← Product Code [DRY](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY) · K133157

# TRI-OPTIC MEASUREMENT CELL; WITH BALANCE BIOSURFACE, WITH CARMEDA BIOACTIVE SURFACE, WITH TRILLIUM BIOSURFACE, WITH 6 (K133157)

_Medtronic, Inc. · DRY · Oct 23, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY/K133157

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [DRY](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY.md)
- **Decision Date:** Oct 23, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The BioTrend oxygen saturation and hematocrit system measures percent oxygen saturation and hematocrit of the blood in the extracorporeal circuit. The extracorporeal circuit is used for, but is not limited to, cardiopulmonary bypass, closedchest support, and limb perfusion.

## Device Story

Tri-optic Measurement Cell is a sterile, single-use disposable component of the BioTrend Oxygen Saturation and Hematocrit System. Device integrates into extracorporeal circuits (cardiopulmonary bypass, closed-chest support, limb perfusion). Cell connects to a sensor cable linked to the BioTrend instrument; system measures blood oxygen saturation and hematocrit in real-time. Subject device adds Balance Biosurface (heparin-free, thromboresistant coating) to blood-contacting surfaces. Device used by perfusionists or clinicians in clinical settings. Output provides continuous monitoring of blood parameters, assisting clinicians in managing patient perfusion and oxygenation status during procedures.

## Clinical Evidence

Bench testing only. Testing included coating coverage, leaching, and measurement accuracy verification. No clinical data was required or provided.

## Technological Characteristics

Sterile, single-use disposable cell. Blood-contacting surfaces feature biocompatible coatings: uncoated, Carmeda BioActive, Trillium, or Balance Biosurface. Operates as part of an extracorporeal circuit connected to the BioTrend instrument via sensor cable. No electronic components within the cell itself; sensing is performed by the connected BioTrend monitor.

## Regulatory Identification

A cardiopulmonary bypass on-line blood gas monitor is a device used in conjunction with a blood gas sensor to measure the level of gases in the blood.

## Predicate Devices

- Medtronic MX2 Oxygen Saturation and Hematocrit Monitoring System ([K910421](/device/K910421.md))
- Medtronic BioTrend Oxygen Saturation and Hematocrit System ([K093650](/device/K093650.md))
- Medtronic Tri-optic Measurement Cell with Trillium Biosurface ([K012743](/device/K012743.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, conveying the department's mission of protecting the health of all Americans.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 23, 2014

Medtronic, Inc. Jessica Sixberry Senior Regulatory Affairs Specialist 7611 Northland Drive Minneapolis, Minnesota 55428

Re: K133157

Trade/Device Name: Tri-optic Measurement Cell with Balance Biosurface Regulation Number: 21 CFR 870.4330 Regulation Name: Cardiopulmonary Bypass In-Line Blood Gas Monitor Regulatory Class: II Product Code: DRY Dated: September 18, 2014 Received: September 19, 2014

Dear Ms. Jessica Sixberry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

M.A. Hillebrand

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K133157

Device Name Tri-Optic Measurement Cells

Indications for Use (Describe)

The BioTrend oxygen saturation and hematocrit system measures percent oxygen saturation and hematocrit of the blood in the extracorporeal circuit. The extracorporeal circuit is used for, but is not limited to, cardiopulmonary bypass, closedchest support, and limb perfusion.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CER 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subnart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Medtronic logo. On the left side of the logo is a gray circle with a human figure inside. The human figure is in three different positions, as if it is rotating. To the right of the circle is the word "Medtronic" in blue, bold letters.

# 510(k) Summary

| Date Prepared:  | February 11, 2014                                                                                                                                   |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:      | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establish Registration Number: 2184009           |
| Contact Person: | Jessica Sixberry<br>Senior Regulatory Affairs Specialist<br>Phone: (763) 514-9849<br>Fax: (763) 367-8360<br>Email: jessica.m.sixberry@medtronic.com |

## Device Name and Classification:

| Trade Name:        | Tri-optic Measurement Cell with Balance™ Biosurface |
|--------------------|-----------------------------------------------------|
| Common Name:       | Cardiopulmonary bypass in-line blood gas monitor    |
| Regulation Number: | 21 CFR 870.4330                                     |
| Product Code:      | DRY                                                 |
| Classification:    | Class II                                            |

