EDWARDS PRIMA PLUS STENTLESS BIOPROSTHESIS MODEL 2500P
Applicant
Edwards Lifesciences, LLC
Product Code
DYE · Cardiovascular
Decision Date
Feb 27, 2001
Decision
APWD
Regulation
21 CFR 870.3925
Device Class
Class 3
Attributes
Therapeutic
Intended Use
The Edwards Prima Plus Stentless Bioprosthesis Model 2500P is indicated for patients who require replacement of their native or prosthetic aortic valve using the subcoronary implantation technique.
Device Story
Porcine aortic valve bioprosthesis; preserved in buffered glutaraldehyde; treated with XenoLogiX process (ethanol and polysorbate-80) to reduce antigenicity and increase tissue stability. Designed for aortic position; available in 21, 23, 25, 27mm diameters. Features woven polyester cloth at inflow annulus for suture support; includes green and black marking sutures for anatomical alignment and trim guide for subcoronary implantation. Implanted by cardiac surgeons in clinical settings. Functions as replacement heart valve to restore hemodynamic flow; reduces transvalvular pressure gradients. Benefits include improved functional status (NYHA class) and hemodynamic performance. Clinical evidence from 366 patients demonstrates safety and effectiveness over mean 2.9-year follow-up.
Clinical Evidence
Two prospective, non-randomized clinical studies (N=366 total; 1074.2 patient-years). Primary endpoints: adverse events (mortality, explant, thromboembolism, structural deterioration) and effectiveness (NYHA functional class, echocardiographic hemodynamic assessment). Results: 1-year freedom from valve-related mortality 98.3%; 8-year freedom from reoperation 100%. Significant improvement in NYHA class observed post-operatively. Hemodynamic data (mean gradients and EOA) provided for sizes 19-29mm.
Technological Characteristics
Porcine aortic valve tissue; glutaraldehyde-fixed; XenoLogiX process (ethanol/polysorbate-80). Components: woven polyester cloth, PTFE thread, PTFE-impregnated PET thread, black silk suture. Dimensions: 21-27mm diameters. Sterilization: terminal glutaraldehyde. Standards: ISO 10993-1 (biocompatibility), ISO 5840:1996 (structural performance).
Indications for Use
Indicated for patients requiring replacement of native or prosthetic aortic valve via subcoronary implantation. Patient population includes adults with aortic valve disease (stenosis, regurgitation, mixed disease, or malfunctioning prosthesis).
Regulatory Classification
Identification
A replacement heart valve is a device intended to perform the function of any of the heart's natural valves. This device includes valves constructed of prosthetic materials, biologic valves (e.g., porcine valves), or valves constructed of a combination of prosthetic and biologic materials.
Reference Devices
Carpentier-Edwards Bioprosthesis (CEBP) Aortic Model 2625
Related Devices
P980043 — HANCOCK II PORCINE BIOPROSTHESIS · Medtronic, Inc. · Sep 28, 1999
P970030 — TORONTO SPV VALVE · St. Jude Medical, Inc. · Nov 4, 1997
P010041 — CARPENTIER-EDWARDS S.A.V. BIOPROSTHESIS, MODEL 2650 (AORTIC) · Edwards Lifesciences, LLC · Jun 24, 2002
P040021 — SJM BIOCOR VALVE / SJM BIOCOR SUPRA VALVE · ABBOTT MEDICAL · Aug 5, 2005
Submission Summary (Full Text)
{0}
P000007
Page 1 of 14
# Summary of Safety and Effectiveness Data
## Edwards Prima Plus Stentless Bioprosthesis Model 2500P
### Table of Contents
1. GENERAL INFORMATION ... 2
2. INDICATIONS FOR USE ... 2
3. DEVICE DESCRIPTION ... 2
4. CONTRAINDICATIONS ... 2
5. WARNINGS AND PRECAUTIONS ... 3
5.1. WARNINGS ... 3
5.2. PRECAUTIONS ... 3
6. ALTERNATIVE PRACTICES AND PROCEDURES ... 3
7. MARKETING HISTORY ... 3
8. ADVERSE EVENTS ... 3
9. SUMMARY OF NONCLINICAL STUDIES ... 4
9.1. BENCH TESTING ... 4
9.1.1. Biocompatibility Studies ... 4
9.1.2. Hydrodynamic Performance ... 6
9.1.3. Structural Performance ... 7
9.2. ANIMAL STUDIES ... 7
9.2.1. Valve Implantation Studies ... 7
9.2.2. Subcutaneous Implantation Studies ... 8
9.3. STERILIZATION ... 9
9.4. SHELF LIFE ... 9
9.4.1. Package Integrity ... 9
9.4.2. Product Integrity ... 9
10. SUMMARY OF CLINICAL STUDIES ... 10
10.1. DESCRIPTION OF PATIENTS AND ANALYSIS FOR GENDER BIAS ... 13
11. RISK-BENEFIT ANALYSIS ... 13
12. CONCLUSIONS DRAWN FROM THE STUDIES ... 13
13. PANEL RECOMMENDATIONS ... 13
14. APPROVAL SPECIFICATIONS ... 14
8
{1}
# Summary of Safety and Effectiveness Data
Edwards Prima Plus Stentless Bioprosthesis Model 2500P
## 1. GENERAL INFORMATION
Device Generic Name: Replacement Heart Valve
Device Trade Name: Edwards Prima Plus Stentless Bioprosthesis Model 2500P
Applicant’s Name and Address: Edwards Lifesciences LLC
One Edwards Way
Irvine, CA 92614
PMA Application Number: P000007
Date of Panel Recommendation:
Date of Notice of Approval to the Applicant: FEB 27 2001
## 2. INDICATIONS FOR USE
The Edwards Prima Plus Stentless Bioprosthesis Model 2500P is indicated for patients who require replacement of their native or prosthetic aortic valve using the subcoronary implantation technique.
