Medtronic Model 5392 External Pulse Generator (EPG)
K201011 · Medtronic, Inc. · DTE · May 13, 2020 · Cardiovascular
Device Facts
Record ID
K201011
Device Name
Medtronic Model 5392 External Pulse Generator (EPG)
Applicant
Medtronic, Inc.
Product Code
DTE · Cardiovascular
Decision Date
May 13, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3600
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient. Specific indications for temporary cardiac pacing include, but are not limited to, the following: Complete heart block Sinus bradycardia Sick sinus syndrome Bradycardia with congestive heart failure Atrial and/or ventricular arrhythmias Cardiac arrest Support, management, and evaluation of a patient prior to permanent pacemaker Implantation Support during permanent pacemaker replacement Cardiac complications during invasive or surgical procedures Support following cardiac surgery Acute myocardial infarction complicated by heart block Atrial tachyarrhythmias that require high-rate burst pacing for treatment
Device Story
Battery-powered, dual-chamber external pulse generator (EPG) for temporary antibradycardia pacing. Used in clinical environments by trained personnel. Connects to cardiac pacing lead systems to deliver demand (synchronous) or asynchronous pacing therapy. Optional disposable protective cover (53922) prevents accidental control activation. Clinicians use device to manage cardiac rhythms during surgery, post-operative recovery, or acute cardiac events. Provides life-sustaining or diagnostic pacing support; requires continuous patient monitoring to ensure therapy delivery. Modifications in this submission relate to updated instructions for use regarding cleaning and disinfection protocols.
Clinical Evidence
Bench testing only. Verification testing performed to evaluate the impact of updated cleaning and disinfection instructions on device design specifications. Results met all acceptance criteria.
Technological Characteristics
Battery-powered, dual-chamber external pulse generator. Design, materials, and operating principles are identical to the predicate. Includes optional disposable protective cover. Verification performed via regression bench testing to ensure compatibility with updated cleaning/disinfection agents.
Indications for Use
Indicated for patients requiring temporary single or dual chamber cardiac pacing for therapeutic, prophylactic, or diagnostic purposes, including complete heart block, sinus bradycardia, sick sinus syndrome, bradycardia with congestive heart failure, arrhythmias, cardiac arrest, pre-permanent pacemaker support, surgical/procedural complications, post-cardiac surgery support, acute myocardial infarction with heart block, and atrial tachyarrhythmias requiring burst pacing.
Regulatory Classification
Identification
An external pacemaker pulse generator (EPPG) is a prescription device that has a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. This device, which is used outside the body, is used as a temporary substitute for the heart's intrinsic pacing system until a permanent pacemaker can be implanted, or to control irregular heartbeats in patients following cardiac surgery or a myocardial infarction. The device may have adjustments for impulse strength, duration, R-wave sensitivity, and other pacing variables.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must validate electromagnetic compatibility (EMC) within a hospital environment.
(2) Electrical bench testing must demonstrate device safety during intended use. This must include testing with the specific power source (
*i.e.,* battery power, AC mains connections, or both).(3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following:
(i) Testing must demonstrate the accuracy of monitoring functions, alarms, measurement features, therapeutic features, and all adjustable or programmable parameters as identified in labeling;
(ii) Mechanical bench testing of material strength must demonstrate that the device and connection cables will withstand forces or conditions encountered during use;
(iii) Simulated use analysis/testing must demonstrate adequate user interface for adjustable parameters, performance of alarms, display screens, interface with external devices (
*e.g.* data storage, printing), and indicator(s) functionality under intended use conditions; and(iv) Methods and instructions for cleaning the pulse generator and connection cables must be validated.
(4) Appropriate software verification, validation, and hazard analysis must be performed.
(5) Labeling must include the following:
(i) The labeling must clearly state that these devices are intended for use in a hospital environment and under the supervision of a clinician trained in their use;
(ii) Connector terminals should be clearly, unambiguously marked on the outside of the EPPG device. The markings should identify positive (+) and negative (−) polarities. Dual chamber devices should clearly identify atrial and ventricular terminals;
(iii) The labeling must list all pacing modes available in the device;
(iv) Labeling must include a detailed description of any special capabilities (
*e.g.,* overdrive pacing or automatic mode switching); and(v) Appropriate electromagnetic compatibility information must be included.
