Elite is an on-line, cardiopulmonary bypass, blood gas monitor that is used for extracorporeal monitoring of blood oxygen (arterial and venous) saturation, hematocrit, and hemoglobin levels. The device's enhanced functionality includes the ability to control centrifugal or roller pumps and make blood flow, blood pressure and blood temperature measurements.
The Quantum Workstation 12" Elite consists of a pole-mounted 12.1" landscape high definition touch screen. The touch screen displays individual and trend readings with alarm settings. The Quantum Workstation 12" Elite provides memory storage via an SD (Secure Digital) card. The Quantum Workstation 12" Elite is powered from the AC Mains supply and incorporates a battery
{4}------------------------------------------------
SpectrumMedical
backup that automatically switches on in the event of an interruption to the mains power supply. The battery backup is provided via two (2) lithium-ion batteries with a two-hour minimum life.
The Quantum Workstation 12" Elite includes the following ports / connections:
- One (1) sensor port for the Hb / SO2 sensor
- Two (2) sensor ports for blood flow ●
- Two (2) sensor ports for blood pressure/temperature
- One (1) LAN / Ethernet port ●
- Two (2) USB 2.0 ports ●
- Three (3) Spectrum Medical ports
- One (1) User Serial Port
Accessories for the Quantum Workstation 12" Elite include the power supply, mounting arm (long or short), Hb / SQ2 sensor, flow sensors and pressure/temperature sensors.
## V. INTENDED USE / INDICATIONS FOR USE
The intended use of the Quantum Workstation 12" Elite is for the non-invasive continuous monitoring of oxygen saturation and hematocrit / hemoglobin concentration of the blood in an extracorporeal circuit. When using its range of accessories, the Quantum Workstation 12" Elite is configured to measure and display the following measurements:
| SaO2 | Arterial Saturation (%) |
|--------------------------|--------------------------------------------------------|
| SvO2 | Venous Saturation (%) |
| Hb | Hemoglobin (g/L, gm/dl and mmol/L) |
| Hct | Calculated Hematocrit (%) |
| Blood Flow | 2 channels with arterial and venous flow differentials |
| Pressure and temperature | 4 channels |
The Quantum Workstation 12" Elite provides monitoring information to trained clinicians and can be configured by them to set parameter specific alarms.
The workstation's monitoring and alarm functionality does not directly control patient care. The User makes clinical judgments regarding the treatment of the patient as a result of information displayed by the workstation.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Spectrum Medical Ltd's Quantum Workstation 12.1" (K181923) and Quantum Workstation 12" Elite have the same manufacturer, intended use, clinical application, clinical setting / site of use, target user, target patient population, and principle of operation / mechanism of operation / fundamental scientific technology. They also have the same performance for SO2, Hb, and Hct. They both have an Ethernet port and SD card memory storage. They both are also compatible with the intended environment and with other devices.
Spectrum Medical Ltd's Quantum Workstation 12" Elite and Quantum Workstation 12.1" differ in
{5}------------------------------------------------
SpectrumMedical
that the Quantum Workstation 12" Elite provides additional measurements for blood flow, pressure and bubble detection. The Quantum Workstation 12" Elite can also control centrifugal or roller pumps.
These differences do not raise new issues of safety or effectiveness.
# VII. PERFORMANCE DATA - NON-CLINICAL TESTING
No animal testing was submitted to support the substantial equivalence of the Quantum Workstation 12" Elite to the Quantum Workstation 12.1".
The following non-clinical testing was performed to support the substantial equivalence of the Quantum Workstation 12" Elite to the legally marketed predicate device:
- . Electrical safety
- Electromagnetic compatibility (EMC) ●
- Hardware testing
- Software verification and validation ●
#### VIII. PERFORMANCE DATA - CLINICAL TESTING
No clinical data were submitted to support the substantial equivalence of the Quantum Workstation 12" Elite to the Quantum Workstation 12.1".
#### IX. CONCLUSIONS
Based on the indications for use, technological characteristics, results of non-clinical testing, and comparison to predicate devices, the Quantum Workstation 12" Elite has been shown to be substantially equivalent to a legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.