## 3. DEVICE DESCRIPTION
The Edwards Prima Plus Stentless Bioprosthesis Model 2500P is a porcine valve aortic root cylinder that has been preserved in a buffered glutaraldehyde solution. The bioprosthesis is treated according to the Edwards XenoLogiX process, which uses ethanol and polysorbate-80 (a surfactant), and is packaged and terminally sterilized in glutaraldehyde. Glutaraldehyde is shown to both reduce the antigenicity of tissue xenograft valves and increase tissue stability; however, glutaraldehyde has not been shown to affect or reduce the calcification rate of the valve.
The Edwards Prima Plus Stentless Bioprosthesis Model 2500P is designed for the aortic position and is available in the following implantation diameters: 21, 23, 25, and 27mm.
Woven polyester cloth is sewn with green suture around the inflow annulus to give additional support to the first suture line. Green marking sutures midway around the intercommissural periphery aid in the placement of stitches around the annulus. Black marking sutures at the mid-commissural positions on the inflow rim aid in proper alignment with the patient’s anatomy. A green trim guide placed externally on the valve wall indicates the recommended limit for trimming the valve for subcoronary implantation while maintaining adequate tissue for placement of the second suture line.
## 4. CONTRAINDICATIONS
None known.
P000007
Page 2 of 14
{2}
# 5. WARNINGS AND PRECAUTIONS
Please refer to device labeling for a list of the warnings and precautions.
# 6. ALTERNATIVE PRACTICES AND PROCEDURES
The alternative to the Edwards Prima Plus Stentless Bioprosthesis Model 2500P is surgical replacement of the malfunctioning aortic valve with an allograft or another prosthetic replacement heart valve for which there is an approved premarket approval application (PMA). When a replacement heart valve is chosen as the appropriate therapy, the option of choosing between a mechanical or biological valve prosthesis exists. The choice of replacement heart valve depends on an assessment of patient factors that include age, preoperative condition, anatomy, and the patient’s ability to tolerate long-term anticoagulant therapy.
Other forms of treatment may include the use of cardiac drug therapy or other types of surgical treatment, such as native valve reconstruction or modification.
# 7. MARKETING HISTORY
Currently the Edwards Prima Plus Stentless Bioprosthesis Model 2500P is distributed in Australia, Austria, Belgium, Canada, Chile, China, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, South Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, the United Kingdom, and Uruguay.
The Edwards Prima Plus Stentless Bioprosthesis Model 2500P has not been withdrawn from marketing in any country for any reason relating to the safety and/or the effectiveness of the device.
# 8. ADVERSE EVENTS
Two multi-center, non-randomized, prospective clinical studies were conducted. The first study was a long-term evaluation of 160 patients implanted with the Edwards Prima Stentless Bioprosthesis Model 2500 in the subcoronary configuration and was conducted between 1991 and 1999. The second study was a short-term evaluation of 206 patients implanted with the Edwards Prima Plus Stentless Bioprosthesis Model 2500P in the subcoronary configuration and was conducted between 1998 and 2000. In the long-term study, patients were evaluated preoperatively, intraoperatively/at discharge, at 3 to 6 months, at 1 year, and annually thereafter. In the short-term study, patients were evaluated preoperatively, intraoperatively/at discharge, at 3 to 6 months, and at 1 year.
Table 1 presents the observed rates for early adverse events (≤30 days for valve-related adverse events), the linearized rates for late adverse events (>30 days postoperatively), and the cumulative freedom from adverse event rates at 1, 5, and 8 years postoperatively. The adverse event rates were based on 366 patients at 13 centers, with one center participating in both the long-term and short-term studies. The cumulative follow-up was 1074.2 patient-years with a mean follow-up of 2.9 years (SD=2.9 years, range=0 to 8.2 years).
P000007
Page 3 of 14
{3}
All patients analyzed: N= 366 Cumulative follow-up: 1074.2 patient-years
Table 1: Observed Adverse Event Rates (Subcoronary Implant Technique)
| Complication | Early Events | | Late Events¹ | | Freedom from Event (%) ± 95% CI² | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | n³ | % | n | %/pt.-yr. | 1 year (n = 366) | 5 years (n = 134) | 8 years (n = 56) |
| Mortality (all) | 12 | 3.3 | 42 | 4.0 | 94.6 ± 2.9 | 81.3 ± 6.3 | 66.6 ± 37.7 |
| Vulve-related mortality | | | | | | | |
| Valve-related mortality | 2 | 0.5 | 18 | 1.7 | 98.3 ± 1.7 | 93.1 ± 4.4 | 87.0 ± 30.7 |
| Explant | 0 | 0.0 | 6 | 0.6 | 99.6 ± 0.8 | 96.9 ± 3.0 | 95.1 ± 20.6 |
| Reoperation⁴ | 0 | 0.0 | 0 | 0.0 | 100 ± 0.0 | 100 ± 0.0 | 100 ± 0.0 |
| Bleeding | 7 | 1.9 | 9 | 0.9 | 95.2 ± 2.8 | 95.2 ± 2.8 | 93.5 ± 27.1 |
| Endocarditis | 0 | 0.0 | 9 | 0.9 | 99.2 ± 1.2 | 94.5 ± 4.0 | 94.5 ± 21.7 |
| Hemolysis | 0 | 0.0 | 0 | 0.0 | 100 ± 0.0 | 100 ± 0.0 | 100 ± 0.0 |
| Nonstructural dysfunction⁵ | 5 | 1.4 | 9 | 0.9 | 96.1 ± 2.6 | 96.1 ± 3.6 | 96.1 ± 26.4 |
| Perivalvular leak | 5 | 1.4 | 8 | 0.8 | 96.5 ± 2.5 | 96.5 ± 3.4 | 96.5 ± 25.1 |
| Structural valve deterioration | 0 | 0.0 | 11 | 1.1 | 100 ± 0.0 | 96.0 ± 3.4 | 86.8 ± 30.9 |
| Thromboembolism | 12 | 3.3 | 28 | 2.7 | 95.8 ± 2.7 | 84.8 ± 6.2 | 82.7 ± 38.9 |
| Valve thrombosis | 0 | 0.0 | 0 | 0.0 | 100 ± 0.0 | 100 ± 0.0 | 100 ± 0.0 |
Notes:
1. Late event rates were calculated as linearized rates (%/pt-yr) based on 1044.3 late patient-years (>30 days postoperatively).