Predicate Devices
Medtronic Model 5392 External Pulse Generator (EPG) (K190825)
Related Devices
K162550 — Medtronic Model 5392 External Pulse Generator (EPG) · Medtronic, Inc. · Oct 12, 2016
K150246 — Medtronic Model 5392 External Pulse Generator (EPG) · Medtronic, Inc. · Feb 18, 2015
K190825 — Medtronic Model 5392 External Pulse Generator (EPG) · Medtronic, Inc. · Apr 30, 2019
K162054 — Medtronic Temporary External Pacemaker 53401 · Medtronic, Inc. · Oct 18, 2016
K180873 — Medtronic Model 53401 External Pulse Generator (EPG) · Medtronic, Inc. · Apr 19, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 13, 2020
Medtronic, Inc. Alexandra Theisen Senior Regulatory Affairs Specialist 8200 Coral Sea Street N.E. Mounds View, Minnesota 55112
Re: K201011
Trade/Device Name: Medtronic Model 5392 External Pulse Generator (EPG) Regulation Number: 21 CFR 870.3600 Regulation Name: External Pacemaker Pulse Generator Regulatory Class: Class III Product Code: DTE Dated: April 16, 2020 Received: April 17, 2020
Dear Alexandra Theisen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number: K190825
Device Name: Medtronic Model 5392 External Pulse Generator (EPG)
Indications for Use: The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient.
Specific indications for temporary cardiac pacing include, but are not limited to, the following:
- Complete heart block ●
- Sinus bradycardia ●
- Sick sinus syndrome ●
- Bradycardia with congestive heart failure ●
- Atrial and/or ventricular arrhythmias ●
- Cardiac arrest
- o Support, management, and evaluation of a patient prior to permanent pacemaker Implantation
- Support during permanent pacemaker replacement ●
- Cardiac complications during invasive or surgical procedures
- Support following cardiac surgery ●
- Acute myocardial infarction complicated by heart block ●
- Atrial tachyarrhythmias that require high-rate burst pacing for treatment ●
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### 510(k) Summary
[As required by 21 CFR 807.92]
| Date Prepared: | April 16, 2020 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Medtronic, Inc.<br>Medtronic Cardiac Rhythm Heart Failure<br>8200 Coral Sea Street N.E.<br>Mounds View, MN 55112<br>Establishment Registration Number: 2182208 |
| Contact Person: | Alexandra Theisen<br>Senior Regulatory Affairs Specialist<br>Medtronic Cardiac Rhythm Heart Failure<br>Phone: 763.526.6758<br>Fax: 651.367.0603<br>Email: alexandra.r.theisen@medtronic.com |
| Alternate Contact: | Wendy Saunders<br>Regulatory Affairs Director<br>Medtronic Cardiac Rhythm Heart Failure<br>Phone: 763.526.8134<br>Fax: 651.367.0603<br>Email: wendy.a.saunders@medtronic.com |
### General Information
| Trade Name: | Medtronic Model 5392 External Pulse Generator (EPG) |
|-----------------------|-------------------------------------------------------------|
| Common Name: | External pacemaker pulse generator |
| Regulation Number: | 21 CFR 870.3600 |
| Product Code: | DTE |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
| Special Controls: | Not applicable |
| Predicate Devices: | Medtronic Model 5392 External Pulse Generator (EPG) K190825 |
{4}------------------------------------------------
# Device Description
The Model 5392 EPG is a battery-powered, dual chamber, temporary pacemaker designed primarily for temporary antibradycardia pacing therapy. The 53922 EPG cover accessory is an optional disposable protective cover to reduce accidental activation of the controls of the 5392 EPG.
### Indications for Use
The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient.
Specific indications for temporary cardiac pacing include, but are not limited to, the following:
- Complete heart block
- Sinus bradycardia
- Sick sinus syndrome ●
- Bradycardia with congestive heart failure ●
- o Atrial and/or ventricular arrhythmias
- Cardiac arrest
- o Support, management, and evaluation of a patient prior to permanent pacemaker Implantation
- Support during permanent pacemaker replacement
- Cardiac complications during invasive or surgical procedures
- Support following cardiac surgery
- Acute myocardial infarction complicated by heart block
- o Atrial tachyarrhythmias that require high-rate burst pacing for treatment
# Technological Characteristics
Intended use, design, materials, performance and technological characteristics are substantially equivalent to the predicate devices referenced.
When compared to the predicate device (K190825), the modified Medtronic Model 5392 External Pulse Generator presented in this submission has the same:
- Intended use/indications for use
- Operating principle
- Design features ●
- Device functionality
- Biological safety ●
- Packaging materials ●
- Shelf life
{5}------------------------------------------------
The modified Model 5392 EPG and the predicate device differ in the following:
- Modified 5392 EPG Instructions for Use (IFU) 0
### Substantial Equivalence and Summary of Studies:
Technological differences between the subject and predicate devices have been evaluated through bench tests. The data from this testing showed that the devices that were cleaned and/or disinfected with the additional cleaners/disinfectants in the modified 5392 EPG IFU continue to meet the current 5392 design specifications. Refer to Device Description Information and Labeling for additional details on the change.
The modified Medtronic Model 5392 External Pulse Generator is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. The device design and materials have been verified through the following:
- Regression bench testing
### Conclusion:
The results of the above verification testing met the specified acceptance criteria and did not raise new safety or performance issues. Therefore, the modifications made to the Medtronic Model 5392 External Pulse Generator described in this submission result in a device that is substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.