## Predicate Devices

Medtronic MX2 Oxygen Saturation and Hematocrit Monitoring System (K910421) Medtronic BioTrend Oxygen Saturation and Hematocrit System (K093650) Medtronic Tri-optic Measurement Cell with Trillium Biosurface (K012743)

### Device Description

The Medtronic Tri-optic Measurement Cell is a sterile, single-use cell used with the BioTrend Oxygen Saturation and Hematocrit System. This 510(k) premarket notification was submitted to add Balance Biosurface to the disposable cell as update the labeling information for all Tri-optic Measurement Cell versions. The disposable cell is coated on its blood contacting surfaces with Balance Biosurface. Previously cleared disposable cells are also available in uncoated form, and with Carmeda BioActive Surface coating or Trillium Biosurface coating. A list of the models and descriptions is included in Table 5-1.

{4}------------------------------------------------

| Models                             | Description                                                                                   |
|------------------------------------|-----------------------------------------------------------------------------------------------|
| BBTMC25,<br>BBTMC38,<br>BBTMC50    | Tri-optic Measurement Cell with Balance Biosurface                                            |
| CB4714R1,<br>CB4715R1,<br>CB4716R1 | Tri-optic Measurement Cell with additional tubing extension with<br>Carmeda BioActive Surface |
| TMC25T,<br>TMC38T,<br>TMC50T       | Tri-optic Measurement Cell with Trillium Biosurface                                           |
| TMC25,<br>TMC38,<br>TMC50          | Tri-optic Measurement Cell                                                                    |
| 1473R1,<br>1474R1,<br>1475R1       | Tri-optic Measurement Cell with additional tubing extension                                   |

## Table 5-1: Model Descriptions

## Indications for Use

The BioTrend Oxygen Saturation and Hematocrit System measures percent oxygen saturation and hematocrit of the blood in the extracorporeal circuit. The extracorporeal circuit is used for, but is not limited to, cardiopulmonary bypass, closed-chest support, and limb perfusion.

## Comparison to Predicate Devices

The Tri-optic Measurement Cells have the same intended use, design and materials, and principles of operation and technology when compared to the predicate devices.

- Intended Use: The Tri-optic Measurement Cells have the same intended use as the . currently marketed Tri-optic Measurement Cells.
- Design and Materials: The design and the materials of the Tri-optic Measurement Cells are the same as the predicate devices. The devices are available uncoated and three coating options, Carmeda, Trillium and Balance, which are biocompatible surface coatings that increase the thromboresistance of the blood contact surfaces. Carmeda and Trillium are available coatings on the predicate Tri-optic Measurement Cell devices. Balance is a heparin-free version of Trillium coating. Balance is available on the Affinity Pixie Arterial Filter as well as numerous other previously cleared Medtronic devices.

{5}------------------------------------------------

- . Principles of Operation and Technology: The technology of the subject device and the predicate devices are identical. The disposable cells are included within the cardiopulmonary bypass circuit. The disposable cells are connected to a sensor cable connected to the BioTrend instrument that reads the oxygen saturation and hematocrit of the blood going through the circuit.
- . Performance: The performance of the Tri-optic Measurement Cells remains unchanged and continues to meet the performance specifications.

## Summary of Performance Data

Bench testing was used to demonstrate the performance characteristics of the Tri-optic Measurement Cells with Balance Biosurface coating. The testing summarized in Table 5-2 was performed to demonstrate substantial equivalence:

| Test                | Description                                                                         | Results |
|---------------------|-------------------------------------------------------------------------------------|---------|
| Coverage<br>Testing | Ensures the coating has been<br>applied successfully                                | Pass    |
| Leaching<br>Testing | Ensures the coating adheres<br>properly                                             | Pass    |
| Accuracy            | Ensures coating does not interfere<br>with the accuracy of reading the<br>flow rate | Pass    |

Table 5-2: Testing Summary

Clinical testing was not required to establish substantial equivalence. Performance testing was deemed unnecessary for the labeling changes related to the remainder of the models described in this 510(k).

## Conclusion

Medtronic has demonstrated that the Tri-optic Measurement Cells are substantially equivalent to the predicate devices based upon design, test results, and indications for use.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY/K133157](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DRY/K133157)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