2. Freedom from event rates were calculated using the Kaplan-Meier method. Greenwood's formula was used for calculation of the 95% confidence intervals.
3. n = number of patients
4. Includes reoperation without valve explant.
5. Nonstructural dysfunction includes perivalvular leak. All operative nonstructural dysfunction events were perivalvular leaks.
## 9. SUMMARY OF NONCLINICAL STUDIES
### 9.1. Bench Testing
In vitro studies were performed for the Edwards Prima Plus Stentless Bioprosthesis Model 2500P as recommended in the FDA’s Draft Replacement Heart Valve Guidance (1994). Although tested in the nonclinical studies, the clinical study (Section 10) did not generate sufficient data to support the safety and effectiveness of sizes 19 and 29 aortic valves. The data from the preclinical testing of these sizes are included in the summaries below since the results were used in the overall evaluation of the approved devices.
### 9.1.1. Biocompatibility Studies
Biocompatibility tests were performed in accordance with the requirements of ISO 10993-1, with the exception of carcinogenicity and hemocompatibility testing. Carcinogenicity testing was determined to be unnecessary because the test articles demonstrated no mutagenic potential at levels at or above those intended for the clinical application. Device hemocompatibility was evaluated and found to be acceptable in animal implantation studies (refer to Section 9.2.1). All studies were performed by Edwards Lifesciences LLC, Irvine, CA in accordance with the FDA GLP Regulations (21 CFR 58). A matrix of the tests performed and the corresponding results are provided in Table 2.
P000007
Page 4 of 14
{4}
Table 2: Biocompatibility Tests and Results
| Test Performed | Test Objective | Samples: Control | Samples: Prima Plus | Results |
| --- | --- | --- | --- | --- |
| In vitro inhibition of cell growth | Assess the effect of the aqueous extract of a material on the normal growth of cells in culture. The sample is considered non-inhibitory to cell growth if the percent of inhibition is equal to or less than 29% | Negative control only: Water | Polyethylene terephthalate (PET) cloth
Polytetrafluoroethylene (PTFE) thread
PTFE impregnated PET thread
Black silk suture thread | Non-inhibitory to cell growth. 0% inhibition.
Non-inhibitory to cell growth. 0% inhibition.
Non-inhibitory to cell growth. 0% inhibition.
Non-inhibitory to cell growth at a concentration representative of that used in the device. Inhibitory to cell growth at elevated sample concentrations. |
| In vitro cytotoxicity (Medium eluate method) | Evaluate the cytotoxic effects of a material growth medium extract on a human fibroblast monolayer. A sample is judged non-cytotoxic if lysis is not greater than the negative control. | Negative Control: Cell growth medium
Positive Control: Approximately 5% Ethanol in water | PET cloth
PTFE thread
PTFE impregnated PET thread
Black silk suture thread | Non-cytotoxic to cells. 0% cell lysis.
Non-cytotoxic to cells. 0% cell lysis.
Non-cytotoxic to cells. 0% cell lysis.
Non-cytotoxic at concentrations representative of that used in the device. Cytotoxic at concentrations above those used on the device. |
| In vitro cytotoxicity (Agar overlay assay) | Evaluate the cytotoxicity of diffusible components of a material through an agar overlay assay. A sample is judged non-cytotoxic if lysis is not greater than the negative control. | Negative control: Polypropylene solid sample
Positive control: Polyvinyl chloride (PVC) with Organotin | PET cloth
PTFE thread
PTFE impregnated PET thread
Black silk suture thread | Non-cytotoxic to cells. 0% cell lysis.
Non-cytotoxic to cells. 0% cell lysis.
Non-cytotoxic to cells. 0% cell lysis.
Moderate to severe cytotoxicity (20 to 60% cell lysis) due to glutaraldehyde and formaldehyde residuals present in these exaggerated sample sizes and under the static environments imposed in this in vitro test. |
| In vitro mutagenicity (Sister chromatid exchange assay) | Detect the presence of mutagenic moieties in biomaterials using activated and non-activated systems. | Negative control: Distilled water or the corresponding medium used for the test article extraction. Positive control (non-activated system): Distilled water with mitomycin C @ 0.005 μg/mL). Positive control (activated system): Distilled water with cyclophosphamide @ 1.0 μg/mL) | PET cloth
PTFE thread
PTFE impregnated PET thread
Black silk suture thread | Non-mutagenic using activated and non-activated systems.
Non-mutagenic using activated and non-activated systems.
Non-mutagenic using activated and non-activated systems.
Non-mutagenic at all concentrations using the activated system and at concentrations representative of the final device using the non-activated system |
| USP mouse systemic injection | Evaluate the systemic effect of a material extract in mice. The sample is considered systemically non-toxic if all the mice treated with the sample extract survive at the end of 72 hours and none shows an outward symptom of greater reaction or weight change than mice treated with the negative control. | Negative control: Normal saline and vegetable oil or the corresponding medium used for the test article extraction | PET cloth
PTFE thread
PTFE impregnated PET thread
Black silk suture thread | All mice normal. Non-toxic.
All mice normal. Non-toxic.
All mice normal. Non-toxic.
All mice normal. Non-toxic. |
| USP rabbit intracutaneous irritation | Evaluate the effects of a material extract in contact with the dermis of rabbits. The sample is considered non-irritating if the average erythema/edema rating for any given time is not remarkably greater than that for the negative control. | Negative control: Normal saline and vegetable oil or the corresponding medium used for the test article extraction | PET cloth
PTFE thread
PTFE impregnated PET thread
Black silk suture thread | All rabbits normal. Non-irritating.
All rabbits normal. Non-irritating.
All rabbits normal. Non-irritating. |
P000007
Page 5 of 14
/2
{5}
Table 2: Biocompatibility Tests and Results (continued)
| Test Performed | Test Objective | Samples: Control | Samples: Prima Plus | Results |
| --- | --- | --- | --- | --- |
| USP rabbit intramuscular implantation test (subchronic and chronic) | Evaluate the effect of direct exposure of the test material when implanted into the paravertebral muscle of rabbits for 7, 30, 60, or 90 days. A material is biocompatible if there is no gross visible evidence of tissue damage and if histopathological examination shows no signs of chemical-induced cytotoxicity | Negative control: Polyethylene 306 | PET cloth
PTFE thread
PTFE impregnated PET thread
Black silk suture thread | Material is biocompatible (sub-chronic and chronic evaluations) with no signs of chemical-induced cytotoxicity.
Material is biocompatible (sub-chronic and chronic evaluations) with no signs of chemical-induced cytotoxicity.
Material is biocompatible (sub-chronic and chronic evaluations) with no signs of chemical-induced cytotoxicity.
Material is biocompatible (sub-chronic and chronic evaluations) with no signs of chemical-induced cytotoxicity. |
| Guinea pig maximization test | Evaluate the potential of a material to produce sensitization when the material saline extract is repeatedly exposed to guinea pigs. A material is considered to possess no apparent sensitizing properties if the erythema and edema score is not remarkably greater than the negative control. | Negative control: Normal saline and vegetable oil or the corresponding medium used for the test article extraction | PET cloth
PTFE thread
PTFE impregnated PET thread
Black silk suture thread | All guinea pigs normal. Non-sensitizing.
All guinea pigs normal. Non-sensitizing.
All guinea pigs normal. Non-sensitizing.
All guinea pigs normal. Non-sensitizing. |
## 9.1.2. Hydrodynamic Performance
*In vitro* hydrodynamic performance studies of the Edwards Prima Plus Stentless Bioprosthesis Model 2500P (sizes 19, 25, and 29mm in the subcoronary configuration) were performed in accordance with a tailored protocol reviewed and approved by FDA. Sizes 19, 25, and 31mm Carpentier-Edwards Bioprosthesis (CEBP) Aortic Model 2625 porcine valves were used as a reference in studies requiring concurrent testing of a tissue valve marketed in the U.S. A matrix of the hydrodynamic tests and results is provided in Table 3.
Table 3: Hydrodynamic Testing and Results
| Test | Sample Size: Prima Plus Stentless | Sample Size: Reference Valve (CEBP) | Results |
| --- | --- | --- | --- |
| Steady Forward Flow Pressure Drop | 3 of each | 1 of each | Pressure drop < reference valve |
| Steady Backflow Leakage Testing | 3 of each | 1 of each | Leakage rates > reference valve |
| Pulsatile Flow Pressure Drop | 3 of each | 1 of each | Relatively low and comparable pressure drops |
| Pulsatile Flow Regurgitation | 3 of each | 1 of each | Relatively low and comparable leakage rates |
| Flow Visualization | 1 - 19 mm | N/A^{1} | Acceptable flow patterns |
| Verification of the Bernoulli Relationship | 3 of each | N/A | Good correlation in transvalvular pressure drop obtained by Doppler ultrasonography and transducer |
Note:
1. N/A = not applicable
*In vitro* hydrodynamic pulsatile flow performance studies of the Edwards Prima Stentless Bioprosthesis Model 2500 (sizes 19, 25, and 29mm in the subcoronary configuration) were also performed in accordance with the tailored protocol reviewed and approved by FDA. Results
P000007
Page 6 of 14
/3
{6}
from these studies, when statistically compared with the corresponding results from the studies above, indicate that no clinically significant differences exist in valve performance.
## 9.1.3. Structural Performance
In vitro structural performance (accelerated wear) studies of the Edwards Prima Plus Stentless Bioprosthesis Model 2500P (sizes 19, 25, and 29mm in the subcoronary configuration) were performed in accordance with testing recommendations outlined in the FDA’s Draft Replacement Heart Valve Guidance (1994), ISO 5840:1996 Cardiovascular Implants - Cardiac Valve Prostheses, and CEN/TC 285 Non-Active Surgical Implants - Part 1. Heart Valve Substitutes. A 31mm Carpentier-Edwards Bioprosthesis (CEBP) Aortic Model 2625 porcine valve was used as a reference in studies requiring concurrent testing of a tissue valve marketed in the U.S. All test and reference valves were final production samples. A matrix of the structural performance tests performed on the device are provided in Table 4.
Table 4: Structural Performance Testing and Results
| Test | Sample Size: Prima Plus Stentless | Sample Size: Reference Valve | Results |
| --- | --- | --- | --- |
| Accelerated Wear Testing | 3 of each | 1-31 mm | None of the Prima Plus stentless valves or the reference valves showed any failure during durability testing out to 5 equivalent years. Valves displayed good opening throughout durability testing.
The visual inspection observations were supported by the valve regurgitation results, which did not increase with durability testing time. The mean pressure drop in Prima Plus stentless and in the control valves decreased after 5 equivalent years durability testing |
## 9.2. Animal Studies
### 9.2.1. Valve Implantation Studies
Two chronic in vivo animal implantation studies were conducted using Edwards Prima Stentless Bioprosthesis Model 2500 valves implanted in a healthy juvenile sheep model. A total of nine valves were implanted in the aortic position using the subcoronary implant technique for a total of five months. All nine animals remained healthy throughout the in-life period. The animals demonstrated no clinical signs indicative of valve-related abnormalities over the five-month (20-week) evaluation period.
Parameters evaluated during the study included physical observations, surgical implant observations, hematology and blood chemistry measurements (prior to implant and at explant), cardiac output and peak transvalvular gradients (at explant only), explant valve analysis for calcium and phosphate content, necropsy observations, and histopathological evaluation of selected organs and of the explanted valve and host tissue.
### Clinical Chemistry and Hematology
Hematology and blood chemistry measurements were within normal limits for the age and size of sheep evaluated.
### Hemodynamic Performance
Cardiac outputs and peak transvalvular gradient measurements conducted at explant were as follows: cardiac output: $4.1 \pm 0.2$ L/min, and peak gradient: $27 \pm 12$ mmHg. Left ventricular catheterization and angiography performed at explant on four sheep showed no detectable
P000007
{7}
regurgitation in two sheep; mild perivalvular regurgitant jets (1+) in one sheep; and mild regurgitant jets at the valvular coaptation (1+) in one sheep.
## Histopathology
All surviving animals were sacrificed at approximately 20 weeks post-implant. Selected systemic organs were grossly examined and microscopically evaluated; no untoward effects were noted. The bioprosthetic valve and sheep host tissue were explanted and x-rayed for appearance prior to being microscopically examined. Histopathologically, there was evidence of calcification in one of the nine sheep. Histologically, this series of explants demonstrated consistent findings with those previously observed in porcine aortic valve bioprostheses. There was evidence of cuspal calcification in one of the nine explanted bioprosthetic valves. The studies also demonstrate that the calcification of the aortic wall tissue is expected to occur at a more rapid rate than that of the cuspal tissue.
## Anticalcification Treatment Effectiveness
Samples of the explanted bioprosthetic valve leaflets and the sheep native tissue were evaluated for calcification by measuring calcium (Ca) and phosphate (PO₄) content. The measured values were not considered significant unless they were 1% or greater over the background measurement. All results were under this threshold except for leaflet samples from one sheep and wall samples from one sheep. Of the nine valves, two valves (20%) had elevated quantitative calcium content in the leaflet or wall tissue versus the remaining seven valves after 20 weeks of implantation. The measured levels (mean ± std. dev.) of calcium and phosphate in the explanted leaflet tissue and wall tissue were 6.3 ± 18 mg calcium/g dry tissue weight and 4.4 ± 1.8 mg PO₄/g dry tissue weight, and 4.2 ± 9.7 mg calcium/g dry tissue weight and 8.2 ± 13 mg PO₄/g dry tissue weight, respectively.
## Handling Characteristics
All valves were sewn in with relative ease and observed to have good coaptation and fit within each annulus.
## 9.2.2. Subcutaneous Implantation Studies
Two in vivo subcutaneous implantation studies in rats and rabbits were performed. Porcine valve leaflet tissue exposed to the Edwards Lifesciences XenoLogiX process (fixation in glutaraldehyde, processing in a solution containing ethanol and polysorbate 80 [a surfactant], and packaging in glutaraldehyde) was tested against tissue exposed to glutaraldehyde only. Samples were implanted into subcutaneous pockets created in weanling rats approximately 24 to 28 days of age and into juvenile rabbits approximately 8 weeks of age. Implant duration ranged from approximately 30 days to 90 days from the date of implantation. After explant, samples were evaluated for x-ray evaluation, histological evaluation, and quantitative elemental results. The results indicate that porcine leaflet tissues exposed to the Edwards Lifesciences process show a statistically significant reduction in calcification potential when compared to samples that are exposed to the glutaraldehyde fixation process alone (p<0.05). The clinical significance of these study results is unknown. A matrix of the subcutaneous implant studies performed is provided in Table 5.
P000007
Page 8 of 14
/5
{8}
Table 5: Subcutaneous Implant Study Results
| Study and Test Parameter | Results: Prima Porcine Leaflet Tissue (n=3) | Results: Glutaraldehyde Porcine Leaflet Tissue (n=3) | Statistical Analysis Results |
| --- | --- | --- | --- |
| 10-Day Oral Subcutaneous Implant Study | | | |
| X-ray evaluation^{1} | 1.1 ± 1.2 | 3.0 ± 0.0 | p<0.05 |
| Histological evaluation^{2} | 1.4 ± 1.4 | 3.7 ± 0.5 | p<0.05 |
| Elemental analyses^{3} | Calcium: 56 ± 69
Phosphate: 74 ± 89 | Calcium: 218 ± 38
Phosphate: 324 ± 38 | p<0.05
p<0.05 |
| 14-Day Oral Subcutaneous Implant Study | | | |
| X-ray evaluation^{1} | 1.9 ± 1.2 | 3.0 ± 0.0 | p<0.05 |
| Histological evaluation^{2} | 1.4 ± 1.4 | 3.3 ± 0.5 | p<0.05 |
| Elemental analyses^{3} | Calcium: 102 ± 77
Phosphate: 126 ± 94 | Calcium: 250 ± 25
Phosphate: 360 ± 23 | p<0.05
p<0.05 |
1 Explanted tissue is examined by x-ray and graded for degree of calcification: 0=none; 1=mild; 2=moderate; 3=severe. Statistical analyses between groups performed using the Wilcoxon rank sum test.
2 Explanted tissue is Von Kassa stained and examined histologically for the presence of calcium phosphate: 0=negative; 1=minimal; 2=mild; 3=moderate; 4=marked; 5=severe. Statistical analyses between groups performed using the Wilcoxon rank sum test.
3 Explanted tissue is analyzed for calcium and phosphate content. Results are reported as mg calcium (or phosphate) per g dry tissue weight. Statistical analyses between groups performed using a two-sided t-test.
## 9.3. Sterilization
The Edwards Prima Plus Stentless Bioprosthesis Model 2500P is terminally sterilized in buffered glutaraldehyde solution. After terminal sterilization, the product is held in quarantine until sterility is verified per process specifications. Requalification of the process is performed quarterly.
## 9.4. Shelf Life
Both packaging and product integrity studies were conducted to ensure that the shelf life for the package and product is maintained for a minimum of four (4) years. Packaging integrity studies consisted of real-time and accelerated aging, whereas product integrity samples underwent real-time aging.
## 9.4.1. Package Integrity
The integrity of the valve packaging components was evaluated after exposure to the maximum steam sterilization cycles and terminal liquid sterilization process. Package integrity testing consisted of physical (leak and glutaraldehyde packaging solution concentration) and sterility testing before and after exposure to glutaraldehyde in an elevated temperature condition, and after a simulated shipping process. Accelerated aging results simulating 0, 1, and 4 years real-time demonstrated package integrity throughout the 4-year shelf life period. Packaging validation studies conducted after maximum exposure to the terminal liquid sterilization process demonstrated that this sterilization method does not adversely affect package integrity.
## 9.4.2. Product Integrity
### Non-biological Component Shelf Life
Thread and cloth components were evaluated by functional testing of the individual non-biological materials after 4 years of real-time storage in glutaraldehyde. Results demonstrate that storage in glutaraldehyde for up to 4 years has minimal effect on the properties and functions of the individual non-biological materials used in the valve.
P000007
Page 9 of 14
{9}
# Tissue Shelf Life
Porcine valve tissue stability and storage solution adequacy were evaluated using three parameters: shrinkage temperature, moisture content, and glutaraldehyde concentration. Tissue samples subjected to real-time aging were evaluated at designated intervals for shrinkage temperature and moisture content. Glutaraldehyde content of the storage solution was determined by glutaraldehyde assay.
The results demonstrated that the tissue shrinkage temperature is stable over time at the recommended storage temperature of $4^{\circ}$ to $25^{\circ}\mathrm{C}$ for a duration exceeding the 4-year shelf life. The effects of storage time on the moisture content were monitored because chemical changes in the tissue could affect the hydration level of the tissue. A gradual decrease in moisture content with time was seen, with a more rapid decline at higher temperatures. Glutaraldehyde assays showed the expected trend of a gradual increase in concentration over time, with a more rapid increase at higher storage temperatures. Acceptable levels of glutaraldehyde concentration were maintained for the 4-year shelf life period in the recommended storage temperature range of $4^{\circ}$ to $25^{\circ}\mathrm{C}$. These results demonstrate product integrity to 4 years.
# 10. SUMMARY OF CLINICAL STUDIES
The safety endpoints captured in the prospective studies were adverse events; blood analyses were used to confirm the absence or presence of hemolysis, hemolytic anemia, and endocarditis. The safety results are provided above in Table 1. Effectiveness endpoints were New York Heart Association (NYHA) functional classification and echocardiographic assessments. Preoperative and operative patient demographics are presented below, followed by the effectiveness results. There were insufficient clinical data to support the safety and effectiveness of this device for root inclusion or full root implantation.
Table 6: Preoperative Patient Demographics
| Variable | Category | Study Results (N=366; 1074.2 total pt-yrs.) | |
| --- | --- | --- | --- |
| | | n | % (n/N)¹ |
| Age at implant | Mean ± SD | 366 | 70.2± 7.1 |
| Gender | Male | 217 | 59.3% |
| | Female | 149 | 40.7 % |
| NYHA Classification | I | 23 | 6.3% |
| | II | 138 | 37.7% |
| | III | 177 | 48.4% |
| | IV | 25 | 6.8% |
| | Not reported | 3 | 0.8% |
| Diagnosis | Stenosis | 243 | 66.4% |
| | Regurgitation | 27 | 7.4% |
| | Mixed Disease | 94 | 25.7% |
| | Malfunctioning prosthesis | 2 | 0.5% |
Note:
1. n = number of patients in each category; N = total number of study patients.
P000007
Page 10 of 14
{10}
Table 7: Operative Patient Demographics
| Variable | Category | Study Results (N=366; 1074.2 total pt-yrs.) | |
| --- | --- | --- | --- |
| | | n | % (n/N)¹ |
| Etiology² | Calcification/degeneration | 297 | 81.1% |
| | Rheumatic heart disease | 33 | 9.0% |
| | Congenital abnormalities | 31 | 8.5% |
| | Other³ | 6 | 1.6% |
| Concomitant Procedures² | None | 230 | 62.8% |
| | CABG⁴ | 122 | 33.3% |
| | AAA⁵ repair | 5 | 1.4% |
| | Mitral valve repair | 3 | 0.8% |
| | Mitral valve replacement | 1 | 0.3% |
| | Other⁶ | 8 | 2.2% |
| Pre-existing Conditions² | None | 143 | 39.1% |
| | TIA/CVA⁷ | 26 | 7.1% |
| | Congestive Heart Failure | 36 | 9.8% |
| | Arrhythmias | 37 | 10.1% |
| | Systemic Hypertension | 88 | 24.0% |
| | CAD⁸/CABG | 133 | 36.3% |
| Valve Size (mm) | 19 | 7 | 1.9% |
| | 21 | 47 | 12.8% |
| | 23 | 85 | 23.2% |
| | 25 | 123 | 33.6% |
| | 27 | 81 | 22.1% |
| | 29 | 23 | 6.3% |
Notes:
1. n = number of patients in each category; N = total number of study patients
2. May be more than one per patient
3. Includes previously failed prosthesis, root dilatation, and ischemic disease
4. CABG = Coronary Artery Bypass Graft
5. AAA=Abdominal Aortic Aneurysm
6. Includes carotid endarterectomy, fistula exploration, ventricular septal defect repair, aortotomy, intra-aortic balloon pump, tumorectomy, and interatrial septum exploration
7. TIA = Transient ischemic attack. CVA = Cerebrovascular accident.
8. CAD = Coronary Artery Disease
Table 8: Effectiveness Outcomes, Functional NYHA
| NYHA Functional Class | Preoperative Assessment | Postoperative Assessments | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | 1 Year | | 4 to 5 Years | |
| | n/N¹ | % | n/N | % | n/N | % |
| I | 17/313 | 5.4% | 184/250 | 73.6% | 58/160 | 36.3% |
| II | 119/313 | 38.0% | 23/250 | 9.2% | 37/160 | 23.1% |
| III | 156/313 | 49.8% | 3/250 | 1.2% | 8/160 | 5.0% |
| IV | 21/313 | 6.7% | 0/250 | 0.0% | 1/160 | 0.6% |
| Not Available | 0/313 | 0.0% | 40/250 | 16.0% | 56/160 | 35.0% |
Notes:
1. n = number of patients in each category; N = total number of study patients
P000007
Page 11 of 14
{11}
Table 9: Effectiveness Outcomes, Hemodynamic Results¹
| Hemodynamic Parameter | Results By Valve Size | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 19mm | 21mm | 23mm | 25mm | 27mm | 29mm |
| Ursults (n=112) | | | | | | |
| Mean gradient² | n = 7 | n = 41 | n = 68 | n = 95 | n = 69 | n = 20 |
| • mean ± sd | 15.4 ± 7.4 | 15.9 ± 7.5 | 13.5 ± 5.6 | 10.7 ± 5.4 | 9.5 ± 5.3 | 7.0 ± 5.2 |
| • min, max | 6.0, 24.0 | 4.0, 37.0 | 2.0, 28.0 | 1.0, 34.0 | 1.0, 30.0 | 1.0, 21.0 |
| EOA³ | n = 6 | n = 37 | n = 62 | n = 80 | n = 55 | n = 16 |
| • mean ± sd | 1.05 ± 0.32 | 1.17 ± 0.33 | 1.35 ± 0.37 | 1.68 ± 0.60 | 1.87 ± 0.54 | 2.71 ± 1.64 |
| • min, max | 0.70, 1.45 | 0.50, 1.96 | 0.64, 2.26 | 0.89, 4.61 | 1.15, 3.60 | 1.20, 8.19 |
| Regurgitation⁴ | n = 7 | n = 43 | n = 70 | n = 98 | n = 72 | n = 21 |
| 0 | 7 (100%) | 33 (76.7%) | 45 (64.3%) | 66 (67.3%) | 50 (69.4%) | 18 (85.7%) |
| 1+ | 0 (0.0%) | 9 (20.9%) | 17 (24.3%) | 25 (25.5%) | 13 (18.1%) | 1 (4.8%) |
| 2+ | 0 (0.0%) | 1 (2.3%) | 8 (11.4%) | 6 (6.1%) | 7 (9.7%) | 2 (9.5%) |
| 3+ | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 1 (1.0%) | 0 (0.0%) | 0 (0.0%) |
| 4+ | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Not available | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 2 (2.8%) | 0 (0.0%) |
| Excess (n=203) | | | | | | |
| Mean gradient² | n = 7 | n = 24 | n = 46 | n = 72 | n = 45 | n = 5 |
| • mean ± sd | 17.8 ± 9.0 | 14.5 ± 6.4 | 11.5 ± 8.7 | 9.5 ± 5.8 | 6.7 ± 3.0 | 4.4 ± 2.7 |
| • min, max | 6.3, 31.0 | 5.0, 29.0 | 2.0, 55.9 | 1.9, 29.0 | 2.0, 17.0 | 1.5, 8.0 |
| EOA³ | n = 7 | n = 21 | n = 44 | n = 61 | n = 37 | n = 4 |
| • mean ± sd | 0.88 ± 0.20 | 1.20 ± 0.48 | 1.43 ± 0.43 | 1.74 ± 0.53 | 2.04 ± 0.62 | 2.64 ± 0.56 |
| • min, max | 0.68, 1.30 | 0.76, 2.50 | 0.60, 2.64 | 0.80, 3.70 | 0.76, 3.42 | 2.12, 3.22 |
| Regurgitation⁴ | n = 7 | n = 26 | n = 48 | n = 74 | n = 45 | n = 5 |
| 0 | 5 (71.4%) | 16 (61.5%) | 29 (60.4%) | 45 (60.8%) | 28 (62.2%) | 3 (60.0%) |
| 1+ | 2 (28.6%) | 9 (34.6%) | 11 (22.9%) | 17 (23.0%) | 15 (33.3%) | 1 (20.0%) |
| 2+ | 0 (0.0%) | 0 (0.0%) | 8 (16.7%) | 11 (14.9%) | 2 (4.4%) | 1 (20.0%) |
| 3+ | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 1 (1.4%) | 0 (0.0%) | 0 (0.0%) |
| 4+ | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Not available | 0 (0.0%) | 1 (3.8%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Excess (n=112) | | | | | | |
| Mean gradient² | n = 2 | n = 13 | n = 26 | n = 38 | n = 21 | n = 1 |
| • mean ± sd | 27.9 ± 5.8 | 15.9 ± 7.0 | 9.9 ± 5.7 | 8.8 ± 4.9 | 6.5 ± 3.9 | 4.2 |
| • min, max | 23.8, 32.0 | 5.5, 32.2 | 3.0, 23.2 | 1.3, 23.0 | 1.0, 14.0 | 4.2, 4.2 |
| EOA³ | n = 1 | n = 12 | n = 23 | n = 30 | n = 16 | n = 0 |
| • mean ± sd | 1.00 | 1.10 ± 0.36 | 1.60 ± 0.51 | 1.91 ± 0.36 | 2.06 ± 0.58 | -- |
| • min, max | 1.00, 1.00 | 0.20, 1.68 | 0.47, 2.60 | 0.93, 4.06 | 1.20, 3.20 | -- |
| Regurgitation⁴ | n = 5 | n = 17 | n = 31 | n = 41 | n = 21 | n = 1 |
| 0 | 4 (80.0%) | 11 (64.7%) | 17 (54.8%) | 24 (58.5%) | 17 (81.0%) | 0 (0.0%) |
| 1+ | 0 (0.0%) | 6 (35.3%) | 9 (29.0%) | 12 (29.3%) | 2 (9.5%) | 0 (0.0%) |
| 2+ | 0 (0.0%) | 0 (0.0%) | 4 (12.9%) | 1 (2.4%) | 1 (4.8%) | 1 (100%) |
| 3+ | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 1 (4.8%) | 0 (0.0%) |
| 4+ | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Not available | 1 (20.0%) | 0 (0.0%) | 1 (3.2%) | 4 (9.8%) | 0 (0.0%) | 0 (0.0%) |
Notes:
1. Hemodynamic evaluations were performed using transthoracic echocardiography (TTE).
2. Mean gradient in mm Hg.
3. EOA: Effective Orifice Area, cm²
4. Regurgitation: 0 = none; 1+ = trivial; 2+ = mild; 3+ = moderate; 4+ = severe
P000007
Page 12 of 14
19
{12}
# 10.1. Description of Patients and Analysis for Gender Bias
A gender bias was not found in the Edwards Lifesciences clinical studies.
Of the 366 patients followed in the clinical studies, 59% were male and 41% were female. This gender distribution is consistent with the incidence of patients presenting for aortic valve replacement in the U.S. The log-rank test was used to compare all adverse event outcomes by gender. No significant difference in outcomes between males and females were noted for any valve-related adverse event. Therefore, the results for valve-related adverse events following aortic valve replacement in this study are representative for both men and women.
# 11. RISK-BENEFIT ANALYSIS
Laboratory and clinical data provide reasonable assurance that the Edwards Prima Plus Stentless Bioprosthesis Model 2500P is safe and effective when used according to the approved labeling.
# 12. CONCLUSIONS DRAWN FROM THE STUDIES
The results from pre-clinical laboratory studies performed on the Edwards Prima Plus Stentless Bioprosthesis Model 2500P for biocompatibility testing, hydrodynamic performance testing (steady forward flow pressure drop, steady backflow leakage testing, pulsatile flow pressure drop, pulsatile flow regurgitation, flow visualization, and verification of the Bernoulli Relationship), and structural performance testing (accelerated wear testing) demonstrate that this device is suitable for long-term implant.
The animal studies show that the Edwards Prima Plus Stentless Bioprosthesis Model 2500P is safe for valve replacement.
The clinical studies submitted in the PMA provide sound scientific evidence that the Edwards Prima Plus Stentless Bioprosthesis Model 2500P is safe and effective for the replacement of native or prosthetic aortic valves using the subcoronary implantation technique.
# 13. PANEL RECOMMENDATIONS
In accordance with the provisions of section 515(c)(2) of the Act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory Systems Device Panel, a FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
# 14. FDA DECISION
The applicant’s manufacturing and control facilities were inspected and the facilities were found to be in compliance with the Quality System Regulation (QSR)(21 CFR Part 820).
FDA issued an approval on February 27, 2001.
P000007
Page 13 of 14
20
{13}
# 15. APPROVAL SPECIFICATIONS
Directions for use: See Final Approved Labeling (Instructions for Use).
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the Final Draft Labeling (Instructions for Use).
Post-approval Requirements and Restrictions: See Approval Order.
P000007
Page 14 of 14
2/
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.